Étienne Létourneau, Fabiola Vallejo Castaneda, Nancy El Bared, Danny Duplan, Martin Hinse Presented by: Étienne Létourneau 2015 Joint Congress, Montreal,

Slides:



Advertisements
Similar presentations
Quality Control in Diagnostic Radiology
Advertisements

In the past few years the usage of conformal and IMRT treatments has been increasing rapidly. These treatments employ the use of tighter margins around.
The Tomotherapy Experience at Advocate Good Samaritan Hospital
10th Annual Lung Cancer Conference Radiation Oncology
RAD309 Patient Dose.
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
5 th September 2005 Paul Collins Computed Tomography Dosimetry Assessment of Effective Dose in Computed Tomography using an Anthropomorphic Phantom Paul.
IGRT Prostate – the ADHB experience Nicola Gordon Imaging Specialist RT ADHB.
The TrueBeam System ™ Clinic Name Presenter’s name Clinic location
Radiation Exposure, Dose and Relative Biological Effectiveness in Medicine Background Image:
tomos = slice, graphein = to write
Tissue inhomogeneities in Monte Carlo treatment planning for proton therapy L. Beaulieu 1, M. Bazalova 2,3, C. Furstoss 4, F. Verhaegen 2,5 (1) Centre.
X-Rays and CT Scans.
Factors affecting the X-Ray output
X-Ray Production & Emission
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
X-Ray Production & Emission
Quality Control Rad T 110.
Dose Distribution and Scatter Analysis
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Measurement of Dose to Critical Structures Surrounding the Prostate from Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional Conformal Radiation.
“Dosimetry and optimisation of Cone Beam CT for dentomaxillofacial applications” Safety and Efficacy of a New and Emerging Dental X-ray Modality Cone Beam.
Научно-практический центр протонной лучевой терапии и радиохирургии (Москва-Дубна) A SYSTEM FOR MEASUREMENT OF A THERAPEUTIC PROTON BEAM DOSE DISTRIBUTION.
External Beam Radiotherapy
H Ariyaratne1,2, H Chesham2, J Pettingell2, K Sikora2, R Alonzi1,2
Brachytherapy Medical radiation.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
Radiographic Dosimetry
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Radiation Monitoring and Protection
In vivo dosimetry Eirik Malinen Eva Stabell Bergstrand Dag Rune Olsen.
CT Instrumentation and X-ray system
Commissioning of an Optically-Stimulated Luminescence (OSL) system for remote dosimetry audits J.F. Aguirre, P. Alvarez, C. Amador, A. Tailor, D. Followill,
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
Araki F. Ikegami T. and Ishidoya T.
Historical Review and Modalities in Diagnostics
S Scarboro 1,2, D Cody 1,2, D Followill 1,2, P Alvarez 1, M McNitt-Gray 3, D Zhang 3, L Court 1,2, S Kry 1,2 * 1 UT MD Anderson Cancer Center, Houston,
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Determining Radiation Intensity
Analysis by Monte-Carlo simulations of the characteristics of nano and micro dosimeters for real time measurements in radiotherapy and medical physics.
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Introduction to Radiation Therapy
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
Karolina Kokurewicz Supervisors: Dino Jaroszynski, Giuseppe Schettino
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Interaction of x-ray photons (and gamma ray photons) with matter.
Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
3/2003 Rev 1 II.3.7 – slide 1 of 40 Session II.3.7 IAEA Post Graduate Educational Course Radiation Protection and Safe Use of Radiation Sources Part IIQuantities.
Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
Radiation Protection RTMR 284 CHAPTER 20 PART II.
Ch 36 Radiation Protection Design
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
Electron Beam Therapy.
EURADOS Working Group 9: Radiation Dosimetry in Radiotherapy
Very High Energy Electron for Radiotherapy Studies
Evaluation Of RTOG Guidelines For Monte Carlo Based Lung SBRT Planning
Linac Commissioning Overview
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
Radiation Protection in Dental Radiology
CBCT Equivalent Source Generation Using HVL and Beam Profile Measurements. Johnny Little PSM - Medical Physics Graduate Student University of Arizona.
Advanced radiotherapy for lung cancer
GHG meeting at ESTRO36 May, 2017
Presentation transcript:

Étienne Létourneau, Fabiola Vallejo Castaneda, Nancy El Bared, Danny Duplan, Martin Hinse Presented by: Étienne Létourneau 2015 Joint Congress, Montreal, Quebec – May 28 CALCULATING DOSAGE FOR CBCT

Overview  Objectives  Methods and Calibration  Measurements and Results  Acquisition adjustments  Future work

Objectives  To measure the Cone Beam Computed Tomography (CBCT) dose-to-organs using Optically Stimulated Luminescence (OSL) dosimeters : We want to know what we are giving to the patient!  To reduce the dose by adjusting the CBCT acquisition parameters.

