IPC Friedrich-Schiller-Universität Jena 1 Reduction of out of focus light  Excitation light excites fluorescence more or less within the whole sample.

Slides:



Advertisements
Similar presentations
Chapter 4 Notes Part II: Microscopy (refer to pg.60-61)
Advertisements

Today’s Topic (Slight ) correction to last week’s lecture
Bio-optics & Microscopy (MEDS 6450) 11/16/2010 Presented by: Mathilde Bonnemasison Leia Shuhaibar Steve Pirnie Ronghua (Ronnie) Yang Neil MAA, Juskaitis.
Lecture 11. Microscopy. Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single or.
Microscopy Outline 1.Resolution and Simple Optical Microscope 2.Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3.Newer.
ERT107 MICROBIOLOGY FOR BIOPROCESS ENGINEERING Pn Syazni Zainul Kamal PPK Bioprocess.
Nick Beavers Project Manager Deconvolution from Andy Molnar Software Engineer.
1. Resolution (R) : separation of close objects, light wavelength NA, numerical aperture 2. Contrast : distinction of objects from background “light field”
Dave Piston May 18, 2014 Optical Sectioning 2: Confocal Designs and “Pseudo-confocals” Outline 1.PMT detectors 2.Laser Scanning Design Features 3.Pseudo-confocals.
ABRF meeting 09 Light Microscopy Research Group. Why are there no standards? Imaging was largely an ultrastructure tool Digital imaging only common in.
R. Hui Photonics for bio-imaging and bio- sensing Rongqing Hui Dept. Electrical Engineering & Computer Science, The University of Kansas, Lawrence Kansas.
Lasers and Confocal. Laser Acronym: Light Amplification by Stimulated Emission of Radiation Ordinary light emission: Comes from spontaneous decay of excited.
Comparison and Implications of Three Optical Microscopy Data Acquisition Modalities James Butler Ph.D. Nikon Instruments, Inc. Widefield Fluorescence Confocal.
1 Confocal Microscopy David Kelly November 2013 Handbook of Biological Confocal Microscopy. Ed. J. Pawley, Plenum Press Fundamentals of Light Microscopy.
Fluorescence microscopy – Principle and practical consideration Hiro Ohkura.
Short pulses in optical microscopy Ivan Scheblykin, Chemical Physics, LU Outline: Introduction to traditional optical microscopy based on single photon.
Computational Biology, Part 22 Biological Imaging I G. Steven Vanni Meel Velliste Robert F. Murphy Copyright  1998, All rights reserved.
P. Moghe, 125:583 1 Microscopy Techniques for Biomaterials and Cell Based Interfaces Professor Prabhas V. Moghe October 26, :583 Fall 2006.
USE AND CARE OF THE MICROSCOPE LECTURE 1. MICROSCOPY u Light Microscopy: any microscope that uses visible light to observe specimens u Compound Light.
Media Cybernetics Deconvolution Approaches and Challenges Jonathan Girroir
Study of Protein Association by Fluorescence-based Methods Kristin Michalski UWM RET Intern In association with Professor Vali Raicu.
Light Microscopy Sarah Heintz. Compound Microscope The microscope uses a lens that is close to the object and uses light to focus on the real image of.
Biophotonics lecture 16. November Fourier plane Point object Image f f f f  ’’ Magnification: m=1 Angles:sin(  ’)=sin(  ’=  Magnification.
Content of the lecture Principles of confocal imaging. Different implementations/modes. Primer on multi-photon (MPH) and multi-harmonic (MHG) generation.
Microscope.
Practical aspects of fluorescence microscopy
Microscopy.
Biophotonics lecture 23. November Last week: -Fluorescence microscopy in general, labeling, etc… -How to do optical sectioning and fill the missing.
DIC, fluorescence and confocal microscopy Department of Mechatronics GIST Yong-Gu Lee References: 1.“Fundamentals of light microscopy and electronic imaging,”
TIRF Total Internal Reflection Fluorescence Microscopy specialized fluorescence microscopy technique specifically images a very thin optical section (50-250nm)
Naomi Kinjal Asaad Binoy
Fluorescence Techniques
Agenda How to make the specimen visible – Definition of Contrast
Triplet Blinking - stable triplet state - S 1 T 1 due to solvent -halogens -transition metals Transition to the Dark Side.
By: Rob Page and Tara Trovarello
IPC Friedrich-Schiller-Universität Jena 1 From Wednesday, 9 November 2011 the weekly lectures will be every Wednesday, 8:30 Humboldtstr. 8, Seminar Room.
May Chuck DiMarzio, Northeastern University ECE-1466 Modern Optics Course Notes Part 7 Prof. Charles A. DiMarzio Northeastern University.
BIO 224 Intro to Molecular and Cell Biology. Microscopes are tools frequently used in cell biology Type of microscope used depends on the specimen being.
Biophotonics lecture 11. January Today: -Correct sampling in microscopy -Deconvolution techniques.
Epi-illumination is form of Kohler Illumination:
Microscopes Compound Bright-Field Light Microscope
Chapter 6 Cellular Measurements in Biomaterials and Tissue Engineering.
Developments in Imaging Technology and Staining Techniques Unit C: Section 1.3 Science 10.
3. Optical Coherence Tomography (OCT)
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
The Microscope and Forensic Identification. Magnification of Images A microscope is an optical instrument that uses a lens or a combination of lenses.
Literature Exam: Monday the 7th of Febraury 2011 Prev week's lecture: Thursday.
1 Correct Sampling. What is SAMPLING? Intensity [a.u.] X [µm] 1.
Use Snell’s law to determine the angle of total internal reflection in the coverslip (without oil). Oil immersion objectives More light (no total internal.
Date of download: 5/27/2016 Copyright © 2016 SPIE. All rights reserved. (a) Schematic of the light path. The galvo scanners are mounted above periscopes.
Designing a Microscopy Experiment Kurt Thorn, PhD Director, Image from Susanne Rafelski, Marshall lab.
Bruno De Fanti Anel EE230 University of California, Santa Cruz December 2011 LASER SCANNING CONFOCAL MICROENDOSCOPY.
Confocal Microscopy Kurt Thorn NIC.
Dr R.Jayaprada CONFOCAL MICROSCOPY.
Imaging.
Microscopy.
Reporter: Chien-Chung Tsai
Date of download: 7/5/2016 Copyright © 2016 SPIE. All rights reserved. Spatial light modulator-two-photon microscope (SLM-2PM) scheme: (1) Ti:Sa laser.
Microscope.
Lasers and Confocal.
3.3 Other types of microscopy
Looking at the small stuff
What Is Spectral Imaging? An Introduction
MFM setup. MFM setup. The excitation lasers are combined in a fiber through an acousto-optic tunable filter, collimated, reflected on a dichroic mirror.
LIGHT MICROSCOPY variations
Volume 99, Issue 8, Pages (October 2010)
Photographic Image Formation I
Fluorescence microscopy with super-resolved optical sections
Transition to the Dark Side
Presentation transcript:

IPC Friedrich-Schiller-Universität Jena 1 Reduction of out of focus light  Excitation light excites fluorescence more or less within the whole sample  Out of focus fluorescence light is not imaged sharply  Out of focus fluorescence reduces especially for thick samples the image quality 7.2 Confocal fluorescence microscopy 7. Fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 2 Reduction of out of focus light  A confocal microscope uses focused laser illumination and a pinhole in an optically conjugate plane in front of the detector to eliminate out-of- focus blur  As only light produced by fluorescence close to the focal plane is detected, the contrast is much better than that of wide- field microscopes.  Allows recording individual optical sections or three dimensional reconstruction of objects 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 3 The Widefield Microscope xyz Pinhole Pinhole PMT The Confocal Microscope ZI(Z) Standard Lightsource Camera Standard Lightsource Camera Lightsource with Pinhole plane spread function:

Scanning in a CLSM PinholePMT Image of the Object: xyz Lightsource with Pinhole Sample Scanning Object scanning versus Beam scanning

IPC Friedrich-Schiller-Universität Jena 5 Reduction of out of focus light  In contrast to widefiled fluorescence microscopy where the whole sample is illuminated in confocal microscopy only one point in the sample is illuminated at a time  2D or 3D imaging requires scanning over a regular raster (i.e. a rectangular pattern of parallel scanning lines) in the specimen: raster-scan  Comparison widefiled vs. confocal Linewise scanned image Cell in its meta-/ana-phase. Plasma membrane is stained with a red fluorescing antibody while the spindle apparatus is labeled with a green fluorescent marker 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 6 Reduction of out of focus light Resolution in confocal microscopy Comparison of axial (x-z) point spread functions for widefield (left) and confocal (right) microscopy 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 7 Confocal OTF k x,y kzkz Excite AND Detect: P(r) = P Excitation (r) P Detection (r) PSF(r) = PSF Excitation (r) PSF Detection (r) OTF(k) = OTF Excitation (r)  OTF Detection (r) k x,y kzkz  Increasing the aperture angle (  ) enhances resolution !!

