April 24 - 27, 2007 The Effect of Temperature on the Uptake Rates of a New PDMS-Based Permeation Passive Sampler for VOCs in Air Suresh Seethapathy, Tadeusz.

Slides:



Advertisements
Similar presentations
Sampling and Measurement for Volatile Organic Compounds
Advertisements

Gas Chromatography Introduction 1.) Gas Chromatography
Gas Chromatography 427 PHC.
Gas Chromatography.
Field Evaluation of Diffusive Samplers for Indoor Air VOC Measurements
The Foundations: Classical Split and Splitless Injection
New Groundwater Techniques and Technologies 17 th Annual RETS REMP Conference June 25-27, 2007 Eric L. Darois, CHP EPRI Consultant/RSCS Inc.
Preparation of a haloalkane. Haloalkanes can be made by a substitution reaction with an alcohol. Tertiary alcohols are the most reactive, and therefore.
Week 3 Collection and Analysis of Biodiesel
Dale T Littlejohn Senior Geologist. What is fate and transport in the vadose zone? Vadose Zone Hydrocarbon release from buried pipeline Aquifer Surface.
ERT 313/4 BIOSEPARATION ENGINEERING MASS TRANSFER & ITS APPLICATIONS
Passive Sampling of Nitrogen Dioxide and Sulfur Dioxide in Ambient Air of Chiang Mai Shakya, K.M. 1, P. Thavornyutikarn 1, S. Chantara 1, W. Saipunkaew.
InfraCal TOG/TPH Analyzer, Model CVH Extraction Procedure.
Analysis of Residues. Laboratory Analysis of Debris and Other Samples 1. Preparation of Liquid Samples: Liquid samples are simply drawn into a special.
Modeling Wing Tank Flammability Dhaval D. Dadia Dr. Tobias Rossmann Rutgers, The State University of New Jersey Piscataway, New Jersey Steven Summer Federal.
I n t e g r i t y - S e r v i c e - E x c e l l e n c e Headquarters U.S. Air Force 1 Direct Sampling Ion Trap Mass Spectrometry Rob R. Smith Oak Ridge.
COMPARISON OF DRYING KINETICS OF SPENT GRAIN DRIED ON INERT MATERIAL OF DIFFERENT HEAT CAPACITY M. Zielinska a,b S. Cenkowski b a Department of Agro-Food.
1 Soil Vapor Extraction Limitations and Enhancements LeeAnn Racz AgE 558 Semester Project April 2001.
How long will the test last?
Presented by the members of Superior Engineering Solutions: Mike Najera Dong Hoon Kim Seth Maher Priya Heerwani Thermal Desorption.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Partitioning of Volatile Organic Carbon Compoundss.
Methods Monitoring Polar Compounds Using Membrane Extraction and High-Speed Gas Chromatography Authors: Jonathan Maurer, Dr. Anthony J. Borgerding* Department.
Partitioning of VOCs: Why do we care? ä Determines how best to treat a site ä vapor extraction ä pump and treat ä remove contaminated soil ä Determines.
Gas Liquid Chromatography
Pervaporative membrane filtration for subsurface irrigation Tom Bond 1 *, May N. Sule 1, Lindsay C. Todman 1, Michael R. Templeton 1 and Jonathan A. Brant.
LECTURE 4: CHROMATOGRAPHY Content: - Chromatographic separation - classifying analytical separations; column chromatography, planar chromatography - gas.
7th Passive Sampling Workshop and Symposium RESTON, VA, USA, NEW APPROACH IN CALIBRATION OF PASSIVE PERMEATION SAMPLERS USED FOR VOC’s.
Technical meeting December 1 st 2014 Dr. Daniele Spinelli Laboratory Manager
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
 Introduction  Theory  Advantages/Disadvantages  Applications  Conclusions.
AUTOANALYZER T. Elsarnagawy 22/12/ Elsarnagawy, MASH
Experiment 2 DISTILLATION AND GAS CHROMATOGRAPHY OF ALKANES.
Personal Samplers Instructional Goal The participants will see the various methods used to collect personal samples and how that may impact them on the.
