輻射劑量學之品質保證 講者:蕭安成 物理師 參考資料: 1. The Physics of Radiation Therapy Faiz M. Khan 2. Introduction to Radiological Physics and Radiation Dosimetry.

Slides:



Advertisements
Similar presentations
Chapter 4 Radiation Dosimeters
Advertisements

RADIATION PROTECTION IN RADIOTHERAPY Part 10: Optimization of protection in External Beam Radiotherapy PRACTICAL EXERCISE IAEA Post Graduate Educational.
NE Introduction to Nuclear Science Spring 2012 Classroom Session 9: Radiation Interaction with Matter  Absorbed Dose (D), Kerma (K) Gray (Gy) =
Tenth lecture Last lecture.
NE Introduction to Nuclear Science Spring 2012
Radiation Units & Quantities
Radiologic Units. Intensity Radiation intensity is the amount of energy passing through a given area that is perpendicular to the direction of radiation.
Charged Particle Radiation
Radiation Exposure, Dose and Relative Biological Effectiveness in Medicine Background Image:
Dose. Energy Gained Particles lose energy in matter. Eventually energy loss is due to ionization. An important measure is the amount of energy gained.
Measurement of Absorbed Dose (6)
DOSIMETRY PROTOCOLS.
Quantities and Measurements - 2 Dosimetric Quantities
Kilovoltage X-ray Dosimetry
Radiation Dosimetry Dose Calculations D, LET & H can frequently be obtained reliably by calculations: Alpha & low – Energy Beta Emitters Distributed in.
NPL’s activities dosimetry for proton and ion therapy Hugo Palmans Radiation Dosimetry Team, National Physical Laboratory, Middlesex, UK
Alexander Brandl ERHS 630 Exposure and Dose Environmental and Radiological Health Sciences.
TRAINING COURSE ON RADIATION DOSIMETRY: Instrumentation 1 – Gas detectors / Part 1 Anthony WAKER, University of Ontario Instutute of Technology Wed. 21/11/2012,
ACADs (08-006) Covered Keywords Roentgen, gray, exposure rates, absorbed dose, dose equivalent, quality factors, linear energy transfer, relative biological.
Radiation Dosimetry (Measurement of Absorbed Dose)
Radiology is concerned with the application of radiation to the human body for diagnostically and therapeutically purposes. This requires an understanding.
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
BIOLOGICAL EFFICIENCY OF A THERAPEUTIC PROTON BEAM: A STUDY OF HUMAN MELANOMA CELL LINE I. Petrović 1, A. Ristić-Fira 1, D. Todorović 1, L. Korićanac 1,
Photon and Energy Fluence
Electron Beams: Physical Principles and Dosimetry
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Dose Distribution and Scatter Analysis
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
CT physics and instrumentation
Internal Radiation Dosimetry Lab 9. Radiation Measurement We use different terms depending on whether: 1.The radiation is coming from a radioactive source.
APHY398C 6/4/ Dosimetry   Quantifying the incidence of various biological changes as a function of the radiation dose.   Exposure Ratio of total.
Health Physics 1b: Dosimetric Quantities (and terms)
What is Radiation? The transfer of energy in the form of particles or waves from one object to another though a medium. Module #2.
Interactions of radiation with Matter
3/2003 Rev 1 II.1.2 – slide 1 of 32 IAEA Post Graduate Educational Course Radiation Protection and Safe Use of Radiation Sources Session II.1.2 Part IIQuantities.
Chapter 5 Interactions of Ionizing Radiation. Ionization The process by which a neutral atom acquires a positive or a negative charge Directly ionizing.
Determining Radiation Intensity
PHYSICS 225, 2 ND YEAR LAB NUCLEAR RADIATION DETECTORS G.F. West Thurs, Jan. 19.
What are the dose quantities in CT ? IAEA/RCA Kampala Kampala.
Radiation Units. 1-Radioactivity Units n Physical Units – Becquerel n Amount of radioactive sample s.t. there is 1 atomic decay per second n Henri Becquerel:
Calibration of an Ionisation Chamber for use in Megavoltage Dosimetry
3/2003 Rev 1 II.2.9a – slide 1 of 25 IAEA Post Graduate Educational Course Radiation Protection and Safe Use of Radiation Sources Part IIQuantities and.
INTERACTIONS OF RADIATION WITH MATTER. twCshttp:// twCs
1 A protocol for the determination of absorbed dose from high energy photon and electron beams AAPM TG-21 Protocol (Med Phys 10: , 1983) 和信治癌中心醫院.
Dr Rupak Sethuraman Radiation Biology – 1. FORMAT Introduction Sources of radiation Methods of interaction of radiation with the human body Dosimetry.
Adapting A Clinical Medical Accelerator For Primary Standard Dosimetry
AAPM TG-51 Protocol (Med Phys 26: , 1999)
Qing Liang, PhD Medical Physicist Mercy Health System, Janesville, WI
Louisiana State University Radiation Safety Office
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
INTERACTION OF PARTICLES WITH MATTER
Chapter 5 Interactions of Ionizing Radiation
Electron Beam Therapy.
Dr: Mohamed Afifi By Lecturer Radiological Science
Dose Equivilant Rad Pro III NUCP 2331.
X-Radiation.
Absolute Dose Measurement
A system of dosimetric calculations
Chapter 3. Fundamentals of Dosimetry
Arghya Chattaraj, T. Palani Selvam, D. Datta
Annihilation This interaction in matter most often involves a positron (positive electron) and an electron (negatron). After a positron has transferred.
Outside the nucleus, the beta decay {image} will not occur because the neutron and electron have more total mass than the proton. This process can occur.
Ch 10. A System of Dosimetric Calculations
Chapter 8 (Part 1) Measurement of Absorbed Dose
Ch6 Measurement of Ionizing Radiation
Principles and Practice of Radiation Therapy
Presentation transcript:

