1.1 Range of molar absorptivity

Slides:



Advertisements
Similar presentations
SPECTROSCOPY.
Advertisements

7. Beer’s Law and It’s Implications for Instrument Construction.
An introduction to Ultraviolet/Visible Absorption Spectroscopy
Chapter 14 APPLICATION OF ULTRAVIOLET/VISIBLE MOLECULAR ABSORPTION SPECTROMETRY Absorption measurements based upon ultraviolet and visible radiation find.
8-1 Ultraviolet-Visible Spectroscopy Introduction to UV-Visible §Absorption spectroscopy from 160 nm to 780 nm §Measurement of transmittance àConversion.
Molecular Spectroscopy Visible and Ultraviolet Spectroscopy
What is Spectroscopy? The study of molecular structure and dynamics through the absorption, emission and scattering of light.
Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh
Chapter 14 Applications of Ultraviolet-Visible Molecular Absorption Spectrometry.
Chapter 19 Spectrophotometry : Instruments & Application.
UV-vis Absorption (Extinction) Spectroscopy Single-Beam or Double-Beam Fixed or Dispersive Common: Source – Tungsten Halogen Lamp ( nm) Sample.
Spectrophotometric Analysis
Outline Final Comments on Titrations/Equilibria Titration of Base with a strong acid End-point detection Choice of indicators Titration Curve method Start.
Introduction to Spectrophotometry
Essential Components of a UV-vis Spectrophotometer Monochromator Signal Processor Display Source Sample Transducer.
The Electromagnetic Spectrum
Absorbance of Electromagnetic Radiation
UV-vis Absorption (Extinction) Spectroscopy Single-Beam or Double-Beam Fixed or Dispersive Common: Source – Tungsten Halogen Lamp ( nm) Sample.
UV/Visible Spectroscopy
Spectrophotometry: An Analytical Tool. PGCC CHM 103 Sinex IoIo I Cell with Pathlength, b, containing solution light source detector blank where I o =
Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Absorption measurements based upon ultraviolet and visible radiation.
Molecular Spectrometry (UV and Visible) Part 1: Absorption.
Application of UV/Vis Spectroscopy
AN INTRODUCTION TO ULTRAVIOLET/VISIBLE ABSORPTION SPECTROSCOPY
UV–Vis Molecular Absorption Spectrometry Chapter 13.
Ultraviolet – Visible Spectrophotometer. What is UV – Vis spectroscopy? Ultraviolet – visible spectroscopy is a measure of the absorbance of radiation.
Introduction to Instrumental Analysis - Spectrophotometry
Chapter 26 Molecular Absorption Spectrometry Molecular spectroscopic methods are among the most widely used of all instrumental analytical methods. Molecular.
1 Spectroscopic ANALYSIS Part 5 – Spectroscopic Analysis using UV-Visible Absorption Chulalongkorn University, Bangkok, Thailand January 2012 Dr Ron Beckett.
10/11/ ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY.
Spectrophotometry: An Analytical Tool
Chapter 13 An Introduction to Ultraviolet- Visible Molecular Absorption Spectrometry.
Microplate reader spectrophotometer. The Beer-Lambert Law A=abc Now let us look at the Beer-Lambert law and explore it's significance. A is absorbance.
Spectrophotometry.
Chapter 6 An Introduction to Spectrometric Methods Spectrometric methods are a large group of analytical methods that are based on atomic and molecular.
Introduction to Spectroscopic Methods of Analysis (part 2)
Lecture 6c. Introduction Electromagnetic spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy.
1 UV-Vis Absorption Spectroscopy Lecture Measurement of Transmittance and Absorbance: The power of the beam transmitted by the analyte solution.
Asa Arjoon U6 Chemistry Presentation on Visible and Ultraviolet Spectroscopy Name : Form : Subject :
Introduction to Spectrochemical Methods
UV-VIS Spectroscopy Dr. AKM Shafiqul Islam.
Spectrophotometer.
Outline Start Chapter 18 Spectroscopy and Quantitative Analysis.
1.1 What’s electromagnetic radiation
UV SPECTROSCOPY Absorption spectra.
Summary: (Last lecture) Absorption spectroscopy definition electromagnetic spectroscopy matter absorption spectroscopy fundamental terms (transmittance,
Chapter 26 Molecular Absorption Spectrometry. 26 A Ultraviolet and visible molecular absorption spectroscopy Ultraviolet and visible radiation absorption.
Spectrophotometry at a Glance
Instrumental Analysis
ULTRAVIOLET AND VISIBLE SPECTROSCOPY
11 Instrumental Analysis Tutorial Use mathematical formulae to calculate absorbance, transmittance of a sample and wave parameters. Determine factors.
1 Instrumental Analysis Tutorial 2. 2 Objectives By the end of this session the student should be able to: 1.Describe the grating principle of work. 2.Describe.
Introduction to Spectrophotometry
UV/Visible Spectroscopy
Chem. 133 – 3/14 Lecture.
Spectroscopy Techniques
Chapter 13 Chapter 13 Chapter 13 Chapter 13 Chapter 13 Chapter 13
Spectrophotometer Dr . S. Jayakumar.
Spectrophotometry: An Analytical Tool
1901 Application of Spectrophotometry
Measurements involving light –A Basic Principles and Instrumentation
Dnyanasadhana College, Thane. Department of Chemistry T. Y. B. Sc
UV SPECTROSCOPY Absorption spectra.
UV-VIS Molecular Spectroscopy
Analytical methods Prepared By Dr. Biswajit Saha.
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
A SEMINAR ON Ultraviolet-Visible Spectroscopy
Spectrophotometric Analysis
Presentation transcript:

