ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging

Slides:



Advertisements
Similar presentations
Pulse Timing parameters & Weighting
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Imaging Sequences part I
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Parameters and Trade-offs
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
Principles of MRI: Image Formation
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
Magnetic Resonance Imaging
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Lecture 2 1 H Nuclear Magnetic Resonance. Gas Chromatograph of Molecular Hydrogen at –100 °C Thermoconductivity Detector 12.
MRI. Magnetic Resonance 1.Principle first observed in Used for spectroscopy and imaging 3.Imaging techniques are a form of tomography, where slices.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Nuclear Magnetic Resonance Spectrometry Chap 19. Absorption in CW Experiments Energy of precessing particle E = -μ z B o = -μ B o cos θ When an RF photon.
Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)
Basics of Magnetic Resonance Imaging
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Resonance condition. Pulse A coil of wire placed around the X axis will provide a magnetic field along the X axis when a direct current is passed through.
G Practical MRI 1 Magnetization, Relaxation, Bloch Equation, Basic Radiofrequency (RF) Pulse Shapes.
Radiofrequency Pulse Shapes and Functions
MRI Image Formation Karla Miller FMRIB Physics Group.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Physics of Magnetic Resonance Chapter 12
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
Medical Imaging Systems: MRI Image Formation
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
September, 2003BME 1450 Introduction to NMR 1 Nuclear Magnetic Resonance (NMR) is a phenomenon discovered about 60 years ago.
fMRI Methods Lecture2 – MRI Physics
Nuclear Magnetic Resonance I Magnetization properties Generation and detection of signals.
Allen W. Song, PhD Brain Imaging and Analysis Center Duke University MRI: Image Formation.
Magnetic Resonance Imaging – Basic Principles –
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
MRI. Vector Review x y z Vector Review (2) The Dot Product The Cross Product (a scalar) (a vector) (a scalar)
BME1450 Intro to MRI February 2002 The Basics The Details – Physics The Details – Imaging.
Lecture 3: The MR Signal Equation We have solved the Bloch equation and examined –Precession –T2 relaxation –T1 relaxation MR signal equation –Understand.
V.G.Wimalasena Principal School of Radiography
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
MRI: Contrast Mechanisms and Pulse Sequences
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
1 Nuclear Magnetic Resonance ANIMATED ILLUSTRATIONS MS Powerpoint Presentation Files Uses Animation Schemes as available in MS XP or MS 2003 versions A.
Charged particle. Moving charge = current Associated magnetic field - B.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Lecture 1: Magnetic Resonance
Lecture 3: The Bloch Equations Solved & the MR Signal Equation Solve the Bloch equation –Precession –T2 relaxation –T1 relaxation MR signal equation –Understand.
FMRI data acquisition.
MRI Physics in a Nutshell Christian Schwarzbauer
Where Mt is the magnetization at time = t, the time after the 90o pulse, Mmax is the maximum magnetization at full recovery. At a time = one T1, the signal.
Magnetic Resonance Imaging: Physical Principles
Assume object does not vary in y
Physics of Magnetic Resonance Imaging
Spatial Encoding: Sub mm from meter sized RF
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Basic MRI I Chapter 3 Notes.
10.5 Fourier Transform NMR Instrumentation
10.6 Fourier Transform Mass Spectrometry
Introduction to MR Robert Frost
(4)ELECTRONIC SUPPORT SYSTEM
T2 Relaxation Time T2 relaxation time is defined as the time needed to dephase up to 37% of the original value. T2 relaxation refers to the progressive.
Introduction to MR Robert Frost
Presentation transcript:

ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging Mark Mirotznik, Ph.D. Associate Professor The University of Delaware

Physics of Magnetic Resonance Summary Protons and electrons have a property called spin that results in them looking like tiny magnets. S N In the absence of an external magnetic field all the magnetic dipole are oriented randomly so we get zero net magnetic field when we add them all up. Random Orientation = No Net Magnetization

Physics of Magnetic Resonance Summary When we add a large external magnetic field we can get the protons to line up in 1 of 2 orientations (spin up or spin down) with a few more per million in one of the orientations than the other. This produces a net magnetization along the axis of the applied magnetic field. When we add a large external magnetic field we cause a torque on the already spinning proton that causes it to precess like a top around the applied magnetic field. The frequency is precesses , called its Larmor frequency., is determined from Larmor’s equation.

