IAEA International Atomic Energy Agency Module 2.7: Error in TPS data entry (Panama) IAEA Training Course.

Slides:



Advertisements
Similar presentations
ITU-T Technical Session on EMF
Advertisements

Paolo Gemma, Senior Expert, Huawei
IAEA International Atomic Energy Agency Module 3.3: Incidents in any clinic IAEA Training Course.
بسم الله الرحمن الرحيم ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
IAEA Quality Audits in Radiotherapy
CLINICAL WORKSHOP Image-Guided Adaptive Brachytherapy for Gynaecology Hospital name Participants’ names and roles Vienna, November 2013.
RADIOTHERAPY ACCIDENT IN COSTA RICA - CAUSE AND PREVENTION OF RADIATION ACCIDENTS IN HOSPITALS Module XIX.
FDA-QA-DAS/2010 FDA’s Public Meeting: Device Improvements to Reduce the Number of Under-doses, Over-doses, and Misaligned Exposures from Therapeutic Radiation.
Auditing Computer Systems
IAEA International Atomic Energy Agency Module 2.2: Erroneous use of TPS (UK) IAEA Training Course.
Gauges and well logging
Quality Assurance: Manufacturer & Clinical Aspects  Alan Cohen, M.S. DABR  Paul Naine, MSc. MIPEM  Jim Schewe, PhD, DABMP Accuray Incorporated Elekta.
IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.
Accidents in Radiotherapy
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 3: Analysis of causes and contributing factors IAEA Training.
Radiation Protection in Radiotherapy
WELCOME!!! FDA & MITA RT Meeting
IAEA International Atomic Energy Agency International Reporting Systems Ola Holmberg, PhD Head, Radiation Protection of Patients Unit Radiation Safety.
MEDICAL RECORDS MANAGEMENT IN EYE CARE SERVICES 6.International classification of Disease & Procedures and the method of Indexing data.
Radiotherapy for Kidney cancer
11. – , Athens 8th European Conference on Medical Physics DOSIMETRY AUDITS IN RADIOTHERAPY IN THE CZECH REPUBLIC Irena Koniarová Daniela Ekendahl.
Incidents reported to the FANC: Review and conclusions Brussels, September 3 rd 2010 Karen Haest Health Protection.
Perma Permanent Brachytherapy Medical Events: Experiences of Pennsylvania Frank Costello Organization of Agreement States August 28, 2012 Milwaukee, Wisconsin.
Introduction to Nuclear Medicine
Simulating Potential Layouts for a Proton Therapy Treatment Center Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil-
Incident Reporting Procedure
Quality Assurance for a modern treatment planning system
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 5: Reporting, investigating and preventing accidental exposures.
FDA Public Meeting Device Improvements to Reduce the Number of Under-Doses, Over-Doses, and Misaligned Exposures From Therapeutic Radiation James M. Galvin,
Knowledge of radiation exposure in common radiological examinations amongst radiology department staff AL Chang, LH Cope, DH Keane, S Wood Presented by.
Instructions and Reporting Requirements Module 7 Electronic Reporting For Facilities March 2014 North Carolina Central Cancer Registry State Center for.
Adverse Event/Unanticipated Problems Policy and Procedures November 2007.
IAEA International Atomic Energy Agency Module 3.2: Other cases (brachytherapy) IAEA Training Course.
FDA Public Meeting: Device Improvements to Reduce the Number of Under-doses and Misaligned Exposures from Therapeutic Radiation.
Validity of more than 30Gy radiation therapy for long-surviving patients with painful bone metastases E.Katayama 1,2, H.Okada 1, I.Asakawa 2, T.Tamamoto.
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
IAEA International Atomic Energy Agency Module 2.6: Miscalibration of beam (Costa Rica) IAEA Training Course.
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
Chapter 5 Technology in the Business Office Copyright © 2011, 2006 Mosby, Inc., an affiliate of Elsevier. All rights reserved.
Part VIII:Medical Exposures in Radiotherapy Lecture 6: Determination of dose to the patient in Radiotherapy II IAEA Post Graduate Educational Course on.
IAEA International Atomic Energy Agency Module 6.2: Source not under control (Mexico) IAEA Training Course.
IAEA International Atomic Energy Agency PREVENTION OF ACCIDENTAL EXPOSURE IN RADIOTHERAPY Part 0: Introduction IAEA Training Course.
Radiation Sources in medicine diagnostic Radiology
CBER Common Problems on Source Plasma Inspections Judy Ellen Ciaraldi BS, MT(ASCP)SBB, CQA(ASQ) CBER, OBRR, DBA September 16, 2009.
Authorization and Inspection of Cyclotron Facilities Radiation Protection of Staff.
Biological Effects of Ionizing Radiation Deterministic effects
Introduction to Radiation Therapy
IAEA International Atomic Energy Agency Industrial Radiography Emergency Preparedness Day 5 – Lecture 8.
IAEA International Atomic Energy Agency Radiation protection of the public IAEA Regional Training Course on Radiation Protection of patients for Radiographers,
IAEA Regional Meeting June 07 National Workshop June 07 Expert Visit to Facilities 29 June-1 July 07 Orientation 15 July 07 Competency Training.
COUNTRY REPORT: PHILIPPINES Dr. Miriam Joy C. Calaguas.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Component D: Activity D.3: Surveys Department EU Twinning Project.
Chief Medical Physicist
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
IAEA E-learning Program
IAEA E-learning Program
IAEA E-learning Program
IAEA E-learning Program
IAEA E-learning Program
IAEA E-learning Program
IAEA E-learning Program
أجهزة العلاج الإشعاعي Clinical Radiation Generator
Designing and Debugging Batch and Interactive COBOL Programs
How to Create and Start a Test Session
HOW RO-ILS HELP DOSIMETRISTS TO IMPROVE PATIENT SAFETY
Hot and cold spots are common problems associated with planning:
Prevention of Accidental Exposures to Patients Undergoing Radiation Therapy INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION ——————————————————————————————————————
GHG meeting at ESTRO36 May, 2017
Minjmaa Minjgee, MD, National Cancer Centre, MONGOLIA
Presentation transcript:

IAEA International Atomic Energy Agency Module 2.7: Error in TPS data entry (Panama) IAEA Training Course

IAEA Prevention of accidental exposure in radiotherapy2 Brief history of the event

IAEA Prevention of accidental exposure in radiotherapy3 Background information Year 2000, the radiation therapy department of ION was divided between two different hospitals and a total of 1100 patients received radiotherapy. Justo Arosemena hospital (External beam therapy) Gorgas hospital (Brachytherapy and hospitalization of in-patients)

IAEA Prevention of accidental exposure in radiotherapy4 l Equipment for external beam therapy (EBT) in Justo Arosemena hospital: Cobalt-60 unit (Theratron 780C) Cobalt-60 unit (ATC/9 Picker) Orthovoltage unit (Siemens Stabilipan) TPS (RTP/2 Multidata v.2.11) One 60 Co unit and the orthovoltage unit were decommissioned and not in use at the time of the accident. EBT given from 6 a.m. to 9 p.m. on Theratron (in two shifts). Background information

IAEA Prevention of accidental exposure in radiotherapy5 l Staff of ION: Five radiation oncologists Two of these radiation oncologists (one in the morning and one in the evening) assigned to Justo Arosemena hospital on a monthly rotation Four radiotherapy technologists Two medical physicists One dosimetrist Background information

IAEA Prevention of accidental exposure in radiotherapy6 l Factors influencing workload in Justo Arosemena hospital: 70 to 80 patients treated per day Many of these patients treated during the evening with only a single therapist present Team divided between two sites Multiple fields (SSD set-up technique) with beam modifying devices (blocks and wedges) utilized Background information

IAEA Prevention of accidental exposure in radiotherapy7 Multidata TPS (2D) used to plan treatment The TPS allowed four shielding blocks to be entered in any field for calculation of dose distribution Brief description of the event

IAEA Prevention of accidental exposure in radiotherapy8 In April 2000 one of the oncologists required one additional block for some treatments in the pelvic region Brief description of the event

IAEA Prevention of accidental exposure in radiotherapy9 In order to overcome the limitation of four blocks imposed by the TPS, … … a new way of entering data was tried (August 2000): to enter several blocks “at once”. The TPS accepted the data entry, without giving a warning, but calculated incorrect treatment times Brief description of the event

IAEA Prevention of accidental exposure in radiotherapy10 In November 2000 radiation oncologists observed unusual reactions in some patients (unusually prolonged diarrhoea). The physicists checked the patient charts but did not find any abnormality (the computer calculations were not questioned) Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy11 Patient charts checked but computer calculations not questioned

IAEA Prevention of accidental exposure in radiotherapy12 In February 2001 the error in dose calculations was finally determined The treatment was simulated on a water phantom and dose measurements were made, which confirmed higher dose … treatment of relevant patients was suspended. Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy13 The resulting treatment plan The computer printout provides slightly distorted isodoses but the icon with the blocks was correct. The treatment time indicated was approximately twice the intended.