1.1 Range of molar absorptivity Chapters 13&14 Ultraviolet-Visible Molecular Absorption Spectrometry 1 Absorbing Species 1.1 Range of molar absorptivity Electronic excitation of outer valence (i.e. bonding) electron How probable for this electronic excitation? (allowed transition, or forbidden) judged by the range of   = 104 -105 L mol-1 cm-1, strong absorption  < 103 L mol-1 cm-1, low intensity

1.2 Which electron get excited? 1.2.1 Organic molecules ,  (bonding) and n (non-bonding) orbitals *, * (anti-bonding) orbitals   * E large ( < 150 nm, out of range)  = 10 -10,000 Lmol-1cm-1 n  * E smaller ( = 150 - 250 nm)  = 200-2000 Lmol-1cm-1   * n  * E smallest ( = 200 - 700 nm)  = 10-10,000 Lmol-1cm-1 Ideal for UV-Vis spectrometry of organic chromophore

1.2.2 Inorganic molecules Most transition metal ions are colored (absorption in Vis) due to d  d electronic transition Fig. 14-3 (p.370) Fig. 14-4 (p.371)

1.2.3 Charge-transfer absorption A: electron donor, metal ions D: electron acceptor, ligand  > 10,000 Fig. 14-5 (p.371)

2 Beer’s Law 2.1 Approximation of T and A A = -log T = log (P0/P) = ··b·c : molar absorptivity at one particular wavelength (L·mol-1cm-1) b: path length of absorption (cm) c: molar concentration (mol·L-1) Fig. 6-25 (p.158) Light loss due to reflection (17.3%), scattering, … Fig. 13-1 (p.337)

= 1bc1 + 2bc2 A2 = 1,2·b·c1 + 2,2·b·c2 2.2 Application of Beer’s law to mixtures Absorbance is additive Atotal = A1 + A2 + … = 1bc1 + 2bc2 For a 2-component mixture, we measure the absorption at two different wavelength, respectively A1 = 1,1·b·c1 + 2,1·b·c2 A2 = 1,2·b·c1 + 2,2·b·c2

A = -log T = log (P0/P) = ··b·c 2.3 Limitations of Beer’s law 2.3.1 Real deviations At low concentration A = -log T = log (P0/P) = ··b·c At c > 0.01 M solute-solute interaction, hydrogen-bond, … can alter the electronic absorption at a given wavelength  dilute the solution

2.3.2 Chemical effects analyte associates, dissociates or reacts with a solvent to give molecule with different  Example: acid-base equilibrium of an indicator 430 570 HIn 6.30 x102 7.12x103 (measured in HCl solution) In- 2.06 x 104 9.61 x102 (measured in NaOH solution) What’s the absorbance of unbuffered solution at c = 2 x 10-5M?

