ICT – Intraoperative Corneal Topography for image registration purposes S.Schründer
Model eye 6mm optical zone 8mm corneal radius 6mm pupil, 4mm below apex 5mm to rotation axis 12mm radius bulbus
Tilted eye
Motivation Aberrometer Ablation pattern Tilted eye 1st wrong pattern TreatmentDiagnosis Ablation pattern Tilted eye 2nd wrong pattern Eyetracker
Solution Registration in 3d space
Requirement 3d objects
Height map pre ablation* *Difference to 8.4mm sphere Range 100µm
Height map post ablation* *Difference to 8.7mm sphere Range 100µm
Fine registration: Iterative Closest Point Find corresponding 3d points (via Hough tables) Minimize their distance Distance Minimization: 1.Levenberg-Marquardt for fast convergence 2.Simmulated annealing to overcome local minima
Registration points area Detected flap edge
Difference map after registration 7.5mm circle Range 105µm
Line scans Dimensions in mm
3d Flap thickness 9.5mm circle Range 290µm
Line scans Dimensions in mm
Next steps Rigid incorporation into a laser system Extended clinical trials Stand alone device
Centered cornea
0.5mm lateral shift
Corresponding tilt 0.22mm decentration
Linear function* *8mm corneal radius, 4mm apex/pupil, 13mm apex/rotation axis
Andreas Dankert Hendrik Fuchs Saso Spasovski Coworkers at BioShape
Thank you. bioshape.com