輻射劑量學之品質保證 講者:蕭安成 物理師 參考資料: 1. The Physics of Radiation Therapy Faiz M. Khan 2. Introduction to Radiological Physics and Radiation Dosimetry Frank H. Attix,

醫用直線加速器品質保證作業項目

品保作業儀器設備 Daily : Sun Nuclear Daily QA 2

品保作業儀器設備 Beam uniformity

品保作業儀器設備 Output calibration

Measurement of Absorbed Dose

The Roentgen The roentgen is an unit of exposure ( X ). The ICRU defines X as the quotient of dQ by dm where dQ is the absolute value of the total charge of the ions of one sign produced in air when all the electrons ( + or - ) liberated by photons in air of mass dm are completely stopped in air. X = dQ / dm The SI unit is C/kg but the special unit is roentgen ( R ) 1R = 2.58 × 10-4 C/kg

The Roentgen Charged Particle Equilibrium (CPE ) : Electron produced outside the collection region, which enter the ion-collecting region, is equal to the electron produced inside the collection region , which deposit their energy outside the region.

Radiation Absorbed Dose Exposure: photon beam, in air, E<3MeV Absorbed dose: for all types of ionizing radiation Absorbed dose is a measure of the biologically significant effects produced by ionizing radiation Absorbed dose = dE/dm dE is the mean energy imparted by ionizing radiation to material of dm The SI unit for absorbed dose is the gray (Gy) 1Gy = 1 J/kg ( 1 rad=100ergs/g=10-2J/kg, 1cGy=1rad )

Relationship Between Kerma, Exposure, and Absorbed Dose Kerma ( K ): Kinetic energy released in the medium. K = dEtr / dm dEtr is the sum of the initial kinetic energies of all the charged particles liberated by uncharged particles ( photons) in a material of mass dm The unit for kerma is the same as for dose, that is, J/kg. The name of its SI unit is gray (Gy)

