BIOMECHANICS OF SPINAL INSTRUMENTATION

Slides:



Advertisements
Similar presentations
Robots In Spine Biomechanics
Advertisements

Lumbar Segmental Instability The Role of Interbody Fusion zRole of Interbody Spacers are to provide a means to stabilize the anterior column zAnterior.
Emergency Spinal Radiological Assessment
Re-written by: Daniel Habashi Intertrochanteric Hip Fractures.
Olecranon fracture Lonnie Froberg, MD, Ph.D Odense University Hospital.
An Experimental Study and Fatigue Damage Model for Fretting Fatigue
“A Controlled Randomized Outcome Study of Femoral Ring Allograft versus BAK Instrumentation in Anterior Interbody Fusion” Dr. Donald W. Kucharzyk Dr. Michael.
Beams and Frames.
Coen A. Wijdicks, Ph.D. Chief Operating Officer Director of the Department of BioMedical Engineering Senior Staff Scientist Steadman.
NUMERICAL METHODS THAT CAN BE USED IN BIOMECHANICS 1)Mechanics of Materials Approach (A) Complex Beam Theory (i) Straight Beam (ii) Curved Beam (iii)
Avoiding Pitfalls: Testing Considerations
Prediction of Load-Displacement Curve for Weld-Bonded Stainless Steel Using Finite Element Method Essam Al-Bahkali Jonny Herwan Department of Mechanical.
Chap. 5 Muscular Fitness Chap. 6 Flexibility. Health Benefits Increased bone density Increased HDL-C Increased muscle mass which increases BMR Decreased.
Objective Measurement for Lumbar Spinal Angels Submitted To Prof. Dr. Maher El-keblawy Professor of Basic Science Department Faculty of Physical Therapy.
 Finite Element modeling to predict the implant holding strength of osteoporotic bone Bhishan Kayastha ( ) Supervisor: Karen Reynolds.
1/27Optimum DesignOptimum Topology Design of an Interbody Fusion Implant Optimum Topology Design of an Interbody Fusion Implant for Lumbar Spine Fixation.
1/24OPTI Design of an Interbody Fusion Implant Optimum Topology and Shape Design of an Interbody Fusion Implant for Lumbar Spine Fixation Andrés.
Lifting Lifting process is applied frequently by most of the people, so it’s very necessary to know the loads during lifting, include the weight of the.
Fracture Mechanisms (Long Bones) n Direct and Indirect n Direct –Magnitude and area distribution of the force –Rate at which force acts on the bone.
Daniel Boedeker, MD Spinal Hardware Extraction.  Spinal instrumentation has been utilized since the 1940s  Pedicle screws became increasingly popular.
Bridge vs. Tension Band Construct for Spinal Fractures Fixation: A Biomechanical Analysis C. Richards, J. Ouellet, M. Fouse, N. Noiseux, P. Jarzem, R.
1/23DETC Design of an Interbody Fusion Implant Optimum Topology and Shape Design of an Interbody Fusion Implant for Lumbar Spine Fixation Andrés.
Decompression Surgery
Lower Lumbar Fractures Wayne Cheng, MD. Duke University Medical Center.
ATC 222 The Spine Chapter 25 Natasha Tibbetts, ATC.
Biocompatibility of Titanium Implants in Beagle Dogs Chuck Rosenwasser Mentors: Frank Raia & Mel Rosenwasser Columbia Presbyterian Medical Center.
SPINAL FUSION AND INSTRUMENTATION
In Vivo Loads on the Lumbar Spine Standing and walking activities: 1000 N Supine posture: ~250 N Standing at ease: ~500 N Lifting activities: >>
Engineering Doctorate – Nuclear Materials Development of Advanced Defect Assessment Methods Involving Weld Residual Stresses If using an image in the.
Biomechanical Properties of Formalin Fixed Lumbar Intervertebral Discs Emily Brown Advisor: Dr. Gary Bledsoe REU Summer 2009 Saint Louis University.
Chapter 15 The Spine Impairments, Diagnosis, and Management Guidelines.
IMPACT Phase II – 9/13/00 Activity Report Slide 1/20 University of Louisville IMPACT Architecture Team Glen Prater, Jr., Associate Professor Ellen G. Brehob,
BIOMECHANICS OF FUSION
Techniques used for Functional and Multi-segmental Spinal Unit Functional Spinal Unit (Motion Segment) –the smallest segment of the spine that exhibits.
Chapter Five Vibration Analysis.
Overview of Mechanical Engineering for Non-MEs Part 2: Mechanics of Materials 6 Introduction – Concept of Stress.
Failed Medical Device:
Probabilistic Analysis: Applications to Biomechanics Students: Saikat Pal, Jason Halloran, Mark Baldwin, Josh Stowe, Aaron Fields, Shounak Mitra Collaborators:
In the name of God H. Moin M.D, F. R.C.S Oct
John T. Wilkinson m. d. , Chad E. Songy m. d. , Frances l
LATHE VIBRATIONS ANALYSIS ON SURFACE ROUHHNESS OF MACHINED DETAILS LATHE VIBRATIONS ANALYSIS ON SURFACE ROUHHNESS OF MACHINED DETAILS * Gennady Aryassov,
by Robert W. Gaines J Bone Joint Surg Am Volume 82(10):
Biomechanics of Human Spine
KIN 330 Structural and Functional Analysis of Human Movement.
EGM 5653 Advanced Mechanics of Materials
Posterior Cervical Pedicle Screw versus 360° stabilization of the Subaxial Cervical Spine in a Multilevel Vertebral Body Failure Model. An In vitro Biomechanical.
North-West Medical University named after I.I.Mechnikov
Universitätsklinikum Carl Gustav Carus an der Technischen Universität Dresden Preoperative simulation reduces surgical time and radiation exposure for.
Lordosis (Depuy Bengal® Stackable Cage System)
Chapter 9: The Biomechanics of the Human Spine
Defect-Defect Interaction in Carbon Nanotubes under Mechanical Loading Topological defects can be formed in carbon nanotubes (CNTs) during processing or.
OUTCOME OF SPINE SURGERY IN ELDORET
 Dual energy x-ray absorptiometry (DXA) is the current standard for measuring bone mineral density (BMD) as it offers quick scan times with low radiation.
Introduction In the last years several interspinous prostheses have been used for the treatment of several degenerative disc diseases involving the lumbosacralregion;
Olson EM, Sturm PF, Jain VV, Schultz LR, Glos DL, Bylski-Austrow DI
Objective Measurement for Lumbar Spinal Angels
Pediatric orthopedic surgeon – Ramallah - Palestine
Facet sparing laminectomy
Cervical Spine Trauma Odontoid fractures Anatomic pathology
Patterns of Failure in the Distal Radius Following Treatment for AO 23-A3.2 Fractures Using Two-Column Volar Plates Deana Mercer, MD; Christina Salas,
Pediatric orthopedic surgeon – Ramallah - Palestine
Changes in Adjacent Segment Biomechanics After Laminectomy and Laminotomy in Lumbar Spine 1Elmasry, S; 1Asfour, S; 2Gjolaj, J; 3Latta, L; 2Eismont, F;
R. Vazquez B.Ing. C.E. Aubin Ph.D., H. Labelle M.D.
Kinetics: def. describes the effect of forces on the body (applies to musculoskeletal system) -force: consider as a push or pull that produce: arrest,
Garrido E†, Bermejo F†, Tucker SK†‡, Noordeen HNN†‡, Morley TR‡
BIOLOGICAL IMPLICATIONS OF
Anterior instrumentation and correction
Scoliosis surgery with hybrid system in osteogenesis imperfecta (OI)
Coombs MT, De Carvalho MF, Glos DL, Kim J, Wall EJ, Bylski-Austrow DI
Presentation transcript:

