External Beam Radiotherapy

Slides:



Advertisements
Similar presentations
Radiation biology and protection in dental radiology
Advertisements

Beam Therapy Equipment 2 Linear Accelerator Head.
Overview of Industrial Radiography Sources and Accidents
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
Industrial Radiography
Radiation Occupational exposures and protection A. H. Mehrparvar, MD Occupational Medicine department Yazd University of Medical Sciences.
Radiation Sources in medicine diagnostic Radiology
Introduction to Radiation Therapy
Radiation Protection in Radiotherapy
Surgery Surgery is the initial therapy for nearly all patients with brain tumors and can cure most benign tumors, including meningiomas Goal : to remove.
Factors affecting the X-Ray output
X-Ray Production & Emission
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
X-Ray Production & Emission
Industrial Radiography Basic Safety Requirements
Beam Therapy Equipment 3 Patient Treatment and Accessories.
Ch .4 Clinical Radiation Generators
Quality Control Rad T 110.
Dose Distribution and Scatter Analysis
Linac Beam.
Arif Raza Radiotherapy Department. Introduction Clinical speciality for treatment of cancer Tissue destroying procedure Use of ionizing radiation Single.
Introduction to Nuclear Medicine
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
ANALYTICAL X-RAY SAFETY User Training Centre for Environmental Health, Safety and Security Management.
Научно-практический центр протонной лучевой терапии и радиохирургии (Москва-Дубна) A SYSTEM FOR MEASUREMENT OF A THERAPEUTIC PROTON BEAM DOSE DISTRIBUTION.
Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Radiation Protection in Radiotherapy
IAEA International Atomic Energy Agency Dental Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 2(2)
INVERSE SQUARE LAW. The picture above demonstrates the typical x-ray tube used to produce a point source of x-rays. Then as radiation exits the tube it.
Using Radiation in Medicine. There are 3 main uses of radiation in medicine: Treatment Diagnosis Sterilization.
Institute for Advanced Radiation Oncology
RADIATION SAFETY Mrs. Brinston. Introduction As a healthcare worker, you know that radiation is an important tool for detecting and treating diseases.
30.1X-rays and radioactivity X-rays are a kind of electromagnetic wave of high frequency (i.e. short wavelength). Nature of X-rays radio wave micro- wave.
Brachytherapy and GYN malignancy
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
30.1 X-rays and radioactivity
Determining Radiation Intensity
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Radiation Sources in medicine diagnostic Radiology
Introduction to Radiation Therapy
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
The Radiotherapy Application NMT 232 L 8 1. Radiotherapy Generators Kilovoltage X-rays are generally used in the treatment of skin cancers and superficial.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Interaction of x-ray photons (and gamma ray photons) with matter.
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
Treatment Machines for External Beam Radiotherapy
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
Development of elements of 3D planning program for radiotherapy Graphical editor options  automated enclose of contour  correction of intersections 
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 12.1 : Shielding and X-ray room design Practical.
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
Ch 36 Radiation Protection Design
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
Electron Beam Therapy.
Ch 4. Clinical Radiation Generators
Optimisation of Patient Protection for Radiography
Khan, The Physics of Radiation Therapy, Ch7
Ch 7. Quality of X-ray Beams
Very High Energy Electron for Radiotherapy Studies
Part No...., Module No....Lesson No
Factors Effecting the Production
أجهزة العلاج الإشعاعي Clinical Radiation Generator
Resident Physics Lectures (Year 1)
Radiosurgical Management of Brain Metastases
Ch 10. A System of Dosimetric Calculations
Resident Physics Lectures (Year 1)
Hot and cold spots are common problems associated with planning:
Principles and Practice of Radiation Therapy
Presentation transcript:

External Beam Radiotherapy Radiation Sources in Radiotherapy External Beam Radiotherapy Day 7 – Lecture 4

Objective To become familiar with the radiation sources, devices and ancillary equipment used in external beam radiotherapy.

Contents Treatment planning systems; Radiotherapy simulators; Superficial / orthovoltage units; Cobalt-60 units including Gamma-knife; Linear accelerators; Computed Tomography (CT) scanners for radiotherapy; Multileaf Collimators (MLC).

