W. A. Maxwell, MD, PhD ASCRS 2008 Comparison of the Optical Image Quality for Presbyopia Correcting IOLs using Modulation Transfer Function Testing W.

Slides:



Advertisements
Similar presentations
Chapter 4 Companion site for Light and Video Microscopy Author: Wayne.
Advertisements

1 Comparison of Distance Image Quality with Accommodating and Multifocal Aspheric Presbyopia-Correcting IOLs John F. Doane, M.D., F.A.C.S. Discover Vision.
Theoretical Optical Performance of an Equal Conic Intraocular Lens and Comparison to Spherical and Aspheric IOLs Edwin J. Sarver, PhD.
Dr H. Razmjoo Isfahan University of Medical Sciences Multifocal IOLs
LIGHT AND THE RETINAL IMAGE: KEY POINTS Light travels in (more or less) straight lines: the pinhole camera’s inverted image Enlarging the pinhole leads.
Blizard Advanced Light Microscopy club Making friends with your microscope.
Foundations of Physics
R. Lehmann, MD ASCRS 2008 Clinically Relevant Advantages in the Functional Performance of the AcrySof ® IQ IOL Robert P. Lehmann, MD, FACS Lehmann Eye.
The Eye. Eye Dissected  The eye is an optical lens system. Two single lensesTwo single lenses  The goal of the eye is to focus an image on the retina.
Ch 25 1 Chapter 25 Optical Instruments © 2006, B.J. Lieb Some figures electronically reproduced by permission of Pearson Education, Inc., Upper Saddle.
... M A K E S Y O U R N E T W O R K S M A R T E R Lenses & Filters.
Multifocal Intraocular Lenses & Contrast Sensitivity
Aspheric IOL’s: clinical benefits and customizing cataract surgery Bojan Pajic, MD, PhD, FEBO Swiss Eye Research Foundation, ORASIS, Reinach, Switzerland.
Myoung Joon Kim, MD Len Zheleznyak, MS2
1 OPTICAL QUALITY OF DUAL- OPTIC ACCOMMODATING INTRAOCULAR LENS Andrea Galvis, MD 1,3 a Ivan Ossma, MD MPH MSc 1,2, a Sanjeev Kasthurirangan, PhD 4, b.
Ruth Lapid-Gortzak MD PhD 1,2, Jan Willem van der Linden BOpt 2, and Ivanka J. van der Meulen MD 1,2 1 Department of Ophthalmology, Academic Medical Center,
Chapter 17 Optics 17.1 Reflection and Refraction
Optics and the Eye. The Visible Spectrum Some similarities between the eye and a camera.
1 Clinical Performance of the Crystalens® AO Guy M. Kezirian, MD, FACS.
Retrospective Comparison of 3177 Eyes Implanted with Presbyopic IOLs Carlos Buznego MD Elizabeth A. Davis MD, FACS Guy M. Kezirian MD, FACS William B.
1© Manhattan Press (H.K.) Ltd. The human eye Visual defects and their corrections Visual defects and their corrections 12.3 Properties of vision Visual.
Visualized light paths in different multifocal
Mitchell A Jackson MD Founder/Director, Jacksoneye Lake Villa, IL USA
Axial color. Lateral color Plane parallel plate.
Module 1-4 Basic Geometrical Optics. Image Formation with Lenses Lenses are at the heart of many optical devices, not the least of which are cameras,
Chapter 25 Optical Instruments.
W. Andrew Maxwell, MD, PhD California Eye Institute; Fresno, CA and Billy R. Hammond, Jr, PhD Vision Sciences Laboratory, University of Georgia; Athens,
Cameras Course web page: vision.cis.udel.edu/cv March 22, 2003  Lecture 16.
Lenses in Combination The analysis of multi-lens systems requires only one new rule: The image of the first lens acts as the object for the second lens.
Advanced Optics Lab at San Jose State University Ramen Bahuguna Department of Physics.
Fundamental of Optical Engineering Lecture 3.  Aberration happens when the rays do not converge to a point where it should be. We may distinguish the.
Recent Advances in Intraocular Lenses Jim Schwiegerling, PhD Ophthalmology & Vision Sciences Optical Sciences.
W. Maxwell, MD, PhD California Eye Institute Fresno, California
Theme 2: The theoretical eye Form and dimensions of the eye Optical constants of the eye Optical Geometry applied to the eye Schematic model of the eye.
W. Andrew Maxwell, MD, PhD Rob Gray, PhD Disclosures: This study was funded by Alcon Research, Ltd, which also assisted with the preparation of these slides.
A Fellow Eye Comparison of Aberrations, Modulation Transfer Function and Contrast Sensitivity After AcrySof IQ and AcrySof Natural IOL Implantation. Mayank.