BIOMECHANICS OF SPINAL INSTRUMENTATION

SPINAL INSTRUMENTATION Goal: To maintain anatomic alignment of injured spinal segments by sharing the loads acting on the spine until a solid biologic fusion takes place Prerequisite Understandings: Pathology Spine Biomechanics Biomechanics of implant constructs

SPINAL INSTRUMENTATION ADVANTAGES - Promote fusion - Enhance early mobilization DISADVANTAGES - Adverse effects on the adjacent segments - Stress-shielding effects on the stabilized segment - Hardware Failure

HARDWARE FAILURE Breakage of Rods or Screws Hook Dislodgement Loosening Screw Pull-out in cases of extreme Osteoporosis

Biomechanical Consideration Factors Biomechanical Strength (Static and Dynamic) Impact strength Surgical construct strength Metal-metal interface strength Bone-metal interface strength Stability Segmental stiffness/flexibility Load Sharing Implant survival Stress shielding Fusion rate and quality

Implant Design Factors Mechanical strength: Profile: Low profile preferred particularly in anterior instrumentation User Friendliness: Top loading Poly-axial screw insertion Lateral-medial ajustment Easy-to-use surgical insruments Versatility:

Biomechanical Consideration Factors Fretting Corrosion: Any damage to the material that takes place at the edge of contacting parts or within the local contact area Associated with wear, surface damage and accumulated debris The rate of fretting corrosion is mostly governed by micromotion between component surfaces within an interconnection Cause of late infection and loosening of interconnections Greater tendency for: designs with lower strength interconnections Improper assembly such as misalignment and incomplete seating Higher applied loads

Christopher Evanich, M.D. PULL-OUT AND CYCLIC FAILURES OF THE ANTERIOR VERTEBRAL SCREW FIXATION IN RELATION TO BONE MINERAL DENSITY Howard S. An, M.D. Tae-Hong Lim, Ph.D. Christopher Evanich, M.D. Toru Hasegawa, M.D. Kaya Y. Hasanoglu, M.D. Linda McGrady, B.S.

Anterior Spinal Instrumentation Fractures Tumors Correction of Deformities

Hardware Failure Breakage of rods or screws Screw loosening Screw pullout in cases of extreme Osteoporosis

PREVIOUS STUDIES Screw Pullout Strength in relation to Design Variables and Insertion Methods of Pedicle Screw - Zindrick et al., 1986; Krag et al., 1986; Skinner -et al., 1990; Daftari et al., 1992 Screw Pullout Strength in relation to Bone Mineral Density (BMD) - Coe et al., 1990; Soshi et al., 1991; Yamagata et al., 1992; An et al., 1993

PREVIOUS STUDIES Loosening of the Pedicles Screw in relation to Design Variables and Insertion Methods of Pedicle Screw - Zindrick et al., 1986 No BMD Measurement

Anterior Vertebral Screw Fixation Failure Bone Mineral Density (BMD) Correlation between Anterior Vertebral Screw Fixation Failure And Bone Mineral Density (BMD) Of the Verebral Body

PURPOSE OF THE STUDY Relationships between BMD and: - Pull-out Strength - Cyclic Screw Loosening

PART II: Cyclic Loosening Test PART I: Pull-out Test (6 fresh-frozen lumbar spines) PART II: Cyclic Loosening Test (5 fresh-frozen lumbar spines)

MEASUREMENT OF BMD (g/cm2) Dual Energy X-ray Absorptiometry (DEXA) - Scan Speed: 60 mm/sec - Resolution: 1 x 1 mm Lateral view of each lumbar spine

Kaneda Screw (6.5 mm diameter; 55 mm long) Measurement of Torque (Nm) and Vertebral Width (mm) for screw insertion

Pull-out Strength Test Screw was pulled out along its long axis. Loading Rate: 10 mm/min (Displacement Control)

DATA ANALYSIS (Pull-out Strength Test) Maximum pull-out force in the load displacement curve Regression Analyses

Cyclic Loading Test to Induce Screw Loosening Cyclic loading was applied to the screw in the cephalad-caudal direction using an MTS machine. Loading Frequnecy: 0.5 Hz Loading Amplitude: 200 N  100 N 

DATA ANALYSIS (Cyclic Loosening Test) Number of Loading Cycles (NLC) to Induce the Screw Loosening: when the displacement > the displacement at the first peak load + 1 mm. Regression analysis NLCs for specimens with BMD < 0.45 vs. BMD > 0.45.

