Tabuk University Hematology – 2, MLT 307

Slides:



Advertisements
Similar presentations
What happens in the aperture?
Advertisements

HEMATOLOGY WHAT IT IS : Study & measurement of individual elements of Blood. WHAT IT’S COMPOSED OF. SHOW SLIDES FROM PERIPHERAL BLOOD TUTOR CD OR USE PLATE.
Analytic Phase of Laboratory Testing
MLAB Hematology Keri Brophy-Martinez
Complete Blood Count ( CBC). Complete Blood Count ( CBC)
MLAB 1415: Hematology Micros 60.
Automated Hematology Cell Counters
Buffy coat (leukocytes and platelets) - <1%
Automated Hematology Cell Counters
Laboratory Quality Control
Quality Assurance.
Automation in hematology Until Slide 23
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Quality Assurance in Haematology
Quality Assurance in the clinical laboratory
-Automation blood count -Red and White blood count and differential count (Manual blood count) Prepared by: Mona alharbi Experiment:
Quality Assurance / Quality Control
Chemometrics Method comparison
Unit #7 - Basic Quality Control for the Clinical Laboratory
Physiology Presentation Roll No.# 218, 224, 230, 236, 242, 248 Muhammad Mohsin Ali Dynamo.
-Automation blood count -Red and White blood count and differential count (Manual blood count) Experiment:
Objectives 1. Methods used to measure the different hematological values, and compare it with the normal values. 2. The calculation of RBC indices, their.
Quality Assurance.
Quality Assessment 2 Quality Control.
A compact 5-diff solution
Agile - affordable - accurate. Hematology, is the part of internal medicine, that is concerned with the study of blood, the blood-forming organs, and.
Clinical Haematology & Cell Counting
Quality WHAT IS QUALITY
CLS 541-Advanced Medical Haematology M. sc. Program Lab-2-
Sarmishtha Ghosh Physiology
Automation in haematology
Automated CBC Parameters
Pictures/Diagrams Copyright © University of Nebraska
Automated CBC Parameters
Complete Blood Count.
Quality Control Lecture 5
Laboratory QA/QC An Overview.
Clinical pathology: Complete Blood count
An Overview. Definitions (1)  Quality Control - QC refers to the measures that must be included during each assay run to verify that the test is working.
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
Red Cell Indices Nada Mohamed Ahmed, MD, MT (ASCP)i.
Course title: Hematology (1) Course code : MLHE-201 Supervisor :Prof.Dr Magda Sultan Date : Outcome : The student will be able to use the automated.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
L ABORATORY Q UALITY C ONTROL. INTRODUCTION _A major role of the clinical laboratory is the measurement of substances in body fluids or tissues for the.
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
Preparing a Blood Smear. Samples for Hematology Capillary blood and venous blood can be used for hematology tests. Capillary blood is beneficial if a.
HAEMATOLOGY LINE A.C.KAROBIA
Flags, Codes and Messages on the COULTER® LH 750/755
Blood Testing. Learning Objective  To identify the types of tests performed on blood  To explain why these tests are useful  To discover the normal.
Diagnostic clinical chemistry
Quality Assurance.
MLAB Hematology Keri Brophy-Martinez
Quality is a Lousy Idea-
Automatic Electronic Blood Cell Counting
Medical Laboratory Instrumentation
MLAB Hematology Keri Brophy-Martinez
Unit #6 - Basic Quality Control for the Clinical Laboratory
Automated Hematology Cell Counters
Quality Assurance in the clinical laboratory
Micros 60 & Quality Control
Laboratory Quality Control
Practical clinical chemistry
Quality is a Lousy Idea-
Supplementary Figure S7
Chapter 41 Hematology.
Introduction To Medical Technology
Quality Control Lecture 3
▪Internal quality control:
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

Tabuk University Hematology – 2, MLT 307 Faculty of Applied Medical Sciences Department Of Medical Lab. Technology 3rd Year – Level 5 – AY 1434-1435 Hematology – 2, MLT 307

Quality Assurance and Automation in Hematology By/ Dr WalidZAMMITI; Phd; M.Sc; MLT

Objectives Describe the electrical impedance and light scatter principles for performing cell counts. Utilize quality control procedures to determine if patient results are acceptable. Explain histograms and their indications. Concentrate on some parameters and indices. Identify the major components of a quality assurance program. Be able to distinguish between quality assurance & quality control. Define and give examples of each of the following terms: Accuracy-Calibration-Control-Standard-Precision. Understand the concepts of internal & external control.

Quality system begins and ends with the patient

Quality Assurance vs. Quality Control An overall management plan to guarantee the integrity of data (The “system”) A series of analytical measurements used to assess the quality of the analytical data (The “tools”)

Quality Assurance in Hematology QA includes all aspects of laboratory activities that affects the results produced, from the choice of methods, to the education of personnel, to the handling of specimens and reporting results. The real purpose of QA activities is to determine how correct or incorrect the results emanating from the lab are, and to allow those managing the lab to determine whether or not the lab is fulfilling its functions satisfactorily.

