“Evaluating and Defining the Sharpness of IOLs: Microedge Structure of Commercially Available Square-Edge Hydrophobic IOLs” Matthias Müller, PhD, 1 Liliana.

Slides:



Advertisements
Similar presentations
VisTor The new Toric IOL by Hanita Lenses
Advertisements

YAG capsulotomy K.P.SHANTHA SORUBARANI.
“Cataract formation after implantation of phakic posterior chamber intraocular lenses: A brief overview and review of the literature” Liliana Werner, MD,
Straylight (disability glare) results in case of a diffractive multifocal IOL design with apodization pattern adjusted to reduce glare. Ruth Lapid-Gortzak.
Efficacy of 360-Degree Square Edge on PC IOLs: Laboratory Study
Guy Kleinmann Kaplan Medical Center, Rehovot, Israel Financial disclosure: Dr. Kleinmann is a consultant for Hanita Lenses, Israel.
Visual outcome & subjective visual symptoms of the Tecnis ZM900 multifocal intraocular lens in Asian eyes Dr Colin S.H. Tan MBBS, MMed (Ophth), FRCSEd.
NEW TRULIGN™ TORIC IOL Surgeon Training
Forms of Cataract 1) Developmental (in general non progressive)
Astigmatism Following 2 IOL Injection Techniques: Wound Assisted Versus Wound Directed Jay J. Meyer, MD Hart B. Moss, MD Kenneth L. Cohen, MD University.
Ruth Lapid-Gortzak MD PhD 1,2, Jan Willem van der Linden BOpt 2, and Ivanka J. van der Meulen MD 1,2 1 Department of Ophthalmology, Academic Medical Center,
Intraocular lenses for small incision surgery
Mayank A. Nanavaty, DO, MRCOphth, MRCS(Ed) David J. Spalton, FRCP, FRCS, FRCOphth James F. Boyce, PhD Thomas J. T. P. Van den berg, PhD St. Thomas’ Hospital,
Choosing the Proper Power for the IOL
Clinical evaluation of foldable acrylic phakic IOL (fP) implantation ASCRS, San Diego, 2011 A.John Kanellopoulos, MD Professor NYU Medical School, NY Director,
Anterior Chamber Depth, Iridocorneal Angle Width, and Intraocular Pressure Changes After Phacoemulsification: Narrow vs Open Iridocorneal Angles Huang.
“Calcification of Different Designs of Silicone Intraocular Lenses in Eyes with Asteroid Hyalosis” Liliana Werner, MD, PhD Jack Stringham, MS, Bryan Monson,
Second generation: Collamer™ (?) Fyodorov lens. Third generation: Staar ICL -Material: Hydrophilic collagen polymer (63% hydroxy- ethyl-methyl-acrylate;
Intraocular lens dislocation secondary to haptic torsion Lawrence E. Lohman, MD FACS Matthew C. Willett, MD.
Adhesion of Human Lens Epithelial Cells on Hydrophilic Acrylic IOLs Coated With Polyethylene Glycol in Culture Authors: Antonio Carlos Lottelli Rodrigues,
Capsular Tension Rings: Current Indications and Outcomes Maryam Mokhtarzadeh, MD Jayne S. Weiss, MD John M. Ramocki, MD No financial conflicts to disclose.
Comparison of Nd:YAG Laser Capsulotomy Rates after implantation of 2 IOLs: One Hydrophilic Acrylic and the other Silicone with sharp posterior optic edge.
“Pathological Assessment of the First Human Eyes Obtained Postmortem Implanted with the Bag-In-The-Lens Design” Liliana Werner, MD, PhD 1 Marie-José Tassignon,
Repeatability of Anterior and Posterior Corneal Higher- Order Aberrations in 4 mm, 6 mm and 8 mm diameters measured by Pentacam System Jadwiga Wojtowicz.
PIGGYBACK IMPLANTATION OF FOR HIGH HYPEROPIA 24 MONTHS FOLLOW UP JL. FEBBRARO I.VIELPEAU, F. KRIFA, S.BERBACHE, S. GANEM Department of Ophthalmology S.
Recent Advances in Intraocular Lenses Jim Schwiegerling, PhD Ophthalmology & Vision Sciences Optical Sciences.
ASCRS/ESCRS Survey on Foldable IOLs Requiring Explantation or Secondary Intervention 2005 Update.
Phaco in post- vitrectomy cataracts George Kampougeris MD, MRCSEd, PhD Consultant Ophthalmic Surgeon
“ ” “Clinico-pathologic correlation of capsulorhexis phimosis with anterior flexing of single-piece hydrophilic acrylic intraocular lens haptics” Liliana.
Unilateral multifocal lens implantation in patients with a contralateral monofocal or phakic eye is a viable presbyopic correction option Robert J. Cionni,