RESULTS

Means (SD) of Measured Parameters (Pull-out Test) BMD: 0.58 g/cm2 (STD 0.14) Torque: 0.86 Nm (STD 0.2) Depth: 41.0 mm (STD 3.28) Pull-out Strength: 211 N (STD 124)

CORRELATION COEFFICIENTS (Pull-out Test) Strength BMD Torque (T) Width (W) Pull-out Strength 1.0 0.85 0.47 No relation BMD - 1.0 0.58 No relation Torque (T) - - 1.0 No relation Width (W) - - - 1.0

Pull-out Strength vs. BMD Pull-out Strength = -226 + 774 x BMD CORRELATION Pull-out Strength vs. BMD R = 0.83 (p < 0.0002) Pull-out Strength = -226 + 774 x BMD

Pull-out Strength vs. BMD Pull-out Force (N) BMD (g/cm2)

Means (SD) of Measured Parameters (Cyclic Loosening Test) BMD: 0.32 g/cm2 (STD 0.10) Torque: 6.9 Kg-cm (STD 2.8) NLC: 149 (STD 234) (5 to 960)

CORRELATION (Cyclic Loosening Test) NLC vs. BMD Second Order Polynomial Relationship NLC = -1190 BMD + 3168 BMD2 R = 0.80 (p < 0.01)

NLC vs. BMD

CORRELATIONS (Cyclic Loosening Test) NLC vs. Screw Insertion Torque: Second Order Polynomial R = 0.68 (p < 0.01) BMD vs. Screw Insertion Torque: Linear R = 0.51 (p < 0.02)

MEAN NLCs (SD) 12 Specimens with BMD<0.45: 18.0 (20.1) 270.8 (278.8) These were significantly different. (p=0.003)

BMD: correlation with pull-out strength as well as cyclic loosening failure of the anterior vertebral screw a useful means to evaluate the severity of osteoporosis DEXA: measurement of BMD - Low-radiation dose - Short scanning time - Increased image resolution - Improved precision

CONCLUSION Quantitative assessment of BMD using a DEXA unit may be a good predictor of anterior vertebral screw fixation failure. BMD < 0.45 g/cm2 may be the critical value for osteoporosis in screw loosening.

PREDICTION OF FATIGUE LOOSENING FAILURE OF THE PEDICLE SCREW FIXATION Tae-Hong Lim, Ph.D. Lee H. Riley, III, M.D. Howard S. An, M.D. Linda M. McGrady, B.S. John Klein, M.S.

Pullout Strength of the Pedicle Screw In relation to Pedicle Screw Design Variables - Zindrick et al., 1986; Krag et al., 1986; Skinner -et al., 1990; Daftari et al., 1992 Effects of screw insertion torque, screw hole preparation method, and screw angulation - Skinner et al. 1991; Daftari et al. , 1992; Zdeblick et al. , 1993 In relation to Bone Mineral Density (BMD) - Coe et al., 1990; Soshi et al., 1991; Yamagata et al., 1992

Loosening Failure of the Screw Fixation Pedicles Screw Loosening in relation to Design Variables and Insertion Methods of Pedicle Screw - Zindrick et al., 1986 Anterior Vertebral Screw Loosening in relation to the Bone Mineral Density of the Vertebral Body - Lim et al., 1994

PURPOSE Fatigue Loosening of the Pedicle Screw in relation to Bone Mineral Density of the Vertebral Body

Pedicle Size and Screw Insertion Torque PURPOSE Fatigue Loosening of the Pedicle Screw in relation to Pedicle Size and Screw Insertion Torque

MATERIALS Three fresh frozen human lumbar spines (L1 - L5) were used in this study. Anterior-posterior and lateral radiographs were taken to exclude spines with gross pathology.

MEASUREMENT OF BMD (g/cm2) Dual Energy X-ray Absorptiometry (DEXA) - Scan Speed: 60 mm/sec - Resolution: 1 x 1 mm Lateral view of each lumbar spine

Pedicle Size Measurement Pedicle Height (PH): Long Axis (cephalad-caudal direction) Pedicle Width (PW): Short Axis (medial-lateral direction) Pedicle Screw Placement A 3.5 mm drill hole and 6.25 mm tapper A 6.25 mm Steffee Screw (40 mm long) Screw Insertion Torque (TQ) Measured using a Torque Wrench

Cyclic Loading Test to Induce Screw Loosening Cyclic loading was applied to the screw in the cephalad-caudal direction using an MTS machine. Loading Frequency: 0.5 Hz Loading Amplitude: 200 N  100 N 

Cyclic Test Set-up MTS Load Cell MMA Loading Direction Fixture RAM

DATA ANALYSIS Number of Loading Cycles (NLC) to Induce the Screw Loosening: when the displacement > the displacement at the first peak load + 1 mm. Regression analysis Relationships between NLC, BMD, PH, PW, and TQ

Means (SDs) of the Measured Parameters BMD: 0.465 g/cm2 (0.187) PH: 15.57 mm (4.97) PW: 12.08 mm (2.44) TQ: 9.25 Kg-cm (3.14) NLC: 300 (377)

Correlation between the Measured Parameters BMD PH PW TQ NLC BMD 1.0 - - - - PH 0.54 1.0 - - - PW *0.07 0.51 1.0 - - TQ *0.44 0.59 *0.11 1.0 - NLC 0.58 0.72 *0.42 0.50 1.0 * indicates relationship with no statistical significance (p > 0.05).

Linear Relationship between NLC and BMD BMD (g/cm/cm)

Linear Relationship between NLC and PH PH (mm)

Multiple regression analysis demonstrated a significant correlation between NLC and other measured parameters. (R = 0.79, p = 0.02) NLC = 775.8 x BMD + 22.4 x PH - 44.3 PW + 15.5 TQ - 17.2

The stepwise regression analysis revealed that PH was the most significant predictor of pedicle screw loosening.

