Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA

Slides:



Advertisements
Similar presentations
High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging
Advertisements

Protein Quantitation II: Multiple Reaction Monitoring
Oligonucleotide analysis by Linear MALDI-TOF mass spectrometry Stephen J. Hattan , Kenneth C. Parker, Marvin L. Vestal; SimulTOF Systems, 60 Union Ave,
Introduction to MALDI-TOF MS
Lecture 9. MALDI/TOF. Introduction Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique used in mass spectrometry, allowing.
Novel 3-D Sample Plate using Monolithic Capture Media in Collimated-Hole Structures for Interfacing High Capacity Separations with MALDI-TOF Poster Number.
Lecture Three: "MALDI-TOF MS, Up Close, and Personal Review of principles Delayed Extraction Reflector / reflectron Post-source Decay (PSD)
High Performance MALDI-TOF Mass Spectrometry Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA.
Primary Ionization Techniques for Molecules
Development of TOF-MS from intellectual curiosity to practical technique Marvin Vestal SimulTOF Systems The Jim Waters Symposium at PittCon March 2, 2014.
How to identify peptides October 2013 Gustavo de Souza IMM, OUS.
Peptide Mass Fingerprinting
Chem. 133 – 5/7 Lecture. Announcements I Exam 3 on Tuesday (will give summary of material to know later) Format will be similar to other exams I will.
Chem. 133 – 4/28 Lecture. Announcements Lab Report 2.3 due Today Pass back graded materials (lab reports 2.2, Q5, and AP3.1) Today’s Lecture Mass Spectrometry.
Chapter 20 Molecular Mass Spectrometry. Introduction... Mass spectroscopy is perhaps one of the most widely applicable of all the analytical tools available.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform.
PROTEIN IDENTIFICATION BY MASS SPECTROMETRY. OBJECTIVES To become familiar with matrix assisted laser desorption ionization-time of flight mass spectrometry.
Mass spec lecture. What kind of info can mass spec give you? Molecular weight Elemental composition (low MW with high resolution instrument) Structural.
Mass Labels While Mass Spectroscopy is not covered in our texts, other mass/density dependent technologies including X-ray Spectroscopy, Gamma Spectroscopy,
LC-MS Lecture 7.
Mass Spectrometry 101 An Introductory Lecture On Mass Spectrometry Fundamentals Presented to the Sandler Mass Spectrometry Users’ Group University of California.
MALDI-TOF-TOF with High Resolution Precursor Selection and Multiplexed MS-MS Poster Number ThP 618 Kevin Hayden, Stephen C. Gabeler, Mark Dahl and Marvin.
Proteomics Informatics – Overview of Mass spectrometry (Week 2)
Comparison of chicken light and dark meat using LC MALDI-TOF mass spectrometry as a model system for biomarker discovery WP 651 Jie Du; Stephen J. Hattan.
Introduction Recent research has proposed rapid and robust identification of intact microorganisms using matrix assisted laser desorption/ ionization time-of-flight.
Tissue Imaging with New MALDI-TOF(TOF) Instrumentation Presented at MSIA 2013 Vanderbilt University April 17,2013 Marvin Vestal, CSO.
Molecular mass spectrometry Chapter 20 The study of “molecular ions” M + e -  M e -
Tissue Imaging by 5 kHz High-Performance MALDI-TOF Poster Number TP191 Christina Vestal 1, Kenneth Parker 1, Kevin Hayden 1, George Mills 1, Marvin Vestal.
Mass Spectrometry I Basic Data Processing. Mass spectrometry A mass spectrometer measures molecular masses. The mass unit is called dalton, which is 1/12.
INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable.
Based on ionization of gas phase molecule followed by analysis of the masses of the ions produced. The Mass Spectrum: Graph of ion intensity versus mass-to-charge.
-Aluminum housing with o-ring seal allows for washing (salt removal), plate cleaning and regeneration enabling for multiple use -Sample elution for MS.
