Background information for users of STIS Charles R. Proffitt.

Slides:



Advertisements
Similar presentations
A NEW CTE PHOTOMETRIC CORRECTION FORMULA FOR ACS Marco Chiaberge TIPS meeting 05/16/2012.
Advertisements

SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Science Calibration & Instrument Status TIPS 20 Nov 2003 Last COS TIPS Aug 2003.
Post-SM4 Sensitivity Calibration of the STIS Echelle Modes Azalee Bostroem With help from: Ralph Bohlin, Alessandra Aloisi, Charles Proffitt, Kenneth Hart,
Spectroscopy of Saturn
Echelle spectra reduction with IRAF*
Chandra X-Ray Observatory CXC Paul Plucinsky EPIC Cal Nov ACIS Calibration: Planned Updates & Future Issues 2.ACIS Operations: Controlling the.
STIS 1 ST Order Spectroscopic Point Source Flux Calibration Charles R. Proffitt STScI and CSC.
NICMOS IntraPixel Sensitivity Chun Xu and Bahram Mobasher Space Telescope Science Institute Abstract We present here the new measurements of the NICMOS.
Memorandam of the discussion on FMOS observations and data kicked off by Ian Lewis Masayuki Akiyama 14 January 2004 FMOS Science Workshop.
1 TIPS 2011 May Persistence in the WFC3 IR detector Knox S. Long.
NGC 6217 in DSI mode with F658N Mosaic before new superbias why we don’t release images until after SMOV! ACS CCD monitoring and pipeline calibration review.
TIPS presentation 19 July 2007Charles Proffitt COS-STIS TeamPage 1 STIS FUV MAMA Dark Current Charles Proffitt COS-STIS Team.
Beating Noise Observational Techniques ASTR 3010 Lecture 11 Textbook.
CCDs in space: the effects of radiation on Hubble’s Advanced Camera for Surveys (ACS) Max Mutchler, David Golimowski (Space Telescope Science Institute),
STIS Closeout Plan Paul Goudfrooij 2005 HST Calibration Workshop, 10/26/2005.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA TIPS / COS TV Update 21 December 2006 Agenda Agenda Testing Setup and Status Testing Setup.
CCD Detectors CCD=“charge coupled device” Readout method:
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Status FUV Detector “1-bounce design” NUV Detector HST aberration fully-corrected Calibration.
NIRSpec Operations Concept Michael Regan(STScI), Jeff Valenti (STScI) Wolfram Freduling(ECF), Harald Kuntschner(ECF), Robert Fosbury (ECF)
ACS Drizzling Overview J. Mack; DA Training 10/5/07 Distortion Dither Strategies MultiDrizzle ‘Fine-tuning’ Data Quality Photometry.
 PLATO PLAnetary Transits & Oscillations of stars Data onboard treatment PPLC study February 2009 on behalf of Reza Samadi for the PLATO data treatment.
TIPS COS Overview - Keyes : 18 May 2000 COS Overview v CDR: April 2000 v FUV delivery: January 2001 v Environmental (Thermal-Vac): spring 2002 v Launch:
The Status of COS Flat Fields Tom Ake TIPS 21 August 2008.
Selection of the New COS/FUV Lifetime Position Cristina Oliveira Jan TIPS Meeting - COS/FUV Lifetime1.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA TIPS: STIS Report Paul Goudfrooij 1.Unusual Target ACQ Failures: Update & Resolution 2.Calibration.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Pipeline Calibration Outline of CALCOS Processing Steps and Keywords Reference Files Associations.
Performance of the Space Telescope Imaging Spectrograph after SM4 Charles R. Proffitt, A. Aloisi, K. A. Bostroem, C. R. Cox, R. I. Diaz, W. V. Dixon, P.
1-D Flat Fields for COS G130M and G160M Tom Ake TIPS 17 June 2010.
Two-Gyro Science Impact and Observer Information Ken Sembach STUC Meeting 18-November November-2004.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Monthly Status Review 18 January 2007.
Observational Astrophysics I
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Pipeline Calibration Goals of CALCOS Association Table Input and Output Files High Level.