Rumex International Co. Bridge between innovative technology and perfect surgery Rumex International Company th Street North, Suite 317 Clearwater,
INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY Burak Bilgin**, M.D.,
Phakic posterior chamber IOLs 1.Cells involved in crystalline lens and capsular bag opacification 2.Evolution of designs of PPCIOLs 3.Relevant aspects.
Long term follow up of Impact of Cortico Cleaving Hydrodissection on Posterior Capsule Opacification (PCO) after paediatric cataract surgery Samaresh Srivastava,
Rumex International Co. Bridge between innovative technology and perfect surgery Rumex International Company th Street North, Suite 317 Clearwater,
Authors: Lawrence Strenk, PhD, 1 Liliana Werner, MD, PhD, 2 Nick Mamalis, MD, 2 Susan Strenk, PhD 1 From: 1) MRI Research, Inc., Cleveland, OH; 2) Moran.
Analysis of Intraoperative Capsular Trauma Induced by a New Single-Use Polymer Irrigation/Aspiration Tip During Posterior Capsule Polishing Don Davis,
Adriana S. Forseto1, MD Walton Nosé1,2, MD
Influence of IOL optic material on posterior capsule opacification and visual function Ken Hayashi, MD Hideyuki Hayashi, MD Hayashi Eye Hospital, Fukuoka,
Edge profile of commercially available square-edge intraocular lenses
Futoshi Taketani, MD,PhD,
Johns Hopkins Hospital
Early capsular bag contraction with haptic dislocation following implantation of a flexible hydrophilic acrylic “psuedoaccomodating” IOL Nigel Morlet FRACS.
LogMAR-Analysis of multifocal intraocular lenses: Clinical performance A. Mannsfeld, I.-J. Limberger, A. Ehmer, M.P. Holzer, G. U.Auffarth International.
Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses. Rodrigo.
Minimizing Risk in Visian ICL Implantation.
9-Month Results after Implantation of a new accommodative IOL that works with one focus Mark Tomalla M.D.* Clinic for Refractive and Ophthalmic Surgery,
Simulated Experiments on IOL Power Calculation Using Anterior Segment OCT Dong Hyun Jo, M.D., 1,2 Mee Kum Kim, M.D., 1,2 Won Ryang Wee, M.D. 1,2 1 Department.
Examination of posterior capsular opacity under intraocular tamponade using silicone oil. Norihito Gotoh, Hiroyuki Matsushima, Koichiro Mukai, Tadashi.
Precision with the VisTor IOL Prof. Med. Manfred Tetz ESCRS 2015 Barcelona.
Dissatisfication After Multifocal Intraocular Lens Implantation in Taiwan Yu Wei Lin, MD (Presenting Author); Ching-Ju Hsieh; Lin-Chung Woung The authors.
Liquifaction Method and Extent of Posterior Capsule Opacification: Two-Year Follow-up Marie Kalfertova, Mariya Burova, Pavel Rozsival, Nada Jiraskova Nada.
Kavita Gala David Spalton Mayank Nanavaty St Thomas’ Hospital , London
TEK-CLEAR ACCOMMODATIVE IOL Dr. Mark Wevill, Ultralase, UK Presbyopia International Barcelona 2009 No financial disclosures.
OUR EXPERIENCE WITH PRELOADED IOL CT LUCIA 601P(Y)
Department of Ophthalmology, University of Ulsan College of Medicine,
Postoperative Refraction and Patient Satisfaction after Bilateral Implantation of Presbyopia-Correcting Intraocular Lenses Robert Cionni, MD Financial.
Invest. Ophthalmol. Vis. Sci ;53(1): doi: /iovs Figure Legend:
Invest. Ophthalmol. Vis. Sci ;53(1): doi: /iovs Figure Legend:
Opacification of the optic of an Akeos Adapt® intraocular lens
MI60 INTRAOCULAR LENSES – OUR EXPERIENCE
A presentation to: Meeting name Date
The authors have no financial interest
성모병원 안센터 CHANGES IN ASTIGMATISM RELATIVE TO IOL HAPTIC INSERTION AXIS IN WITH-THE-RULE AND AGAINST-THE-RULE ASTIGMATISM PATIENTS Hyun Seung Kim, M.D.
Hayashi Eye Hospital, Fukuoka, Japan
Long-Term Quantitative Analysis of Posterior Capsule Opacification After Implantation of Dual-Optic Accommodating IOLs Andrea Galvis, MD 1 , 3 Ivan.
Long term PCO preventive effect of capsular bending ring
Five-Year Experience With Routine Use of Healon5 in Cataract Surgery
Presentation transcript:

“Evaluating and Defining the Sharpness of IOLs: Microedge Structure of Commercially Available Square-Edge Hydrophobic IOLs” Matthias Müller, PhD, 1 Liliana Werner, MD, PhD, 1,2 Manfred Tetz, MD 1 1 Berlin Eye Research Institute (BERI), Berlin, Germany 2 John A. Moran Eye Center, Salt Lake City, UT, USA -The authors have no financial or proprietary interest in any product mentioned in this poster. -Supported in part by unrestricted research grants to the BERI from Alcon, AMO, Wavelight, Hoya, and Advanced Vision Science, and by a 2007 Research Grant from the ESCRS (Werner). -Some of the IOLs described here are not FDA approved.

Background Posterior chamber intraocular lenses (IOLs) with a square posterior optic edge were found to be associated with better results in terms of posterior capsule opacification (PCO) prevention, regardless of the material used in their manufacture. This IOL design feature can be appropriately assessed in morphological studies using scanning electron microscopy (SEM). However, SEM studies of new IOLs have generally focused on the quality of their optic surface, or their optic finishing, with no specifications made on how sharp the optic edge needs to be to effectively prevent lens epithelial cells (LECs) from growing onto the posterior capsule. 1 Tetz and Wildeck performed the first attempt to evaluate and quantify the edge structure of IOLs at the microscopic level (Part 1). 2 They experimentally evaluated the optimum microedge profile of an IOL to prevent LEC migration in cell culture. Experimental PMMA lenses with different edge profiles were imaged under SEM, and the area above the edge, representing the deviation from an ideal square was calculated with a digital system (EPCO). 3 The objectiveThe objective of this poster was to describe the findings of our current study (Part 2), for which we used an improved methodology to evaluate the optic microedge structure of currently available, hydrophobic IOLs marketed as square edge lenses. The experimental square edge PMMA lens in Part 1, with the edge design that effectively stopped LEC growth in culture was used as the reference lens against which the currently available square edge lenses were compared. 4

