X-RAY INTERACTIONS Objectives: To understand;

Slides:



Advertisements
Similar presentations
X Rays Medical Physics Notes.
Advertisements

X-rays : Their Production Their Interaction with Matter
X-Ray & γ-Ray Interactions with Matter
Intermediate Physics for Medicine and Biology Chapter 15: Interaction of Photons and Charged Particles with Matter Professor Yasser M. Kadah Web:
Gamma-Ray Spectra _ + The photomultiplier records the (UV) light emitted during electronic recombination in the scintillator. Therefore, the spectrum collected.
X-Ray Interaction with Matter & Human Biology
Interactions of Radiation With Matter
Physics of Radiography
Rutherford Backscattering Spectrometry
Physics of Radiotherapy Lecture II: Interaction of Ionizing Radiation With Matter.
Medical Imaging X-Rays I.
Radiation Interaction Q&A
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
Rutherford Backscattering Spectrometry
Interactions with Matter
Particle Interactions
Interaction of radiation with matter - 3
Radiation Dosimetry Dose Calculations D, LET & H can frequently be obtained reliably by calculations: Alpha & low – Energy Beta Emitters Distributed in.
Physics of Radiography Interaction with matter. By the end of the first part of the session you should be able to: 1.Understand what can happen as x-ray.
Interaction of Gamma-Rays - General Considerations uncharged transfer of energy creation of fast electrons.
1 Photon Interactions  When a photon beam enters matter, it undergoes an interaction at random and is removed from the beam.
Assumptions: 1)Matter is composed of discrete particles (i.e. electrons, nucleus) 2)Distance between particles >> particle size 3)X-ray photons are small.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Dose Distribution and Scatter Analysis
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
DMI 261 Radiation Biology AndProtection. Unit 2 RADIATION Electromagnetic Spectrum –The frequency range of electromagnetic radiation and the photon wavelength.
Radiation Interactions
Interactions of Radiation With Matter
Resident Physics Lectures
1 Patient Interactions 2010FINAL. 2 1.______________ 2.______________ 3.______________ 4.______________ 5.______________.
Lecture 1.3: Interaction of Radiation with Matter
Patient Interactions Photoelectric Classic Coherent Scatter
Space Instrumentation. Definition How do we measure these particles? h p+p+ e-e- Device Signal Source.
Mammography Questions and Answers
1 Dosimetry part 1: X-ray interactions with matter G.Haddadi, PhD, MSc Associated prof. of Medical Physics Fassa University of Medical Sciences.
Photon Tissue Interactions
Resident Physics Lectures Christensen, Chapter 4 Basic Interactions Between X-Rays and Matter George David Associate Professor Medical College of Georgia.
Radiation and radiation dosimetry Spring 2006 Introduction Audun Sanderud Department of Physics University of Oslo.
Medical Imaging Radiation I. Naked to the Bone: Medical Imaging in the Twentieth Century (Paperback)by Bettyann Kevles Bettyann Kevles E=mc2: A Biography.
Radiation Quality Chapter 4. X-ray Intensity Intensity: the amount of energy present per unit time per unit area perpendicular to the beam direction at.
© Jimoid.com 2005 Ionising Radiation There are two types of radiation; ionising and non-ionising. Radiation Ionising Non-ionising Indirectly ionising (neutral.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
Gamma ray interaction with matter A) Primary interactions 1) Coherent scattering (Rayleigh scattering) 2) Incoherent scattering (Compton scattering) 3)
Chapter 5 Interactions of Ionizing Radiation. Ionization The process by which a neutral atom acquires a positive or a negative charge Directly ionizing.
Interactions of EM Radiation with Matter
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Appendix G1: Chapter 12: X-Ray Interaction with Matter
Resident Physics Lectures Christensen, Chapter 4 Basic Interactions Between X-Rays and Matter, Grid Attenuation and Filtration George David Associate.
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Interaction of x-ray photons (and gamma ray photons) with matter.
RTT 425 Radiation Therapy Physics Radiation Quality, Chapter 4 From Stanton and Stinson: Applied physics for Radiation Oncology.
Barium & Iodine Imaging Physics
Interactions of Ionizing Radiation
Interaction of Radiation with Matter
Chapter 9 Concepts you already know!!! INTENSITY = ROENTGENS AKA EXPOSURE l mAs and mR proportional? l kVp and mRproportional? l Distance and mRproportional?
Chp 2 : Interaction of radiation with matter Radiation Contrast
Chapter 5 Interactions of Ionizing Radiation
Interactions of Radiation With Matter
RAD 254 Chapter 10 X-ray interaction with matter
X-ray Interactions with Matter
X-Ray Physics Assumptions:
Resident Physics Lectures (Year 1)
Interaction of gamma rays with matter
X Rays Medical Physics Notes.
Resident Physics Lectures (year 1)
Resident Physics Lectures (Year 1)
Chapter 5 - Interactions of Ionizing Radiation
INTERACTION OF X-RAYS WITH MATTER
Computed Tomography (C.T)
Presentation transcript:

X-RAY INTERACTIONS Objectives: To understand; 1. Derivation of the attenuation law 2. Half value layer Contributions to the attenuation coefficient and their energy dependence 4. Beam hardening calculations 5 . K-edge filters 6. Representation of attenuation coefficients as linear combinations of a basis set X-RAY INTERACTIONS

Attenuation Law - Monoenergetic Case Consider x-rays having an incident fluence N(0) which penetrate a distance x into a subject as shown in Figure 1. x N(0) N(x) N(x + ∆x) ∆x

The fluence at x + ∆x is related to that at x by N( x + ∆x) = N(x) + ∆N where ∆N < 0.

∆N = - total( cm2 / el ) * ( D∆x) (el/cm2) * N(x) (1) The change in the fluence due to the attenuation in x is given by ∆N = - total( cm2 / el ) * ( D∆x) (el/cm2) * N(x) (1) = prob. of attenuation per incident x-ray * N(x) where total = total attenuation cross-section = material density (gm/cm3) D = electron density in (electrons/gram) = NA (6x1023 atoms / mole) *Z (e l / atom) / A(grams/mole) ∆x = differential thickness traversed(cm)

The linear attenuation coefficient µL is defined as, µL = total  D (c m - 1) (2) Then from equation 1 ∆N = - µL ∆x N ( x ) For small ∆x dN = - µL dx N(x) dN/N = -µL dx

Then from equation 1 ∆N = - µL ∆x N ( x ) For small ∆x dN = - µL dx N(x) dN/N = -µL dx

Integrating both sides ln N = -µL x + C elnN = N = e -(µL x + C) N = e c *e -µL x at x = 0 N = N(o) so ec = N ( 0 ) N(x) = N(o) e - µLx (3)

( µL /  ) = µm mass attenuation coefficient (cm2 / gm) This can also be expressed as N(x) = N(o) e - ( µL/ )( x ) (4) Where ( µL /  ) = µm mass attenuation coefficient (cm2 / gm) and px = t thickness expressed in (gm / cm2) It is useful to use g / cm2 when absorbing materials are mixed rather than separated in x.

For a monoenergetic beam, the intensity and fluence are related by, I = k N (5) For that case, I(x) = I(o) e - µLx (6)

Half Value Layers The penetration of an x-ray beam is often defined in terms of the half value layer (HVL), the distance it must travel to decrease in intensity by a factor of two. This is illustrated in Figure 2 for the case of a monoenergetic beam.

Since the half value layer is the distance required to cut the initial number of photons in half, e - (µL HVL1) = 1/2 e + ( µL HVL1 ) = 2 HVL1 = ln(2) / µL = 0.693 / µL (8)

For a monoenergetic beam HVL1 = HVL2 and, as shown in Figure 3, µL is a constant ln N(0) Slope = - L Figure 3 ln N(x) x

Contributions to UL There are five 5 possible photon interactions: 1. Rayleigh (coherent scattering) 2. Photoelectric effect Compton scattering Historically, low energy Compton scattering (~ < 20 keV) was known as Thompson or classical scattering. 4. Pair and triplet production 5. Photonuclear interactions The fourth and fifth of these are not relevant for diagnostic x-ray imaging. We will just summarize the first three of these which are the only important interactions in the diagnostic x-ray energy range.

Rayleigh (coherent) scatter bout 5% of cross-section in the Dx region Elastic event, photon loses essentially none of its energy Photon interacts with the entire atom Atom recoils just enough to conserve momentum Photon redirected through a small angle Photons are only removed from beam in narrow geometry Contributes nothing to dose Contribution to µm ~ Z / k2 (cm2 / g) (9) Rayleigh (coherent) scatter

Photoelectric effect kc e- e- k The basic interaction is shown in Figure 4 Photoelectric effect e- e- • + k kc The incoming photon is absorbed, the bound electron is ejected and the e- hole vacancy is filled with outer shell electrons producing characteristic photons and Auger electrons (electrons knocked out by characteristic x-rays).

The photoelectric cross section is sketched in Figure 5. X-ray Energy k

contribution to µm =  ~ ( Z / k )3 ( cm2 / gm ) (10) L and K-edge discontinuities occur when the photon energy is just large enough to knock out an L or K shell electron respectively. Away from discontinuities,the photoelectric contribution to the mass attenuation coefficient behaves as contribution to µm =  ~ ( Z / k )3 ( cm2 / gm ) (10) This is the dominant contribution to the attenuation coefficient at high Z and low energy.

