1 ECG Signal Make-up
2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit must have isolation and protection circuits to protect the patient n Output Impedance must be low compared to the load impedance n Proper bandwidth range. Not too narrow and not too wide n Gain at least 1000 times n For bipolar electrodes, use differential amplifier
3 ECG Functional Blocks 1. Protection circuit 2. Lead selector 3. Calibration signal 4. Preamplifier 5. Isolation circuit 6. Driven right leg system 7. Driver amplifier 8. Memory System 9. Microcomputer 10. Recorder - Printer
4 The Electrocardiogram n Outline n The ECG n Specific Requirements n Functional Blocks n Frequent Problems n Some Solutions
5 Heart Dipole
6 Cardiac Vector
7 Eindhoven’s Triangle
8 Wilson’s Central Terminal Connection
9 Augmented Leads
10 Comparing Augmented leads to Standard Leads
11 Precordial Leads
12 Problems Frequently Encountered n Frequency Distortion n Saturation or Cutoff Distortion n Ground Loops n Open Lead Wires n Artifacts from Large Electric Devices n Interference From Electric Devices n Electromagnetic Interference
13 Frequency Distortion
14 Saturation - cutoff Distortion
15 Ground Loops
16 Ground Loops - 2
17 Artifacts from large electric devices
18 (a) 60 Hz Line Interference (b) EMG interference
19 Mechanism of electric-field pickup from Power line
20 Current flows from the power line into the body
21 Magnetic-field Pickup by ECG
22 Voltage protection scheme at input
23 Voltage limiting device
24 Electrostatic Shield - to eliminate electric-field interference
25 Driven-Right-Leg System
26 Equivalent Ct. of a driven right leg system
27 Bio-Signals Charachteristics
28 ECG Amplifier
29
30
31
32
33
34
35
36
37
38
39