MIRI Dither Patterns Christine H Chen. Dithering Goals 1.Mitigate the effect of bad pixels 2.Obtain sub-pixel sampling 3.Self-calibrate data if changing.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA COS Monthly Status Review 18 January 2007.
TIPS COS Status: SMOV update III STScI/CU COS Team 17 September 2009.
COS Training Series II. Optimizing Observations --- David Sahnow February mm.
TIPS - Oct 13, 2005 M. Sirianni Temperature change for ACS CCDs: initial study on scientific performance M. Sirianni, T. Wheeler, C.Cox, M. Mutchler, A.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA SMOV4 Requirements Review Cosmic Origins Spectrograph Scott D. Friedman STScI 30 July 2003.
MOS Data Reduction Michael Balogh University of Durham.
Practical applications: CCD spectroscopy Tracing path of 2-d spectrum across detector –Measuring position of spectrum on detector –Fitting a polynomial.
NICMOS Calibration Challenges in the Ultra Deep Field Rodger Thompson Steward Observatory University of Arizona.
STIS Closeout Plan Paul Goudfrooij Space Telescope Users Committee, 11/18/2004.
1 JULIA ROMAN-DUVAL & COS/STIS TEAM AUGUST 12, 2014 MOVING COS/FUV TO LIFETIME POSITION 3 See also Poster by Proffitt et al.
14 January Observational Astronomy SPECTROSCOPIC data reduction Piskunov & Valenti 2002, A&A 385, 1095.
11-Jun-04 1 Joseph Hora & the IRAC instrument team Harvard-Smithsonian Center for Astrophysics The Infrared Array Camera (IRAC) on the Spitzer Space Telescope.
NIRISS NRM bad pixel tolerance analysis David Lafrenière 2012 February 21.
The Status of HST/COS and HST/STIS Cristina Oliveira 8/12/14COS and STIS Cal Workshop1 SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA.
SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA STIS Status Report Paul Goudfrooij Group News / Reorganization Unusual Target ACQ Failure Items.
RAW DATA BIAS & DARK SUBTRACTION PIXEL-TO-PIXEL DQE CORR. LOCATE EXTR. WINDOW THROUGHPUT CORRECTION (incl. L-flat, blaze function, transmission of optics,
CCD Image Processing: Issues & Solutions. CCDs: noise sources dark current –signal from unexposed CCD read noise –uncertainty in counting electrons in.
ACS WFC Flat-Field Changes Temperature change from -77 C to -81 C on July 4, 2006 leads to expected changes for flat fields. Are L-flat measures stable.
1 Notes about eis_prep Alessandro Gardini University of Oslo Oslo, July 9th, 2009.
STIS Status TIPS, September 17, 2009 Charles Proffitt for the STIS team.
The STIS NUV-MAMA objective prism … … and looking beyond for HST UV slitless spectroscopy Jes ú s Ma í z Apell á niz HST Calibration worskhop 26 October.
In conclusion the intensity level of the CCD is linear up to the saturation limit, but there is a spilling of charges well before the saturation if.
Application of a Charge Transfer Model to Space Telescope Data Paul Bristow Dec’03
A. Ealet Berkeley, december Spectrograph calibration Determination of specifications Calibration strategy Note in
From the NGSL to Absolute Flux Sara Heap, NASA/Goddard Space Flight Center Don Lindler, Sigma Space Corporation Phase 1: NGSL observations + in situ calibration.
CCD Calibrations Eliminating noise and other sources of error.
Single Object Slitless Spectroscopy Simulations
A.Zanichelli, B.Garilli, M.Scodeggio, D.Rizzo
NAC flat fielding and intensity calibration
Single Object & Time Series Spectroscopy with JWST NIRCam
NIRSpec pipeline concept Guido De Marchi, Tracy Beck, Torsten Böker
Charge Transfer Efficiency of Charge Coupled Device
COS FUV Flat Fields and Signal-to-Noise Characteristics
ESAC 2017 JWST Workshop JWST User Documentation Hands on experience
Summary Single Object & Time Series Spectroscopy Jeff Valenti JWST Mission Scientist Space Telescope Science Institute.
UVIS Calibration Update
UVIS Calibration Update
Presentation transcript:

Background information for users of STIS Charles R. Proffitt

STIS Presentation 2 Outline of Topics More on Bright and Faint limits with STIS Calibration Lamps Wavelength Calibration Target Acquisitions CCD Operations and Characteristics MAMA Characteristics Time Resolved Observations Observing Overheads Summary of Data Products Spatial Undersampling of STIS Data A few selected data artifacts

STIS Presentation 3 Bright Object Limits STIS IHB gives tables of worst-case limiting magnitude as a function of grating and source spectrum. Normalization can vary enormously, depending on grating, aperture, source SED, reddening, etc. Cool stars especially tricky  NUV flux very sensitive to all stellar parameters esp. metallicity  FUV flux often dominated by chromospheric emission  Not included in Kurucz or other photospheric models  Strongly affected by stellar activity Epsilon Eri: Kurucz model vs. observed spectrum

STIS Presentation 4 STIS Spectroscopic BOP Limits Limit for CENWAVE with highest countrate Assumes slitless 1st order; 0.2X0.2 for echelles

STIS Presentation 5 Bright and Faint Limits - example Example: bright and faint limits for an A0 star  Faint limit defined as S/N=10 in one hour  For CCD bright limit will saturate CCD in 0.1  For MAMA bright limit determined by local or global BOP limits

STIS Presentation 6 Approximate Bright and Faint Limiting Mag for A0V star at a single wavelength using typical clear apertures (don’t take exact numbers too seriously) GratingWavelengthMag to give S/N=10 in 1 h Bright limit mag delta G750L G750M G430L G430M G230LB G230MB G230L G230M G140L G140M E230M E230H E140M E140H NUV-PRISM

Calibration Lamps

STIS Presentation 8 STIS Calibration Lamps Cal Insert Platform  Flatfielding lamps  Tungsten (4 lamps)  Krypton ( nm)  Deuterium ( nm)  Echelle wavelength cal  PtCr/Ne (LINE) Cal insert mechanism (CIM) blocks external light & acts as additional external shutter Hole in the Mirror (HITM)  PtCr/Ne ( HITM1/2)  1st order wavecals  Locate aperture during target ACQ

STIS Presentation 9 Flat fielding Lamps For small scale pixel-to-pixel flat fielding Krypton for FUV Deuterium for NUV Tungsten for CCD  Also used for IR fringe flats for G750L & G750M at > 7500 Å

STIS Presentation 10 LINE and HITM lamps spectra Low dispersion STIS spectra of LINE and HITM1 lamps

Wavelength Calibration

STIS Presentation 12 Wavelength calibration Causes of Wavelength mis-alignments  MSM positioning does not repeat exactly.  Projection of target/aperture shifts by a few pixels  Thermal flexure of STIS bench can also shift projection on detector by a couple of pixels  Any drift/mis-centering of target in aperture will cause corresponding shift in wavelength scale

STIS Presentation 13 Wavelength calibration  Wavecal observations must be adjacent to science  No intervening MSM motions because of non-repeatability  Recommend repeating wavecals every 40 minutes  Slit-to-slit alignment & repeatability is good  No need to use same slit for science and wavecal  For best alignment do ACQ/PEAK in small aperture G430L Wavecal with 52X0.2 Aperture - Aperture bars allow offsets in cross dispersion direction to also be determined. E140M Echelle Wavecal

STIS Presentation 14 Wavelength calibration AUTO-Wavecals meet needed requirements  May not always schedule at most efficient time AUTO-WAVECALS may be turned off for visit  GO-WAVECALS may then be specified by observer  No automatic enforcement of timing requirements for GO- WAVECALs

Target Acquisitions

STIS Presentation 16 Need for STIS Onboard Acquistions With GSC1 typical rms pointing errors were ~ 1”  GSC2 is more accurate - 0.1”-0.3” accuracy expected Many STIS apertures smaller than this Pointing errors along dispersion direction, translate directly to wavelength errors Basic STIS ACQ procedure desiged to centroid to ~1/5 CCD pixel or about 0.01”

STIS Presentation 17 Target ACQ exposures ACQ procedure does the following: 1) Images target using 5”x5” subarray 1 2) For point source ACQ, use flux weighted centroid around brightest 3x3 checkbox (extend source algorithm also available). 3) Move spacecraft to put target at reference location on CCD 4) Re-image target 1 & centroid again 5) Image reference aperture using HITM1 lamp & locate aperture 6) Move spacecraft to put target at center of 0.2X0.2 reference aperture First image Second image Lamp image of 0.2X0.2 aperture 1 Each external ACQ image is actually made from 2 subarray images dithered by 3 pixels in x and y. They are shifted into alignment and then combined by taking the minimum value at each pixel to eliminate cosmic ray hits and hot pixels.

