Without reference, identify principles about Electrosurgical Units with at least 70 percent accuracy.

Slides:



Advertisements
Similar presentations
Effect of Electricity on Living Tissue
Advertisements

Manufacturing Processes Lab I MET 1321 Stick Welding
ELECTRICITY 1. Electricity is a form _________ 2. _____________ is a good conductor. 3. ___________ _______________: path electric current from generating.
Two-mode Safety and Battery System Service
KENTAMED Radio B is a cutting-edge technology 4MHz radiofrequency surgery unit. It combines the well-known tissue spearing effects of the radiofrequency.
Surgical Diathermy Machine
Kenny Low Dec 2009 Urology registrar Thursday teaching
SMAW (Stick Welding) SECTION OVERVIEW:
EE 4BD4 Lecture 25 Electrosurgery Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
Introduction to (SMAW) Shielded Metal Arc Welding
Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.
Electrocautery.
Section 2 Safety, Tools and Equipment, Shop Practices Unit 4 General Safety Practices.
Frog Dissection.
3.03 Understand support services
Surgical Instruments and Aseptic Technique
Electrical Surgical Unit
© 2005 Pearson Prentice Hall This work is protected by United States copyright laws and is provided solely for the use of instructors in teaching their.
General Industrial Battery Safety Changing and Charging Storage Batteries (g)  Battery charging installations shall be located in areas designated.
What is a Watt? Everything you ever wanted to know about Electrosurgery. Fahad bamehriz, MD.
Unit 4, Lesson 9 Keeping You and Your Computer Safe AOIT Computer Systems Copyright © 2008–2013 National Academy Foundation. All rights reserved.
Physiological Effects of Electricity on Human Body by Mohd Yusof Baharuddin.
SITE SAFETY HAZARDS AND PROCEDURES. ELECTRICAL HAZARDS  Electricity is a serious workplace hazard.  Employees can be exposed to dangers such as : 
What is electrosurgery?
Diathermy.
Current Sensing for Navigated Electrosurgery K. Carter, A. Lasso, T. Ungi, E. Morin, G. Fichtinger.
Copyright © 2008–2011 National Academy Foundation. All rights reserved. Unit 4, Lesson 9 Keeping You and Your Computer Safe AOIT Computer Systems.
HAZARD OF MEDICAL INSTRUMENT II by Mohd Yusof Baharuddin.
Chapter 28 Surgical Instruments and Aseptic Technique.
BASICS OF ELECTROSURGERY
Electrosurgery & Gyn Surgery: Get the Point Across
Chapter 7 Safety. Introduction This chapter covers the following topics: Dangers of electricity Preventive measures Electrostatic discharge Safety practices.
Computer controlled I NSTANT R ESPONSE ™ system provides safer and more precise electrosurgery. TM.
Argon Enhanced Electrosurgery. Characteristics of argon gas (helium, neon, freon etc. also possible, but.....) inertnon-explosive heavier than air not.
Equipment Source- Alexander’s Care of the Patient in Surgery Patty Pavlikowski RN CST MA.
PREPAIRED AND PRESNETED BY: DR.HASAN AL-SHEHRI GS RESIDENT UNDER SUPERVISION OF : PROF : MOHMMED ALAM GS CONSULTANT PRINCIPLES AND COMPLECATIONS DIATHERMY.
Electro surg i cal unit  PURPOSE ESU are used for surgical cutting and for controlling bleeding by causing coagulation “ hemostasis ” at the surgical.
Dr.Moallemy Radiofrequency Lesioning. Dr.Moallemy  Radiofrequency (RF) current is used in pain medicine to make discrete therapeutic lesions in various.
Electro surgery in Gynaecology
Fractional Ablation Course V Ver 1.0
RDT Hospital, Dept of Gen Surgery
Technological Sciences for the Operating Room
Electrosurgical Unit.
PHYSICS – Dangers of electricity. LEARNING OBJECTIVES Core State the hazards of: – damaged insulation – overheating of cables – damp conditions State.
GMAW (MIG Welding). 2 GMAW Unit Topics During this overview, we will discuss the following topics: Safety GMAW Basics Equipment Set-Up Welding Variables.
VACUUM CIRCUIT BREAKER. WHAT IS A CIRCUIT BREAKER A circuit breaker is an equipment that breaks a circuit either manually or automatically under all conditions.
Electrosurgery Units ELECTROSURGICAL UNITS (ESUs)
SAFE USE OF DIATHERMY RACHEAL AFOLABI, RN, RM, H. Dip. PERI-OPERATIVE NURSING, H.Dip. HEALTH SYSTEM MANAGEMENT.
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN LECTURE NOTE 7 Electrosurgical Devices BME 312-BMI II-L7- ALİ IŞIN 2014 FACULTY OF ENGINEERING.
UNIT-5. ELECTRICAL SAFETY, WIRING & INTRODUCTION TO POWER SYSTEM
Electro surgical unit.
Biomedical Instruments Design Biopotential Amplifiers
Basics of Medical Equipment Lecture 6
Welding Safety – The Basics
Welding Safety – The Basics
General Safety Practices
Electrosurgical/Surgical Diathermy Units
Electro surgery in Gynaecology
A. In monopolar electrosurgery, current from an electrosurgical generator passes from an active electrode (the “Bovie” tip) through the patient to a return.
Electrosurgery in Periodontics
SURGICAL DIATHERMY.
SURGICAL DIATHERMY.
Diathermy and its safe use
Biomedical Instruments Design Biopotential Amplifiers 1.
A. Touching both terminals with the hands can cause electrical shock
Fundamental Use of Surgical Energy
PHYSICS – Dangers of electricity. LEARNING OBJECTIVES Core State the hazards of: – damaged insulation – overheating of cables – damp conditions State.
Presentation On Schottky Diode. Course Code:3208 Course Title : Microwave radar and satellite communication lab Presented By Salma Akter BKH F.
Presentation transcript:

Without reference, identify principles about Electrosurgical Units with at least 70 percent accuracy.

 Purpose of an Electrosurgical Unit (ESU) ◦ A system that uses radio frequency to  Sever tissue  Stop bleeding (hemeostasis) ◦ Units operate in the range of 300 kHz to 3 MHz

 Types ◦ Monopolar  Active electrode - one side of the generator  The electrosurgical effect takes place at the active electrode only  Typically offers the best alternative for wide surface coagulating and dissection of tissue  Return electrode  The other side is known as o patient plate o Reference electrode o dispersive electrode

 No electrical effect is intended or desired  Needed to complete the electrical circuit  Placement  As close as possible to the surgical site  Minimize the distance that the current will need to travel through the body ◦ Bipolar  Both sides of the generator are connected to a surgical handpiece  Allows for better current control  The current passes between two poles of a forceps rather than through the entire body

 Surgical Effects and Operational Modes ◦ Coagulation  Used to control or stop bleeding  It is a low power technique  Uses an interrupt current to supply controlled dehydration  Can also be used for destroying undesirable tissue masses ◦ Cutting  Heats tissue so rapidly that cells explode and contents vaporize leaving an incision  When used, the electrode is separated from the tissue by a thin layer of steam  Used like a stainless steel scalpel  Little or no control over bleeding

◦ Blend  Hybrid of cut and coagulate  Used when surgeon desires more hemeostasis effect than cut provides  This mode saves time, as only one pass with electrode is necessary ◦ Spray (fulguration)  Used for rapid control of bleeding across a wide area  A non-contact type of coagulation  The current sparks or jumps from active electrode to the tissue  Uses  To destroy surface layers of cells in the bed from which a lesion has been removed  Ideal for sealing off hidden bleeders due to their path of least resistance

 Power ◦ Can be varied for all modes ◦ Setting depends on tissue type, procedures, and the type of electrode used ◦ Low power  Neurosurgery  Oral ◦ Medium power  General surgery  Abdominal opening ◦ High power  Open the chest  Mastectomies

 Precautions ◦ Return plate burns  Protect the patient from being burned  The pad must have as much direct contact with the patient as possible  Edge effect – use a pad that has a larger edge to lower current density  Monitor Systems  Used to prevent a patient from a pad related burn  The ESU is designed to send a small current up one side of the pad through the patient’s skin and back through the other side of the pad  If the ESU senses a change in resistance from the pad to patient contact it will set off an alarm

◦ Alternate current path burns  When the pencil is activated there is opportunity for RF leakage interference  Wrapping surgical cable around metallic material and securing it to a flammable material may lead to a fire hazards  You need to be aware of the hazards of RF leakage and choose products that are shielded properly from these hazards  Prevent electrical shock  Know the basic principles of electricity, current density, and the path of least resistance  Never activate in open circuit