Screw loosening: a significant complication in spinal fixation. DISCUSSION Screw loosening: a significant complication in spinal fixation. Pedicle screw loosening was experimentally induced under cyclic loading in this study. Application of the cyclic load in the cephalad-caudal direction on the screw in order to simulate the load transferred to the screw by the connecting rod or plate

DISCUSSION NLC was significantly correlated with BMD, PH, and TQ, while PH was the most significant predictor of screw loosening failure. Honl et al. also found that the amount of loosening correlated more with pedicle geometry than BMD. (Proceedings, 2nd World Congress of Biomechanics, 1994) Limitations: Small number of Specimens; No studies on the optimal ratios of the screw diameter to PH for the best fixation strength

A significant relationship between the fatigue loosening of the pedicle screw and PH and BMD was found. The assessment of the pedicle geometry as well as the vertebral body BMD may provide valuable preoperative information to predict the early loosening failure of the pedicle screw fixation.

BIOMECHANICAL EVALUATION OF ANTERIOR SPINAL INSTRUMENTATION Jae Won You, M.D., Ph.D. Tae-Hong Lim, Ph.D. Howard S. An, M.D. Jung Hwa Hong, M.S. Jason M. Eck, B.S. Linda McGrady, B.S.

Anterior Spinal Instrumentation Plate Type Syracuse I-Plate, CASP, Z-Plate, University Plate, etc. Rod Type Zielke, Kaneda, TSRH instrumentation, etc.

Purpose of the Study Compare the biomechanical Stability of anterior fixation constructs, particularly comparing the use of plate vs. rod type anterior fixators.

METHODS

Specimen Preparation 20 Fresh Calf Lumbar Spines (L2-L5) The lumbar spine dissected free of muscles, leaving ligaments, capsules, and disc intact L5 fixed to the loading rig, and L2 mounted in an unconstrained loading setup. 3 markers reflecting infra-red light attached to L3 and L4 vertebral body.

Loading Setup Pure Moments: 6 Directions: Incremental Loading: using dead weights, unconstrained 6 Directions: FLX, EXT, RLB, LLB, RAR, and LAR Incremental Loading: to a maximum of 6.4 Nm in 6 steps

Motion Measurements 3-D Motion Analysis System: 3 Vicon Cameras (Oxford, England) Micro-Vax Workstation (DEC, Maynard, MA) Marker position data were transformed to rotational angles in FLX/EXT, LBs, and ARs.

Testing Constructs Intact Spine Anterior fixation with an interbody graft following total discectomy and endplate excision of L3-L4 PMMA block was used as a bone graft Anterior fixation only

Tested Anterior Fixators Plate Type: University Anterior Plating System (UNIV): AcroMed Corp. Z-Plate (ZP): Danek, Inc. Rod Type: Kaneda device with transfixator (KAN): AcroMed, Corp TSRH vertebral body screw construct: Danek, Inc.

Data Analyses Stabilizing Effect: Role of Graft: % change of motion as compared with the intact motion Role of Graft: % change of motion between with and without graft cases ANOVA with Tukey’s multiple mean comparison

Results

In Lateral Bending and Flexion: Stabilizing Effect compared with Intact Motion (Anterior Fixation with Graft) In Axial Rotation: Similar to the Intact Motion in UNIV, ZP, TSRH Fixation Significantly stabilized by KAN (p < 0.05) In Lateral Bending and Flexion: Singnificantly stabilized by all tested devices (p < 0.05) No significant difference between devices

Stabilizing Effect compared with Intact Motion (Anterior Fixation with Graft) *

Stabilizing Effect compared with Intact Motion (Anterior Fixation with Graft) In Extension: Singnificantly stabilized by all tested devices (p < 0.05) Stabilizing effect of KAN > ZP (p = 0.05), but no difference between any other devices

Stabilizing Effect compared with Intact Motion (Anterior Fixation with Graft) *

Stabilizing Effect compared with Intact Motion (Anterior Fixation only) In Axial Rotation: Similar to the intact motion in UNIV, ZP, TSRH Fixation Significantly stabilized by KAN (p < 0.05) In Lateral Bending: Singnificantly stabilized by all tested devices (p < 0.05) No difference between devices (p > 0.05)

Stabilizing Effect compared with Intact Motion (Anterior Fixation only) In Flexion: Significantly stabilized by UNIV, KAN, TSRH(p < 0.05), but not in ZP fixation Stabilizing Effect of ZP < KAN (p = 0.01) and TSRH (p = 0.05) In Extension: Singnificantly stabilized by KAN (p < 0.05) Stabilizing Effect of KAN > TSRH (p = 0.005)

Stabilizing Role of Interbody Graft In Axial Rotation: No significant stabilizing effect In Lateral Bending: Significantly increase the stabilizing effect in all tested cases In Flexion and Extension: Significantly increase the stabilizing effect in ZP fixation only

Stabilizing Role of Interbody Graft

DISCUSSION

Biomechanical Spinal Implant Testing Protocols Measuring Construct Stiffness in Axial Load, Torsion, Flexion, and Extension in nondestructive manner Ashmann et al., 1989 Measuring flexibility in terms of uncostrained motion under pure moment application Panjabi, 1988

Previous Studies Abumi et al. 1989: Gaines et al., 1991: Kneda device provides good stability in FLX and EXT, but not in AR. Gaines et al., 1991: Kaneda device provided good stabilization and resistance to torsional and lateral bending loads. Zdeblick et al., 1993: Kaneda and TSRH devices are effective in restoring acute stability to the lumbar spine after corpectomy.

In this study Measure the unconstrained motion under pure moments application. Findings: Kaneda device provides effective stabilization in all directions, particularly with the use of interbody graft. UNIV, ZP, TSRH are also effective in stabilizing LB, FLX, and EXT with an interbody graft, although they restore to the intact motion in AR.

CONCLUSION Modern plate and rod type anterior fixation devices are effective in restoring stability of anterior and middle column defects in flexion, extension, and lateral bending beyond the intact specimen.

CONCLUSION Interbody grafting may be important in anterior instrumentation for effective stabilization immediately after surgery. Anterior fixation restores torsional stability relatively less, although Kaneda device was the best among the tested devices.

Percentage AR Motion Changes compared to the Intact Case

Percentage LB Motion Changes compared to the Intact Case

Percentage FLX Motion Changes compared to the Intact Case

Percentage EXT Motion Changes compared to the Intact Case

Effects of Crosslinking Devices in Pedicle Screw Instrumentation: A Biomechanical and Finite Element Modeling Study Tae-Hong Lim, Ph.D. Howard S. An, M.D. Jason C. Eck, B.S. Jae Y. Ahn, M.D.