STIS Presentation 18 Selecting Target ACQ parameters For point-source ACQ exposure S/N > 40:1 suggested  More is better. If 40:1 SN needs < 0.1 s minimum exposure time, see if 0.1 s exposure is unsaturated before switching to less sensitive setting.  But don’t let ACQ saturate. Allowing central pixel to overfill and bleed along columns may affect centroid in y direction. If there are multiple close stars, be sure which one will be brightest in chosen ACQ filter. Point source ACQ accurate to 1/5 pixel or 0.01” Diffuse source ACQ algorithms also available Larger checkboxes (up to 101x101 pixels or ~ 5”x5”) Choice of flux Weighted or geometric centering

STIS Presentation 19 ACQ/Peak exposures Peakups recommended for apertures ≤ 0.1” in size Do after ACQ Always done using CCD Peakups measure flux through small aperture and move spacecraft to maximize flux Need to peakup in both directions for small & short apertures (0.1X0.09) Special procedures for 0.1X0.03 peakups Peakups can use images or dispersed light Accuracy ~ 5% of slit width

STIS Presentation 20 Fixing Orientation on Sky STIS long slit can be oriented to put extended or multiple targets in aperture Orient in APT should be (degrees east of N) + 45 Usually 180 degree alternative is just as good

STIS CCD Operations

STIS Presentation 22 CCD Operations STIS CCD Format AMP D Bias and dark correction  Daily dark and bias observations and more intensive pre-and post anneal observations used to create weekly superbias and superdark images used for OTFR pipeline reduction.  Super-bias image subtracted from science image.  Serial and parallel overscan regions used to provide 2D correction to bias levels of image.  Superdark is subtracted from science image.  For side 2 data, superdark scaled for CCD housing temperature CCD includes physical and virtual overscan regions. Four amps, but most science uses AMP D.

STIS Presentation 23 CCD Operations Science data also divided by pixel-to-pixel flat field images based on data collected in yearly campaigns. Some models also have low order flat field images to correct for vignetting. Monthly anneals warm CCD from ~-85 to ~ +5 C  Heals ~80% of transient hot pixels;  Increasing numbers of permanent ones accumulate.

STIS Presentation 24 CCD Dark Current & Hot Pixels Initial dark current low: median value ~ e - /s  Extrapolation predicts cnts/pixel/s for Cycle 17 Increased over time due to radiation damage On side-2 no closed loop T control  CCD temperature & dark current varies with T  Use housing temperature to scale dark current before dark subtraction  Inexact scaling is an additional source of noise  Monthly anneal (warm from -85 C to + 5C) to heal hot pixels

STIS Presentation 25 CCD Read Noise AMP D has always had lowest read-noise and is used for science  At Gain=1, read-noise initially ~ 4 e -  Increased to 4.5 e - after SMOV3a  After switch to STIS side-2, additional kHz electronic noise increased read noise to ~ 5.5 e - (herring bone pattern) careful Fourier filtering can sometimes remove this  Gain=4 showed pick-up noise even on side-1  ~7.3 e - on side 1; ~ 7.7 e - on side 2 From STIS ISR By Tom Brown

STIS Presentation 26 CCD Options Gain: 1, 2, 4, or 8  Only Gains values of 1 and 4 supported for GO observations.  Gain=1 has lower read-noise, but amps saturate at ~33,000 e -.  Gain=4 has higher read-noise, but allows full well of CCD to be used (144,000 e - at center, ~ 120,000 e - at edges)  In saturated GAIN=4 images, electrons bleed to other pixels (perpendicular to dispersion direction), but are not lost. Total response remains linear, allowing very high S/N with special processing techniques. Binnng at readout by 1, 2, or 4 in either axis or both  Binning data during read-out reduces read-noise and file size  Increases impact of bad pixels and cosmic rays.  With older, noisier detector, usually not worthwhile

STIS Presentation 27 CCD Options - cont CCD Sub-arrays  Can save only part of image array on read-out  Reduces file size and number of buffer dumps required  Decreases readout time, allowing increased cadence.  For GOs only support reducing AXIS2 size (perpindicular to dispersion direction)  Discards virtual overscan in parallel direction, but retains physical overscan in serial direction to aid in bias removal  Lack of virtual overscan does make bias subraction more difficult  Reducing AXIS1 is an available-but-unsupported mode.  Reducing both discards all overscan regions, greatly increasing difficulty of accurate bias removal.  Different clocking patterns used by any CCD sub-arrays may introduce artifacts, and invalidate assumptions of empirical CTI corrections algorithms. Cosmic Ray rejection is normally done by taking multiple images.  CRREJECT=2 is default AXIS1 AXIS2

