B. Spatial coherancy & source size Spetial coherancy is related to the size of the source. Source size governs spatial coherancy and maller source sizes.

Slides:



Advertisements
Similar presentations
Chapter 4 Companion site for Light and Video Microscopy Author: Wayne.
Advertisements

Electron Optics Basic Introduction Bob Ashley
Electron Optics.
Do it with electrons ! II.
Scanning Electron Microscope (SEM)
Groups: WA 2,4,5,7. History  The electron microscope was first invented by a team of German engineers headed by Max Knoll and physicist Ernst Ruska in.
 A. An optical instrument that uses a lens or combination of lenses to magnify and resolve the fine details of an object.
Components of Initial Eprobe Startup Electron Gun –turning on beam, setting beam current, acceleration, saturation level Focusing lenses –Permit focussed.
Chapter 23 Mirrors and Lenses.
OPTICS. I. IMAGES A. Definition- An image is formed where light rays originating from the same point on an object intersect on a surface or appear to.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
Electron Optics Two essential components: 1)Electron source (gun) 2)Focusing system (lenses) Add scanning apparatus for imaging Electron gun Cathode Anode.
Chapter 23 Mirrors and Lenses.
PRINCIPLES OF MEASUREMENT AND INSTRUMENTATION EKT 112
Lecture 5.1 Scanning Electron Microscopy (SEM)
Electron Microscope. Light Resolution  The resolution of a microscope is limited by the diffraction of light. Single diffractionSingle diffraction 
Scanning Electron Microscope Jamie Goings. Theory Conventional microscopes use light and glass lenses SEM uses electrons and magnetic lenses to create.
The TEM system and components:
Transmission Electron Microscopy (TEM) By Austin Avery.
Optical Instruments Chapter 25.
Chapter 25: Interference and Diffraction
Chapter 23 Mirrors and Lenses.
Introduction to scanning electron microscopy
Example: What is the ratio of the irradiances at the central peak maximum to the first of the secondary maxima?
Ch 25 1 Chapter 25 Optical Instruments © 2006, B.J. Lieb Some figures electronically reproduced by permission of Pearson Education, Inc., Upper Saddle.
Transmission Electron Microscopy
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
Microscope.
Chapter 25 Optical Instruments.
Electrons Thermionic Emission
Visual Angle How large an object appears, and how much detail we can see on it, depends on the size of the image it makes on the retina. This, in turns,
Lenses Chapter 30.
Lenses in Combination The analysis of multi-lens systems requires only one new rule: The image of the first lens acts as the object for the second lens.
Chapter 23 Mirrors and Lenses.
Introduction to transmission electron microscopy
Chapter 34 Lecture Eight: Images: II. Image Formed by a Thin Lens A thin lens is one whose thickness is small compared to the radii of curvature For a.
Scanning Electron Microscope (SEM)
Chapter 13 The Characteristics of light. Objectives Identify the components of the electromagnetic spectrum. Calculate the frequency or wavelength of.
Lesson 25 Lenses Eleanor Roosevelt High School Chin-Sung Lin.
Scanning Electron Microscope (SEM)
Transmission Electron Microscope Basic premise of a TEM is to project a magnified image of the specimen onto a fluorescent screen where it can be viewed.
Lecture 26-1 Lens Equation ( < 0 )  True for thin lens and paraxial rays.  magnification m = h’/h = - q/p.
The Nature of Light. Light Can Act Like Waves or In 1801 Thomas Young an English scientist did an experiment. –Double slit experiment Passed a beam of.
Thin Lenses A lens is an optical device consisting of two refracting surfaces The simplest lens has two spherical surfaces close enough together that we.
The law of reflection: The law of refraction: Image formation
Transmission Electron Microscope
1 The law of reflection: The law of refraction: Snell’s Law Image formation.
Electrons Thermionic Emission
SARDAR PATEL INSTITUTE OF TECHNOLOGY E.NO : Guide By:- V.N.Thakkar.
Part 10 Optics --Mirrors and Lenses Chapter 24 Geometric Optics.
REFRACTION OF LIGHT & OPTICAL INSTRUMENTS Chapter 14.
Copyright © 2009 Pearson Education, Inc. Chapter 35-Diffraction.
Do it with electrons !. Microscopy Structure determines properties We have discussed crystal structure (x-ray diffraction) But consider now different.
METHODOLOGY Nanotechnology Prof. Dr. Abdul Majid Department of Physics
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Electron sources and guns
Introduction to scanning electron microscopy
Microscope.
Introduction to scanning electron microscopy
Chapter 35-Diffraction Chapter 35 opener. Parallel coherent light from a laser, which acts as nearly a point source, illuminates these shears. Instead.
Laboratory equipment Lecture (3).
METALLURGICAL MICROSCOPE
TEM and SEM.
Chapter 35-Diffraction Chapter 35 opener. Parallel coherent light from a laser, which acts as nearly a point source, illuminates these shears. Instead.
Do it with electrons !.
Electrons Thermionic Emission
SEM Imaging - Magnification and Deflection System
Microscopy.
Presentation transcript:

b. Spatial coherancy & source size Spetial coherancy is related to the size of the source. Source size governs spatial coherancy and maller source sizes give better coherancy. we define the distance dc, the effective source size for coherent illumination to be, D c << / 2  Where, = electron wavelength  = angle subtended by the source at the specimen Small beams are more spatially coherent than large beams. The more coherent the beam is, the better the quality of the high resolution phase contrast images, the sharper the diffraction patterns, better the diffraction contrasts in images of crystalline specimens. c. Stability The thermionic sources are generally very stable when compared to field emission sources.It is important that the electron current coming from the source should be stable, with stable high- voltage supply, otherwise, the screen intensity will vary, making it difficult to take correct exposed images & making microanalysis impossible in many cases. In summary, A smaller source size gives higher brightness & better spatial coherancy but less stability. 2.3 ELECTRONIC GUNS a.Thermionic gun Schematic diagram of a thermionic electron gun.A high voltage is placed between the filament and the anode,modified by a potential on the wehnelt which acts to focus the electrons into the crossover. The three major parts of a thermionic gun,from top to bottom: Cathode,the wehnelt cylinder and anode shown separately