PURPOSE To evaluate the stabilizing effect of transfixation in flexion, extension, lateral bending, and axial rotation modes To determine the optimal position of transfixation to achieve greater stabilizing effect

Finite element studies Flexibility tests Unstable Calf Spine Model Finite element studies

FLEXIBILITY TESTS 10 Ligamentous calf spines - 5 spines for L3-L4 stabilization - 5 spines for L2-L4 stabilization Maximum pure moment of 6.4 Nm in FLX, EXT, LB, and AR Vicon 3-D motion analysis system was used to measure the resultant segmental motions.

Tested Constructs One Segment (L3-L4) Stabilization - Intact - Instrumentation without Transfixation after total discectomy - Instrumentation with Single Transfixation

Tested Constructs Two Segment (L2-L4) Stabilization - Intact - Instrumentation without Transfixation after L3 corpectomy - Instrumentation with Single Transfixation at the Middle Points - Instrumentation with Double Transfixation at the proximal and distal 1/3 points

Finite Element Studies To determine the optimal position of transfixation Boundary and Loading Conditions: - Nodes in lower vertebra were held fixed. - Axial compression (445 N), FLX, EXT, LB, and AR Moments (5 Nm) at the middle point of the vertebra element MENTAT II Finite Element Analysis Package (Marc Analysis Research Corp.)

Finite Element Models Axial Compression force and Moment ISOLA System Transfixators Vertebrae Fixed Nodes One Segment Stabilization Two Segment Stabilization

RESULTS

Rotational Motions (deg) responding to Applied Moments of 6 Rotational Motions (deg) responding to Applied Moments of 6.4 Nm in One Segment Stabilization p<.01

Rotational Motions (deg) responding to Applied Moments of 6 Rotational Motions (deg) responding to Applied Moments of 6.4 Nm in Two Segment Stabilization p = .01 p = .06

Single Transfixation (FE Model Predictions) Compared with no TF case: No additional stabilizing effect in axial compression, FLX, and EXT In one segment stabilization: - Up to 10.8% LB motion decrease - Up to 25.0% AR motion decrease In two segment stabilization: - Up to 13.0% LB motion decrease - Up to 18.2% AR Motion decrease

Stabilizing Effect of 1 TF in AR Mode with respect to the implanting Positions 2 Segment Stabilization 1 segment stabilization

Double Transfixation (FE Model Predictions) Compared with no TF case: No additional stabilizing effect in axial compression, FLX, and EXT Up to 14.6% LB motion decrease Up to 30.3% AR motion decrease

Stabilizing Effect of 2 TF in AR Mode with respect to the Implanting Positions

CONCLUSION Flexibility Tests: Significant stabilizing effect of TF in AR only as shown in previous studies 2 TFs can provide more AR stability than one TF. With 2 TFs, the restored AR stability is similar to the intact specimen.

CONCLUSION Finite Element Model Predictions: Model prediction correlated well with the experimental results. The greatest AR stability can be obtained by implanting two TFs, one at the proximal 1/8 points and the other at the midpoints of the longitudinal rods in two level stabilization.

A BIOMECHANICAL COMPARISON BETWEEN MODERN ANTERIOR VERSUS POSTERIOR PLATE FIXATION OF UNSTABLE CERVICAL SPINE INJURIES Tae-Hong Lim, Ph.D. Howard S. An, M.D. Young Do Koh, M.D. Linda M. McGrady, B.S.*

Unstable Cervical Spine Injuries Flexion-distraction Injury Three column injury Unilateral or bilateral facet dislocation Burst Fracture of the Vertebral Body produce severe instability by destroying the anterior and middle column of the cervical spine

Surgical Treatments Closed or open reduction and fusion with Anterior Fixation anterior plate Posterior Fixation Wiring Posterior lateral mass screw system Combined Fixation

Surgical Treatment Goals Reduction Maintenance of alignment Early rehabilitation Enhancement of fusion Decreased use of external orthoses Biomechanical stability provided by various fixation methods is an important factor to achieve these goals.

Purpose of the Study To determine and compare the biomechanical stability provided by modern posterior, anterior, and combined screw-plate fixation in flexion-distraction and corpectomy models simulated in human cadaveric cervical spines.

METHODS

Flexibility Tests 10 cadaveric cervical spines (C2-T1) Group I (n = 5): One-level 3-column injury at c4-5 Group II (n =5): Corpectomy of C5 vertebral body Pure moment of 2.45 Nm was achieved in five steps using dead weights in FLX, EXT, LB, and AR directions. Vicon 3-D motion analysis system was used to measure the resultant segmental motions. Stability was quantified in terms of the percent changes in rotational motions of the stabilized segment compared with those in the intact segment.

Tested Constructs Group I (Flexion Distraction Model) Intact Posterior fixation without interbody grafting after 3-column injury Posterior fixation + an interbody graft (PMMA block) Combined anterior and posterior fixation + PMMA block Anterior fixation + PMMA block 3-Column injury was made by cutting supraspinous, interspinous, capsular and posterior longitudinal ligaments, ligamentum flavum, and the posterior half of the intervertebral disc.

Tested Constructs Group II (Corpectomy Model) Intact Posterior fixation without interbody grafting after corpectomy Posterior fixation + an interbody graft (PMMA block) Combined anterior and posterior fixation + PMMA block Anterior fixation + PMMA block C5 vertebral body was removed (corpectomy) to simulate a very unstable burst fracture. Axis plate and Orion plate systems (Danek Inc., Memphis, TN) were used for posterior and anterior fixation, respectively.

Axial Rotation (Flexion-Distraction Model) The posterior fixation (cases 1, 2, and 3) decreased the motion significantly. Anterior fixation alone provided the rigidity similar to the intact specimen.

(fracture-dislocation)

Lateral Bending (Flexion-Distraction Model) Posterior fixation with or without PMMA block reduced motion as compared with the intact case, although the differences were not statistically significant. When anterior fixation was used alone, the lateral bending became significantly greater than the intact motion (p < 0.05).