STIS Presentation 28 CCD Charge Transfer Inefficiency During parallel transfers some electrons get trapped Trapped e - be released later during read, causing extended “tail”. Number of free traps depends on flux level that has moved through that pixel. CTI gets worse with increasing radiation damage No sufficient pixel based physical model, so need empirical corrections. Loss increases with # of transfers (1-CTI) n  Putting target near readout amp reduces losses  E1 positions defined near row 900  Typical exposures of faint targets with the STIS CCD in cycle 17 might experience 20-30% CTI losses when target at center of the detector, but only 5-8% if at E1 near row 900.

STIS MAMAs

STIS Presentation 30 FUV MAMA Dark Current FUV MAMA initially had very low dark current (7x10-6 counts/lo-res-pixel/s), but occasionally showed enhanced glow. Initially glow present only rarely  became more frequent over time Lower edge & lower right hand corner remains mostly dark (near original 7 x counts/lo-res-pixel/s). Physical basis of FUV dark current glow is unclear

STIS Presentation darks Apr 1997 Aug s Mean 0.21 Glow 0.37 D.C x c/p/s (hi-res-pixel) 2048 x darks Aug 1998 Nov s Mean 0.45 Glow 1.08 D.C x c/p/s 141 darks Dec 1999 May s Mean 0.56 Glow 1.24 D.C x c/p/s 125 darks May 2003 Aug s Mean 0.65 Glow 1.65 D.C x c/p/s Dark Corner Glow region

STIS Presentation 32 FUV MAMA Dark Current FUV MAMA dark current increases ~ linearly with time since HV turn-on Increases faster at higher T Rate of increase has gone up over the years Hot pixels also increasing

STIS Presentation 33 FUV MAMA Dark Current Mitigation strategies  Use only first orbit of each SAA free period for observations that need low dark current. (only 1 orbit per day).  Keep FUV HVPS off when detector not in use (ops change).  Cool detector (NUV MAMA off).  Place target on darker part of detector.  New D1 aperture position defined near bottom edge of detector The count rate summed in each column over a seven pixel high region of the mean dark image covering the period between May 2003 and August The dotted line gives the results for a region near the standard 1st order spectral location, and the solid line gives the results at the new D1 position located near the bottom edge of the detector.

STIS Presentation 34 NUV MAMA Dark Current NUV MAMA dark current dominated by a different physical mechanism than the FUV MAMA Meta-stable states with lifetimes of days to weeks are populated by high-energy particle impacts, leading to a phosphorescent window glow. Long term trend depends on low-earth orbit radiation environment

STIS Presentation 35 NUV MAMA Dark Current Effect of temperature changes on dark current is complex Short term changes lead to a large increase in the dexcitation rate, leading to a large, but temporary, increase in the dark current, including daily cycling as MAMA warms up. Over the long term, a smaller equilibrium number of populated states partially balances the higher excitation rate caused by higher average T. If detector cold for long time, large but temporary increase until a new equilibrium is reached.

STIS Presentation 36 MAMA Pipeline Dark Images Low dark rates require averaging hundreds of images to make useful dark image. NUV darks semi-empirically scaled for time and temperature changes and subtracted in pipeline.  Secular changes are seen in shape of NUV dark current over time. Unpredictable nature of FUV glow makes subtracting it in OTFR pipeline impractical  Only base dark current and hot pixels subtracted by pipeline - users need to to custom extraction of glow.  In background limited observations, FUV hot pixels should just be masked out because poor statistics makes subtraction difficult.

STIS Presentation 37 MAMA Flat Fields On-orbit lamp images used to provide MAMA pixel-to- pixel flats collected during occasional campaigns.  MAMA flats very stable once data binned to lo-res (1024x1024)  Can use same pixel-to-pixel flats for essentially all data.  Flat fielding of unbinned 2048x2048 hi-res images not repeatable - significant structure remains  hi res mostly useful for filtering out hot pixels. Low order MAMA flatfields provided for selected modes (mostly FUV modes).