Purpose  In radiation therapy (RT), the CBCT is used for patient positioning prior to treatment.  CBCT dose seems negligible compared to the treatment dose. Why bother? –A.L.A.R.A –CBCT gives dose to healthy tissues –A patient can receive as many CBCT scans as treatment fractions (sometimes more) –In some cases, the treatment dose to a given organ is already reaching the recommended limits –The goal of CBCT is matching (most of the time using bones), and matching doesn’t require high quality

Methods and Calibration  The Elekta X-ray Volumetric Imaging (XVI) system uses divergent x-rays with a pyramid shape. 1.kV X-ray beam reference axis 2.X-ray beam (followed by a collimator cassette) 3.Projected X-ray field 4.kV detector panel

Methods and Calibration  Important CBCT parameters –Energy –# of frames –mA per frame –ms per frame –Start/Stop angles –Filter All influence the total number of photons. These are the parameters we’ll adjust in order to reduce the dose-to-patient.

Methods and Calibration  Optically Stimulated Luminescence Dosimeters (OSLD) work in a similar way than Thermoluminescent Dosimeters (TLD) using the electrons trapped in a crystalline structure.  Electrons are released using light instead of heat.  Landauer Nanodots have a useful energy range from 5 keV to 20 MeV.

Methods and Calibration  How did we calibrate OSLD?

Methods and Calibration  Ok, so where did we start? –Measuring the Half Value Layer (HVL) using : Aluminum RTI Barracuda Dose detector (also estimated kVp and mAs) –With the HVL, we get the air-kerma calibration factor N K from the NRC for a Farmer chamber in the kV range

Methods and Calibration

8x8 cm 2 4x4 cm cm

Measurements and Results  We filled an heterogeneous anthropomorphic adult phantom with OSLs.

Measurements and Results  We measured for the following XVI default acquisitions: Head Chest Pelvis

Measurements and Results

 These results match similar studies: –Song et al. Med. Phys. 35: (2008) XVI dose from 0.1 to 3.5 cGy/scan Using ionization chamber in phantom –Hyer et al. Appl. Clin. Med. Phys (2010) XVI dose from 0.1 to 3.0 cGy/scan Fiber-Optic-Coupled devices –Alei, Ding and Guan. Med Phys. 37(1) (2007) XVI dose from 0.1 to 2.8 cGy/scan Using TLD and Monte-Carlo simulation

Measurements and Results  These results need to be interpret carefully : –Depending on the patient size: Since the number of photons are always the same, a larger patient might receive less dose and vice versa. –OSLD response depends on the radiation angle. –The absorbed dose in a given material depends on the electronic density and the energy of the interacting photon. –However, these results still a good approximation

Acquisition adjustments  We’ve adjusted the CBCT acquisition parameters (mA per frame, ms per frame and total number of frames). Acquisition Maximum dose (cGy) % of daily treatment dose Ratio with default dose Head0.11/2 Pelvis1.01/4 Chest (low dose)0.51/8 Chest (ultra low dose) 0.21/22

Acquisition adjustments  We’ve created low dose breast scans: –The imaged breast received a maximum of 0.7 cGy. –Considering we are giving a maximum of 25 fractions for a breast treatment (most of time 16), the patient will not receive more than 20 cGy from CBCT which correspond to less than 0.4 % of the total dose.

Acquisition adjustments  A clinical study using the CBCT dose-to-organs… –The maximum dose to the heart by our half CBCT left breast protocol is 0.6 cGy. –10 patients with left sided breast cancer receiving 50 Gy in 25 fractions and getting daily HCBCT were randomly chosen and analyzed retrospectively –Assuming daily HCBCT, the heart will receive a maximum extra dose of 15 cGy. HCBCT frequency % of the target volume covered by isodose 95% Daily100% Weekly97% None95%

Future work  Do a similar experiment on CT scan  Compared with different CBCT scans such as the Varian On-Board Imaging (OBI)  Encourage other centers to adjust their acquisition parameters even if they can’t measure the dose-to-organs

Acknowledgement  Martin Hinse, Nancy El Bared, Danny Duplan, Fabiola Vallejo and all the CICL staff. This research is a beautiful example of interdisciplinary team work. Merci à vous!  Be there for my presentation on dose to cardiac implantable electronic device, Thursday June 11 AM, Dosimetry and Quality Control session

References [1] C.-M Ma, Chair et al. AAPM protocol for kV x-ray beam dosimetry in radiotherapy and radiobiology. Medical Physics, 28(6), June 2001 [2] Hyer, D.E et al. An organ and effective dose study of XVI and OBI cone-beam CT systems. Journal of applied clinical medical physics (2010), 11(2), 3183, April 2010 [3] El Bared, Nancy et al. Daily Half Cone Beam Computed Tomographies for left sided breast cancer: Is the added dose worth it?. CARO- ACRO, St.John’s, August 2014

THANK YOU! QUESTIONS?