Poor Reproducibility of HIV­1 Low-level Viraemia Results with 3 Commercial Real-time PCR Assays Jean Ruelle 1, Laurent Debaisieux 2, Ellen Vancutsem 3,

Slides:



Advertisements
Similar presentations
Analytical Performance of the APTIMA HPV Assay on the PANTHER Analyzer
Advertisements

Statistical Reasoning for everyday life
Utilizing a Non-Commercial Real- Time PCR to Detect HIV-1 RNA in HIV Antibody Negative Diagnostic Sera Submitted to The Maryland Public Health Laboratory.
We processed six samples in triplicate using 11 different array platforms at one or two laboratories. we obtained measures of array signal variability.
Technical and Operational Considerations for Scaling Up
Simplification, cost-reduction strategies and examples from the field Teri Roberts Diagnostics Advisor Médecins Sans Frontières, Access Campaign 7th.
SPH 247 Statistical Analysis of Laboratory Data 1April 16, 2013SPH 247 Statistical Analysis of Laboratory Data.
Karen Cristiano Biologicals Unit, CRIVIB Calibration against the WHO Standards of National Reference Preparations for detection of blood viruses by NAT:
Linearity Panels HIV RNA, HCV RNA, HBV DNA, and CMV DNA
Cost-effectiveness of different starting criteria of antiretroviral therapy in Mexico. Caro Y., Colchero A., Valencia A., Bautista-Arredondo S., Sierra.
COBAS TaqMan 48 Sales Brochure
Chemometrics Method comparison
Detection of HIV-1 RNA in seminal plasma samples from treated patients with undetectable HIV-1 RNA in blood plasma Marcelin AG 1, Tubiana R 1, Lambert-Niclot.
PROFICIENCY TESTING OF IN-HOUSE NAT ASSAYS USED FOR BLOOD SCREENING XXI SoGAT International Working Group Meeting on the Standardization of NAT for the.
Persisting long term benefit of genotypic guided treatment in HIV infected patients failing HAART and Importance of Protease Inhibitor plasma levels. Viradapt.
Detection of parvovirus B19 and novel human parvoviruses in high-risk individuals Ashleigh Manning 1, Kate Templeton 2, Ed. Gomperts 3, Peter Simmonds.
HIV, HCV, and HBV NAT Controls Formulation, Stability and Performance Mark Manak BBI Diagnostics, Inc. A Division of SeraCare Life Sciences, Inc. SoGAT.
Transmissibility of B19 by transfusion: evaluation of donor viremia prior to performing a linked donor-recipient study M Busch, representing REDS-II Parvo.
Lara Isobel Compston, Daniel Candotti, Jean-Pierre Allain
Performance Characteristics of the FARRZYME™ High Avidity Anti-dsDNA IgG Antibody ELISA Brenda B. Suh-Lailam 1, Tyson R. Chiaro 1, K. Wayne Davis 1, Andrew.
Immune correlates of unusual control of viral replication after cessation of HAART Ellen Van Gulck 1, Leo Heyndrickx 1, Céline Merlin 1, Sandra Coppens.
1 Treatment Failure HAIVN Harvard Medical School AIDS Initiative in Vietnam.
Occult hepatitis B virus (HBV) and hepatitis C virus (HCV) viremia in women with and at-risk for HIV/AIDS Taylor L, Gholam P, Delong A, Rompalo A, Klein.
Real-Time Quantitative PCR Basis
Stability of HCV, HIV-1 and HBV nucleic acids in plasma samples stored at different temperatures Marta José, Rodrigo Gajardo and Juan I. Jorquera Instituto.
HCV PCR By Henrietta Orji July 31 st, 2010 Hepatitis C Virus by Polymerase Chain Reaction.
Effect of 24 Week Intensification with a CCR5-Antagonist on the Decay of the HIV-1 Latent Reservoir IAS HIV RESERVOIRS WORKSHOP, 16 & 17 JULY 2010, VIENNA.
The Positive Predictive Value of World Health Organization (WHO) Immunologic Criteria for Treatment Failure in a Public Health Antiretroviral Delivery.
National Institute for Biological Standards and Control Assuring the quality of biological medicines Human Cytomegalovirus (HCMV) Proposed 1 st International.
The Use of Pooled Viral Load Testing to Identify Antiretroviral Treatment Failure Davey Smith 1, Susanne May 2, Josué Perez-Santiago 1, Matthew Strain.
LOWERING THE DETECTION LIMITS OF HIV-1 VIRAL LOAD USING REAL-TIME IMMUNO-PCR FOR HIV-1 P24 ANTIGEN Niel T. Constantine, Ph.D., Daniel Edelman, M.S., Janet.
Development of Standard Reagents for WNV NAT M. Rios, A. Grinev, K. Sirnivasan, O. Wood, S. Daniel, I. Hewlett CBER/FDA.