© Eyes and cameras – what am I? Read the descriptions and see if you can name the part of the eye or camera being described!
P91: Clinical Performance of Phakic Angle-Supported Investigational IOL in Prospective Global Trials, ASCRS 2010, Boston P91: Clinical performance of phakic.
The Hong Kong Polytechnic University Optics 2----by Dr.H.Huang, Department of Applied Physics1 Diffraction Introduction: Diffraction is often distinguished.
Reproducibility of Optical Measurements in Pseudophakic eyes Using Double-Pass Measurement System St Thomas Hospital, London Anish Dhital, David Spalton,
DGII 2008 Comparison of Aspheric ReSTOR and Tecnis multifocal IOL Dongho Lee MD, PhD Yonsei eye center, Seoul, South Korea No Financial Interest.
AcrySof ® ReSTOR ® Aspheric IOL. Aspheric IOL AcrySof ® ReSTOR ® 2 AcrySof ® ReSTOR ® Aspheric IOL SN6AD3 Add Power: +4 D Spectacle Plane: 3.2 D Range:
IFS prototype – PM3 LAM, 13/06/2003 Prototype testing at CRAL Tests at room temperature in the visible.
Ophthalmology & Vision Sciences
Futoshi Taketani, MD,PhD,
LogMAR-Analysis of multifocal intraocular lenses: Clinical performance A. Mannsfeld, I.-J. Limberger, A. Ehmer, M.P. Holzer, G. U.Auffarth International.
Financial Disclosure: Medical Director–Galilei R&D Consultant, Ziemer Group AG, Port, Switzerland Consultant & Territory Manager for.
Lenses Properties, Characteristics & Ray Diagrams.
Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses. Rodrigo.
Investigation of Multifocal Toric IOLs to Compensate for Corneal Astigmatism and to Provide Near, Intermediate, and Distance Vision José L. Rincón, MD.
Preliminary Results after Cataract Surgery with the Aspheric Acrysof ReSTOR IOL to Correct Presbyopia Meeting of the ASCRS Chicago 8-10 February 2007 R.M.M.A.
Warren Hill, MD, FACS East Valley Ophthalmology 5620 E. Broadway Road
Inadvertent Insertion of an Opposite- Side Tecnis ZM900 Multifocal IOL Wilson Takashi Hida, M.D. Celso Takashi Nakano; Jonathan Lake;
Toric IOLs: wavefront aberrometry and quality of life Mencucci Rita Giordano Cristina, Stiko Ermelinda, Miranda Paolo, Eleonora Favuzza, Ugo Menchini Authors.
Global Meta-Analysis on Visual and Optical Quality Comparison for Aspheric vs Spherical IOL Technology James P. McCulley, MD Department of Ophthalmology.
Date of download: 5/30/2016 Copyright © 2016 American Medical Association. All rights reserved. From: Quantitative Performance of Bifocal and Multifocal.
Example: What kind of lens must be used, in order to give an erect image 1/5 as large as an object placed 15 cm in front of it? M = -q/p  -q/p=1/5 So.
Multifocal lenses analysis with OQAS system Dr Alfredo Castillo Hospital Quirón Madrid ASCRS Chicago.
Mirrors and Lenses How do eyeglasses correct your vision? When you look in a mirror, where is the face you see? In which ways is a cell phone camera similar.
Outcomes Update for an Aspheric Accommodating IOL John A. Hovanesian, M.D. Clinical Instructor, UCLA Jules Stein Eye Institute Harvard Eye Associates,
Vinohrady Teaching Hospital, Prague, Czech Republic Vinohrady Teaching Hospital, Prague, Czech Republic M. Vokrojova MD, M. Vokrojova MD, D. Sivekova MD,
Postoperative Refraction and Patient Satisfaction after Bilateral Implantation of Presbyopia-Correcting Intraocular Lenses Robert Cionni, MD Financial.
Multi focal IOL.
Hayashi Eye Hospital, Fukuoka, Japan
Source: Figure 1. Image of colors with their different wavelengths refracted at different angles.
Eye clinic of the 3rd Faculty of Medicine, Prague, Czech Republic
Measurements of Multifocal Intraocular Lenses Using a Fluid-Filled Model Eye Charles Campbell This presentation is the result of work done with the assistance.
Thomas Kohnen, MD Department of Ophthalmology
Comparison of vision with an accommodating IOL versus a multifocal IOL
Comparative Global Literature Review of Visual and Optical Quality of Refractive, Diffractive, and Hybrid IOL Designs James P. McCulley, MD Department.
Presentation transcript:

W. A. Maxwell, MD, PhD ASCRS 2008 Comparison of the Optical Image Quality for Presbyopia Correcting IOLs using Modulation Transfer Function Testing W. Maxwell, MD, PhD California Eye Institute Fresno, California Financial Disclosure: Consultant & Travel (Alcon Research, Ltd.)

W. A. Maxwell, MD, PhD ASCRS 2008 Testing Methodology Study Objective To compare the image quality of various Presbyopia Correcting IOLs with different designs ( accommodative or multifocal; aspheric or spherical) using the modulation transfer function (MTF) testing method and U.S. Air Force 1951 Resolution Target (AFT) testing Modulation Transfer Function (MTF) Testing Objective method of measuring image contrast degradation at different spatial frequencies 1 Modified ISO model eye The standard ISO Model Eye 1 configuration incorporates a 35 mm doublet lens to converge light to simulate the refractive power of a human cornea The model cornea in the ISO Model Eye is an aberration free lens, and is not representative of a typical human cornea with its inherent positive spherical aberration Thus, a modified ISO model eye, representing an average human cornea, was used in this study Study Objective To compare the image quality of various Presbyopia Correcting IOLs with different designs ( accommodative or multifocal; aspheric or spherical) using the modulation transfer function (MTF) testing method and U.S. Air Force 1951 Resolution Target (AFT) testing Modulation Transfer Function (MTF) Testing Objective method of measuring image contrast degradation at different spatial frequencies 1 Modified ISO model eye The standard ISO Model Eye 1 configuration incorporates a 35 mm doublet lens to converge light to simulate the refractive power of a human cornea The model cornea in the ISO Model Eye is an aberration free lens, and is not representative of a typical human cornea with its inherent positive spherical aberration Thus, a modified ISO model eye, representing an average human cornea, was used in this study Ref 1: ISO Ophthalmic implants – Intraocular lenses – Part 2 Optical properties and test methods, 2000.