IAEA Prevention of accidental exposure in radiotherapy14 Remark: findings from quality audits First audit: February 1999 l Quality controls were made, but written procedures were missing Second audit February 2001 l Procedures were in place, but no procedure for the use of TPS l The auditor was not notified of the new approach for data entry l Tests were performed but not for the specific conditions of this event

IAEA Prevention of accidental exposure in radiotherapy15 Technical description of the problem

IAEA Prevention of accidental exposure in radiotherapy16 l The treatment planning system (TPS) at ION: Multidata RTD/2 Version 2.11 System installed in Beam data for 60 Co entered and verified at this stage. This is a 2D TPS. It allows shielding blocks to be entered and taken into account when calculating treatment time and dose distribution. Treatment planning

IAEA Prevention of accidental exposure in radiotherapy17 l Two of the modules in the Multidata TPS: “Dose chart calculator” for calculation of treatment time to a given point “External beam” for calculation of treatment time to a given point AND calculation of isodoses Treatment planning

IAEA Prevention of accidental exposure in radiotherapy18 l Restriction of the treatment planning system: Maximum 4 blocks can be digitized for a field in the “External beam” module. In the “Dose chart calculator” module, there is no such restriction. Treatment planning

IAEA Prevention of accidental exposure in radiotherapy19 l Treatments in the pelvic region were performed using “the box technique”. l Up to four blocks per field were often used for these fields. Treatment planning

Treatment Planning Entering blocks separately Menu: 1.Add 1 block 2.Type transmission factor 3.Digitize contour 4.Repeat the procedure for next block

IAEA Prevention of accidental exposure in radiotherapy21 Entering four shielding blocks correctly

IAEA Prevention of accidental exposure in radiotherapy22

IAEA Prevention of accidental exposure in radiotherapy23 l For some cervix patients, a central shielding was added to the four blocks. l Since no isodoses were requested for these cases, the “Dose chart calculator” module was used. This allows for more than four blocks. l Treatment time was correctly calculated. Treatment planning

IAEA Prevention of accidental exposure in radiotherapy24 l One of the oncologists started to request isodoses for these patients with five blocks. l The “External beam” module had to be used for this. Because of the four block limitation, initially four or less blocks were digitized. l Treatment time was slightly incorrect due to this. The effect was understood. Treatment planning

IAEA Prevention of accidental exposure in radiotherapy25 l Staff came up with an approach to enter multiple blocks simultaneously. l This approach was used for fields with four or more blocks. Even though the method was incorrect, the TPS was essentially able to handle this method. l Treatment time was essentially correctly calculated. Treatment planning

IAEA Prevention of accidental exposure in radiotherapy26 Entering several blocks as one

IAEA Prevention of accidental exposure in radiotherapy27

IAEA Prevention of accidental exposure in radiotherapy28 Variation to new approach This worked well, but, as the procedure was not written… …another physicist entered the data in a similar but slightly different way. This variation causes wrong isodoses and the wrong treatment time.

IAEA Prevention of accidental exposure in radiotherapy29

IAEA Prevention of accidental exposure in radiotherapy30 Computer printouts

IAEA Prevention of accidental exposure in radiotherapy31

IAEA Prevention of accidental exposure in radiotherapy32

IAEA Prevention of accidental exposure in radiotherapy33

IAEA Prevention of accidental exposure in radiotherapy34

IAEA Prevention of accidental exposure in radiotherapy35

IAEA Prevention of accidental exposure in radiotherapy36

IAEA Prevention of accidental exposure in radiotherapy37 The distortion is not so obvious for a four field treatment. l The icon does not indicate that the TPS is incorrectly used l Calculated treatment time approximately TWICE AS LONG AS INTENDED Second variation – multiple fields

IAEA Prevention of accidental exposure in radiotherapy38

IAEA Prevention of accidental exposure in radiotherapy39

IAEA Prevention of accidental exposure in radiotherapy40

IAEA Prevention of accidental exposure in radiotherapy41 Calculated treatment time l The calculated treatment time was approximately twice the intended l Example: Treatment time on similar patients had been 0.6 min (one field). Now it had become more than 1.2 min (one field).