Relationship Between Kerma, Exposure, and Absorbed Dose Kerma ( K ): Kcol and Krad are the collision and the radiation parts of kerma K = Kcol + Krad the photon energy fluence, Ψ averaged mass energy absorption coefficient, men / r

Relationship Between Kerma, Exposure, and Absorbed Dose Exposure and Kerma : Exposure is the ionization equivalent of the collision kerma in air. X = (Kcol)air · ( e/w ) w/e = 33.97 J/C

Relationship Between Kerma, Exposure, and Absorbed Dose Absorbed Dose and Kerma :

Relationship Between Kerma, Exposure, and Absorbed Dose Absorbed Dose and Kerma : Suppose D1 is the dose at a point in some material in a photon beam and another material is substituted of a thickness of at least one maximum electron range in all directions from the point, then D2 , the dose in the second material, is related to D1 by

Calculation of Absorbed Dose from Exposure Absorbed Dose to Air : In the presence of charged particle equilibrium (CPE), dose at a point in any medium is equal to the collision part of kerma. Dair = ( Kcol )air = X · ( w/e ) Dair(rad) = 0.876 ( rad/R) · X (R)

Calculation of Absorbed Dose from Exposure Absorbed Dose to Any Medium : Under CPE Dmed / Dair = (men/r)med / (men/r )air · A A = med / air Dmed(rad) = fmed · X (R) · A fmed : roentgen-to-rad conversion factor

Calculation of Absorbed Dose from Exposure Absorbed Dose to Any Medium :

Calculation of Absorbed Dose from Exposure Dose calculation with Ion Chamber In Air For low-energy radiations, chamber wall are thick enough to provide CPE. For high-energy radiation, Co-60, build-up cap + chamber wall to provide CPE.

Farmer Chamber

Parallel-Plate Chamber

Electrometer

Calculation of Absorbed Dose from Exposure Dose calculation with Ion Chamber In Air X = M · Nx ; D f.s. = ftissue· X · Aeq Nx is the exposure calibration factor for the given chamber

Calculation of Absorbed Dose from Exposure Dose Measurement from Exposure with Ion Chamber in a Medium Dmed = M · Nx ·W/e ·[(men/r)med / (men/r)air] ·Am

The Bragg-Gray Cavity Theory Limitations when calculate absorbed dose from exposure: Photon only In air only Photon energy <3MeV The Bragg-Gray cavity theory, on the other hand, may be used without such restrictions to calculate dose directly from ion chamber measurements in a medium

The Bragg-Gray Cavity Theory Bragg-Gray theory The ionization produced in a gas-filled cavity placed in a medium is related to the energy absorbed in the surrounding medium. When the cavity is sufficiently small, electron fluence does not change. Dmed / Dgas = ( S / r )med / ( S / r )gas (S / r)med / (S / r)gas = mass stopping power ratio for the electron crossing the cavity

The Bragg-Gray Cavity Theory Bragg-Gray theory Dmed / Dgas = ( S / r )med / ( S / r )gas Jgas : the ionization charge of one sign produced per unit mass of the cavity gas

The Bragg-Gray Cavity Theory The Spencer-Attix formulation of the Bragg-Gray cavity theory Φ(E) is the distribution of electron fluence in energy L/r is the restricted mass collision stopping power with Δ as the cutoff energy

Effective Point of Measurement Plane Parallel Chambers at the inner surface of the proximal collecting plate Cylindrical Chambers Shift proximal to the chamber axis by 0.75r for an electron beam (TG-21) 0.5r for an electron beam (TG-25) 0.6r for photon beams, 0.5r for electron beams(TG-51)

CALIBRATION OF MEGAVOLTAGE BEAMS: TG-21 PROTOCOL

Cavity-Gas Calibration Factor (Ngas) The AAPM TG-21 protocol for absorbed dose calibration introduced a factor (Ngas ) to represent calibration of the cavity gas in terms of absorbed dose to the gas in the chamber per unit charge or electrometer reading. For an ionization chamber containing air in the cavity and exposed to a Go-60 g ray

