F luoroscopy. 2 Fluoroscopy system 3 Different fluoroscopy systems  Remote control systems Not requiring the presence of medical specialists inside.

Slides:



Advertisements
Similar presentations
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Advertisements

Lecture 11. Microscopy. Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single or.
Digital Radiography.
RAD 354 Chapt. 28 The Digital Image Spatial resolution Contrast resolution Contrast-detail curve Pt. dose considerations.
X-RAYS IN MOTION “Viewing dynamic studies of the human body”
Digital FLUORO Summer 2008.
Scanning Electron Microscope Jamie Goings. Theory Conventional microscopes use light and glass lenses SEM uses electrons and magnetic lenses to create.
Digital Technology 14.2 Data capture; Digital imaging using charge-coupled devices (CCDs)
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Fluoroscopy: Viewing Systems
Fluoroscopy Equipment
Quality Control Rad T 110.
RAD 350 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion Invented by Thomas Edison.
DIGITAL FLUOROSCOPY.
Special Imaging Systems
Fluoroscopy Intro to EQUIPMENT
ECE 5th SEMESTER Subject Consumer Electronics (CE)
FLUOROSCOPY EQUIPMENT
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
The Television Picture
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
Image Quality Radiographic Resolution.
Digital Radiology. 2 Aim : To become familiar with the digital imaging techniques in projection radiography and fluoroscopy.Aim : To become familiar with.
Fluoroscopy Equipment. Introduction Fluoro: is dynamic radiographic examination Fluoroscopy is primarily domain of the radiologist The role of radiographer.
Scanning Electron Microscope (SEM)
Control room High resolution flat screen monitors CCD image intensifiers are easily identified by their shape A modern fluoroscopic suite.
Fluoroscopy Q&A Robert L. Metzger, Ph.D..
Angiographic facilities Tortorici Chapter 11. Designed around exam table 500 sq feet min. (25 x 20) 60 sq feet in control room Wide doors, (multiple)
Pulse-Height Analyzers Basic Functions Single Channel Analyzers Time Methods Multi-channel Analyzers.
Introduction to Graphical Hardware Display Technologies
Appendix H: Chapter 40: Fluoroscopy
Chapter 3: Factors of Image Quality 1. Interlaced vs. progressive scanning 2. Matrix size 5. Field of view (FOV) 3. Vertical resolution 4. Horizontal resolution.
What Exactly is Television?  A process of transmitting images through a signal from one place or another.
DIGITAL CAMERAS Prof Oakes. Overview Camera history Digital Cameras/Digital Images Image Capture Image Display Frame Rate Progressive and Interlaced scans.
March 2004 Charles A. DiMarzio, Northeastern University ECEG287 Optical Detection Course Notes Part 15: Introduction to Array Detectors Profs.
Sounds of Old Technology IB Assessment Statements Topic 14.2., Data Capture and Digital Imaging Using Charge-Coupled Devices (CCDs) Define capacitance.
TELEVISION CAMERA TUBES
RAD 254 Chapter 27 Digital Fluoroscopy
Imaging System Components M A Oghabian Medical Physics Group, Tehran University of Medical Sciences
RAD 150 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic (motion)
RAD 354 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Doses in Fluoroscopy factors influencing patient doses Colin Martin and David Sutton.
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Digital Radiology Dr M A Oghabian Medical Physics Group Tehran University of Medical Sciences
FLUOROSCOPY & IMAGE INTENSIFICATION Sergeo Guilbaud, RT School of Radiologic Sciences.
The Cathode Ray Tube Monitor
T ELEVISION Simplified cross-sectional view of a Vidicon TV camera tube 2Communication Systems (EC-326)
Electronics Lecture 5 By Dr. Mona Elneklawi.
Imaging Characteristics
Resident Physics Lectures
Laboratory equipment Lecture (3).
TELEVISION Camera Tube
IMAGE ORTHICON.
Fundamentals of video and TV
Imaging System Components
Fluoroscopic Unit Thomas Edison 1896.
Fluoroscopy: Dynamic Imaging Systems
Resident Physics Lectures
Fluoroscopic Image Intensifier Image production
Fluoroscopy Dynamic Image Recording Systems
Fluoroscopy: Viewing Systems
Fluoroscopy – Viewing Systems TV Monitors
Fluoroscopy – Viewing Systems Optical Mirrors TV Camera & TV Camera Tubes Charge Coupled Devices (CCD) Based on: Principles of Radiographic Imaging, 3rd.
Fluoroscopy – Viewing Systems TV Monitors
Flouroscopic imaging.
Fluoroscopic Image Quality Considerations
Fluoroscopic Image Intensifier Image Production
Fluoroscopy: Static Image Recording Systems
Fluoroscopic Image Quality Considerations
Presentation transcript:

F luoroscopy

2 Fluoroscopy system

3 Different fluoroscopy systems  Remote control systems Not requiring the presence of medical specialists inside the X- ray room  Mobile C-arms Mostly used in surgical theatres.

4 Different fluoroscopy systems  Interventional radiology systems Requiring specific safety considerations. Interventionalists can be near the patient during the procedure.  Multipurpose fluoroscopy systems They can be used as a remote control system or as a system to perform simple interventional procedures

Topic 2 : Image Intensifier component and parameters

+ Input Screen Output Screen Photocathode Electrode E 1 Electrode E 3 Electrode E 2 Electrons Path

7 Image intensifier systems

Image intensifier component * Input screen: conversion of incident X-rays into light photons (CsI) * 1 X-ray photon creates  3,000 light photons * Photocathode: conversion of light photons into electrons * only 10 to 20% of light photons are converted into photoelectrons * Electrodes : focalization of electrons onto the output screen * electrodes provide the electronic magnification * Output screen: conversion of accelerated electrons into light photons

Image intensifier parameters (I) * Conversion coefficient (Gx): the ratio of the output screen brightness to the input screen dose rate [cd.m -2 Gys -1 ] 3 Gx depends on the quality of the incident beam (IEC publication 573 recommends HVL of 7  0.2 mm Al) 3 Gx is directly proportional to: * the applied tube potential * the diameter () of the input screen * input screen of 22 cm  Gx = 200 * input screen of 16 cm  Gx = 200 x (16/22) 2 = 105 * input screen of 11 cm  Gx = 200 x (11/22) 2 = 50

Image intensifier parameters (II)  Brightness Uniformity: the input screen brightness may vary from the center of the I.I. to the periphery Uniformity = (Brightness (c) - Brightness (p) ) x 100 / Brightness (c)  Geometrical distortion: all x-ray image intensifiers exhibit some degree of pincushion distortion. This is usually caused by either magnetic contamination of the image tube or the installation of the intensifier in a strong magnetic environment.

Image distortion

Image intensifier parameters (III)  Spatial resolution limit: the value of the highest spatial frequency that can be visually detected  it provides a sensitive measure of the state of focusing of a system  it is quoted by manufacturer  it can be measured optically  it correlates well with the high frequency limit of the Modulation Transfer Function (MTF)  it can be assessed by the Hüttner resolution pattern

Line pair gauges

Line pair gauges GOOD RESOLUTION POOR RESOLUTION

Image intensifier parameters (IV)  Overall image quality: threshold contrast-detail detection  X-ray, electrons and light scatter process in an I.I. can result in a significant loss of contrast of radiological detail.  The degree of contrast is effected by the design of the image tube and coupling optics.  Spurious sources of contrast loss are:  accumulation of dust and dirt on the various optical surfaces  reduction in the quality of the vacuum  aging process (destruction of phosphor screen)  Sources of noise are:  X-ray quantum mottle  photo-conversion processes

 Overall image quality can be assessed using: A contrast-detail detectability test object (array of disc- shaped metal details which gives a range of diameters and X-ray transmission  Sources of image degradation such as contrast loss, noise and unsharpness limit the number of details that are visible.  Image quality can be detected as a reduction in the number of low contrast and/or small details. Image intensifier parameters (V)

Overall image quality

Topic 3 : Image Intensifier and TV system

Image intensifier - TV system  Output screen image can be transferred to different optical displaying systems:  conventional TV  Generating a full frame of 525 lines (in USA)  625 lines and 25 full frames/s up to 1000 lines (in Europe)  interlaced mode is used to prevent flickering  cinema  35 mm film format: from 25 to 150 images/s  photography  rolled film of 105 mm: max 6 images/s  film of 100 mm x 100 mm