IAEA Prevention of accidental exposure in radiotherapy42 Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy43 l In November 2000, radiation oncologists were observing unusually prolonged diarrhoea in some patients. l On request, physicists reviewed charts (double checked). TPS output was not questioned. No anomaly was found. Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy44 l In Dec. 2000, similar symptoms were observed. In Feb. 2001, physicists initiated a more thorough search for the cause. l In March 2001, physicists identified a problem with computer calculations. Treatment was suspended. Nov’00 Dec’00 Jan’01 Feb’01 Mar’01 Symptoms Chart checks Symptoms More thorough checks Problem found Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy45 l Isodoses and treatment time were re-examined closer and anomalies were found. l The treatment was simulated on a water phantom and measurement of doses were made, which confirmed higher dose. Nov’00 Dec’00 Jan’01 Feb’01 Mar’01 Symptoms Chart checks Symptoms More thorough checks Problem found Discovery of the problem

IAEA Prevention of accidental exposure in radiotherapy46 Estimation of dose to patients

IAEA Prevention of accidental exposure in radiotherapy47 Some of the patients treated in the abdominal region were affected In total: 28 patients affected. Brain: 4.3% Head and neck: 12.1% Mamma: 16.8% Lung: 7.9% Cervix uteri: 15.5% Endometrium: 1.5% Prostate: 9.3% Rectum: 3.9% Others: 28.7% Treatments performed at INO

IAEA Prevention of accidental exposure in radiotherapy48 Dose estimation Dose to the 28 affected patients was estimated retrospectively Dose to prescription point for multiple fields was estimated Based on the patients’ charts: dose rate under reference conditions beam set up (depth, effective field, and beam modifiers), and treatment times

IAEA Prevention of accidental exposure in radiotherapy49 Biologically effective dose Since the dose per fraction was much higher than standard, the biologically effective dose (BED) and the dose equivalent to a treatment of 2 Gy/fraction were also calculated, using the linear quadratic model (α/β = 3 for intestine was used for evaluation of late effects).

IAEA Prevention of accidental exposure in radiotherapy50 Number of patients and their dose (equivalent to 2 Gy/fraction) (as of May 30, 2001)

IAEA Prevention of accidental exposure in radiotherapy This accidental exposure

IAEA Prevention of accidental exposure in radiotherapy This accidental exposure up to 160

IAEA Prevention of accidental exposure in radiotherapy53 Skin changes even though multiple fields used

IAEA Prevention of accidental exposure in radiotherapy54

IAEA Prevention of accidental exposure in radiotherapy55

IAEA Prevention of accidental exposure in radiotherapy56

IAEA Prevention of accidental exposure in radiotherapy57

IAEA Prevention of accidental exposure in radiotherapy58 Dilated air-filled loops of small bowel from a distal small bowel obstruction likely secondary to radiation induced stenosis

IAEA Prevention of accidental exposure in radiotherapy59 Effects on patients Effects at the moment of the evaluation mission (May 30, 2001) 8 deaths of 28 patients 5 of these deaths radiation related 2 unknown (not enough data) 1 due to metastatic cancer 20 surviving patients of the affected

IAEA Prevention of accidental exposure in radiotherapy60 Lessons and recommendations

IAEA Prevention of accidental exposure in radiotherapy61 Initiating event and contributory factors The event was triggered by The search for a way to overcome the limitation of the TPS (four blocks only) Contributory factors The computer presented the icon as if the blocks were correctly recognized The procedure was not tested The trick “worked” and was time-saving

IAEA Prevention of accidental exposure in radiotherapy62 Initiating event and contributory factors Contributory factors (continued) Treatment times were longer than usual but no one detected it workload limited interaction (radiation oncologists, medical physicists and radiotherapy technologists) computer calculations in general were not verified Patient reactions were realized but the follow-up was insufficient

IAEA Prevention of accidental exposure in radiotherapy63 Lessons for manufacturers Avoid ambiguity in the instructions Thorough testing of software, also for non-intended use Guide users with warnings on the screen for incorrect data entry Be readily available for consultation, especially when a change in the way of use is intended

IAEA Prevention of accidental exposure in radiotherapy64 Lessons for radiotherapy departments TPS is a safety critical piece of equipment Quality control should include TPS Procedures should be written Change in procedures should be validated before being put into use Computer calculation should be verified (manual checks for one point)

IAEA Prevention of accidental exposure in radiotherapy65 Lessons for radiotherapy departments Awareness of staff for unusual treatment parameters should be stimulated and trained Communication should be favoured Unusual reactions should be completely investigated and dosimetry data tested

IAEA Prevention of accidental exposure in radiotherapy66 Panama incident summary ‘Minor’ change of practice in use of a treatment planning system Not systematically verified 8 patients dead

IAEA Prevention of accidental exposure in radiotherapy67 Reference IAEA: Investigation of an accidental exposure of radiotherapy patients in Panama (2001)

IAEA Prevention of accidental exposure in radiotherapy68 Postscript Towards the end of 2004, two physicists involved in this event were sentenced to four years in prison respectively, as well as a period of seven years when they were not allowed to practice in the profession.

IAEA Prevention of accidental exposure in radiotherapy69 Postscript According to the court, they did not inform their superiors regarding the modifications in practice in relation to the use of the treatment planning software.