Physics of Magnetic Resonance Summary from Last Lecture The net magnetization vector is the sum of all of these little magnetic moments added together. This is what we measure. x y z x y z x y z z z x y x y Net Magnetization Vector

Physics of Magnetic Resonance Summary from Last Lecture However since all the spinning protons are precessing out of phase with each other this results in zero net magnetization in the transverse plane. = 0 This is bad news since is where are signal comes from! Somehow we need to get these guys spinning together!

Physics of Magnetic Resonance Summary from Last Lecture To get them to all spin together we add a RF field whose frequency is the same as the Larmor resonant frequency of the proton and is oriented in the xy or transverse plane. B1 RF Excitation time B1

Physics of Magnetic Resonance Summary from Last Lecture x y z B1 a a B1Dt Time of Application of RF Pulse Tip Angle Amplitude of RF Pulse

Physics of Magnetic Resonance Summary from Last Lecture x y z a = envelope of the RF signal In general

Physics of Magnetic Resonance Summary from Last Lecture To get the signal out we place a coil near the sample. A time-varying transverse magnetic field will produce a voltage on the coil that can be digitized and stored for processing. recall and

Relaxation Processes

Physics of Magnetic Resonance Relaxation After the RF field is removed over time the spin system will return back to it’s equilibrium state due to several relaxation processes.

Physics of Magnetic Resonance Relaxation After the RF field is removed over time the spin system will return back to it’s equilibrium state due to several relaxation processes. These are: Spin-Spin relaxation (also called the T2 relaxation): Due to random processes in which neighboring proton spins effect each other spin system will lose coherence and Mxy will decay. This is an irreversible process. Spin-Lattice relaxation (also called T1 relaxation): Due to another random process the Mz will begin to recover back to it’s original equilibrium state. Also irreversible. T2* relaxation: Due to inhomogenities in the external Bo field Mxy will decay much faster than T2. This is a reversible process.

T1 Relaxation

T2 Relaxation (FID)

T2 Decay RF RF

T2* Relaxation

T2* Decay: Dephasing due to field inhomogeneity z' y' Mxy = 0 x' T2* relaxation is dephasing of transverse magnetization too but it turns out to be reversible

Animation of T2* Dephasing

Spin- Echo

Spin Echo

Summary of Relaxation Processes

MRI Image An MRI image is determined by two things three intrinsic properties of the tissue. These are: T1, T2 and Pd. (two relaxation time constants and the density of protons) the details of the external magnetic fields (Bo, B1 and the gradient magnetics (have not talked about these yet)). How they are configured and how we turn them on and off (pulse sequence) effects what the image looks like. By varying the pulse sequence we can control which of the intrinsic properties to emphasize in the image.

Tissue Contrast

TR TE TE 0.5TE 0.5TE 0.5TE 0.5TE 180 degree RF pulses 90 degree White matter T1=813 ms T2=101 ms

TR TE TE CASE I: TR>>T1 , TE<T2 What do we measure? 180 degree RF pulses CASE I: TR>>T1 , TE<T2 What do we measure? 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

CASE I: TR>>T1 , TE<T2 Short TE means that the signal has not decayed much due to T2relaxation. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? TR 180 degree RF pulses 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

CASE I: TR>>T1 , TE<T2 Short TE means that the signal has not decayed much due to T2relaxation. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? PD Weighted imaging! TR 180 degree RF pulses 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

CASE II: TR>>T1 , TE~T2 180 degree RF pulses CASE II: TR>>T1 , TE~T2 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