ZIMBABWE AIDS CARE FOUNDATION NEWLANDS CLINIC Virological Outcomes in Adult Patients on Second Line ART, at Newlands Clinic Dr S. Bote.
History of ILC Involvement in NAT Standardization 1998: Inter-Organization Discussion for the Establishment of Standard Reference Material for Nucleic.
Evaluation of the Presage™ ST2 ELISA Jun Lu 1, David G. Grenache 1,2 1 ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 2 Department.
NUCLEIC ACID AMPLIFICATION TECHNOLOGY HCV-RNA / HBV-DNA / HIV-RNA testing blood and blood components for transfusion Italian External Quality Assessment.
1 Exercise 7: Accuracy and precision. 2 Origin of the error : Accuracy and precision Systematic (not random) –bias –impossible to be corrected  accuracy.
For further information please contact: ERNDIM QUANTITATIVE ORGANIC ACIDS IN URINE: HIGHLIGHTS AND RECOMMENDATIONS Martens.
Audit of outcomes in HIV BHIVA Audit and Standards Sub-Committee E Ong (chair), J Anderson, D Churchill, M Desai, S Edwards, S Ellis, A Freedman, P Gupta,
Molecular Monitoring in CML the process and questions we can answer Letizia Foroni.
DISCONTINUATION OR NO DISCONTIUNATION: COMPARISON OF SINGLE UNIT HIV ANTIGEN TESTING VS. POOLED NAT TESTING Gerald Schochetman, Ph.D. Abbott Laboratories.
Development and Characterization of Synthetic Standards: HCV RNA Susan E. Bromley, Ph.D. Director, Product Development Bayer Healthcare Diagnostics 24.
Treatment Failure HAIVN Harvard Medical School AIDS Initiative in Vietnam.
Phar. Nhat Mang/ Roche Vietnam
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Copyright © 2015, TestAmerica Laboratories, Inc. All rights reserved. 1 EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows.
Relative Sensitivities of US Licensed NAT Assays for Detection of Viremia in Early HIV and HCV Infection MP Busch, SA Glynn, DJ Wright, D Hirschkorn, ME.
BioPlex 2200 HIV Ag-Ab Assay
Paz Bailey G 1, Sternberg M 1, Puren AJ 2, Markowitz LE 1, Ballard R 1, Delany S 3, Hawkes S 4, Nwanyanwu O 1, Ryan C 1, and Lewis DA 5 1. NCHHSTP, CDC.
HBV viral load 측정 및 임상적 의의 진단검사의학과이희주 Content v 임상적 의의 v 측정법.
Cepheid Symposium IAS 2017 Gwynn Stevens, PhD Director of Virology
John Frater University of Oxford
Abbott HCV core Ag and HCV RNA Comparison Study
Find the lower and upper quartiles for the data set.
Hepatitis C Virus RNA Real-Time Quantitative RT-PCR Method Based on a New Primer Design Strategy  Lida Chen, Wenli Li, Kuo Zhang, Rui Zhang, Tian Lu,
Evaluation of Alere q HIV-1/2 VL Plasma
ASM 2017 Microbe Annual Meeting Sunday-507
Hepatitis C Virus RNA Real-Time Quantitative RT-PCR Method Based on a New Primer Design Strategy  Lida Chen, Wenli Li, Kuo Zhang, Rui Zhang, Tian Lu,
Immunologic and virological response to ART among HIV infected individuals at a tertiary hospital in Ghana ¹Obiri-Yeboah Dorcas, School of Medical Sciences,
Evaluation data on the new Roche cobas s 201 system with the cobas TaqScreen MPX test Lutz Pichl, Volkmar Schottstedt German Red Cross Blood Transfusion.
Diagnostic Testing for Pandemic Influenza in Singapore
Nat. Rev. Nephrol. doi: /nrneph
How to make a box and whisker plot
Disclaimer: I have no conflicts of interest
International Standards and Reference Materials for Quantitative Molecular Infectious Disease Testing  Roberta M. Madej, Jack Davis, Marcia J. Holden,
Trust is good, control is better!
Joseph Perriëns World Health Organization Geneva
Evaluation of the Abbott Investigational Use Only RealTime HIV-1 Assay and Comparison to the Roche Amplicor HIV-1 Monitor Test, Version 1.5  Michael T.
Performance of mPima for quantification of HIV Viral load for pregnant and post-partum woman in primary health care clinics in Mozambique Team: B. Meggi.
