Fundamentals of Digital Radiology

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Comparison of Film v. Digital Image Display. Process of data capture All image recording systems rely on differential absorption within the patient to.
The Field of Digital Radiography
Quality Assurance and Digital Radiography
Digital Radiography.
RAD 354 Chapt. 28 The Digital Image Spatial resolution Contrast resolution Contrast-detail curve Pt. dose considerations.
Resident Physics Lectures
Digital Radiographic Imaging 101
1 DIGITAL RADIOGRAPHY. 2 Digital Radiography A “filmless” imaging system introduced in 1987 Digital radiography uses an electronic sensor, instead of.
Digital Image Manipulation I By Professor Stelmark.
Digital Radiography Chapter 22. History of Digital Radiography Slower process of conversion because no pressing need to convert to digital radiography.
Direct Digital Radiography or Direct Capture Radiography Bushong Ch. 27.
Dawn Guzman Charman, M.Ed., R.T. RAD TECH A
Chapter 1 Introduction to Digital Radiography and PACS
Conventional and Computed Tomography
Digital Image Characteristic
Digital Radiography.  Invention of digital imaging  Fundamentals of digital imaging & equipment  Radiation exposure  Advantages/disadvantages  Infection.
Computed Radiography and Digital Radiography
Unit III Creating the Image. Unit III Creating the Image.
Chapter 6: Digital Radiographic Imaging
Computers in Imaging Q & A Robert Metzger, Ph.D..
Quality Control.
Image Quality Radiographic Resolution.
The Digital Image Dr. John Ryan.
Digital Radiology. 2 Aim : To become familiar with the digital imaging techniques in projection radiography and fluoroscopy.Aim : To become familiar with.
DIGITAL RADIOGRAPHY DA 105.
1 Digital Radiography Fall filmless’ radiology departments Diagnostic radiographers have traded their film and chemistry for a computer mouse.
Seeram Chapter #3: Digital Imaging
George David Associate Professor of Radiology Medical College of Georgia.
Rad Tech 265 More Digital Imaging. Digital Fluoroscopy Radiation dose –Patient dose for DF is significantly less than conventional fluoro At 7.5 pulses/second.
RAD 254 Digital Imaging Basic Elements of Digital Imaging CR/DR.
Digital Imaging. acronyms 1.PSP 2.CRT 3.ADC 4.IP 5.CR 6.DR 7.PACS 8.SNR 9.CNR 10. CCD 11. FOV 12. LUT 13. DICOM 14. RIS 15. HIS 16. TFT 17. DQE 18. DAP.
Factors affecting CT image RAD
Computed Radiography (CR), & Digital Radiography (DR) Images
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
Things they never tell you when the equipment is installed REV 11/2008
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun
Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15
Computer History Earliest computer was the abacus abacus invented around 3,000 B.C. other calculating machines developed in 1600’s digital computer started.
Radiographic Quality Visibility and Sharpness
Digital Imaging.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
RAD 254 Chapter 27 Digital Fluoroscopy
CT Chapter 4: Principles of Computed Tomography. Radiography vs. CT Both based on differential attenuation of x-rays passing through body Radiography.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
CR Image Acquisition By Professor Stelmark.
DIGITAL IMAGING.
Digital Radiographic Image Processing and Manipulation Chapter 7.
Digital Radiology Dr M A Oghabian Medical Physics Group Tehran University of Medical Sciences
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
DIGITAL RADIOGRAPHY.
Direct Digital Radiography or Direct Capture Radiography
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. 1 Digital Imaging Instead of Radiography.
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS &
{ Digital Image Processing By Professor Stelmark.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Fundamentals of Digital Radiology
Computed Tomography Basics
Optimisation of Patient Protection for Radiography
Rad Tech 265.
Computed Radiography Feb.
Fundamentals of Digital Radiology
Digital Imaging CHAPTERS 1, 4-7 CARTER.
Digital Fluoroscopy PPT created by: Jed Miles, BSRS, RT(R), CRT-CA
Fluoroscopy: Static Image Recording Systems
Digital radiography.
Presentation transcript:

Fundamentals of Digital Radiology George David Medical College of Georgia

So what is “Digital”?