CASE II: TR>>T1 , TE~T2 TE on the order of T2means that the signal is proportional to the T2relaxation constant. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? 180 degree RF pulses 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

CASE II: TR>>T1 , TE~T2 TE on the order of T2means that the signal is proportional to the T2relaxation constant. Long TR means that by the next pulse the system is back at equilibrium (Mz due to T1 relaxation has fully recovered) So what are we measuring? T2 Weighted imaging! TR 180 degree RF pulses 90 degree RF pulses TE TE White matter T1=813 ms T2=101 ms

TR TE CASE III: TR~T1 , TE<T2 What do we measure? 180 degree RF pulses 90 degree RF pulses TE

TR TE CASE III: TR~T1 , TE<T2 TE is shorter than T2means that the signal is not heavily weighted on the T2relaxation constant. TR on the order of T1 means that by the next pulse the system is not back at equilibrium (Mz due to T1 relaxation has not fully recovered) So what are we measuring? TR 180 degree RF pulses 90 degree RF pulses TE

TR TE T1 Weighted imaging! CASE III: TR~T1 , TE<T2 TE is shorter than T2means that the signal is not heavily weighted on the T2relaxation constant. TR on the order of T1 means that by the next pulse the system is not back at equilibrium (Mz due to T1 relaxation has not fully recovered) So what are we measuring? T1 Weighted imaging! TR 180 degree RF pulses 90 degree RF pulses TE

Tissue Contrast Summary TE TR PD weighted T1weighted T2 weighted TE<T2 (short TE) TR>>T1 (long TR) T1weighted TR~T1 (short TR) T2 weighted TE~T2 (long TE)

Bloch Equations

Full Bloch equation including relaxation precession, RF excitation transverse magnetization longitudinal magnetization includes Bo and B1

Example: Solve for the transverse components of M after a 90 degree pulse.

Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero

Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero Initial conditions for 90 degree pulse:

Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero Solutions

Example: Solve for the transverse components of M after a 90 degree pulse. After 90 degree pulse the RF field is shut down and only Bo is non-zero Solutions

Example: Solve for the transverse components of M after an arbitrary flip angle (a) After an arbitrary RF pulse the RF field is shut down and only Bo is non-zero

Example: Solve for the transverse components of M after an arbitrary flip angle (a) After an arbitrary RF pulse the RF field is shut down and only Bo is non-zero Initial conditions for 90 degree pulse:

Example: Solve for the transverse components of M after an arbitrary flip angle (a) Solutions

Solve full Bloch equation with only B=Bo Solution for transverse components Mx and My Where a is the flip angle after RF excitation

Signal Detection

Signal Detection via RF coil

Signal Detection via RF coil Transverse magnetization at t=0. Coils oriented as shown above will only respond to changes in the transverse magnetic field (this is what we want) Assuming the magnetic fields are homogenous the signal will be a weighted integration of all the protons within the coil. The waiting will be based on the total magnetization at location x,y,z at the start of the pulse (Mxy(x,y,z,0)) and the tissue decay time T2(x,y,z) This is not an image!!

Signal Detection via RF coil After demodulation:

Creating an Image

Creating an Image To create an image using NMR we need to figure out a way to encode the proton spins spatially in three dimensions. But how?

Frequency and Phase Are Our Friends in MR Imaging w q = wt The spatial information of the proton pools contributing MR signal is determined by the spatial frequency and phase of their magnetization.

Gradient Coils X gradient Y gradient Z gradient x y z Gradient coils generate spatially varying magnetic field so that spins at different location precess at frequencies unique to their location, allowing us to reconstruct 2D or 3D images.

Gradient Coils Purpose: Spatially alter magnitude of B0 (not direction) Sounds generated during imaging due to mechanical stress within gradient coils.

Vector Notation

Larmor frequency within a gradient field

Slice Selection

Slice Selection Gradient BG Coil 1 Coil 2

Helmholtz Coils

Z-Gradient Fields By adding a z-gradient field we cause a variation in the resonant frequency from head to toe.