QA in Haematology Laboratory QA in haematology lab is intended to ensure the reliability of the lab tests. The objective is to achieve precision and accuracy 4 components of QA programme : 1 ) Internal Quality Control ( IQC ) 2 ) External Quality Control ( EQC ) 3 ) Standardization 4 ) Proficiency surveillance

Accuracy vs. Precision Accuracy Precision How well a series of measurements agree with each other: Is the closeness of agreement between independent test results obtained under stipulated conditions. Accuracy How well a easurement agrees with an accepted value: is the closeness of the agreement between the result of a measurement and a true value of the measurand.

Accuracy vs. Precision

Internal Quality Control Internal Quality Control Internal quality control is set up within a laboratory to monitor and ensure the reliability of test results from that laboratory. The primary tool for internal quality control is called a control. A control is a specimen with a predetermined range of result values, called control values, that is processed in the same manner as a patient sample. Control samples are processed with each series or run of patient samples. If the result of a test on a control sample is different from its known value, this indicates a problem in the equipment or the methods being used.

External Quality Control ( EQC ) is the objective evaluation by an outside agency of the performance by a number of laboratories on material which is supplied specially for the purpose is usually organized on a national or regional basis analysis of performance is retrospective the objective is to achieve comparability with results of other labs.

Standardization Refers to both materials and methods. A material standard or reference preparation is used to calibrate analytic instruments and to assign a quantitative value to calibrators. A reference method is an exactly defined technique which provides sufficiently accurate and precise data for it to be used to assess the validity of other methods

Proficiency surveillance Implies critical supervision of all aspects of laboratory tests: collection, labelling, delivery, storage of specimens before the tests are preformed and of reading and reporting of results. Also includes maintenance and control of equipment and apparatus.

Control  What is a Control? QC programs require the same sample to be tested every day testing is done. This type of sample is called a control. Controls, which are often purchased from manufacturers, use a human base to ensure the analyses being tested parallel human ranges. Manufacturers pool together many human blood samples to create the large volume needed for a lot number of control

Tools for Validation of QC results Control Charts: A Control Chart depend on the use of IQC specimens and is developed in the following manner +3 sd +2 sd +1 sd Target value -1 sd -2 sd -3 sd Assay Run

Control Charts Samples of the control specimen are included in every batch of patients’ specimens and the results checked on a control chart Check precision: it is not necessary to know the exact value of the control specimen Value has been determined reliably by a reference method, the same material can be used to check accuracy or to calibrate an instrument Controls with high, low and normal values should be used Advisable to use at least one control sample per batch even if the batch is very small The results obtained with the control samples can be plotted on a chart

How to calculate SD 1. Get the Mean. 2. Get the deviations. (each value minus the mean) 3. Square these. 4. Add the squares. 5. Divide by total numbers less one. 6. Square root of result is Standard Deviation

Types Of Errors An error which varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions.

Systematic vs.Random Errors Systematic Error Avoidable error due to controllable variables in a measurement. Random Errors Unavoidable errors that are always present in any measurement. Impossible to eliminate

Random Error Random errors create a characteristic spread of results for any test method and cannot be accounted for by applying corrections. Random errors are difficult to eliminate but repetition reduces the influences of random errors. Examples of random errors include errors in pipetting and changes in incubation period. Random errors can be minimized by training, supervision and adherence to standard operating procedures.

Random Errors

Systematic Error An error which, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions, or varies according to a definite law when conditions change. Systematic errors create a characteristic bias in the test results and can be accounted for by applying a correction. Systematic errors may be induced by factors such as variations in incubation temperature, blockage of plate washer, change in the reagent batch or modifications in testing method.

Systematic Errors

Automation in Haematology

Automated techniques of blood counting Semi-automated instruments Require some steps, as dilution of blood samples Often measure only a small number of variables Fully automated instruments Require only that an appropriate blood sample is presented to the instrument. They can measure 8-20 variables including some new parameters which do not have any equivalent in manual methods.

The accuracy of automated counters is less impressive than their precision. In general automated differential counters are favourable to the manual in 2 conditions Exam of normal blood samples Flagging of abnormal samples

CBC : Complete Blood Count The complete blood count is performed as an automated procedure. A sample of blood is placed in an analyzer and the cells are sorted by a laser according to size, granularity, and shape.