Cavity-Gas Calibration Factor (Ngas) Ngas is derived from Nx and other chamber-related parameters, all determined for the calibration energy, e.g., Co-60

Cavity-Gas Calibration Factor (Ngas) Once Ngas, is determined, the chamber can be used as a calibrated Bragg-Gray cavity to determine absorbed dose from photon and electron beams of any energy and in phantoms of any composition Ngas, is unique to each ionization chamber, because it is related to the volume of the chamber

Cavity-Gas Calibration Factor (Ngas) Nx = XM-1 Dgas = Jgas ( W/e ) Ngas = D gas Aion M-1 Assume Aion =1 Ngas = D gas M-1 D gas = M × ( W/e ) / (rair × Vc ) Ngas = ( W/e ) / (rair × Vc ) if the volume of the chamber is 0.6 cm3, its Ngas will be 4.73 × 107 Gy/C

Cavity-Gas Calibration Factor (Ngas)

Chamber as a Bragg-Gray Cavity Photon Beams Suppose the chamber, with its build-up cap removed (it is recommended not to use buildup cap for in-phantom dosimetry), is placed in a medium and irradiated by a photon beam of given energy

Chamber as a Bragg-Gray Cavity Photon Beams Dose to medium at point P corresponding to the center of the chamber will then be P’ corresponding to the chamber's effective point of measurement

Chamber as a Bragg-Gray Cavity Photon Beams Pion correction factor for ion recombination losses Prepl corrects for perturbation in the electron and photon fluences at point P as a result of insertion of the cavity in the medium Pwall accounts for perturbation caused by the wall being different from the medium

Chamber as a Bragg-Gray Cavity Photon Beams The AAPM values for Prepl and Pwall have been derived with the chamber irradiated under the conditions of transient electronic equilibrium (on the descending exponential part of the depth dose curve )

Chamber as a Bragg-Gray Cavity Electron Beams When a chamber, with its build-up cap removed, is placed in a medium and irradiated by an electron beam usually assumed that the chamber wall does not introduce any perturbation of the electron fluence thin-walled (≦0.5 mm) chambers composed of low atomic number materials (e.g., graphite, acrylic) Pwall = 1

Chamber as a Bragg-Gray Cavity Electron Beams For an electron beam of mean energy Ez , at depth Z of measurement

Chamber as a Bragg-Gray Cavity Electron Beams Prepl fluence correction increases the fluence in the cavity since electron scattering out of the cavity is less than that expected in the intact medium Gradient correction Displacement in the effective point of measurement, which gives rise to a correction if the point of measurement is on the sloping part of the depth dose curve

Chamber as a Bragg-Gray Cavity Electron Beams Recommends that the electron beam calibration be made at the point of depth dose maximum Because there is no dose gradient at that depth, the gradient correction is ignored Prepl , then, constitutes only a fluence correction for cylindrical chambers as a function of mean electron energy at the depth of measurement and the inner diameter of ion chamber

Chamber as a Bragg-Gray Cavity Electron Beams a depth ionization curve can be converted into a depth dose curve using A A A B B B

Chamber as a Bragg-Gray Cavity Electron Beams The gradient correction, however, is best handled by shifting the point of measurement toward the surface through a distance of 0.5r For well designed plane-parallel chambers with adequate guard rings, both fluence and gradient corrections are ignored, i.e., Prep, = 1; the point of measurement is at the front surface of the cavity

Calibration Phantom The TG-21 protocol recommends that calibrations be expressed in terms of dose to water polystyrene, or acrylic phantoms may be used, but, requires that the dose calibration be reference to water Scaling factors SF = d plastic / d water = m water / m plastic

Calibration Phantom A calibration phantom must provide at least 5 cm margin laterally beyond field borders and at least 10 cm margin in depth beyond the point of measurement Calibration depths for a megavoltage photon beams are recommended to be between 5- and 10-cm depth, depending on energy For electron beams, the calibration depth recommended by TG-21 is the depth of dose maximum for the reference cone

品保作業儀器設備 Monthly : Keithley 35-040 electrometer + NE 2571 Farmer chamber Victoreen 530 electrometer + PTW N30001 Farmer chamber Solid phantom : Acrylic, Polystyrene, and solid water Sun Nuclear Daily QA 2 et al.