Compton Scattering __________________ The Compton interaction is shown in Figure 6. This is essentially a collision with the loosely bound outer orbital electrons. The incoming photon is deflected by an angle . Some of its energy is transferred to the “free electron”.The electron is absorbed almost immediately because of the short range of low energy charged particles. The scattered photon retains a large percentage of the incident energy and goes on to interact again.

Figure 6 T = electron kinetic energy e-  e- k  k’ = X-ray energy

The contribution m is usually called and behaves as  NA Z / A = D = electron density (11) Z / A is unity for hydrogen and 0.4 - 0.5 for all other elements. Compton scattering is the dominant effect at low Z and high energy. The expression for the scattered x-ray energy is k’ = k / {1 + (k / moc2) (1 - cos )} (12) with moc2 = 0.511 MeV

cot  = (1 + k / moc2) tan ( / 2) (13) Notice that it is very difficult to discriminate against scattered radiation on the basis of the x-ray energy loss since K / moc2 is on the order of 0.1 and 1 - cos  is very small. The electron energy is given by T = k - k’ The electron scattering angle is given by cot  = (1 + k / moc2) tan ( / 2) (13)

Compton scattering decreases image contrast as illustrated in Figure 7 where the recorded image intensity for the case of primary (unscattered) x-rays (7A) and primary + scatter (7B) are separately shown.

Figure 7A Detected Primary intensity

Figure 7B Detected Primary Intensity + scattered intensity

Compton scattering decreases contrast by permitting x-rays from unrelated regions of the anatomy to be detected behind the contrast producing object. Scatter grids which preferentially transmit primary x-rays and attenuate predominantly scattered x-rays are used in a wide variety of clinical imaging situations.

The total mass attenuation coefficient µm is equal to the sum of the individual coefficients. µm = µL/ p = R  +  (14) The Raleigh scattering contribution is often ignored for the purpose of approximate calculations.

Once you determine the ratio of  and  for one element, you can predict the attenuation coefficient at another energy, e.g., if for substance x,  =  at 20 keV and µ(20) = (20) +(2 0 ) = 2 (2 0 ) then at 40 keV µ(40) =(40) + (40) ≈ (20) + (2 0 ) x (2 0 / 4 0 )3 =(20) + (20) / 8 = 9/8 (2 0 ) = 9 / 1 6 µ(20)

The mass attenuation coefficients in units of cm2 / gm are shown in Figure 8 for bone, muscle, fat and iodine. Note that since bone has a higher atomic number, it provides higher attenuation at low energy. At higher energies, where the Compton effect is dominant, bone, fat and muscle provide the same attenuation for the same thickness measured in gm / cm2.

Since bone obscures chest images, chest radiography is typically done at 120-140 kVp in order to reduce the relative contrast of bone. For mammography where soft tissue contrast is important, energies in the 20 keV range are used. For angiography, the imaging of blood vessels, optimal image contrast is obtained when the beam energy is predominantly in the region of the iodine k-edge where, at the threshold for ejection of k shell electrons, there is an abrupt discontinuity in the iodine attenuation coefficient. It is useful to remember that  =  at 40 keV for bone 20 keV for tissue (15)

Polyenergetic X-Ray Beams: Let us examine the changes in the x-ray spectrum for the case of polyenergetic beams. We saw previously that for a monoenergetic spectrum the log of the attenuation was a straight line. In the polyenergetic case the dependence of the x-ray fluence is shown in Figure 9. Monoenergetic slope = - L polyenergetic ln N(x) x Figure 9

The deviation from linearity is due to the fact that the attenuation coefficient decreases with energy. This is due to the preferential removal of low energy x-rays from the beam as sketched in Figure 10 x = 0 dI/dk Figure 10 Large x k

The value x indicated is the total amount of attenuating material in the beam. As x increases, intensity decreases, but average energy, k, increases and µeff decreases. This is "beam hardening." Because the photoelectric cross section varies as 1 / k 3, low energy x-rays are preferentially removed. This causes the second half value layer to be larger than the first as shown in Figure 11. Figure 11

The homogeneity factor H is defined as H = HVL1 / HVL2 < 1 (16) A homogeneity factor H ≥ 0.6 is required for patient use to insure that the beam is adequately hardened.