Lateral Bending (fracture-dislocation) Instrumentation 300 200 % Change of Motion 100 -100 1 2 3 4 Instrumentation

Flexion (Flexion-Distraction Model) The posterior fixation (cases 1, 2, and 3) significantly reduced the flexion motion from the intact case. No significant difference was found between the posterior fixation constructs. In case of anterior fixation (case 4), the flexion motion was significantly larger than the intact motion (128%, p < 0.01).

(fracture-dislocation)

Extension (Flexion-Distraction Model) Extension motion was significantly less than the intact motion in the combined anterior posterior fixation only (case 3). There were no significant differences between the posterior constructs with and without PMMA block. Extension motion was also not significantly different between the posterior and anterior fixation constructs.

(fracture-dislocation)

Axial Rotation (Corpectomy Model) Axial rotational motion was significantly reduced from the intact motion only when the combined anterior and posterior fixation was used (case 3). Axial rotational motions in other constructs were not significantly different from the intact motion.

Axial Rotation (Corpectomy) 90 70 50 30 10 % Motion Change -10 -30 -50 -70 -90 -110 -130 1 2 3 4 Instrumentation

Lateral Bending (Corpectomy Model) Lateral bending motions in posterior fixation cases 1, 2, and 3 were significantly less than the intact motion. Anterior fixation alone (case 4) did not provide the rigidity beyond the intact specimen even with the anterior grafting material.

(corpectomy)

Flexion (Corpectomy Model) Flexion motion was significantly reduced from the intact motion in all tested constructs (p < 0.05). The combined anterior and posterior fixation provided most rigid fixation than the other constructs.

(corpectomy)

Extension (Corpectomy Model) Extension motion was significantly less than the intact motion in all tested constructs except for the posterior fixation with PMMA block (case 2).

(corpectomy)

DISCUSSION

Human cadaveric cervical spines were used to minimize the anatomical difference between the in vitro and in vivo cases. OrionTM and AxisTM plate system were chosen for testing because they represent relatively rigid modern anterior posterior plates constructs in cervical spine fixation. A PMMA block was used to simulate the interbody grafting technique while eliminating potential variation in inherent material properties of bone graft. Limitations old age specimens and Bone quality variations Exclusion of supporting structures such as muscle, fascia, and ligamentum nuche

SUMMARY Anterior fixation system is biomechanically inferior to the posterior lateral mass screw-plate fixation, particularly in flexion-distraction injury in which posterior ligamentous structures are disrupted. The anterior fixation seems to be more suitable for anterior middle column injury where the posterior ligamentous elements are intact. Postoperative use of external orthoses should be considered when the anterior plate is used alone for the treatment of unstable cervical spine injuries with disruption of posterior stabilizing elements.

SUMMARY Anterior fixation was particularly ineffective to prevent the flexion motion in the flexion-distraction injury model. This seems to occur mostly due to no rigid connection between the plate and screws. Rigid connection at this junction may significantly improve the fixation. Combined fixation seemed to improve the stability as compared with anterior or posterior fixation alone. However, the difference in stability was not always significant and the combined fixation requires additional incision. These suggest that a combined anterior and posterior fixation should be carefully indicated.

Findings of Finite Element Studies Stress-shielding effects exist in the surgical construct due to the presence of spinal fixation devices and healed bone mass. The fixation device may transmit 9 to 40% of the applied compression load depending upon the stabilized technique used. Semi-rigid fixation may reduce the stress-shielding effect and incidence of hardware failure.

Finite Element Models Intact L3-L4 (INT) Bilateral Fixation (STVSP & PVSP) Unilateral Fixation (UVSP)

Important Factors in Finite Element Modeling Adequate Assumptions Use of Accurate Input Data Geometry; material properties; Element types Proper Solution Procedures Linear or nonlinear analysis; viscoelasticity; poroelasticity, etc. Complete Understanding of Results

EFFECT OF INSTRUMENTED FUSION ON THE BIOMECHANICS OF ADJACENT SEGMENT: AN IN VIVO CANINE STUDY Tae-Hong Lim, PhD, Avinash G. Patwardhan, PhD,* Jung Hwa Hong, MS, Howard S. An, MD, Lee H. Riley III, MD, Scott Hodge, MD,* Jason Eck, BS, Michael M. Zindrick, MD*

Complications in the Segments Adjacent to Fusion Degenerated Disc Stenosis Segmental Instability Osteoarthritis

Previous Studies Ex-vivo and In-vivo Studies of Post-fusion Mechanics: Increased motion and loads at the adjacent segment. In-vivo Animal Studies: Hypoactive metabolism in the adjacent discs; Significant biochemical changes indicating a degeneration process. Changes in the biomechanical response of the adjacent segment resulting from theses alterations in the disc were not investigated.

Previous Studies It is believed that a concentration of load can cause degeneration at the adjacent segment. There are little data on the long-term changes in the biomechanical properties of the adjacent segment.

OBJECTIVE Quantify the long-term changes in the flexibility and viscoelastic properties of the intervertebral disc at the adjacent segment due to the instrumented lumbar spinal fusion in a canine model.

In-vivo Canine Model Flexibility Tests Relaxation and Cyclic Loading Tests

In-vivo Canine Model Development Control Group: 5 adult mongrel dogs (age: 2 yr and weight: 25-30 kg) Euthanized at the beginning of the study Experimental Group: 5 adult mongrel dogs (age: 1.5 yr and weight: 25-30 kg) Posterior instrumented fusion surgery across L5-S1 levels using ISOLA system (AcroMed, Cleveland, OH) Follow-up Period: 30 weeks

Flexibility Tests Ligamentous lumbar spines (L2-S1) Maximum pure moment of 2.0 Nm applied in FLX, EXT, LB, and AR Vicon 3-D motion analysis system was used to measure the resultant segmental motions at L4-5, L5-L7, and L7-S1 levels.