IPC Friedrich-Schiller-Universität Jena 8 We have circumvented Abbe:

IPC Friedrich-Schiller-Universität Jena 9 Confocal OTFs:  WF 1 AU 0.3 AU in-plane, in-focus OTF 1.4 NA Objective WF Limit New Confocal Limit Almost no transfer

IPC Friedrich-Schiller-Universität Jena 10 Confocal laser scanning microscopy  In confocal laser scanning microscopy laser light is focused to a small point at the focal plane of the specimen and moved / scanned by a computer controlled scanning mirror in the X-Y direction at the focal plane.  The fluorescent emission is sent through a pinhole and recorded by a photon multiplier tube (PMT)  An image is assembled with the help of a computer  Advantages:  Good axial out-of focus suppression  Quantification of fluorescence intensity  Simultaneous recording of different dyes in different channels  Disadvantages:  High costs (why?)  Artifacts due to coherence of laser and laser fluctuations  High amount of photo-bleaching 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 11 Confocal laser scanning microscopy  Experimental Setup  Scanning and Descanning by same element 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy Excitation Fluorescence Transmission Detector

IPC Friedrich-Schiller-Universität Jena 12 Confocal laser scanning microscopy  Scan Head:  Excitation filter / Wavelength selection  Scan-System  Beamsplitter  Pinhole  Detectors (photomultiplier) Acousto-optic tunable filter (AOTF) for laser intensity control and wavelength selection in confocal microscopy. Acousto Optic Tunable Filter (AOTF) 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy dichromatic beamsplitters excitation filter

IPC Friedrich-Schiller-Universität Jena 13 Confocal laser scanning microscopy  Scan System:  Mirror system is used to scan laser beam line by line over the sample  Mirror system consists of two rotating mirrors; one for scanning the laser in x- direction and the other for movement in the y-direction (almost parfocal, f  -lens, 4-Galvo idea)  Beam separation  In confocal microscopy several wavelength bands can be detected in parallel. Beam splitting is performed by dichroitic mirrors + filters, prisms, diffraction gratings + apertures. 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy dichroitic beam splitter more detectors variable apertures Diffraction Grating pinhole

IPC Friedrich-Schiller-Universität Jena 14 Confocal laser scanning microscopy  Pinhole:  Pinhole in the optically conjugate sample plane in front of the detector to eliminate out-of-focus blur can be adjusted continuously in its size  Pinhole size determines how much out-of-focus light is eliminated and how much light reaches the detector  The smaller the pinhole the better the axial resolution the smaller the brightness  Pinhole diameter = 1 Airy disc: Pinhole diameter corresponds to diameter of dark ring  Size of this maximum depends on magnification of objective and wavelength of light  Pinhole diameter needs to be adjusted on experimental parameters < 1 Airy Disc  Improved z-resolution  Signal losses > 1 Airy Disc  Improved brightness  Partial loss of confocal effect 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 15 Confocal laser scanning microscopy  Photomultiplier:  As detectors photomultipliers (PMT) are used 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy High dynamic range (Voltage can be adjusted) Multiplication noise Multiplicative noise dark noise (cooling) cosmic radiation

IPC Friedrich-Schiller-Universität Jena 16 Confocal laser scanning microscopy  Photomultiplier:  PMT collects and amplifies incoming photons / electrons and reacts quickly and sensitive on incoming lights  PMTs do not generate an image! Image is generated by a computer PMTs amplify brightness i.e. intensity of incoming light  PMTs see black and white! Wavelength of incoming light is irrelevant for PMTs  In order to measure different wavelengths the light must be filtered and distributed onto several detectors. Every single detector displays the intensity of the selected wavelength area. 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 17 Confocal laser scanning microscopy  Modern detectors:  GAsP PMTs, high efficiency  avalanche photo diodes (APDs), extremely efficient, small area, low maximum rate  APD arrays (expensive)  APD/PMT Hybrid detectors 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy

IPC Friedrich-Schiller-Universität Jena 18 Widefield vs. confocal Widefield Confocal Comparison of widefield (upper row) and laser scanning confocal fluorescence microscopy images (lower row). (a) and (b) Mouse brain hippocampus thick section treated with primary antibodies to glial fibrillary acidic protein (GFAP; red), neurofilaments H (green), and counterstained with Hoechst (blue) to highlight nuclei. (c) and (d) Thick section of rat smooth muscle stained with phalloidin conjugated to Alexa Fluor 568 (targeting actin; red), wheat germ agglutinin conjugated to Oregon Green 488 (glycoproteins; green), and counterstained with DRAQ5 (nuclei; blue). (e) and (f) Sunflower pollen grain tetrad autofluorescence. 7. Fluorescence microscopy 7.2 Confocal fluorescence microscopy Mouse Brain Hippocampus Smooth Muscle Sunflower Pollen Grain