Clinical Objectives To deliver a dose and dose distribution that is adequate for tumor control but which also minimizes complications in normal tissue. Note: It is not the role of the Regulatory Body to evaluate the clinical decisions of medical practitioners authorized to prescribe radiotherapy treatments. MODIFIED

Very important for optimization of protection in medical exposures Treatment Planning Prescription Very important for optimization of protection in medical exposures Planning Planning is the most important link between clinical intention and the practical treatment. Treatment

Treatment Planning About 1/3 of problems are directly related to treatment planning; Problems may affect an individual patient or cohort of patients. IAEA Safety Report Series 17; 2000 “Lessons learned from accidental exposures in radiotherapy”

Radiotherapy Simulator

External Beam Equipment Therapeutic x-ray equipment operates in the range of: 10 kVp - 150 kVp (superficial); 150 kVp - 400 kVp (orthovoltage / deep); Radioactive sources ( γ ray equipment). Cobalt 60 & Caesium 137 Megavoltage electron accelerators for X and electron therapy Linear accelerator

Typical Radiation Levels Cobalt-60 teletherapy Source activity may be around 400 TBq (~10,000 Ci); Average radiation leakage (beam off) should not exceed 0.02 mGy/h at 1 m i.e. it would take 50 hours exposure for 1 mSv; In general, minimize the time spent in the treatment room.

Typical radiation levels (cont) Linear accelerator turned off There is no useful radiation beam when turned off; However, immediately after higher energy beams (> 10 MeV) are turned off there may be induced radioactivity but typically with very short half lives (seconds to minutes); It is suggested that room entry be briefly delayed, especially after long exposures.

Superficial and Orthovoltage x-ray equipment 40 kVp to 120 kVp Orthovoltage (“deep”) 150 kVp to 400 kVp treat small skin lesions to a depth of ~ 5 cm maximum applicator size typically < 7 cm diameter typical SSD < 30 cm beam quality (HVL) typically 0.5 to 8 mm Al treat skin lesions, bone metastases to a depth of ~ 20 cm use applicators or diaphragm SSD 30 to 60 cm beam quality (HVL) typically 0.2 to 5 mm Cu

Superficial x-ray equipment Interlocks prevent inappropriate combinations of kVp and filtration. Electron contamination from the applicator can be significant.

Superficial x-ray equipment (cont) Dose is highly dependent on source-skin distance, filtration and applicator area.

Superficial x-ray equipment (cont) Provides a range of kVp, mA and filtration Filters are used to absorb low energy photons which otherwise may unnecessarily increase skin dose.

Issues with Superficial radiotherapy Short focus to skin distance (FSD) and hence high output and large influence of inverse square law Calibration difficult due to strong dose gradient i.e. dose fall off and electron contamination The lecturer may particularly emphasize the first point. The strong dose gradients are due to FSDs of 10 to 25cm in practice. Shorter the treatment distance higher the influence of inverse square law and hence to ensure accurate dose delivery the treatment distance should be accurate Calibration of beams is typically done at surface, therefore electron contamination from lead glass cones can be of concern.

Issues with superficial therapy Dose determined by a timer on/off effects must be considered Photon beams may be contaminated with electrons scattered from the applicator Control panel

Orthovoltage (deep) x-ray equipment

Deep X-ray therapy (Orthovoltage) Uses conventional X-ray tube Energy range 150- 400 kV X-rays Mostly used around 250 - 300 kVp Treatment depths of around 20 mm Applicators are used in superficial therapy In order to reach deep seated tumours higher energy x-rays of 150-400 kVp were used. As it could reach deeper tissue than superficial x-rays were termed deep x-ray therapy. The usual depth of treatment with this is about 20 mm. Applicators / cones are used in this type of treatment also. X-ray tube Applicator

Deep X-ray therapy (Orthovoltage) Penetration sufficient for palliative treatment of bone lesions relatively close to the surface (ribs, spinal cord) Largely replaced by megavoltage treatment modalities for treatment of other lesions Deep X-ray therapy was used for treatment close to the surface such as ribs and spinal cord. However once the mega-voltage therapy was started, this was withdrawn from use.