Intensity versus Thickness for a Broad Spectrum Now let’s consider the mathematical behavior of the intensity as a function of thickness for a broad spectrum as shown in Figure 12. dI(k,0) dk k Figure 12

dI / dk (k,t) = dI / dk (k, 0) e-µ(k)t (17) The spectrum at depth t can be expressed in terms of the initial spectrum as dI / dk (k,t) = dI / dk (k, 0) e-µ(k)t (17) The integrated intensity is expressed as I (t) = ∫ dI / dk (k,t) dk = ∫ dI / dk (k, 0 ) e-µ(k)t dk (18)

Sometimes the intensity at depth t is expressed in terms of an effective attenuation coefficient µeff defined by the equation I(t) = I(o) e-µeff(t)t (19) µeff(t) is the coefficient you would have to use to predict the intensity if you were to pretend that the beam were monoenergetic.

Note that µeff(t) ≠ µ(t) (effective) ( “instantaneous” ) µeff(t) describes attenuation from 0 to t. µ(t) describes attenuation from t to t + ∆t. The instantaneous coefficient, which is most important for estimating the contrast produced by a differential thickness placed in the patient, is defined by ∆I(t) / I(t) = -µ(t) ∆t (20)

Example Now let’s do an example of a beam hardening calculation. Calculate µ(t) for very large t for the case of a thin target spectrum which is shown in Figure 13 at t=0. Example C dI dk k  km Figure 13

I(t) = ∫ (dI (k, 0 ) / dk) e-µ(k)t dk where the integral goes from 0 to km. Assuming very large t, the beam will be very narrow in energy Pull out e- µ(km)t giving I(t) = Ce-µ(km)t ∫ e- [µ(k) - µ(km)] t dk µ (k) = +  = C 1 + C2 / k3 where C1 is fairly independent of energy.

µ(k) -µ(km) =C2 (1/k3 - 1/km3) = C2 ((km3 - k3) / k3km3) The term e- [µ(k) - µ(km)]t is negligible except at k ≈ km Therefore, let k = km - ∆

Plug in, omit 2 terms, set km~k in the denominator, and do the algebra Plug in, omit 2 terms, set km~k in the denominator, and do the algebra. This results in µ(k) - µ(km) = 3C2(km - k) / km4 = (Km - K) where  = 3C2/km4. Then I(t) I(t) = Ce-µ(km)t ∫ e -  km - k)t dk

Let y = -(k m - k) t dy = t d k Then I(t) = [Ce- µ(km)t ∫ e ydy ] /  where the integral is performed from - km t to 0. This integrates to I(t) ={ (C/) e-µ( km ) t} / t at very large t.

(t)∆t = ∆I(t) / I(t) = - ( µ (km) + 1/t ) ∆t The instantaneous µ(t) is obtained by taking the log and applying equation 20, ln I(t) = ln (C/ - µ(km)t - ln(t) Taking the differential (t)∆t = ∆I(t) / I(t) = - ( µ (km) + 1/t ) ∆t So we get µ(t) = µ(km) + 1/t

As shown in one of the homework problems for the case of thick target bremsstrahlung the result turns out to be µ(t) = µ(km) + 1/t2 Several years ago Peter Joseph used this relationship along with attenuation measurements at very large thicknesses as a means for determining kVp. This is done by plotting (t) as a function of t and determining (km).

K-Edge Filters The absorption properties of various elements having k-edges in the diagnostic region can be used to alter spectra for special purposes. These k edge filters may decrease the average beam energy in some cases leading to “beam softening”. This is illustrated in Figure 14 for the case of iodine and cerium filters which have been used to provide spectra for dual energy imaging where it is desired to have spectra peaked above and below the k-edge of iodine, the element to be imaged in angiography.

Figure 14

Linear combination of attenuation coefficients Since all attenuation coefficients in the diagnostic x-ray range may be expressed in terms of the basic Compton and photoelectric cross sections, any two material coefficients can serve as a basis set from which all other attenuation coefficients may be formed by linear combination. For example the bone attenuation coefficient may be expressed as µb = aP + bC (21) where P and C are the photoelectric and Compton cross sections respectively.

The attenuation coefficients for plastic and aluminum are given by µpl = cP + dC and µal = eP + fC Clearly we could solve for P and C in terms of µpl and µal to obtain µb = gµpl + hµal

Because the P and C have different energy dependences, P proportional to 1/k3 and C independent of energy, by making measurements at two different energies it is possible to solve for the separate contributions of P and C to each part of the image and then to recombine these into tissue, bone, aluminum and plastic images. The advantage of describing tissue and bone in terms of aluminum and plastic is that aluminum and plastic phantoms can be used to calibrate various system parameters or remove nonlinearities.

For example c,d, e and f can be forced to provide the right answers for the known plastic and aluminum coefficients. Using these corrected values, corrected values for g and h, which are functions of c,d,e,and f, can be used to accurately calculate µb which, for example, varies with the amount of bone mineral. The same sort of process is used for tissue. The exact procedure involving the appropriate calibration constants has not been presented above and depends on the details of the image acquisition apparatus.