Flexibility Tests Control Group Experimental Group Control-Intact Control-Implant Isola instrumentation following a partial dissection of the L5-6, L6-7, and L7-S1 facets Experimental Group Fusion + Instrumentation Fusion Alone After removing Isola instrumentation

Viscoelastic Property Measurement MTS Load Cell Relaxation Test A constant strain of 8% were applied using a ramp function in 1 minute. Relaxation Period: 1 hour L4-5 V B - Disc - VB MTS Ram

Viscoelastic Property Measurement Cyclic Deformation Tests in Axial Compression Test Parameters: Test Step Control Type Pre- Strain Maximum Loading Frequency # of Cycles 1 8% 12% 0.1 Hz 50 2 1.0 Hz Strain (%) Time

Measured Parameters Disc Height (mm) using AP and Lateral radiographs before relaxation and cyclic deformation tests Gross Anatomic or Degenerative Changes in the L4-5 Discs dissection of L4-5 VB-Disc-VB unit after mechancial tests Disc Cross-sectional Area (mm2) using an image processor 3-D angular displacements of L4 relative to L5, L5 relative to L7, and L7 relative to S1 in response to a 2.0 Nm

Measured Parameters E1 (equilibrium modulus); E2 (instantaneous modulus); and  (damping coefficient) using a three-parameter standard linear solid (SLS) model Relaxation Time Constant ( = /(E1 +E2)) Dynamic Stiffness (MN/m): peak-to-peak load/peak-to-peak deformation from the last three load-unload cycles Hysteresis

RESULTS

General Observations No complications were observed in the experimental dogs during the follow-up period. At sacrifice, the loosening of the rod-screw connection was observed at the sacrum level in four out of five experimental dogs.

Flexibility Testing Results Significant acute stabilization in all loading modes in the control-implant group. FLX/EXT motion of the experimental groups was statistically smaller than the intact control group, but it was nearly 12 degrees of motion. No significant difference in LB and AR motion between the experimental groups and the intact control group. No solid fusion was achieved across the L7-sacrum level.

L7-S1 Motions FLX EXT LB AR Rotation Angles (deg) Control-Intact Control-Implant Fusion alone Fusion+Implant 28 21 Rotation Angles (deg) 14 7 FLX EXT LB AR

Flexibility Testing Results Significant acute stabilization in all loading modes in the control-implant group. As compared to the intact control group, motions of the experimental groups were significantly reduced in all loading modes. Solid fusion across the L5-L7 segments was achieved at the end of 30 weeks of follow-up.

L5-L7 Motions FLX EXT LB AR Rotation Angles (deg) Control-Intact 16 4 8 12 16 Control-Intact Control-Implant Fusion alone Fusion+Implant Rotation Angles (deg) FLX EXT LB AR

Flexibility Testing Results No significant L4-L5 motion changes were found among the tested groups in all loading modes. The flexibility of the L4-5 segment (adjacent to instrumentation and fusion) was affected either immediately following instrumentation across L5-sacrum or after the 30 weeks follow-up.

L4-L5 Motions (Adjacent to Fusion) Control-Intact Control-Implant Fusion alone Fusion+Implant 12 9 Rotation Angles (deg) 6 3 FLX EXT LB AR

SLS Model Parameters (mean ± SD) Control Group (n=5) Experimental Group (n=5) E1 (MPa) 1.69 ± 0.38 1.46 ± 0.82 E2 (MPa) 3.40 ± 1.59 3.21 ± 1.99  (MPa-min) 47.5 ± 22.3 39.7 ± 22.1 No significant differences in SLS model parameters between the control and experimental groups /(E1+E2) 9.14 ± 2.55 8.99 ± 1.25

Relaxation of the L4-5 Disc

Cyclic Load-Displacement Testing Results (mean ± SD) 0.1 Hz 1 Hz Measured Quantity Control Experimental Control Experimental 1.46 ± 0.89 0.98 ± 0.54 1.65 ± 1.33 1.46 ± 0.54 13.4 ± 4.22 18.0 ± 8.75 14.4 ± 5.29 14.3 ± 5.34 Dynamic Stiffness (MN/mm) Hysteresis (%) No significant changes in measured quantities between the control and experimental groups.

Steady State Load-Unload Curves

Observations of L4-5 Discs No morphological changes as a result of instrumentation and fusion across L5-sacrum. Similar Disc Heights: 3.21 (SD 0.4) mm for the control group 3.17 (SD 0.5) mm for the experimental group Similar cross-sectional areas 303.5 (SD 53.2) mm2 for the control group 335.1 (SD 24.3) mm2 for the experimental group

Observations of L4-5 Discs No visual signs of disc degeneration were observed in both the control and experimental L4-5 intervertebral discs.

An in-vivo canine model: Discussion An in-vivo canine model: It has been frequently used for the studies of spinal instrumentation. In the dog spine, the L7-sacrum segment is most mobile in FLX-EXT as compared to the other lumbar segments. This may have contributed to the rod loosening at the sacrum level and the subsequent development of nonunion at the lumbosacral junction.

Discussion Results of this study indicate that the solid fusion across the L5-L7 segments did not alter the biomechanical properties of the adjacent segment at 30 weeks postoperatively in the canine spine. Fusion may induce changes in the biochemical and nutritional environments, that may indicate the degeneration process. However, no biochemical analysis was performed in this study.

What causes degeneration at the adjacent segment? Increased motion and loads at the the adjacent segment due to: Solid fusion at the lumbosacral junction (Elimination of the most mobile segment) vs. floating fusion Loss of lordotic curves due to instrumentation Metabolic changes due to the presence of screws in the spine Genetic factors? Other factors?

CONCLUSION Solid fusion across the L5-L7 segments but not solid fusion across the lumbosacral junction could be achieved in the dog model even using a rigid Isola pedicle screw instrumentation across the L5-sacrum levels. The flexibility of the L4-5 segment was not changed due to the instrumentation across L5-sacrum in the control as well as the experimental dogs. Fusion of the L5-L7 segments did not significantly affect the viscoelastic properties of the adjacent disc at the end of 30 weeks.