Filmless Department What we mean by Digital Digital Radiographs PACS Picture Archival & Communication Systems Reading from Monitors

What we really mean by Digital No more File Room!!!

Digital Image Formation Place mesh over image

Digital Image Formation Assign each square (pixel) a number based on density Numbers form the digital image 194 73 22

Digital Image Formation The finer the mesh, the better the digital rendering

What is this? 12 X 9 Matrix

Same object, smaller squares 24 X 18 Matrix

Same object, smaller squares 48 X 36 Matrix

Same object, smaller squares 96 X 72 Matrix

Same object, smaller squares 192 X 144 Matrix

Numbers / Gray Shades Each number of a digital image corresponds to a gray shade for one picture element or pixel

So what is a digital image? Image stored as 2D array of #’s representing some image attribute such as optical density x-ray attenuation echo intensity magnetization 125 25 311 111 182 222 176 199 192 85 69 133 149 112 77 103 118 139 154 120 145 301 256 223 287 225 178 322 325 299 353 333 300

Computer Storage 125 25 311 111 182 222 176 199 192 85 69 133 149 112 77 103 118 139 154 120 145 301 256 223 287 225 178 322 325 299 353 333 300 125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300

Digital Copies = If you’ve got the same numbers ... 125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300 125, 25, 311, 111, 182, 222, 176, 199, 192, 85, 69, 133, 149, 112, 77, 103, 118, 139, 154, 125, 120, 145, 301, 256, 223, 287, 256, 225, 178, 322, 325, 299, 353, 333, 300 = If you’ve got the same numbers ...

Digital Copies then you have an identical copy =

= Digital Copies Digital copies are identical All digital images are originals =

Image Matrix Doubling the matrix dimension quadruples the # pixels 125 25 311 111 199 192 85 69 77 103 118 139 145 301 256 223 111 87 118 155 2 X 2 Matrix 4 pixels 4 X 4 Matrix 16 pixels

Doubling the matrix dimension quadruples # pixels Image Matrix Doubling the matrix dimension quadruples # pixels A 10242 matrix compared to a 5122 matrix quadruples disk storage requirements image transmission time digital image manipulation Matrix # Pixels 512 X 512 => 262,144 1024 X1024 => 1,048,576 2048 X2048 => 4,194,304

Matrix Size & Resolution More pixels = better spatial resolution

The Bit Fundamental unit of computer storage Only 2 allowable values 1 Computers do all operations with 0’s & 1’s BUT Computers group bits together

Special Binary Digit Grouping Terms Nibble 4 binary bits (0101) Byte 8 binary bits (1000 1011) Word 16 binary bits (1100 0100 1100 0101) Double Word 32 binary bits (1110 0100 0000 1011 0101 0101 1110 0101)

Abbreviations Review Bit (binary digit) Byte Kilobyte Megabyte Smallest binary unit; has value 0 or 1 only Byte 8 bits Kilobyte 210 or 1024 bytes sometimes rounded to 1000 bytes Megabyte 213 or 1,048,576 bytes or 1024 kilobytes sometimes rounded to 1,000,000 bytes or 1,000 kilobytes

# of unique values which can be represented by 1 bit 2 unique combinations / values 1 2

# of unique values which can be represented by 2 bits 1 2 4 unique combinations / values 3 4

# of unique values which can be represented by 3 bits 5 1 6 2 7 3 8 4 8 unique combinations / values

Digital Image Bit Depth the number of computer bits (1’s or 0’s) available to store each pixel value Values Bits # Values 1 2 3 . 8 0, 1 00, 01, 10, 11 000, 001, 010, 011, 100, 101, 110, 111 . 00000000, 00000001, ... 11111111 2 1 = 2 2 2 = 4 2 3 = 8 . 2 8 = 256

Digital Image Bit Depth bit depth indicates # of possible brightness levels for a pixel presentation of brightness levels pixel values assigned brightness levels brightness levels can be manipulated without affecting image data window level

Bit Depth & Contrast Resolution The more bits per pixel the more possible gray shades and the better contrast resolution. 2 bit; 4 grade shades 8 bits; 256 grade shades