Add module code number and lesson title20 VIDICON FILM PM REFERENCEkV CONTROLLER X-RAY TUBE kV GENERAL SCHEME OF FLUOROSCOPY

Type of TV camera  VIDICON TV camera  improvement of contrast  improvement of signal to noise ratio  high image lag  PLUMBICON TV camera (suitable for cardiology)  lower image lag (follow up of organ motions)  higher quantum noise level  CCD TV camera (digital fluoroscopy)  digital fluoroscopy spot films are limited in resolution, since they depend on the TV camera (no better than about 2 lp/mm) for a 1000 line TV system

TV camera and video signal (I)  Output phosphor of image intensifier is optically coupled to a TV.  A pair of lenses focuses the output image onto the input surface of the television camera.  Often a beam splitting mirror is used in order to reflect part of the light onto a 100 mm camera or cine camera.  Typically, the mirror will reflect 90% of the incident light and transmit 10% onto the television camera.  Older fluoroscopy equipment have a television system using a camera tube with a conductive layer.  In a PLUMBICON tube, this layer is made of lead oxide, whereas in a VIDICON, antimony trisulphide is used

Photoconductive camera tube Focussing optical lens Input plate Steering coils Deviation coil Alignement coil Accelarator grids Control grid Electron beam Video Signal Signal electrode Field grid Electrode Electron gun Iris Photoconductive layer

 The surface of the photoconductor is scanned with an electron beam and the amount of current flowing is related to the amount of light.  The scanning electron beam is produced by a heated photocathode.  Electrons are emitted into the vacuum and accelerated across television camera tube by applying a voltage.  Electron beam is focussed by a set of focussing coils. TV camera and video signal (III)

 This scanning electron beam moves across the surface of the TV camera tube in a series of lines by a series of external coils.  In a typical television system, on the first pass the set of odd numbered lines are scanned followed by the even numbers (interlaced).  The purpose of interlacing is to prevent flickering of the television image on the monitor, by increasing the apparent frequency of frames (50 half frames/second).  In Europe, 25 frames are updated every second. TV camera and video signal (IV)

Different types of scanning INTERLACED SCANNING PROGRESSIVE SCANNING lines in 40 ms i.e. : 25 frames/s

 The video signal comprises a set of repetitive synchronizing pulses. In between there is a signal that is produced by the light falling on the camera surface.  The synchronizing voltage is used to trigger the TV system to begin sweeping across a raster line.  Another voltage pulse is used to trigger the system to start rescanning the television field.  A series of electronic circuits move the scanning beams of the TV camera and monitor in synchronism.  The current, which flows down the scanning beam in the TV monitor, is related to that in the TV camera.  Consequently, the brightness of the image on the TV monitor is proportional to the amount of light falling on the corresponding position on the TV camera. TV camera and video signal (V)

TV camera and video signal (VI)  On most fluoroscopy units, the resolution of the system is governed by the number of lines of the television system.  Thus, it is possible to improve the high contrast resolution by increasing the number of television lines.  Some systems have 1,000 lines and prototype systems with 2,000 lines are being developed.

 Many modern fluoroscopy systems used CCD (charge coupled devices) TV cameras.  The front surface is a mosaic of detectors from which a signal is derived. TV camera and video signal (CCD)

Schematic structure of a charged couple device (CCD)

TV image sampling SYNCHRO 12 µs LIGHT INTENSITY SAMPLING 64 µs VIDEO SIGNAL (1 LINE) DIGITIZED SIGNAL ONE LINE IMAGE 512 x 512 PIXELS HIGHT 512

Digital radiography principle Clock Memory ADC I Iris t t ANALOGUE SIGNAL DIGITAL SIGNAL

33 Where to Get More Information  Physics of diagnostic radiology, Curry et al, Lea & Febiger, 1990  Imaging systems in medical diagnostics, Krestel ed., Siemens, 1990  The physics of diagnostic imaging, Dowsett et al, Chapman&Hall, 1998