Correcting confocal acquisition to optimize imaging of fluorescence resonance energy by sensitized emission Jacco van Rheenen Kees Jalink
outline Confocal FRET Introduction Major sources of error and variability in confocal acquisition Solutions Examples
Fluorescence Resonance Energy Transfer Energy Transfer CFPYFP Efficiency (E): (#quanta transferred)/(#quanta absorbed): E=R 0 6 /(R 6 + R 0 6 ) (Förster eq.)
FRET measurements FRET and E spatial and temporal
Energy Transfer Leakthrough complicates sensitized emission measurements Sensitized emission = acceptor fluorescence resulting from energy transfer from excited donor molecules CFPYFP Donor: CFP, excitation 430 nm Acceptor: YFP, excitation 514 nm Indirect excitation YFP Leak-through Donor CFP
PH CFP CFP emission in Indirect YFP channel CFP 2 YFP PH Energy Transfer 430 nm =CFP (in Indirect YFP Channel) CFP (in CFP Channel) YFP
CFP emission in Indirect YFP channel CFP PH CFP 2 YFP PH Energy Transfer 430 nm =CFP (in Indirect YFP Channel) CFP (in CFP Channel) YFP
Indirect YFP excitation YFP PH CFP 2 YFP PH 430 nm 514 nm =YFP (by 430 nm excitation source) YFP (by 514 nm excitation source)
514 nm Sensitized emission YFP PH CFP 2 YFP PH =YFP (by 430 nm excitation source) YFP (by 514 nm excitation source)
430 nm / Donor Excitation 430 nm / Donor 514 nm / Acceptor Emission collection CFP YFP M Donor M IndAcc M DirAcc Calculating Sensitized emission F sen =(M IndAcc -M Donor -M DirAcc ( )/(1- ) Gordon et al., 1998, Nagy et al., 1998, Hoppe et al., 2002 and van Rheenen et al., 2004
Sensitized emission and FRET efficiency 1000 CFP 1000 YFP 8 nm so 5%, 50 Sens 10 CFP 10 YFP 5 nm so 50%, 5 Sens 5 Sens / 10 CFP = Sens / 1000 CFP = 0.05
van Rheenen et al., 2004, Biophys. J. 86:
Computer saves us lots of time
Confocal vs Wide-field ConfocalWide-field Mercury lamp 2 laser lines Single detectorTwo PMT Axial resolutionNo axial resolution
1.31 Corr unCorr. 430 nm YFPSen 514 nm YFP unCorr Lateral spatial correction = YFP (430 nm excitation) YFP (514 nm excitation)
Axial spatial correction Chromatic aberration
Direct YFP Iono +Ca CFP Sensitized YFP FRET efficiency Laser fluctuation correction 2% 30 min 2% 30 min = YFP (430 nm excitation) YFP (514 nm excitation)
Procedure 1. Donor excitation, collection Donor (M Donor ) and Acceptor M IndAcc 2. Refocusing preparation to minimize chromatic aberration effects 3. Acceptor excitation, collection Acceptor (M IndAcc ) 4. Shade correction 5. Calculating correction factors 6. Calculation sensitized emission, E D and E A
Procedure 1. Donor excitation, collection Donor (M Donor ) and Acceptor M IndAcc 2. Refocusing preparation to minimize chromatic aberration effects 3. Acceptor excitation, collection Acceptor (M IndAcc ) 4. Shade correction 5. Calculating correction factors 6. Calculation sensitized emission, E D and E A
CFP Membrane Cytosol PH PIP YFP PH YFP PH CFP Energy Transfer Homogeneous FRET efficiency reveals homogeneous PIP 2 distribution CFP-PH SensEd
PIP 2 gradients? CFP-PH SensEd
EPAC, a cAMP sensor Ponsioen et al., EMBO reports 2004
Conclusion Easy way to measure FRET Great resolution Cheap way to measure FRET
M. Langeslag B. Ponsioen G. Van der Krogt K. Jalink NWO grant