W. A. Maxwell, MD, PhD ASCRS 2008 The optical bench setup Pinhole Target IOL Light Source CCD Camera Optikos Video MTF measurement system 1024 element detector, a 550 nm narrow band filter, and a 20X microscope objective IOLs were mounted in a lens holder under 10 mm of compression, and aligned on the optical axis in a wet cell filled with deionized (DI) water at 35 ± 2 °C with the anterior side of the IOL facing the incident light 3 and 5 mm apertures used to simulate pupil size Green light illuminates a single slit object located on the front focal plane of a 180 mm collimator The collimated light passes through the model cornea and IOL system that simulates the vergence of a human eye Optikos Video MTF measurement system 1024 element detector, a 550 nm narrow band filter, and a 20X microscope objective IOLs were mounted in a lens holder under 10 mm of compression, and aligned on the optical axis in a wet cell filled with deionized (DI) water at 35 ± 2 °C with the anterior side of the IOL facing the incident light 3 and 5 mm apertures used to simulate pupil size Green light illuminates a single slit object located on the front focal plane of a 180 mm collimator The collimated light passes through the model cornea and IOL system that simulates the vergence of a human eye MTF Bench IOL Models Tested  ReSTOR SN6AD3 (Aspheric)  ReSTOR SA60D3 (Spherical)  Tecnis ZM900  Acri.Lisa 366D  ReZoom NXG1  Crystalens AT-50SE

W. A. Maxwell, MD, PhD ASCRS 2008 Modulation Transfer Function 5 mm Aperture, Distance Focus Spatial Frequencies of 0 to 100 lp/mm Frequency (lp/mm) MTF %

W. A. Maxwell, MD, PhD ASCRS 2008 Modulation Transfer Function 5 mm Aperture, Distance Focus, Discrete frequency of 100 lp/mm

W. A. Maxwell, MD, PhD ASCRS 2008 Modulation Transfer Function 5 mm Aperture, Distance Focus, Discrete frequency of 50 lp/mm

W. A. Maxwell, MD, PhD ASCRS 2008 MTF of ReSTOR ® & ReSTOR ® Aspheric IOLs 3 mm Aperture

W. A. Maxwell, MD, PhD ASCRS 2008 Modulation Transfer Function 3 mm Aperture, Distance Focus, Discrete frequency of 100 lp/mm

W. A. Maxwell, MD, PhD ASCRS 2008 Modulation Transfer Function 3 mm Aperture, Distance Focus, Discrete frequency of 50 lp/mm

W. A. Maxwell, MD, PhD ASCRS 2008 U.S. Air Force 1951 Resolution Target Test (AFT) Optikos Optical Testing System as described for the MTF test The setup for the AFT test was identical to that of the MTF test, except the single slit object was replaced with the AFT An image of the AFT was photographed through each IOL The AFT test includes paired 3-bar target sets used to qualitatively measure resolution efficiency U.S. Air Force 1951 Resolution Target Test (AFT) Optikos Optical Testing System as described for the MTF test The setup for the AFT test was identical to that of the MTF test, except the single slit object was replaced with the AFT An image of the AFT was photographed through each IOL The AFT test includes paired 3-bar target sets used to qualitatively measure resolution efficiency

W. A. Maxwell, MD, PhD ASCRS 2008 U.S. Air Force 1951 Resolution Target test 5 mm Aperture, Distance Focus Crystalens AT-50SE ReSTOR Spherical SA60D3 ReSTOR Aspheric SN6AD3 Tecnis ZM900 ReZoom NXG1 Acri.Lisa 366D

W. A. Maxwell, MD, PhD ASCRS 2008 Conclusions The AcrySof ReSTOR SN6AD3 aspheric IOL produced the highest overall image quality for MTF and AFT testing compared to several presbyopia- correcting IOLs The Acri.Lisa IOL produced the second highest optical quality, while the ReZoom IOL had the lowest optical quality Clinical investigation is needed to determine whether superior IOL optical quality demonstrated in optical bench testing results in measurable visual improvements in clinical practice The AcrySof ReSTOR SN6AD3 aspheric IOL produced the highest overall image quality for MTF and AFT testing compared to several presbyopia- correcting IOLs The Acri.Lisa IOL produced the second highest optical quality, while the ReZoom IOL had the lowest optical quality Clinical investigation is needed to determine whether superior IOL optical quality demonstrated in optical bench testing results in measurable visual improvements in clinical practice