STIS Presentation 38 MAMA Observation Modes ACCUM mode  Keeps track of how many events fall on each pixel.  For medium and high dispersion modes, the pixel locations are corrected for spacecraft doppler motion as image is accumulated.  STIS data buffer can hold  1 hires (2048x2048) image or  up to 7 lowres MAMA + CCD full frame images or  1 hires image + 3 lowres or CCD full frame images  Hires format default for MAMA science, lores for wavecals

STIS Presentation 39 MAMA Observation Modes TIME-TAG mode  Records x-y location and time of each event with 125 micro- second resolution.  Corrections for spacecraft Doppler motion done on ground, not on spacecraft  STIS buffer divided into two sections for time-tag  Each half of buffer can hold 2 x 10 6 events.  One half of buffer can be dumped while other half is recording.  User must predict rate and specify buffer time so that buffer is dumped before one half fills, otherwise gaps will appear in sequence.  If global rate < 20,000 counts / s, continuous observations can be sustained for extended periods (up to 30 buffer dumps). For some projects needing time resolved data, a series of ACCUM observations may be better than time-tag mode.  For CCD observations, the use of subarrays may increase cadence.

STIS Presentation 40 Time resolved STIS Observing DetectorMode Minimum Sample time (texp) Time between samples (dt) Max time for uninterrupted time series MAMAsTime-tag 125  s 0 6e7/R s for R < 20,000 cnts/s 4e6/R s for R > 20,000 cnts/s (R=global count rate) MAMAsHi-res ACCUMs 0.1 s 30s if texp > 3 m 2 m if texp < 3 m No limit MAMAsLo-res ACCUMs 0.1 s 30s if texp > 3 m 1 m if texp < 3 m No Limit 7 CCDFull Frame ACCUMs 0.1 s45 sNo limit for texp > 3 m (texp + dt) x 7 CCD1060x32 subarray ACCUMs 0.1 s20 sNo limit for texp > 3 m (texp + dt) x 256

STIS Presentation 41 Other MAMA Constraints STIS MAMAs cannot be used in any SAA impacted orbit  Optical isolators scintillate from cosmic rays and can cause random bit flips in MAMA electronics  STIS low & high voltage turned off during deepest SAA passages; not practical to turn on MAMA for only part of individual orbits.  Allows use during only one ~ orbit block per day  Observers required to separate CCD and MAMA science observations into separate visits when practical

STIS Presentation 42 Summary of Overheads

STIS Presentation 43 STIS Data Products Selected STIS data file types: opppvvnnd_tag.fits - table of time tag events opppvvnnd_raw.fits - 2d image of unproccesed data opppvvnnd_flt.fits - flat fielded image opppvvnnd_crj.fits - cosmic ray rejected image (CCD) opppvvnnd_x1d.fits - fits table with 1D extracted spectra opppvvnnd_sx1.fits - 1D spectra from summed images opppvvnnd_x2d.fits - 2D spectral image (rectified and flux calibrated) opppvvnnd_sx2.fits - 2D spectral image (rectified and flux calibrated) from summed images

STIS Presentation 44 1D spectral extraction In 1D spectral extraction, an extraction box is centered on spectrum, and summed over cross dispersion direction at each pixel in dispersion direction. Extracted spectrum is then background subtracted and flux calibrated  Corrections for aperture throughputs, time-dependant sensitivity changes and CTI losses (CCD only) are applied. Geometry for extraction of 1st order STIS spectra

STIS Presentation 45 1D spectral extraction - cont For echelle modes, a separate 1D extraction is done for each spectra order Background subtraction is done using a special algorithm that models the scattered light (see STIS ISR by Valenti et al

STIS Presentation 46 2D spectral extraction Image rectified so that wavelength and spatial scales are linear and aligned with x and y coordinates. X2d image is flux calibrated (science images only)  Corrections for aperture width and time-dependant sensitivity changes are applied (no CTI correction).

STIS Presentation 47 Spatial Undersampling of STIS Critical sampling of PSF requires about 2 pixels per PSF FWHM  STIS CCD spatial scale of ~0.051”/pixel undersampled by 5000 Å.  STIS MAMA spatial scale ~0.0245”/pixel undersampled by 2500 Å. Undersampling can produce artifacts when extracting spectra at small spatial scales (affected by tilt of spectrum on detector) Dithering along long slit to sub-sample spatial scale recommended if spatial structure is significant.

STIS Presentation 48 Selected Data Artifacts CCD window reflections. Brightest ring about 1% of flux Some MAMA modes also show imaging ghosts. Airy rings produce spectroscopic fringes IR fringing due to multiple reflections in CCD. Need contemporaneous tungsten fringe flats to correct properly.