Lung Volumes and Gas Distribution - Testing Equipment RET 2414L Pulmonary Function Testing Module 3.1.

Slides:



Advertisements
Similar presentations
Ventilation-Perfusion Relationships
Advertisements

Chapter 7 : Nutrition in Plants 7.1 Photosynthesis
Teledyne Analytical Instruments
Medical Instrumentation Application and Design, 4th Edition
Ventilation / Ventilation Control Tests - Equipment and Equations
ANESTHETIC MACHINES.
CARBON DIOXIDE ABSORPTION
Operating Nitrox Systems 11 Presented by Nuvair. Operating Nitrox Systems2 Membrane systems produce nitrox efficiently Compressed air flows through the.
Flow and Volume of Blood Dept. of Biomedical Engineering YOUNHO HONG.
Anesthesia Machine Presented by Gil Soto C.R.N.A.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 1 of 4.
Anesthesia Apparatus Checkout Recommendations
Capnography Erika A. Gibson, DVM Michigan State University
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
The bird IPPB machine The Bird is pneumatically powered & controlled.
Monitoring Air Quality: Emissions and Odor From A Swine Finishing Livestock Facility.
ACADs (08-006) Covered Keywords analytical balance, conductivity bridge with flow cell, inductively coupled plasma analyzer, in-line sodium monitor, ion.
Check Valves and Manufacturer-Recommended Leak Test
Flame photometry.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Anesthetic Equipment.
1 Oxygen Administration. 2 Introduction  Oxygen is a colorless, odorless, tasteless gas that is essential for the body to function properly and to survive.
Normal Operating Parameters Maintenance Schedule Common Problems Normal Operating Parameters Maintenance Schedule Common Problems ClearWater Tech, LLC.
Wasted Ventilation. Dead Space dead space is the volume of air which is inhaled that does not take part in the gas exchange, either because it (1)
بسم الله الرحمن الرحیم Fariba Rezaeetalab Assistant professor,pulmonologist.
Medical Instrumentation Application and Design, 4th Edition
Module 2 Exchange and transport Measuring lung capacity.
RESPIRATION Dr. Zainab H.H Dept. of Physiology Lec.5,6.
PART 3: Breathing Circuit
Parts of the Anesthetic Machine
Ventilation / Ventilation Control Tests
Scavenging System A&A Pg 356  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected.
Spirometry & Related Testing Equipment RET 2414L Pulmonary Function Testing Module 2.0.
Atomic Emission Spectroscopy
Water Handling Operations
생체계측 II Report # 송성진 Medical Instrumentation II.
8 th 정진웅 MEDICAL INSTRUMENTATION. Chapter. 9 Respiratory System Measurement.
ENTC 4350 Modern Ventilators. Modern Ventillators Ventilation assistance is provided under either of two conditions: (1) breathing initiated by a timing.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3 rd ed. New York: John Wiley & Sons, Figure 9.1 Models of the lungs.
Without reference, identify principles about volume/pressure and high frequency ventilators with at least 70 percent accuracy.
Quality Control Spirometry & Lung Volumes
ANAESTHESIA BREATHING CIRCUITS
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
PART 3: Breathing Circuit
Gases. Properties of Gases Fluidity - Gas particles move around freely with negligible attraction to fill the shape of their container. Fluidity - Gas.
Anesthesia Machine Circuits
The purpose of the breathing rhythm is to ventilate the lungs to allow delivery of oxygen to the alveoli, and elimination of the waste gas carbon dioxide.
Chapter 9 Lung, Kidney, Bone and Skin. Figure 9.1 The spirometer measures lung capacities and lung volumes. Because the subject cannot make the lung volume.
Carbon Dioxide Monitoring
Parts of the Anesthetic Machine A&A Pages
PART 3: Breathing Circuit
Measurement of Human Energy Expenditure
PART 3: Breathing System Parts
Therapeutic equipment
Ch 09 Measurement of the Respiratory System
Hemodialysis Lecture 3.
Pneumatic schematic of the Dräger Infinity V500 intensive care ventilator. A. Gas-mixture and gas-metering assembly. Gas from the supply lines enters the.
“Respiratory equipments”
PNEUMOTACHOMETERS.
The Human Body Chemical Plant reactions membrane separations
Cardiac Output O2 Saturation Capillary Refill
Non-Rebreathing System
(bx•Qb)in Ubx QAWO (AWOx •QAWO)in AWO (bx•Qb)out
Breathing Systems Tom Williams.
PNEUMOTACHOMETERS.
PNEUMOTACHOMETERS.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Diagram of rebreathe apparatus.
Presentation transcript:

Lung Volumes and Gas Distribution - Testing Equipment RET 2414L Pulmonary Function Testing Module 3.1

Lung Volumes / Gas Distribution  Open-Circuit N2 Washout

Lung Volumes / Gas Distribution  Open-Circuit N2 Washout Equipment Required  Spirometer Tissot; a water-sealed spirometer with a volume of approximately 100 liters. Designed for collecting large volumes of exhaled air Pneumotachometer

Lung Volumes / Gas Distribution  Open-Circuit N2 Washout Equipment Required  Free- Breathing valve Valve is opened or closed either manually or by computer control to allow patient to be switched from breathing room air to breathing gas from spirometer or special circuit  Deadspace of valve is determined by closing off all ports and filling the valve with water. The volume of water is the deadspace volume  Oxygen source

Lung Volumes / Gas Distribution  Open-Circuit N2 Washout Equipment Required  Nitrogen analyzer Giesler Tube Ionizer  Measurement on the basis of photointensity (emission spectroscopy)  Sample is pumped (vacuum pump) into an ionizer chamber where a light is emitted and monitored by a phototube. Intensity of light is directly proportional to the percent of N2

Lung Volumes / Gas Distribution  Open-Circuit N2 Washout Equipment Required  Nitrogen analyzer Measurement range (0 – 80%) Calibrate to zero using 100% O2 (0% N2) and a known concentration of N2 Troubleshooting  Sporadic or erratic zero readings which return to the expected value are caused by aspiration of water through the needle valve

Lung Volumes / Gas Distribution  Single-Breath N2 Washout Equipment Required  Same as equipment used for open-circuit N2 washout

Lung Volumes / Gas Distribution  Closed-Circuit He Dilution

Lung Volumes / Gas Distribution  Closed-Circuit He Dilution Equipment Required  Spirometer Volume displacement  Directional breathing circuit  Free breathing valve  He source  O2 source Allows addition of O2 during test to replenish O2 consumed by patient  Maintains constant system volume

Lung Volumes / Gas Distribution  Closed-Circuit He Dilution Equipment Required  He analyzer Thermal conductivity analyzer (wheatstone bridge) Measurement range (0 – 10%) Calibrate to room air (0% He) and a know amount of He (10%), then re-zero. CO2 and water vapor must be removed from gas prior to being analyzed

Lung Volumes / Gas Distribution  Closed-Circuit He Dilution Equipment Required  Fan or Blower Promotes gas mixing within a rebreathing circuit  CO2 absorber (Scrubber) Needed for tests requiring rebreathing Sodium hydroxide (NaOH) – “Soda lime”  Light brown to white - exhausted Barium hydroxide (BaOH2) – “Baralyme”  White to purple – exhausted By product is water  Place upstream from H2O absorber

Lung Volumes / Gas Distribution  Closed-Circuit He Dilution Equipment Required  H2O absorber Removes moisture as patient breaths through the system Moisture adversely affects analyzer Calcium sulfate (CaSO4) Place in-line after CO2 absorber Desiccant color changes from blue to pink when exhausted

Lung Volumes / Gas Distribution  Body Plethysmograph AKA: “Body Box”

Lung Volumes / Gas Distribution  Body Plethysmograph Two types  Constant-volume, variable- pressure “Pressure Plethysmograph”  Flow or variable-volume “Flow Plethysmograph” Both measure thoracic gas volumes and airway resistance and it derivatives Both use a pneumotachometer to measure flow and a mouth transducer with a shutter to measure alveolar pressure

Lung Volumes / Gas Distribution  Body Box Calibration  Mouth pressure is verified with a mercury or water barometer  Flows are verified with a rotometer (flow- metering device) or a 3-liter syringe  Box pressure is calibrated by using a sine-wave rotary pump simulates inspiratory/expiratory volumes