Example A sample is put inside a 1.5T magnet. A z-gradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter?

Example A sample is put inside a 1.5T magnet. A z-gradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter?

Example A sample is put inside a 1.5T magnet. A z-gradient of 3 gauss/cm is applied. If we wish to image a 2 ft in length section of a person what is the range of resonant frequencies we will encounter?

A Field Gradient Makes the Larmor Frequency Depend upon Position B0 63,480,000 Hz 64,260,000 Hz Z Gradient in Z B(Z) = B + G * Z o Z = n(z) g B(z)

Slice Selection (-) 62 MHz 63 MHz 64 MHz G 65 MHz 66 MHz (+)

How do we determine the slice width and center? Slice Selection How do we determine the slice width and center? z After z selection gradient and excitation z-gradient Bo (slice center) (slice width) x

Determining slice thickness Resonance frequency range as the result of slice-selective gradient:

Changing slice thickness There are two ways to do this: Change the slope of the slice selection gradient Change the bandwidth of the RF excitation pulse Both are used in practice, with (a) being more popular

Example Suppose we wish to have a slice thickness of 2 mm and we are using a z-gradient of 1.0 G/cm ? What range of RF frequencies should we use?

Example Suppose we wish to have a slice thickness of 2 mm and we are using a z-gradient of 1 G/cm ? What range of RF frequencies should we use?

Selecting different slices

Selecting different slices In theory, there are two ways to select different slices: Change the position of the zero point of the slice selection gradient with respect to isocenter (b) Change the center frequency of the RF to correspond to a resonance frequency at the desired slice Option (b) is usually used as it is not easy to change the isocenter of a given gradient coil.

RF Excitation (RF Pulse) Fo FT t Fo Fo+1/ t Time Frequency Fo Fo FT DF= 1/ t t

RF Excitation (RF Pulse) Fo FT A Dn t n1 n2

RF Excitation: Flip angle Fo FT A Dn t n1 n2 Envelope of the pulse

RF Excitation: Flip angle Fo FT A Dn t n1 n2

RF Excitation: Flip angle Fo FT A Dn t n1 n2

RF Excitation: Flip angle (truncated sinc)

A potential problem Wait a minute! Dr. M I remember you telling us that if the magnetic field varied from place to place (inhomogeneous) then we would get rapid dephasing of spins. Since some spins are spinning faster than others they quickly get out of phase. That was the whole reason behind the spin echo stuff! Won’t that happen again?

Why yes it will! It is called gradient dephasing Good question! Wait a minute! Dr. M I remember you telling us that if the magnetic field varied from place to place (inhomogeneous) then we would get rapid dephasing of spins. Since some spins are spinning faster than others they quickly get out of phase. That was the whole reason behind the spin echo stuff! Won’t that happen again? Spinning slow Spinning fast Why yes it will! It is called gradient dephasing Good question!

Any ideas on how to get around this? Why yes it will! It is called gradient dephasing. It will quickly kill our signal much faster than T2 or even T2* Spinning slow Spinning fast Any ideas on how to get around this?

Localization in xy plane

Lets Start with a Simple Flat Person (only xz plane) z Bo x

Lets Start with a Simple Flat Person (only xz plane) z After z selection gradient and excitation z-gradient Bo x

Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis Recall that the signal we measure is given by: Now we have selected only a single slice in z (z=zo) and we have no y dependence (flat person) After demodulation (envelope detection)

Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) Let This is what we want to image (called the effective proton density)

Lets Start with a Simple Flat Person (frequency encoding using x-gradient) z After z selection gradient and excitation Bo x

Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis Now lets apply a gradient in the x direction (Gx) After demodulation (envelope detection) What does this look like?

Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) Let The received signal is related to the Fourier transform (THIS IS THE KEY!)