Current Findings of Spinal Instrumentation Rigid Spinal Instrumentation can enhance the solid fusion rate strong enough to allow early mobilization without serious hardware problems Construct stability (Anterior vs. Posterior Fixation) Posterior fixation is superior to anterior fixation in general. Both fixation systems can not provide the axial rotational stability beyond the intact AR stability in most cases. Need to determine the optimum stability of the surgical construct Too much rigid fixation may cause various complications particularly at the adjacent level.

Current Concepts Minimal Invasive Surgery Laprascopic surgery techniques Solid Fusion with minimum use of spinal implants BAK screw system, Metal cages Artificial bone graft use of BMP

Ongoing Debates Use of rigid or semirigid fixation Fixation in more or less lordosis Combination of posterior and anterior fusion Others

Design Factors for Improvement User friendliness fixing all components posteriorly Rigid Fixation between Components rigid connection between screw and rod (or plate), particularly important in anterior cervical fixation Adjustable Connection between Components Poly-axial screw insertion Allow adjustment in medial-lateral direction as well as in vertical direction

Devices needs to be developed Anterior graft device (Cage, Ceramic, etc.) Biological enhancement such as BMP Cervical spine fixation device Instrument for larprascopic surgery Computer-aided surgery techniques Scoliosis reduction system Artificial intervertebral joint

EFFECT OF INTERVERTEBRAL JOINT STIFFNESS CHANGES ON THE LOAD SHARING CHARACTERISTICS IN THE STABILIZED LUMBAR SEGMENT

Department of Orthopaedic Surgery Tae-Hong Lim, Ph.D. Department of Orthopaedic Surgery Rush-Presbyerian-St. Luke’s Medical Center Chicago, Illinois Vijay K. Goel, Ph.D. Department of Biomedical Engineering The University of Iowa Iowa City, Iowa

Load Sharing Ability of the instrumented segment to resist a fraction of the externally applied load Clinical Relevance of Load Sharing: Implant survival Stress shielding of the instrumented segment Fusion/healing rate and quality Load sharing characteristics varies as a function of: Spinal column stiffness Implant stiffness

LOAD SHARING MECHANISM F Spinal Instrumentation When stiffnesses of the spinal column and the screws are not changed; Plate Stiffness Fd FP Fd FP Spinal Segment

Previous Studies 20% of the axial load through the VSP plate in case of PLIF: Goel et al., 1988 10%, 20% and 23% of the axial load through the 4.76mm rods, 6.35mm rods, and VSP plates Duffield et al., 1993

Spinal Column Stiffness Changes In STVSP, Edisc of the elements in dashed area was changed; Edisc= 0.0 MPa (Total Nucleotomy) Edisc= 4.2 MPa Edisc= 8.4 MPa Edisc= 1,000 MPa (PLIF) Edisc= 2,000 MPa (PLIF) Edisc= 3,000 MPa (PLIF) Stabilization with stainless steel VSP system was maintained.

Purpose of This Study To investigate the effects of variations in the intervertebral joint stiffness as well as the implant stiffness on the load sharing characteristics across the stabilized motion segment

Finite Element Analysis METHODS Finite Element Analysis

Finite Element Models Intact L3-L4 (INT) Bilateral Fixation (STVSP & PVSP) Unilateral Fixation (UVSP)

Implant Stiffness Changes While keeping a degenerated intact disc: STVSP: L3-L4 motion segment stabilized by VSP system bilaterally PVSP: L3-L4 motion segment stabilized by non-metal plates and metal screws bilaterally Unilateral: L3-L4 motion segment stabilized by VSP system unilaterally

Boundary & Loading Conditions Boundary Conditions: All nodes in the midsagittal plane were not allowed to move in lateral direction because of assumed midsagittal symmetry, while these constraints were not imposed on UVSP. Nodes in the inferior most surface of the VB were constrained not to move in any direction. Loading Conditions: Axial compressive loads of 200, 413, and 700N Simulated by a uniform distribution of loads on the superior most surface of VB and superior facets of L3

Data Analysis FE models were executed using ANSYS, and solutions were obtained in an iterative manner. Output was processed to obtain: Stresses in various components (axial component of facet contact force) Axial forces transmitted through facets and VSP plate Fp = Plate x-are x sum of axial stresses in the middle of the plate

Axial Forces across the VSP Plates in STVSP Model Axial Forces (N) Applied Axial Compression Load (N)

Axial Forces (N) across the VSP Plates in Case of Varying Spinal Column Stiffness 100% Axial Force (N) 43% 38% 10.9% 10.1% 9.1% Edisc Variations

Axial Forces (N) Across the VSP Plates in Case of Varying Implant Stiffness 38 % 24 % Axial Force (N) 17 % 8% on facets

Average von-Mises Stresses (MPa) in Spinal Segment Cortical Bone Cancellous Bone Disc Annulus INT 1.83 (1.98) 0.11 (0.19) 0.21 (0.29) VSP 1.51 (2.04) 0.07 (0.15) 0.12 (0.18) PVSP 1.91 (2.21) 0.10 (0.18) 0.16 (0.23) UVSP 1.50 (2.03) 0.08 (0.18) 0.15 (0.25) *Maximum stresses are listed in the parentheses.

Maximum von-Mises Stresses in the VSP System Stress (MPa)

A parametric study was conducted to investigate the effect of variations in the intervertebral joint stiffness as well as the implat stiffness on load sharing characteristics across the stabilized motion segment using finite element method. Limitations: Rigid connection was assumed at bone-screw interface and metal-to-metal interfaces. Only a few limited range of variations in the spinal column and implant stiffnesses was simulated without modeling the property changes over the healing process.

Implications of Model Predictions Model Predictions for Spinal Column Stiffness Variations: Importance of preserving the IVD to keep the load on the implants low for reducing the incidence of hardware failure, particularly in case of severe discectomy Benefit of using an interbody graft from the perspective of load sharing Minimal adverse effect of a slight decrease in the graft stiffness on the load sharing when using posterior fixation High potential for subsidence in case of interbody fusion with a too stiff graft even with rigid fixation

Implications of Model Predictions Model Predictions for Implant Stiffness Variations: Greater load on the spinal column in case of using a less rigid fixation Significant stress reduction in the implant components and less stress-shielding effect with decreasing implant stiffness Further studies are required to address how to maintain the surgical construct with the use of less rigid fixation.