Fig. 13-3 (p.340) =[HIn] + [In-]

2.3.3 Instrumental deviations due to polychromatic radiation Beer’s law applies for monochromatic absorption only. If a band of radiation consisting of two wavelength ’, and ” Assuming Beer’s law applies to each wavelength Fig. 7-11 (p.176)

2.3.3 Instrumental deviations due to polychromatic radiation ′ ″ Non-linear calibration curve Fig. 13-4 (p.341)

How to avoid : Select a wavelength band near its maximum absorption where the absorptivity changes little with wavelength Fig. 13-5 (p.341)

2.3.4 Other physical effects stray light – the scattering, reflection radiation from the instrument, outside the nominal wavelength band chosen for measurement mismatched cell for the sample and the blank

3 Noise of Spectrophotometric analyses 3.1 Standard deviation of c

3.2 Sources of instrumental noise Case I Limited readout resolution (31/2-digit displays  0.1% uncertainty from 0%T -100% T) Thermal noise in thermal detector, etc (particularly for IR and neat IR spectrophotometer) Case II Shot noise in photon detector (random emission of photon from the light source or random emission of electrons from the cathode in a detector) Case III Flicker noise, Fail to position sample and blank cells reproducibly in replicate measurements (as a result, different sections of cell window are exposed to radiation, and reflection and scattering losses change)

3.2 Sources of instrumental noise Fig. 13-3 (p.344)

4 Instrumentation 4.1 Designs a. Single beam b. Double-beam-in-space c. Double-beam-in-time Advantage of double beam configuration Compensate for fluctuation in the radiant output, drift in transducer, etc. Continuous recording of spectra Fig. 13-13 (p.352)

Shimadzu UV-2450 Spectrophotometer

Wavelength Range 190 to 900nm (performance guaranteed range). Extendable to 1,100nm through the use of the optional photomultiplier. (The measurable range maybe restricted in the shorter wavelength side depending on the type of photomultipler used.) Monochromator System UV-2450: Single monochromator with a high-performance blazed holographic grating in the aberration corrected Czerny-Turner mounting. Resolution 0.1nm Spectral Bandwidth 0.1, 0.2, 0.5, 1, 2 and 5nm Wavelength Accuracy ±0.3nm Wavelength Repeatability ±0.01nm Wavelength Scanning Speed FAST, MEDIUM, SLOW, and SUPER SLOW Light Source 50W halogen lamp (2,000 hours of life) and D2lamp (500 hours of life) Light Source lamp switching Selectable between 282nm and 393nm Stray Light UV-2450: Less than 0.015% at 220nm and 340nm Detector Photomultiplier R-928 Photometric System Double beam, direct ratio system with dynode feedback Photometric Mode Absorbance (Abs.), transmittance (%), reflectance (%) and energy (E). Photometric Range Abosrbance: -4~5 Abs. (0.001 Abs. increments) Transmittance: 0~999.9% (0.01 increments) Reflectance: 0~999.9% (0.01% increments) Photometric Accuracy ±0.002Abs(0~0.5Abs), ±0.004Abs(0.5~1Abs),±0.3T (0~100%T) (all determined with NIST 930D standard filter) Photometric Repeatability 0.001Abs (0~0.5Abs), ±0.1%T Baseline Correction Selectable with storage in firmware Baseline Flatness Within ±0.001Abs. (excluding noise, 2nm slit width and SLOW wavelength scanning speed) Drift Less than 0.0004 Abs. per hour (after 2 hours warm-up) Dimensions 570 (W) x 660 (D) x 275 (H) (mm) Weight 36kg Power Supply AC 100V/120V/220V/240V, 50/60Hz 250VA (swithc-selectable)

Shimadzu UV-2450 Spectrophotometer

d. Effects of monchromator exit slit width on spectra Narrow exit slit width improves the spectrum resolution but it also significantly reduce the radiant power Trade-off between resolution and S/N ratio

e. Multichannel spectrometer No monochromator, but disperses transmitted light and measures “all wavelength at once” No Scanning-simple and fast More expensive Limited resolution Fig. 13-15 (p.353)

5.1 Quanlitative spectra 5 Applications Solvent effects on the UV-Vis spectra Polar solvents “blur” vibrational features Polar solvents shift absorption maxima n  * blue shift   * red shift - UV-Vis not reliable for qualitative but excellent for quantitative analysis

5.2 Quantitative analysis - Determining the relationship between A and c External Standards Standard-Addition

5.3 Spectrophotometric kinetics Fig. 14-14 (p.382)

Stopped-flow mixing Fig. 14-16 (p.384)