Computer Storage Storage = # Pixels X # Bytes/Pixel Example: 512 X 512 pixels; 1 Byte / Pixel 512 X 512 pixel array # pixels = 512 X 512 = 262,144 pixels Storage = 262,144 pixels X 1 byte / pixel = 262,144 bytes = 256 KBytes = .25 MBytes

Image Size Related to both matrix size & bit depth higher (finer) matrix requires more storage doubling matrix size quadruples image size higher bit depth requires more storage doubling bit depth theoretically doubles image size Computer may require storage in multiples of 8 bits (bytes) 10 or 12 bits stored in 16 bit slot alters image size requirements 1 2 3 4 5 6 7 8 9

Image Compression jpg gif (20) 37’s reduction of digital image storage size by application of algorithm for example, repetitive data could be represented by data value and # repetitions rather than by repeating value jpg 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37, 37 gif (20) 37’s

Image Compression Image Decompression Compression Ratio calculating original digital image from previously compressed data Compression Ratio original image size -------------------------------- compressed image size ratio depends upon data to be compressed algorithm

Compression Types Reversible Compression Non-reversable Compression Image decompresses to original pixel values Low compression ratios only Non-reversable Compression Decompressed image’s pixel values not necessarily identical to original much higher compression ratios possible variation from original image may or may not be visible or clinically significant

Non-Reversable Compression variation from original image generally increases with increasing compression ratio but a higher compression ratio means less storage requirements variation less noticeable for dynamic (moving) images than for still images such as radiographs

Computed Radiography (CR) Re-usable metal imaging plates replace film & cassette Uses conventional bucky & x-ray equipment                                                                  

CR Exposure & Readout

CR Readout

Another View: CR Operation

Computer Radiography (CR) plate is photostimulable phosphor radiation traps electrons in high energy states higher states form latent image H i g h e r E n e r g y - E l e c t r o n S t a t e P h o t o n p u m p s e l e c t r o n t o X - R a y h i g h e r e n e r g y s t a t e P h o t o n - - - L o w e r E n e r g y - - - - - - E l e c t r o n - - - - - - - - - S t a t e - - - - - - - - -

Reading Imaging Plate reader scans plate with laser laser releases electrons trapped in high energy states electrons fall to low energy states electrons give up energy as visible light light intensity is measure of incident radiation Lower Energy Electron State

Reading Imaging Plate Reader scans plate with laser light using rotating mirror Film pulled through scanner by rollers Light given off by plate measured by PM tube & recorded by computer

Laser & Emitted Light are Different Colors Phosphor stimulated by laser light Intensity of emitted light indicates amount of radiation incident on phosphor at each location Only color of light emitted by phosphor measured by PMT

CR Operation after read-out, plate erased using a bright light plate can be erased virtually without limit Plate life defined not by erasure cycles but by physical wear

CR Resolution Small cassettes have better spatial resolution Smaller pixels More pixels / mm

CR Throughput Generally slower than film processing CR reader must finish reading one plate before starting to read the next Film processors can run films back to back                

CR Latitude Much greater latitude than screen/film Plate responds to many decades of input exposure under / overexposures unlikely Computer scale inputs exposure to viewable densities Unlike film, receptor separate from viewer                

Film Screen vs. CR Latitude CR Latitude: .01 – 100 mR 100

Digital Radiography (DR) Digital bucky Incorporated into x-ray equipment

Digital Radiography (DR) Receptor provides direct digital output No processor / reader required Images available in < 15 seconds Much less work for technologist

TFT = THIN-FILM TRANSISTOR ARRAY Direct vs. Indirect TFT = THIN-FILM TRANSISTOR ARRAY

Digital Radiography (DR) Potentially lower patient dose than CR High latitude as for CR Digital bucky fragile First DR portables coming to market

Raw Data Image Unprocessed image as read from receptor CR Intensity data from PMT’s as a result of scanning plate with laser DR Raw Data read directly from TFT array Not a readable diagnostic image Requires computer post-processing Specific software algorithms must be applied to image prior to presenting it as finished radiograph

Enhancing Raw Image (Image Segmentation) * Identify collimated image border Separate raw radiation from anatomy Apply appropriate tone-scale to image Done with look-up table (LUT) This process is specific to a particular body part and projection