Lets Start with a Simple Flat Person Frequency Encoding Mathematical Analysis After demodulation (envelope detection) Let We can now find our image as a function of x by taking an inverse Fourier Transform

A Simple Example of Spatial Encoding with Frequency Encoding Constant Magnetic Field Varying Magnetic Field w/o encoding w/ encoding

A Simple Example of Spatial Encoding with Frequency Encoding Decomposition

Decays faster than T2*

Extend this to a full 3D person

Extend this to a full 3D person After slice selection we need to image in xy plane y x

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply a gradient in the x direction (Gx) After demodulation (envelope detection) Effective proton density

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis After demodulation (envelope detection) Let

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Let Corresponds to a single line or trajectory in the uv plane

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply gradients in both the x direction (Gx) and y direction (Gy) After demodulation (envelope detection) Let Effective proton density

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Now lets apply gradients in both the x direction (Gx) and y direction (Gy) Let

Spatial Encoding in xy plane Frequency Encoding Mathematical Analysis Let Polar Scanning

Gradient Echo (A brief detour)

Ts/2

Spatial Encoding in xy plane Gradient Echo Mathematical Analysis Spins will dephase very quickly (quicker than T2*) due to the gradient fields. After negative x-gradient

Spatial Encoding in xy plane Gradient Echo Mathematical Analysis After negative x-gradient Now impose positive x-gradient for Ts

Spatial Encoding in xy plane Gradient Echo Mathematical Analysis Now impose positive x-gradient for Ts

Phase Encoding

Pulse Repetition

Image Contrast

Image Quality

Field of View and Resolution in MRI Fourier Plane Spatial Domain Du Dx Dv Vcoverage FOVy Dy Ucoverage FOVx

Nyquist Sampling Theorem: Review Assume we have a continuous signal with maximum frequency of fmax To avoid aliasing we must sample the signal at a sampling frequency of fs>=2 fmax The sampling interval T=1/ fs fmax<=1/(2T)

Sampling in MRI Slice selection direction: sampling in z-direction Slice thickness (Dz) controlled by RF excitation bandwidth (Dn) To avoid aliasing Where fmax,z is the highest spatial frequency in along the z-axis

Sampling in MRI Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Rectilinear Scan Du depends on sampling interval T during readout (ADC) Du depends on sampling interval during this time

Sampling in MRI Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Rectilinear Scan Dv depends on spacing between phase encoding Dv depends on the integrated phase shift here

Sampling in MRI Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Polar Scan Angle scan depends on steps in Gy/Gx Angle scan depends on steps in Gy/Gx

Sampling in MRI Within each slice: sampling in xy plan We sample in the Fourier domain (u,v) (called k-space in MRI literature, kx=u, ky=v) Polar Scan Rho spacing depends on sampling interval T during readout r spacing depends on sampling interval during this time

Dv

X-gradient relates dimension x with Larmor freq n by To avoid aliasing only frequency given below are measured

X-gradient relates dimension x with Larmor freq n by To avoid aliasing only frequency given below are measured Field of view in the x-direction (FOVx) is thus given by

Dependant on the phase encoding gradient Gy Dependant on the phase encoding gradient Gy. The amount of phase change is given by Field of view in y

While FOV is limited by the sampling interval in the UV plane (Fourier plane) the resolution is limited by the total extent of the UV plane being sampled. If we ignore high spatial frequency content we will have lower resolution (blur our image). Since MRI scans cover only a finite area of the Fourier space we can expect a finite resolution. Fourier space coverage in MRI

Fourier space coverage in MRI Outside of this range we assume the contributions to be zero. This is equivalent to passing the actual image through a low-pass filter in the uv plane whose transfer function is given by In the spatial domain this is then given by the point spread function (PSF)

In the spatial domain this is then given by the PSF The full width half max (FWHM) resolution is given by the width of the sinc function’s main lobe Increasing the U,V (coverage area in Fourier space) reduces blurring.

Field of View and Resolution in MRI Spatial Domain Fourier Plane Du Dx Dv Vcoverage FOVy Dy Ucoverage FOVx