Laxman Yetukuri T : Modeling of Proteomics Data
Marvin Vestal and Kevin Hayden SimulTOF Systems
Overall efficiency and speed of analysis – key factors are ions per sample molecule desorbed and ions produced per laser pulse Precision and accuracy of.
MS Calibration for Protein Profiles We need calibration for –Accurate mass value Mass error: (Measured Mass – Theoretical Mass) X 10 6 ppm Theoretical.
A new "Molecular Scanner" design for interfacing gel electrophoresis with MALDI-TOF ThP Stephen J. Hattan; Kenneth C. Parker; Marvin L. Vestal SimulTof.
INF380 - Proteomics-51 INF380 – Proteomics Chapter 5 – Fundamentals of Mass Spectrometry Mass spectrometry (MS) is used for measuring the mass-to-charge.
Supplement material to Ref. JASMS Ms. No Detection of an amine impurity and quality assessment related to this impurity of the UV MALDI matrix.
Chemistry 367L/392N Macromolecular Chemistry Lecture 8.
The Future of Time-of-flight Mass Spectrometry
Mass Analyzers : Time-of-flight Mass Spectrometry CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Joel Kimmel Fall 2007.
EBI is an Outstation of the European Molecular Biology Laboratory. In silico analysis of accurate proteomics, complemented by selective isolation of peptides.
Separates charged atoms or molecules according to their mass-to-charge ratio Mass Spectrometry Frequently.
Chem. 133 – 2/16 Lecture. Announcements Lab today –Will cover last 4 set 2 labs + start on set 2 labs –Lab Report on electronics labs – due 2/23 (I planned.
Chem. 133 – 2/18 Lecture. Announcements Homework Set 1.2 (bold problems) – due today Quiz 2 Today Electronics Lab Reports – due Tuesday Seminar this Friday.
Quantitative MALDI-TOF for Clinical Applications.
Deducing protein composition from complex protein preparations by MALDI without peptide separation.. TP #419 Kenneth C. Parker SimulTof Corporation, Sudbury,
Quantification of Albumin and Creatinine in Urine by MALDI-TOF Mass Spectrometry Jane Y. Yang, David A. Herold Department of Pathology, University of California.
A bi-functional glass membrane designed to interface SDS-PAGE separations of proteins with the detection of peptides by mass spectrometry MP 006 Kenneth.
Chem. 133 – 4/26 Lecture. Announcements Return graded quiz and additional problem Lab – Lab report deadlines (2:4 – Thursday) Today’s Lecture – Mass Spectrometry.
2014 생화학 실험 (1) 6주차 실험조교 : 류 지 연 Yonsei Proteome Research Center 산학협동관 421호
What is Mass Spectrometry? Mass spectrometry could be considered as an analytical technique that involves the study in the gas phase of ionized molecules.
Introduction to Liquid Phase Mass Spectrometry
김지형. Introduction precursor peptides are dynamically selected for fragmentation with exclusion to prevent repetitive acquisition of MS/MS spectra.
Chapter 11 Structure Determination: Mass Spectrometry, Infrared Spectroscopy, and Ultraviolet Spectroscopy.
RANIA MOHAMED EL-SHARKAWY Lecturer of clinical chemistry Medical Research Institute, Alexandria University MEDICAL RESEARCH INSTITUTE– ALEXANDRIA UNIVERSITY.
Yonsei Proteome Research Center Peptide Mass Finger-Printing Part II. MALDI-TOF 2013 생화학 실험 (1) 6 주차 자료 임종선 조교 내선 6625.
Target Analyses in Parallel Reaction Monitoring Mode (PRM)
An introduction to Mass spectrometry. What is mass spectrometry? Analytical tool measuring molecular weight (MW) of sample Only picomolar concentrations.
Peptide Mass Finger-Printing Part II. MALDI-TOF
Mass Spectrometry 101 (continued) Hackert - CH 370 / 387D
Acknowledgements Slides and animations were made by Dr. Jon Karty Mass Spectrometry Facility Indiana University, Bloomington.
Mass Spectrometry 101 Hackert - CH 370 / 387D
For Forensic Toxicology use Only
An Introduction to the ProteinChip® Reader
Chem. 133 – 2/16 Lecture.
Figure 1 Schematic representation of a typical MALDI-MSI workflow
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
Presentation transcript:

Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA New MALDI-TOF MS with very high resolving power and mass accuracy Presented at ASMS June 3,2008 Marvin Vestal and Kevin Hayden Virgin Instruments Corp. Sudbury, MA

Objectives Determine optimum design parameters for a high-performance bench-top MALDI-TOF Resolving power and mass accuracy Sensitivity and dynamic range High throughput Efficient sample utilization Simple and reliable Low cost Build prototype instrument and compare experimental results with theoretical predictions

Common Features of Virgin MALDI TOF Simplified sample loading with super microplate format (102x108 mm) Benchtop cabinet 1.75m high x 0.5 m wide x 0.7 m deep 100 kg max weight 5 kHz laser, up to 100 spectra/sec Fully automated, no operator expertise required No joystick, no camera 24/7 operation with minimal preventative maintenance

Sample plate V2 V 2-stage reflector V1 velocity focus Extraction electrode Drift tube Dm d4 d40 D d3 d0 Potential diagram for high-resolution TOF with single-field source and two-field reflector. The field-free distance D is the distance from the source to the mirror plus the distance from the mirror to the detector (not shown).

Instrument Parameters for Reflecting Analyzer (single-field source and two-field mirror) De=3200, d0=6, D=2264, d3=114.2, d40=127.7 all in mm V=8.56, V1=6.31, V2=9.23 all in kV dt=1.5 nsec (0.5 nsec bins, 1 nsec single ion pulse width) Focus mass m*= 3 kDa Dv=6d0 = 36 mm, vn (for m*)=0.0235 mm/nsec Time lag Dt=d0/vn = 250 nsec K=2d0/(Dv-Ds)=0.5 Initial Conditions for MALDI (typical) dv0=400 m/sec, dx=0.01 mm Trajectory error and voltage error assumed to be small

References M. L. Vestal and P. Juhasz, “Resolution and Mass Accuracy in Matrix-Assisted Laser Desorption Time-of-Flight Mass Spectrometry”, J. Am. Soc. Mass Spectrom. 9, 892-911 (1998). M. L. Vestal and K. Hayden, “High-performance MALDI-TOF mass spectrometry for proteomics”, Int. J. Mass Spectrometry 268, 83-92 (2007). M. L. Vestal and K. Hayden, “High-performance MALDI-TOF Mass Spectrometry”, paper TOFam08:15 presented at 55th ASMS Conference on Mass Spectrometry and Allied Topics, Indianapolis, June 3-7, 2007. S. J. Hattan and M. L. Vestal, “Novel 3-D sample plate using monolithic capture media in collimated-hole structures for interfacing high capacity separations with MALDI-TOF”, poster TBP-060 presented at 56th ASMS Conference on Mass Spectrometry and Allied Topics, Indianapolis, June 3, 2008.

Contributions to relative peak width, Dm/m, with 1st and 2nd order velocity focusing Initial position, dx: Rs1 = 2[(Dv-Ds)/2d0y](dx/De) Initial velocity, dv0: Rv1 =(4d0y/De)(dv0/vn)[(1-(m/m*)1/2] Rv2 =2[2d0y/(Dv-Ds)]2 (dv0/vn)2=0 Rv3 =2[2d0y/(Dv-Ds)]3 (dv0/vn)3 Rv = Rv1 + Rv3 Time error, dt: Rt = 2dt/t = 2dtvn/De Trajectory error, dL: RL = 2dL/De Voltage error, dV: RV = dV/V Resolving power: R-1 = [Rs12 + Rv2 + Rt2 + RL2 + RV2 ]-1/2 Neglected in initial calculation

4 GS/s 2 GS/s 6.4 m 4 GS/s 2 GS/s 3.2 m Calculated resolving power as function of m/z, curve 1 corresponds to experimental instrument

BSA digest 13,000 11,600 Resolution is 40% of calculated Spectrum from original version

calculated Comparison of calculated resolving power with experimental results With dm/m=70 ppm due to mass independent errors. Measured 5 mchannel plate ETP MagnetTOFTM (DM167) Original version before trajectory error correction