Image Segmentation Computer then defines anatomic region Computer must establish location of collimated border of image Computer then defines anatomic region Finished image produced by tone scaling Requires histogram analysis of anatomic region

Histogram Graph showing how much of image is exposed at various levels

Tone Scaling Post-Processing Body part & projection-specific algorithms determine average exposure Must correctly identify anatomical region LUT computed to display image with proper Density Contrast

Film/Screen Limited Latitude Film use has little ambiguity about proper radiation exposure

Should I Worry? In CR & DR, image density is no longer a reliable indicator of exposure factor control.

CR / DR Latitude DANGER Will Robinson!!! Almost impossible to under or overexpose CR / DR Underexposures look noisy Overexposures look GOOD!!!

So how do I know if exposure is optimum by looking at my image?

Exposure Index Each manufacturer provides feedback to technologist on exposure to digital receptor Displayed on CR reader monitor Displayed on workstations

Calculated Exposure Index Affected by X-Ray technique selection Improper centering of image on cassette Improper selection of study or projection Placing two or more views on same cassette Can cause image to appear dark

Phototimed Phantom Image 75 kVp 88 mAs 2460 EI

Let’s Approximately Double mAs 75 kVp 88 mAs 2460 EI 75 kVp 160 mAs 2680 EI

Let’s Go Crazy 75 kVp 88 mAs 2460 EI 75 kVp 640 mAs 3300 EI

How Low Can You Go? Cut mAs in Half! 75 kVp 88 mAs 2460 EI 75 kVp 40 mAs 2060 EI

Let’s Go Crazy Low 75 kVp 8 mAs 1380 EI 75 kVp 1 mAs 550 EI

CR Artifacts Physical damage to imaging plates Dirt in reader Cracks, scuffs, scratches Contamination Dust / dirt Dirt in reader Highly sensitive to scatter radiation

DR Artifacts Dead detector elements Spatial variations in background signal & gain Grid interference Software can help correct for above

Shifting Gears: Fluoroscopy Issues

Digital Video Sources DR type image receptor Conventional Image Intensifier with Video Signal Digitized (“Frame Grabber”) I m a g e T u b X-Ray Input Image Tube TV Amplfier Analog to Digital Converter Memory (Computer) Lens System

Digital Spot Film Frame grabber digitizes image Digital image saved by computer Radiographic Technique used required to control quantum noise

Last Image Hold Computer displays last fluoro image before radiation shut off. Image noisier than for digital spot Image made at fluoroscopic technique / intensity Allows operator to review static processes without keeping beam on ideal for teaching environments ideal for orthopedic applications such as hip pinning Less radiation than digital spot

Fluoro Frame Averaging Conventional fluoro only displays current frame Frame averaging allows computer to average current with user-selectable number of previous frames Averages current frame & history

Fluoro Frame Averaging Tradeoff Advantage: Reduces quantum noise Disadvantage Because history frames are averaged with current frame, any motion can result in lag

Other Fluoro Features Real-time Edge Enhancement / Image Filtering Option of using lower frame rates (15, 7.5, 3.75 fps rather than 30) computer displays last frame until next one reduces flicker Lowers patient and scatter exposure Exposure proportional to frame rate dynamic studies may be jumpy

Digital Subtraction Immediate replay of run Free selection of mask before or after bolus >1 frame may be averaged for mask Note subtraction adds noise

Digital Image Manipulations on-screen measurements distances angles volumes/areas stenosis image annotation peak opacification / roadmapping peak opacification displays vessels after a test injection allows visualization of live catheter on top to saved image of test injection

Digital Possibilities Multi-modality imaging / Image fusion PET/CT

DR & Energy Subtraction 2 images taken milliseconds apart at 2 kVp’s Combine / subtract images Soft Tissue Image Bone Image

DR Mobile Units GE Definium AMX 700 See image immediately Wireless transmission of images GE Definium AMX 700

Other Possibilities Tomosynthesis Histogram Equalization Multi-slice linear tomography from one exposure series Histogram Equalization Use computer to provide approximately equal density to various areas of image.

The End ?