Disadvantages of deep x-ray Higher dose to bone - photoelectric absorption Maximum dose on the surface hence higher skin dose Treatment to a depth of only a few centimeters possible Low energy, hence high scattered radiation and larger penumbra The lecturer may point out the drawbacks of using low energy x-rays for the treatment of deep seated tumours.

Gamma ray equipment

Gamma ray equipment (cont) Source head and a typical source transfer mechanism

Why prefer Cobalt unit over Orthovoltage? The above table gives a comparison of Cobalt with an orthovoltage unit (Deep X-ray therapy unit. The lecturer may use this slide to further explain why Telecobalt units are preferred over deep x-ray therapy units.

GAMMA KNIFE The gamma knife device contains 201 cobalt-60 sources of approximately 30 curies each It is placed in a circular array in a heavily shielded assembly. The device aims gamma radiation through a target point in the patient's brain. The patient wears a specialized helmet that is surgically fixed to their skull so that the brain tumor remains stationary at target point of the gamma rays. Therefore it is also known as the stereotactic surgery.

Patient positioning collimator Gamma Knife The Gamma Knife: uses numerous high activity 60Co sources positioned in a device so that the radiation beams converge at the specified point of treatment; is used to treat head tumors. Patient positioning collimator

Linear Accelerator Modern accelerators have a number of treatment options e.g. X-rays or electrons (dual mode); 2 X-ray energies; 5 or more electron energies.

Linear Accelerator (cont) Concept

Linear Accelerator (cont) Radiation exposure: is controlled by two independent integrating transmission ionization chamber systems; one of these is designated as the primary system and should terminate the exposure at the correct number of monitor units; these also steer the beam.

Linear Accelerator (cont) the other system is termed the secondary system and is usually set to terminate the exposure after an additional 0.4 Gy; most modern accelerators also have a timer which will terminate the exposure if both ionization chamber systems fail.

Linear Accelerator (cont) Complex head structure to handle multiple energies and multiple modalities.

Linear Accelerator (cont) Complex control system

Linear Accelerator (cont) Verification systems All accelerator manufacturers produce computer controlled verification systems which provide an additional check that the settings on the accelerator console: are correct for proper accelerator function; and correspond exactly with the parameters determined for the individual patient during the treatment planning process

Linear Accelerator (cont) X-ray Collimators Rectangular (conventional) The transmission through the collimators should be less than 2% of the primary (treatment) beam. Multi-leaf collimators (MLC) the transmission through the collimators should be less than 2% of the primary (treatment) beam. The transmission between the leaves should be checked to ensure that it is less than the manufacturer’s specification.

Linear Accelerator (cont) Electron applicators may be: open sided for modern accelerators using double scattering foils or scanned beams; enclosed for older accelerators using single scattering foils. Both types should be checked for leakage: adjacent to the open beam; on the sides of the applicators.

Linear Accelerator (cont) Neutrons:- should be considered if the x-ray energy is greater than 10 MV Issues which need to be considered when neutrons are presents include: neutron activation shielding problems

A comparison: Cobalt unit Vs Linac The above table gives a comparison of the Telecobalt unit and Linear accelerator.

General Safety Requirements Clear indication shall be provided at the control console and in the treatment room to show when the equipment is in operation. Dual interlocks shall be provided on all doors to the treatment room such that opening a door will interrupt the treatment. It should only be possible to resume treatment from the control console.

General Safety Requirements Warning Signals and Signs

General Safety Requirements “Fail safe” systems There shall be at least two independent “fail safe” systems for terminating the irradiation. These could be: two independent integrating in-beam dosemeters; two independent timers; an integrating dose meter and timer. Each system shall be capable of terminating the exposure.

General Safety Requirements Collimation The exposure shall be limited to the area being examined or treated by the use of collimating devices aligned with the radiation beam. Exposure rates outside the examination or treatment area due to leakage or scatter shall be kept as low as reasonably achievable.