Changes to correct mass independent errors Add voltage regulator at input to mirror HV supplies (removes low frequency noise) Correct trajectory error due to ion deflector Use faster detector (0.5 ns) ETP DM167 in place of 5 mm dual channel plate Use faster digitizer 0.5 ns bins>0.25 ns bins

f1 = f2 Ray diagrams indicating source of trajectory error

f1 = f2 Ray diagram showing general correction to trajectory error due to deflection

Ray diagram of final version employed in the corrected version of the experimental analyzer

100 fm peptide standards Example of spectrum obtained with corrected analyzer

BSA digest Example of spectrum obtained with corrected analyzer

2 4 GS/s 2 GS/s 1 Comparison of experimental results with theoretical calculations. Bars are measured resolving power with standard deviation of 5 replicates. Curve 1 corresponds to digitizer with 0.5 ns bins and curve 2 to 0.25 ns bins.

Calibration Equation: m1/2=D0 +D1t[1+D2t+D3t2] Determine coefficients by least square fit to multiple Peaks covering broad mass range Average=-0.1 mDa RMS =0.99 ppm RMS error for 10 peaks in spectrum of tryptic digest of BSA for all 675 spots on a 102x108 mm sample plate with automatic 2-point internal Calibration (D0 and D1) with D3 and D4 fixed.

Serial Dilution Experiment Peptide Standards Equimolar mixture diluted in 4-hydroxy-a-cyanocinnamic acid from 1 picomole/mL to 1 attomole/mL Automated acquisition and calibration averaging 2000 laser shots/spectrum at 1 kHz No data processing MH+ No. Angiotensin Fragment 1-7 899.47 1 des-Pro2 Bradykinin 963.51 2 Angiotensin II Acetate 1046.54 3 Angiotensin I 1296.69 4 Neurotensin 1672.92 5 ACTH (18-39) 2465.20 6 Glu-1 Fibrinopeptide B 1570.68 7(<10% pure)

100 fm 2 1 4 6 3 5 7 3 1 fm 1 fm 2 4 6 5 7 Examples of spectra from serial dilution of peptide standards

100 am 4 5 6 10 am

5 6 4 1 attomole Threshold 100 ions/peak 5 6

threshold 100 ions/peak 4 Matrix blank, 2V digitizer range, major matrix peaks off-scale

Test Mixture for Determining Dynamic Range Peptide Conc. (fmole/mL) MH+ No. Angiotensin Fragment 1-7 10,000 899.47 1 Angiotensin II Acetate 1,000 1046.54 3 des-Pro2 Bradykinin 100 963.51 2 Angiotensin I 2 1296.69 4 Neurotensin 1 1672.92 5 Glu-1 Fibrinopeptide B (synthetic) 0.2 1570.67 7 ACTH (18-39) 0.01 2465.20 6 BSA digest 0.1-1 1479.79 B 927.49 1283.71 1639.94, etc

Normal laser 1 3 4 B 5 B 7 2 5 Higher laser B B B B 7 B B B B

100 fm/mL peptide standards 200 Hz 5 kHz No detectable effect of laser rate on spectrum quality.

Conclusion and Future Present status Resolving power > 30,000 for peptides Mass error < 2 ppm RMS over entire sample plate Detection limit ~1 attomole/mL Dynamic range ~105 Performance independent of laser rate (to 5 kHz) Future goals Resolving power >100,000 Mass error < 1 ppm RMS 10x improvement in detection limit and dyanamic range Isotopic resolution of proteins to 30,000 Da Multiplexed MS-MS with >4000 resolving power for precursor selection and very high sensitivity and throughput Instruments designed for specific applications, e.g. isotope ratios at ppt level limited by chemical noise

Chris Vestal The Boss Marvin Vestal Head Virgin Steve Hattan Analytical Chemist Mark Dahl EE Matt Gabeler-Lee SW Kevin Hayden Ion Optics ME Joe Valentine EE Joe Fitzpatrick Design Eng. Roger Voyer Machinist Steve Gabeler EE George Mills SW