Physiologic Control Algorithms for Rotary Blood Pumps using Pressure Sensor Input Edward Bullister, Ph.D. Sanford Reich, Ph.D. APEX Medical, Inc. ISRP.

Slides:



Advertisements
Similar presentations
LOW BLOOD PRESSURE. What is blood pressure? Blood pressure is the force exerted by circulating blood by the heart pumping blood into the arteries.
Advertisements

1 CAUTION: Federal Law (USA) restricts this device to sale by or on the order of a physician. Refer to the “Instructions For Use” for complete Indications.
LVAD is a battery-operated, mechanical pump-type device that's surgically implanted. Goal of LVAD: providing the patient with as close to a normal lifestyle.
HEMODYNAMIC SIMULATOR Project Background The objective of this project is to design and build a Hemodynamic Simulator. Hemodynamic Simulator is a modular.
Infusion Pump Controller Requirements Definition A Decision-Table Approach by Richard Riehle.
Viking Pump Flow Manager - Phase 2 Senior Design May
SENSORS AND SCATTERPLOTS Introduction. BODY MASS INDEX (BMI)  BMI is a number calculated from your weight and height. BMI = (weight, kg) (height, m)
Copyright © 2004 Pearson Education, Inc., publishing as Benjamin Cummings Dee Unglaub Silverthorn, Ph.D. H UMAN P HYSIOLOGY PowerPoint ® Lecture Slide.
 Have you ever thought about the sound you hear on a stethoscope. What is the “lub dub” sound?
Sport Books Publisher1 The Heart and Lungs at Work Chapter 7.
INTRODUCTION TO CONTROL SYSTEMS
Blood Pressure. Blood pressure is the force exerted by the blood against the walls of the blood vessels. It is necessary to maintain blood flow though.
1 An FPGA-Based Novel Digital PWM Control Scheme for BLDC Motor Drives 學生 : 林哲偉 學號 :M 指導教授 : 龔應時 IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, VOL.
Pump Controller Team Number: May06-12 Team Members Dwayne Stammer CprE Francois Munyakazi EE Dan Paulsen CprE/EE Faculity Advisor Nicola Elia Client Viking.
CURRENT PROCJECTS The Effect of AVF Size and Position on Distal Perfusion Focus: Alter diameter, length and position of fistula and monitor changes in.
BY IRFAN AZHAR Control systems. What Do Mechatronics Engineers Do?
Plant Modeling for Powertrain Control Design Modelica Automotive Workshop Dearborn, MI November 19, 2002 Dr. Larry Michaels GM Powertrain Controls Engineering.
Frank-Starling Mechanism
 By the end of this lecture the students are expected to:  Define cardiac output, stroke volume, end- diastolic and end-systolic volumes.  Define.
Arterial System and Blood Pressure
SEVERAL MAJOR CAPABILITIES OF CRITICAL CARE INVIRONMENT 1.Process store and intergrate physiologic and diagnostic information from various sources. 2.Accept.
Learning Objectives... To understand the stages of the cardiac cycle.
EML 2023 – Motor Control Lecture 1 – Closed Loop Motor Control.
Blood and Blood Pressure. Components of Blood Plasma – Liquid portion of blood – Contains ions, proteins, hormones Cells – ___________________________________.
Control Theory Control System Objectives  Establish a final condition  Provide safe operation  Eliminate the human element  Assure economical operation.
Cardiovascular Physiology Lab #10 Path of Cardiac Excitation Sinoatrial (SA) Node –pacemaker of the heart Atrioventricular (AV) Node –Delays conduction.
Human Anatomy and Physiology
Networks in Engineering A network consists of a set of interconnected components that deliver a predictable output to a given set of inputs. Function InputOutput.
Copyright © 2004 Pearson Education, Inc., publishing as Benjamin Cummings Dee Unglaub Silverthorn, Ph.D. H UMAN P HYSIOLOGY PowerPoint ® Lecture Slide.
Embedded Control Systems Dr. Bonnie Heck School of ECE Georgia Tech.
The cardiac cycle Ventricular filling the diastole refers to the period of the cardiac cycle during which the ventricles are filling with blood the systole.
Engr. Hinesh Kumar (Lecturer) 1 Phonocardiography.
Blood Pressure and Mean Arterial Pressure It’s all about perfusion!
Copyright © 2004 Pearson Education, Inc., publishing as Benjamin Cummings Ch. 20 The Heart Describe the organization of the cardiovascular system. Discuss.
Proportional and Servo Valves
FloTrac Sensor & Vigileo Monitor
Cardiovascular System
Date of download: 6/9/2016 From: Developing and Implementing Computerized Protocols for Standardization of Clinical Decisions Ann Intern Med. 2000;132(5):
Blood Vessels & Circulation. Recap Learned the structure and function of the heart. Visualized movement of blood through arteries, veins, and capillaries.
Current gas flow scheme Gas handlerTPC 1~2 m Braided plastic tube 1~2 m Alum. tube 5 m Alum. tube Pressure relief valve P10 Rotary Pump (Edwards E2M8)
Ventricular Assist Device Test Loop Client: Minnetronix, Inc. Advisor: Kristyn Masters Team: Joe Bothwell, Dan Carlson, Kayla Ericson, and David Ugai.
Heat to air update August 9th, 2017.
Physiology of Circulation
Date of download: 11/15/2017 Copyright © ASME. All rights reserved.
EHPV® Technology Advanced Control Techniques for Electro-Hydraulic Control Valves by Patrick Opdenbosch Goals Develop a smarter and self-contained valve.
EHPV® Technology Sponsored by HUSCO Intl. & the FPMC Center
Cardiovascular Dynamics
EHPV Technology Auto-Calibration and Control Applied to Electro-Hydraulic Valves by Patrick Opdenbosch GOALS Development of a general formulation for control.
Pulmonary and Systemic Circulation
James A. Magovern, David M. Lasorda 
Siavash S. Asgari, MS, Pramod Bonde, MD 
Hemodynamic effects of partial ventricular support in chronic heart failure: Results of simulation validated with in vivo data  Deborah Morley, PhD, Kenneth.
Ao, aorta; AR, aortic regurgitation; CFD, color-flow Doppler; LV, left ventricle; LV EDP, left ventricular end-diastolic pressure; PW, pulse-wave; SBP,
Linear end-systolic pressure-volume relationship during pulsatile left ventricular bypass represents native heart function  Osamu Kawaguchi, MD, John.
What do I remember? What is the structure of the heart?
Three-dimensional flow characteristics in ventricular assist devices: Impact of valve design and operating conditions  Christoph Benk, MSc, Ramona Lorenz,
Heart Booster: a pericardial support device
Weaning of rotary blood pump recipients after myocardial recovery: a computer study of changes in cardiac energetics  Heinrich Schima, PhD, Michael Vollkron,
Osamu Kawaguchi, MD, Walter E. Pae, MD, FACS, Bill B
CONTROL SYSTEM AN INTRODUCTION.
Ao, aorta; CEOA, corrected effective orifice area; CFD, color-flow Doppler; CW, continuous-wave; DBP, diastolic blood pressure; LV EDP, left ventricular.
Blood Pressure and Mean Arterial Pressure
Clinical experience with an implantable, intracardiac, continuous flow circulatory support device: physiologic implications and their relationship to.
Experimental and numeric investigation of Impella pumps as cavopulmonary assistance for a failing Fontan  Christopher M. Haggerty, BS, Francis Fynn-Thompson,
Guruprasad A. Giridharan, PhD, Steven C
The Heart and Blood Pressure
Blood Pressure.
The cardiac cycle.
The heart’s physiology
A virtual patient sandbox for medical education
Presentation transcript:

Physiologic Control Algorithms for Rotary Blood Pumps using Pressure Sensor Input Edward Bullister, Ph.D. Sanford Reich, Ph.D. APEX Medical, Inc. ISRP August 2001

Why Use Pressure Inputs?  Provides physiologic feedback for pump control.  Provides added-value pump diagnostic and monitoring functions.  Increases capability for patient monitoring.  Potentially increases patient quality of life.

How to Implement?  Control Algorithm Development  Design Strategy to Mimic Patients’ Physiologic Control  Control Algorithm Schematic  Control Algorithm Detail  Control Algorithm Results  Added-Value Diagnostic and Monitoring Functions  Patient Monitoring  Hardware Considerations  Summary

Control Algorithm

Level 1: Basic Control Algorithm  Level 1 Control Input: LVDFP - Left Ventricular Diastolic “Filling Pressure”  Level 1 Control Output: Pump Flow Rate  Proportional Integral Control Algorithm d/dt(Flow) = K * (LVDFP - P desired ) K = 0.1 L/min/mmHg  Flow  Pressure   Simple  Robust

Level 1 Results

Level 2: Exercise Control Algorithm  Level 2 Control Inputs:  Arterial Pressure  Pulse Rate Increase (e.g., during exercise)  Level 2 Control Output:  Desired LVDFP  Level 2 Limits:  Max/min LVDFP  Max/min Arterial Pressure

Level 2 Results

Flow Rate Monitor using Pressure  Pressure Calculated from Pump Speed and Pressure Difference  Independent of Motor Current  Includes High Frequency Content Flow (L/min)  Calculated From Pressure Flow meter Measurement  Time (sec)

Hydraulic Power Monitor  Hydraulic Power (HP) into Blood  Pump: HP pump =  P pump * PumpFlow (continuous)  Heart: HP heart ~  P heart * PumpFlow (measured during systole)

Hardware Considerations  Pressure Sensor Technology  Thin-Film Based  MEMS Based  Any Rotary VAD  Pressure Sensor Placement

Component Analysis Computational Fluid Dynamics (CFD)  Example - Inlet Cannula  Establish optimal location for pressure sensor  Calculate pressure coefficient K for nonlinear relationship:  P = K*V 2

Summary  An initial control algorithm has been implemented to auto- regulate rotary blood pumps using physiological pressure inputs.  Two levels of control for a rotary pump have been tested in a mock loop setup.  The pressure signals produce added-value information.  Additional monitoring and control levels have been conceived.  Goal is to contribute to patient quality of life.