Materials/Methods IOL designs: Sixteen designs of hydrophobic acrylic or silicone IOLs. Reference square edge lens: experimental square edge PMMA lens, with an edge design that effectively stopped LEC growth in culture (Part 1). Control lenses: 2 round edge silicone lenses; IOL powers: D IOL and +0.0 D IOL (or the lowest available plus dioptric power); SEM system: Hitachi S-2700 SEM (X25, X300, X1,000). Analysis from a perpendicular view of the edge; Image analysis system: JPEG photos imported to the AutoCAD LT 2000 system (Autodesk, San Rafael, CA, USA): (Equiconvex) X1,000 Area = μ 2 40-μ radius X25 Ant. surface Post. surface In vivo LEC: 8-21 μ in diameter (larger lengths) Micron scale adjusted (reference bar in SEM photos); Projection of reference circles of 40 and 60 microns of radius on the photos (the area evaluated was therefore the area of interaction of at least 1 or few LECs with the optic edge); Area above the lateral-posterior edge (deviation from a perfect square) was measured in square microns.

Results Reference PMMA lens: 34.0 and 37.5 μ 2 (40- and 60-radius circles, respectively); D control silicone lens: and 1,525.3 μ 2 ; +0.0 D control silicone lens: and 1,512.7 μ 2 ; Acrylic IOLs: 69.5 to μ 2 (40 radius); to μ 2 (60 radius); Silicone IOLs: 4.8 to μ 2 (40 radius); 0.2 to μ 2 (60 radius); Out of 30 lenses, 7 silicone lenses had area values smaller or in the vicinity of the corresponding values of the reference PMMA IOL (+5.0 Z9002, L200, SofPort AO, +0.0 and SoFlex SE, and and AQ310Ai); Acrylic versus silicone lenses: P = for 40- and 60-radius circles (Wilcoxon Two-Sample Test); D versus +0.0 D (or lowest available dioptric power) lenses: P = and P = (NS) for 40- and 60-radius circles.

IOL modelIOL manufacturer Dioptric power Optic materialArea (40 Radius) Area (60 Radius) Matrix AcrylicMedennium0.0Acrylic SA60ATAlcon20.0Acrylic SN60WFAlcon (aspheric)6.0Acrylic HydromaxZeiss10.0Acrylic SA60ATAlcon6.0Acrylic HydromaxZeiss19.0Acrylic Matrix AcrylicMedennium20.0Acrylic SN60WFAlcon (aspheric)20.0Acrylic L450Wavelight20.0Acrylic VA60BBHoya0.0Acrylic ZA9003AMO (aspheric)20.0Acrylic AR40eAMO20.0Acrylic X-60AVS0.0Acrylic ZA9003AMO (aspheric)10.0Acrylic X-60AVS20.0Acrylic MA60MAAlcon0.0Acrylic MA60ACAlcon20.0Acrylic VA60BBHoya20.0Acrylic AR40MAMO0.0Acrylic

Z9003, D Matrix Acrylic, +0.0 D Examples of square optic edge profile and finishing of hydrophobic acrylic IOLs: AR40e, D L450, D VA60BB, +0.0 D SA60AT, D

IOL modelIOL manufacturerDioptric powerOptic materialArea (40 Radius) Area (60 Radius) SoFlex SEBausch & Lomb0.0Silicone SofPort AOB&L (aspheric)20.0Silicone Z9002AMO (aspheric)5.0Silicone AQ310AiStaar (aspheric)20.0Silicone L200Wavelight20.0Silicone AQ310AiStaar (aspheric) 12.5Silicone SoFlex SEBausch & Lomb20.0Silicone Z9000AMO (aspheric)5.0Silicone SofPort AOB&L (aspheric)0.0Silicone Z9002AMO (aspheric)20.0Silicone Z9000AMO (aspheric)20.0Silicone IOLs with deviation areas in the vicinity to the cut off limits (or smaller) had similar values for the 40- and the 60-radius circles, while the others had a tendency to present increasing values with the larger radius, mostly as a function of the convexity of their posterior optic surface.