Parameters : WBC= Total white blood cells RBC= Red blood cell count HGB= Hemoglobin concentration HCT= Hematocrit (PCV) MCV= Mean Cell Volume MCH= Mean Cell Hemoglobin MCHC= Mean Cell Hemoglobin Concentration PLT= Platelets count NEUT%= Percentage Neutrophil count LYMPH%= Percentage Lymphocyte count MONO%= Percentage Monocyte count EO%= Percentage Eosinophil count BASO%= Percentage Basophil count NEUT#= Absolute Neutrophil Count LYMPH#= Absolute Lymphocyte Count MONO#= Absolute Monocyte Count EO#= Absolute Eosinophil Count BASO#= Absolute Basophil Count RDW-SD= Red cell Distribution Width – Standard Deviation RDW-CV= Red cell Distribution Width – Coefficient Variation MPV= Mean Platelet Volume PDW = Platelet Distribution Width Some times other parameters are included; e.g.: Reticulocytes.

Examples of Haematology analysers AcT 5diff (Beckman Coulter ) SE 9000, KX21, XE 2100 (Sysmex) Advia 60 (Bayer) Cell-Dyn 3500 ( Abott)

When to Calibrate You should calibrate your instrument: At installation. After the replacement of any component that involves dilution characteristics or the primary measurements (such as the apertures). When advised to do so by your service representative.

Flagging Condition flags Describes cell population normal abnormal WBC Suspect flags Blasts Immature Grans/Bands 1 Immature Grans/Bands 2 Variant lymphocytes Review Slide

More Flagging RBC Suspect flags NRBCs Macrocytic RBCs Dimorphic RBC population Micro RBCs/RBC fragments RBC agglutination Definitive Flagging Based on predetermined lab limits Provide information for review

Histograms RBC, PLT, and WBC plotted on histogram X-Axis Cell size in femtoliters (fL) Y-Axis # of cells

RBC Histogram As A Quality Control Tool INDICATOR PROBABLE CAUSE COMMENT Left of curve does not touch baseline Schistocytes and extremely small red cells Review smear CBC and Platelet histogram Bimodal peak Transfused cells, therapeutic response Review Smear Right portion of curve extended Red cell autoagglutination Review CBC & Smear Left shift of curve Microcytes Review smear & CBC Right shift of curve Macrocytes

Platelet Histogram As A Quality Control Tool INDICATOR PROBABLE CAUSE COMMENT Peak or spike at left end of histogram (2-8 fl) Cytoplasmic fragments Review smear Spike towards right end of histogram Schistocytes, microcytes, giant platelets Review smear + CBC ( MCV &  RDW) ( MPV &  PDW) Bimodal peak

Histograms - WBCs WBC: Distribution with three individual peaks and valleys at specific regions representing the lymphocytes, monocytes, and granulocytes.

WBC Histogram As A Quality Control Tool INDICATOR PROBABLE CAUSE COMMENT Trail extending downward at extreme left, or lymph peak not starting at baseline NRBC, Plt clumping, unlysed RBC, cryoproteins, parasites Review smear and correct WBC for NRBC Peak to the left of lymph peak or widening of lymph peak towards left NRBC Review smear & correct WBC for NRBC Widening of lymph peak to right Atypical lymphs, blasts, plasma cells, hairy cells, eosinophilia, basophilia Review smear Wider mono peak Monocytosis, plasma cells, eosinophilia, basophilia, blasts

WBC Histogram As A Quality Control Tool INDICATOR PROBABLE CAUSE COMMENT Elevation of left portion of granulocyte Left Shift Review smear Elevation of right portion of granulocyte peak Neutrophilia

RDW-SD RDW is an actual measurement of the width of the erythrocyte distribution curve. It is a measurement of Anisocytosis. May increase before MCV becomes abnormal Reference values: female: 36.4 – 46.3 fL male: 35.1 – 43.9 fL It is increased in many types of anemias to indicate the variation in red cell sizes.

RDW-CV The coefficient of variation (CV) is defined as the % ratio of the standard deviation (x), to the mean (µ) Cv = x/µ Sometimes known as relative standard deviation. Reference values: female: 11.7 – 14.4% male: 11.6 – 14.4 %

MCV = MEAN Cell VOLUME M.C.V. = Hematocrit% X 10 RBC in millions/µl Normal values: Men & women 82 – 97 fl (femtoliters) = cubic microns Increased : Macrocytes Decreased : Microcytes

MCH = Mean Cell Hemoglobin M.C.V. = Hemoglobin g/dl X 10 RBC in millions/µl Normal values: Men & women 27 – 32 pg (pico grams) Increased : Hyperchromic Decreased : Hypochromic

MCHC = Mean Cell Hb Concentration M.C.V. = Hemoglobin g/dl X 100 Hematocrit% Normal values: Men & women 30 – 34 g/dl Increased : Hyperchromic Decreased : Hypochromic

Other Hematology Machines Coagulometers : - Used in Hemostasis studies, and the Endpoint Detection depends on Mechanical, Optical (Photo- optical , Nephelometric , Chromogenic or Immunologic), Electrochemical principles. ESR machines : in 30 minutes. Leucocytes automated Differential Counters : Using cytochemical or image recognition methods.