品保作業儀器設備 Annual : Keithley 35-040 electrometer + NE 2571 Farmer chamber Victoreen 530 electrometer + PTW N30001 Farmer chamber Solid phantom : Acrylic, Polystyrene, and solid water Sun Nuclear Daily QA 2 WellHoffer water phantom + IC 10 chamber et al.

品質保證計畫 醫用直線加速器 每月及年度品質保證作業 物理師執行

INTRODUCTION AAPM TG-51 has recently developed a new protocol for the calibration of high-energy photon and electron beams used in radiation therapy. The formalism and the dosimetry procedures recommended in this protocol are based on the used of an ionization chamber calibrated in terms of absorbed dose-to-water in a standards laboratory’s Co-60 gamma ray beam. This is different from the recommendations given in the AAPM TG-21 protocol, which are based on an exposure calibration factor of an ionization chamber in a Co-60 beam. The purpose of this work is to compare the determination of absorbed dose-to-water in reference conditions in high-energy photon and electron beams following the recommendations given in the two protocols.

METHODS AND MATERIALS Calibrations of photon beams ( nominal energy of 6 and 10 MV ) and electron beams ( nominal energy of 6, 8, 10, 12, 15 and 18 MeV ), generated by a Siemens KDS-2 linac, are performed. Farmer-type ( NE 2571 ) ionization chamber was used for photon beam dosimetry and plane parallel ( PTW Markus ) chambers was used for electron beam dosimetry. Absorbed-dose-to-water calibration factor, ND,w , and exposure calibration factor, Nx , for Farmer-type chamber were provided by NATIONAL RADIATION STANDARD LABORATORY INER. Plane-parallel chamber was calibrated against calibrated cylindrical chamber in a 18 MeV electron beam, as recommended in TG-21, TG-39 and TG-51. 60Co

METHODS AND MATERIALS ND,w = 4.5394 cGy/nC , expanded uncertainty = 1% ( k=2 ). Date of report : 2001/12/24, report No. : NRSL – 90084. Nx = 4.7179 R/nC , expanded uncertainty = 1% ( k=2 ). Date of report : 2001/10/17, report No. : NRSL – 90073. Keithley electrometer and Nucleartron water phantom were used in this study. The depth of clinical dosimetry for electron beams was performed at dref , dref = 0.6R50 – 0.1 ( cm ), as recommended in TG-51. Depth ionization measurements along the central axis were made by using the Markus chamber and referenced to that of a 0.12 cm3 RK chamber mounted on the head of the machine. 60Co

RESULTS 60Co pp 60Co

RESULTS

RESULTS

RESULTS

RESULTS

RESULTS

RESULTS

RESULTS

Discussion and Conclusion The doses at 10 cm in water for 6 MV and 10 MV photon beams and the doses at dref in water for 6 to 18 MeV electron beams determined with TG-51 and TG-21 are within 0.3% and 1.4% . According to TG-51, P-P chambers must be used for reference dosimetry in electron beams of energies 6 MeV or less. In the meantime, NRSL provided ND,W factor for Farmer-type chamber only. So, the ND,W factor of a P-P chamber should be determined by using the cross calibrating method. Measurements at the IAEA Dosimetry Lab. have shown that at Co-60, the absorbed dose to water determined by using the ND,W is about 1% higher than that by using the Nx. But, it is different in this study. Detailed analysis should be done including the data given in the two protocols and the calibration factors provided from air-kerma and absorbed dose to water. Co-60 Co-60 Co-60