Chapter 15 Solutions. Solution types & parts  Solutions can be: Solids – brass, dental fillings, chocolate bar Liquids – sodas, vinegar, salt water Gaseous.
Jet Fuel Vaporization and Condensation: Modeling and Validation C.E. Polymeropoulos Robert Ochs Rutgers, The State University of New Jersey International.
COLUMN CHROMATOGRAPHIC PURIFICATION OF NITROANILINES
Artificial lung Done by : Eng.Samar.T.Alqattan
Experiment 2 DISTILLATION AND GAS CHROMATOGRAPHY OF ALKANES.
6th International Conference and Exhibition on Analytical & Bioanalytical Techniques, Sep 1-3, 2015 Valencia, Spain1 APPLICATION OF PASSIVE SAMPLING TECHNIQUES.
Outdoor (ISO ) and indoor (ISO 11844) classification
Increase of the product recovery of Clostridium acetobutylicum fermentation product by pervaporation P. Izák 1, V. Jarmarová 1,
Distillation ... A Separation Method.
Application of solid-phase microextraction to the recovery of explosives and ignitable liquid residues from forensic specimens K.G. Furton, J.R. Almirall,
1 FUNDAMENTAL PRINCIPALS OF In Situ THERMAL TREATMENT Professor Kent S. Udell Department of Mechanical Engineering Department of Civil and Environmental.
Measurement of Acrylonitrile in Ambient Air AMTAC April 12, 2011 John Bricarello California Air Resources Board.
Gas Chromatography Experiment. Gas Chromatography - Gas Chromatography (GC) is a common technique used to separate and identify volatile organic compounds.
KINETICS OF SURFACE-BOUND BENZO[A]PYRENE AND OZONE ON ORGANIC AND INORGANIC AEROSOLS N.-O. A. Kwamena, J. A. Thornton, J. P. D. Abbatt Department of Chemistry,
GORE, GORE-TEX and designs are trademarks of W. L. Gore & Associates © 2007 W. L. Gore & Associates, Inc. 1 Environmental Investigations Using Versatile,
Experiment 2 DISTILLATION AND GAS CHROMATOGRAPHY OF ALKANES.
Lecture 4c Extraction.
Today’s Lecture  Groundwater sampling. Announcements  Lecture  today: groundwater sampling  Thursday: HAZWOP – quiz (Bruce)  Lab  Lab 1: Phase I.
Jet Fuel Vaporization and Condensation: Modeling and Validation Robert Ochs and C.E. Polymeropoulos Rutgers, The State University of New Jersey International.
Micro Analytical GC System for Tea Aroma in Product Quality Control Fred Lewis, Principal Chemist, Analytical Science Laboratory, Unilever John Crandall,
Correlation Between the Transdermal Permeation of Ketoprofen and its Solubility in Mixtures of a pH 6.5 Phosphate Buffer and Various Solvents Ref.: Drug.
Introduction to Gas Chromatography
Chapter 1: The Nature of Analytical Chemistry
Hot Disk Transient Plane Source (TPS) Technique
Experiments in Analytical Chemistry
THIN LAYER CHROMATOGRAPHY.
Fundamentals of Analytical Chemistry Eighth Edition.
INTRODUCTION Diffusive Samplings Evaluation and Optimization in Low Air Speed Workplaces Williams ESTEVE*, Marianne GUILLEMOT, Eddy LANGLOIS INRS, rue.
FIGURE Selected Hazard Communication Standard Pictograms Associated with the GHS Health Hazards. Some pictograms more clearly suggest.
Fabrication and Characterization of Chemical Vapor Deposition Buckypapers for Use in Air Sampling of Volatile Organic Compounds Jacob S. Shedd, Jonghwa.
Extraction Procedure (2)
SEPARATION TECHNIQUES
SEPARATION OF MIXTURE BY USING COLUMN CHROMATOGRAPHY
SOLUTION AND FILTRATION
High-Performance Liquid Chromatography
GAS CHROMATOGRAPHY.
High-Performance Liquid Chromatography
Presentation transcript:

April , 2007 The Effect of Temperature on the Uptake Rates of a New PDMS-Based Permeation Passive Sampler for VOCs in Air Suresh Seethapathy, Tadeusz Górecki Monika Partyka, Bożena Zabiegała, Jacek Namieśnik Department of Chemistry, University of Waterloo, Waterloo, ON, Canada. Department of Analytical Chemistry, Gdańsk University of Technology, Gdańsk, Poland 7th Passive Sampling Workshop and Symposium Virginia, United States

2 Concentration Profiles for the Two Types of Passive Samplers Sample environment Sorbent Diffusion barrier Distance Concentration Permeation samplers Diffusion samplers Membrane thickness Sample environment Distance Concentration Sorbent

3 Comparison of Diffusion and Permeation Passive Samplers Diffusion samplers –sampling rate can be estimated based on the properties of the analyte –sensitive to moisture, air currents, temperature variations Permeation samplers –effective moisture elimination, much less sensitive to air currents and temperature changes –must be calibrated for each analyte Permeation passive sampling – permeation through a membrane Diffusive passive sampling – diffusion through air

4 Calibration Constant / Uptake rate M = Amount of analyte collected by sorbent D = Diffusion coefficient A = Area of membrane C ma = Concentration of the analyte on the membrane surface in contact with air C ms = Concentration of the analyte on the membrane surface in contact with the sorbent t = Time of sampling L m = Membrane thickness sorbent PDMS membrane C ms C ma CoCo Concentration

5 M = Amount of analyte collected by sorbent D = Diffusion coefficient A = Area of membrane C ma = Concentration of the analyte on the membrane surface in contact with air. C ms = Concentration of the analyte on the membrane surface in contact with the sorbent t = Time of sampling L m = Membrane thickness K = Partition coefficient of the analyte between air and the membrane P= Permeability constant of the polymer towards the analyte C o = Concentration of the analyte in air k = Calibration Constant (time/volume) k -1 = Uptake rate (volume/time) Theory

6 Properties of PDMS Individual segmental motions even at –127 O C !!! Nearly constant diffusion coefficient for most VOCs Relative permeabilities are primarily decided by partition coefficients Temperature effects on permeability increase in temperature results in - decreased partition coefficient - increased diffusivity P = Đ K

7 Energy of Activation of permeation Energy of activation of diffusion Heat of solution For PDMS, towards most volatile organic compounds ΔH s < 0 E d > 0

8 slope Effect of Temperature on the Calibration Constant

9 Passive Sampler Design PDMS membrane Disposable 1.8 mL crimp-top vial based permeation passive sampler Area of PDMS membrane exposed: 38 mm mL crimp top vial Al cap Turned upside down before use

10 1. Remove septum from crimp cap 2. Cut PDMS membrane 3. Add sorbent to the vial 4. Assemble membrane and cap 5. Crimp to seal 6. Attach support/turn upside down for sampling Aluminum crimp cap cut membrane Cutting tool Teflon lined septum sorbent vial PDMS membrane Sampler Preparation

11 Analyte extraction/analysis 7. De-crimp cap, add desorption solvent 8. Crimp with aluminum crimp cap/teflon lined septum and shake 9. Introduce into GC auto sampler – either directly or after transferring the extract into a different vial

12 Design of a Reusable Passive Sampler Aluminum crimp cap PDMS membrane Mouth portion of a screw top vial Mouth portion of a crimp top vial Plastic screw cap Area of PDMS membrane exposed: 38 mm 2 Reusable, 1.8 mL crimp-top vial-based permeation passive sampler

13 Brass ferrule PTFE Plug Neat liquid PTFE tube Experimental Setup Laboratory made permeation tube a a Prof. Jacek Namieśnik, Technical University of Gdańsk, Private communication. Air Purifier Flow controller Standard gas mixture generator Thermostated exposure chamber Activated carbon sampling tube High pressure N 2 source Flow controller and monitor Suction Pump Vent stirrer

14 Effect of Temperature on the Calibration Constant Experiments were conducted over a temperature range of 285 to K Model compounds included n-nonane, n-decane, ethyl acetate, propyl acetate, methyl butyrate, sec-butyl acetate, ethyl butyrate, butyl acetate, propyl butyrate, butyl butyrate. n-nonane and n-decane used as references. Experiments started with esters as they have polarity between that of highly polar alcohols and highly non-polar alkanes and aromatic hydrocarbons.

15 Temperature – Calibration constant relationship for esters

16 Temperature – Calibration constant relationship for esters - average variation for moderately polar esters - 1.4% per o K

17 - non-polar n-nonane and n-decane - 0.9% per o K Temperature – Calibration constant relationship for n-alkanes

18 Energy of Activation of Permeation CompoundSlopeE p (kJ.mole -1 ) R 2 for the 1/T vs ln(k) correlation Propyl acetate Methyl butyrate Sec-butyl acetate Ethyl butyrate Butyl acetate Propyl butyrate n-nonane Butyl butyrate n-decane

19 Conclusions A simple, cost effective passive sampler and sampling procedure have been devised to quantify VOCs in air. Temperature effects on calibration constant have been studied for esters, n-nonane and n-decane. Fundamental transport properties like energy of activation of permeation can be determined using the data. Temperature effects are usually negligible in indoor air environments.

20 Acknowledgments Natural Sciences and Engineering Research Council (NSERC Canada). University Consortium for Field-focused Groundwater Contamination Research. Workplace Safety and Insurance Board of Ontario (WSIB)

21 Questions?