Electrons in the thermionic gun are accelerated across a potential difference of the order 100KV between the cathode (at high negetive potential) and anode (at ground potential). Beam shaping & control of emission is a affected by properties of the shield (gun cop;wehnelt) and acceleration by the anode. In a biased gun the shield is maintained at a potential between 100 – 500 KV negative relative to the filament. This negative voltage serves to repel some emitted electrons back to the filament, and reduces total emission and brightness. The negetive potential of the shield with respect to the filament gives rise to a strong electrostatic field around the shield which acts as an electrostatic lens focusing the true source to form an image of the source below the anode. The gun cross over is used as the actual source of electrons for the electron microscope. b. Field emission gun The FEG is much simpler than the thermionic guns, Anode 1: provides the extraction voltage to pull electrons out of the tip. Anode 2: accelerates the electrons to 100KV or more. The electrons are accelerated through the applied potential by the second anode.The combined fields of the anodes act like a more refined electrostatic lens to produce a crossover. This lens controös the effective source size and position, but it isn‘t very flexible. Incorporating a magnetic lens into the gun gives a more controllable beam and larger brightness. Drawbacks of FEG: The source size is very small needs UHV conditions. UHV technology is expensive & requires much higher level of operator competence. Fig:Electron paths from a FES showing how a fine crossover is formed by two anodes acting as an electrostatic lens.sometimes an extra (gun)lens is added below the second anode

2.4 CONDENSER SYSTEM Operation of a two-lens condenser system for illuminating the specimen a.under focus and c. Over focus b. in-focus operation with condenser lens c2 d. Additional use of condenser lens c1 to demagnify the crossover,the demagnified image then being focused on the specimen with condenser lens C2 The function of the condenser lens system is to focus the electron beam emerging from the electron gun onto the specimen to permit optimal illuminating conditions for visualizing and recording the image A double condenser system adds considerable flexibility to the illuminating system by allowing a wider range of intensities with a given gun adjustment and making it possible to reduce the area of the object which is irradiated. A strong first lens (c1: short focal length) is used to produce a demagnified image (nearly 1  m diameter) of the electron source and a weaker second lens (c2: long focal length) projects this demagnified image onto the specimen plane producing a slight magnification so the final focused beam size is about 2-3  m. The focal length of c2 can be varied to spread the beam over a larger area of the specimen for example to record images at low magnification(<10000X). The fig. Shows the important modes of operation of a two lens condenser system. The beam divergence  0 i,e the divergence of radiation incident onto a specimen point is given by the diameter of the c 2 aperture. The image intensity is proportional to the current density J 0, which is directly related to  0 through brightness  (J 0 =    0 2 ). The current density and beam divergence are maximum when condenser 2 is focused. The diameter of the illuminating object region can be enlarged by the means of over focussing & under focussing of the condenser lens. The beam divergence is now determined by the second image of the crossover. Since this image is much smaller than the c2 aperture, the beam divergence is strongly decreased. At the same time the current density and thus the image intensity are reduced. Small beam divergences of < 1 mrad are required to obtain sharp diffraction spots. The minimum beam diameter which can be generated by the two condenser lenses is order of 1  m, but for analytical investigations for example X-ray microanalysis, it would be useful to have probes with a diameter << 0.1  m. this can be obtained by using condenser objective lenses.

2.5 OBJECTIVE LENS & INTERMEDIATE LENSES The optical enlarging system of an electron microscope consists of an objective lens followed by one or more projector lenses. The objective lens determines resolution & contrast in the image, and all subsequent lenses bring the final image to a convenient magnification for observation and recording. The objectve lens is more critical lens since it determines the resolving power of the instrument and performs the first stage of image forming. The optic of objective lens is shown in figure. A diffraction pattern is formed in the back focal plane of the lens. The first projector lens ( often called the intermediate or diffraction lens ) can usually be switched between two settings. (i) In the image mode it is focused on the image plane of the objective. The magnification of the final image on the microscope screen is then controlled by the strength of the remaining projector lenses. (ii) In the diffraction mode the intermediate lens is focused on the back focal plane of the objective and the diffraction pattern is projected on to the viewing screen. Requirements in the construction of the objective lens 1. The specimen must be situated close to the front focal plane of the objective to provide an initial magnification of 50 – 100 X. 2. Focal length should be as small as practical to insure minimum chromatic and spherical aberration. Since these decrease as the focal length decreases. The specimen has to be placed inside the lens field to obtain the necessary short focal length and this poses a problem of introducing a specimen in to the confined space of the lens. 3. There must be adequate clearance for insertion of specimen, aperture and anticontaminator. 4. There must be provision for inserting electrical or magnetic devices (stigmators) to correct for minute asymmetries in the lens field. Fig:The objective lens and first intermediate lens.the objective lens(OL)is focused on the specimen and forms an intermediate image as shown in (a).In imaging modes the intermediate lens(IL)magnifies this image further and passes it to the projector for display. In order to make the diffraction pattern visible,the intermediate lens is refocused on the back focal plane of the objective lens(BFP)and the diffraction pattern is passed to the projector system.