SoFlex SE, +0.0 D Z9000, D Z9002, DZ9002, +5.0 D Examples of square optic edge profile and finishing of silicone IOLs: SofPort AO, D AQ310Ai, D

Discussion / Conclusions A square edge on the posterior optic surface was found to be the most import IOL-related factor for PCO prevention. According to different experimental studies, this may be due to the mechanical barrier effect exerted by the square edge, 5 to the contact inhibition of migrating LECs at the capsular bend created by the sharp optic edge, 6 to the higher pressures exerted by IOLs with a square edge optic profile on the posterior capsule, 7 or perhaps to varying combinations of all of the above. Analysis of the microstructure of the optic edge of currently available, square edge, hydrophobic IOLs revealed a large variation of the deviation area from a perfect square, as well as of the edge finishing. Both parameters varied not only among different designs, but also between different powers of the same design.

Discussion / Conclusions If IOLs with different square microedge profiles produce similar outcomes in terms of PCO formation, one may conclude that other factors play a role in its prevention. We believe the factor that may play the most important role in even out the differences in the microedge profiles shown in our study is the shrink wrapping of the IOL by the capsular bag, enhancing the contact of the posterior IOL surface with the posterior capsule. The amount of postoperative capsular bag shrinkage has been indirectly determined in clinical studies, by the measurement of the diameter or the area of the capsulorhexis opening at different postoperative time points (13.8 to 14.8%) Central PCO = 0 Soemmering‘s ring

Discussion / Conclusions We believe that existing and future clinical data will help us to better understand the effect of microedge structure and design on reducing PCO. At present, a cut-off value should be sought for to clinically label an IOL as square edged. This study may help on the task for better understanding differences in microedge structures. We only focused on commercially available hydrophobic IOLs. Due to their low water content, we believe the SEM technique used did not cause any significant alteration of the IOL edge profile. Modern hydrophilic lenses, with water contents generally in the vicinity of 26% may have their microedge structure significantly modified during the vacuum required in standard SEM procedures. Therefore, we are currently evaluating the microedge structure of hydrophilic lenses by using an environmental SEM technique, which operates in low vacuum and does not require any prior coating. Results of this evaluation are the object of an upcoming report.

1.Kohnen T, Maqdowski G, Koch DD. Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses. J Cataract Refract Surg 1996; 22: Tetz M, Wildeck A. Evaluating and defining the sharpness of intraocular lenses: Part 1: Influence of optic design on the growth of the lens epithelial cells in vitro. J Cataract Refract Surg 2005; 31: Tetz MR, Auffarth GU, Sperker M, Blum M, Volcker HE. Photographic image analysis system of posterior capsule opacification. J Cataract Refract Surg 1997; 23: Werner L, Müller M, Tetz M. Evaluating and defining the sharpness of intraocular lenses: Part 2: Microedge structure of commercially available square-edged hydrophobic lenses. J Cataract Refract Surg 2008; 34: Werner L, Mamalis N, Pandey SK, et al. Posterior capsule opacification in rabbit eyes implanted with hydrophilic acrylic intraocular lenses with enhanced square edge. J Cataract Refract Surg 2004; 30: Nishi O, Nishi K. Preventing posterior capsule opacification by creating a discontinuous sharp bend in the capsule. J Cataract Refract Surg 1999; 25: Boyce JF, Bhermi GS, Spalton DJ, El-Osta AR. Mathematic modeling of the forces between an intraocular lens and the capsule. J Cataract Refract Surg 2002; 28: Gonvers M, Sickenberg M, van Melle G. Change in capsulorhexis size after implantation of three types of intraocular lenses. J Cataract Refract Surg 1997; 23: Kimura W, Yamanishi S, Kimura T, et al. Measuring the anterior capsule opening after cataract surgery to assess capsule shrinkage. J Cataract Refract Surg 1998; 24: Joo CK, Shin JA, Kim JH. Capsular opening contraction after continuous curvilinear capsulorhexis and intraocular lens implantation. J Cataract Refract Surg 1996; 22: Hayashi K, Hayashi H. Intraocular lens factors that may affect anterior capsule contraction. Ophthalmology 2005; 112: References