Presentation transcript:

Poor Reproducibility of HIV­1 Low-level Viraemia Results with 3 Commercial Real-time PCR Assays Jean Ruelle 1, Laurent Debaisieux 2, Ellen Vancutsem 3, Annelies De Bel 3 and Patrick Goubau 1. 1: UCLouvain, AIDS Reference Laboratory, Avenue Hippocrate 54 - B , 1200 Brussels, Belgium. 2: Hôpital Universitaire Erasme, AIDS Reference Laboratory, Route de Lennik 808, 1070 Brussels, Belgium. 3: Vrije Universiteit Brussel, AIDS Reference Laboratory, site Universitair Ziekenhuis Brussel, Laarbeeklaan 101, 1090 Brussels, Belgium. Contact author: Meeting 2011 September 30, Leuven 1. Background and Objectives - Ultra-sensitive real-time PCR-based HIV-1 viral load assays theoretically allow the detection of any RNA copy present in the sample, allowing the detection of residual viral replication in patients previously defined as virologically suppressed. - Recently, studies on the clinical outcome of HIV‐1 patients experiencing low-level replication (<50 cop/ml) in comparison to fully suppressed patients were published. We hypothesised that this exceeds the technical limit of PCR. Hence, we tested the analytical variability below 50 cop/ml of three commercial real‐time assays. 2. Methods Assays tested: - Versant HIV-1 RNA 1.0 kPCR (Siemens), limit of quantification (LOQ) 37 cop/ml. - Realtime HIV-1 (Abbott), LOQ 40 cop/ml. - Cobas Ampliprep/Cobas Taqman HIV-1 v2 (Roche Diagnostics), LOQ 20 cop/ml. Analysis were performed in the clinical routine settings of Belgian AIDS reference laboratories, following the assay manufacturers’recommandations (one assay per site). All assays showed important variability; it was the highest for the Roche assay. The Abbott assay was here less sensitive for very low target concentrations, and the Siemens assay quantified under 50 cop/ml some aliquots of the 100 cop/ml target dilution. 4. Conclusions The most recent versions of widespread commercial real‐time assays show an important variability at VL below 50 cop/ml. Patient outcome studies based on very low‐level viraemia are biased. Blips under 100 cop/ml are mostly due to analytical assay variability. 4. Conclusions The most recent versions of widespread commercial real‐time assays show an important variability at VL below 50 cop/ml. Patient outcome studies based on very low‐level viraemia are biased. Blips under 100 cop/ml are mostly due to analytical assay variability. Acknowledgements: Companies provided free assay kits for the evaluation. Abbott and Roche diagnostics offered support to calculate quantitative results from the PCR Cq values under the LOQ. Samples: - Prospective collection in each participating laboratory from HIV-1 positive patients. - 3 aliquots of 1.1 ml of plasma from fresh EDTA-blood samples were stored at -80°C, no supplementary thaw/freeze cycle was applied. - The selection was made at random: the only criteria applied was a detectable plasma viral load (VL) under 50 cop/ml with the first aliquot. - The same assay lot was used throughout the study, except for the re-evaluation of blips. 3. Results Dilutions of a reference sample A shared sample used as a quality control tested on each run was used as a reference. Its absolute value was defined as the mean of all the results obtained in routine settings with the present control lot in the 3 labs. Ten replicates at 4 target dilutions of 100, 50, 25 and 12.5 cop/ml were tested on each platform. In the boxplot, the bottom and the top of the box represent the lower and upper quartiles, respectively. The triangle in the box is the median. The ends of the whiskers represent the minimum and the maximum of all data. No assay gave signal for 10 replicates of the HIV negative plasma used for dilutions. Si: Versant HIV-1 RNA 1.0 kPCR (Siemens) ; Ab: Realtime HIV-1 (Abbott); Ro: Cobas Ampliprep/Cobas Taqman HIV-1 v2 (Roche) Intra-assay variability Each participating laboratory tested 3 times 40 samples with detectable VL below 50 cop/ml, in separate runs. Variability was assessed qualitatively with a discrepancy score: a score of +1 was attributed if any of the replicates crossed the thresholds of 50 cop/ml. When a detectable sample with VL under 50 cop/ml was tested in triplicate, the result is not reproduced in more than 50% of cases where at least one replicate was not detected. As a consequence, when the VL is not detected in clinical routine settings, we should not consider it as a proof of full viral suppression. Inter-assay variability 61 clinical samples with detectable VL below 50 cop/ml were tested in triplicate, one time on each platform. Mean score was 0.7, 52% of the samples were not detected by at least one assay, 11.5% detected by one assay only. VL assay Mean discrepancy score Samples where at least one replicate is not detected Samples where at least one replicate is higher than 100 cop/ml Versant kPCR (Si)0.5954%2.5% Realtime HIV-1(Ab)0.857%5.7% CAP/CTM v2 (Ro)0.7355%2.5% Blips As variability is high round the clinically relevant limit of 50 cop/ml, we reanalysed 26 blips, defined as a VL between 50 and 100 cop/ml experienced by patients whose previous and next sample was <50 cop/ml. A low but detectable VL under therapy may be considered as treatment failure, or a parameter used in clinical studies to consider a patient as non-responder. The VL blips could only be reproduced in 19.2% of cases. In the remaining cases, the sample had a VL below 50 or completely undetectable. In 1 case (4 %), the VL was above 100 cop/ml. Therefore, the variability measured here is in agreement with the 2011 revision of DHHS guidelines which define the virological failure threshold as 200 cop/ml.