High Sensitivity EPMA: Past, Present and Future John Donovan CAMCOR University of Oregon (541) 346-4632 camcor.uoregon.edu.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

Saeedeh Ghaffari Nanofabrication Fall 2011 April 15 1.
PROVIDED BY EMAD BEHDAD SPECIALIZED PROTECTION AND CORROSION OF MATERIALS LECTURERS H.MONAJATI,PHD JAFARI,PHD ISLAMIC AZAD UNIVERSITY OF NAJAFABAD BARANCH.
AMCF Materials Characterization School 2012 X-Ray Photoelectron Spectroscopy Tim Morgan.
AA and Atomic Fluorescence Spectroscopy Chapter 9
Spectrum Identification & Artifacts Peak Identification.
The world leader in serving science Principles of Operation.
Chemical Analysis with the SEM Qualitative analysis Quantitative analysis Mapping of locations of elements.
Comparison of Hg EDS and WDS Analysis on Commercial Plastic Material By John J. Donovan University of Oregon, Department of Chemistry.
X-Ray Photoelectron Spectroscopy (XPS)
Electron Probe Microanalysis. A technique to quantitatively analyze samples for their chemical composition on a micro-scale (~1μm) Instrument: Known as.
ED and WD X-ray Analysis
X-Ray Microanalysis – Precision and Sensitivity Recall… K-ratio Si = [I SiKα (unknown ) / I SiKα (std.) ] x CF CF relates concentration in std to pure.
Electron Beam MicroAnalysis- Theory and Application Electron Probe MicroAnalysis - (EPMA) Statistics: Precision and Accuracy UofO- Geology 619 Modified.
Ge 116 Module 1: Scanning Electron Microscopy
Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.
NON DESTRUCTIVE CHEMICAL ANALYSIS
Thin Film Quantitation of Chemistry and Thickness Using EPMA John Donovan Micro Analytical Facility CAMCOR (Characterization of Advanced Materials in Oregon)
Cross section measurements for analysis of D and T in thicker films Liqun Shi Institute of Modern Physics, Fudan University, Shanghai, , People’s.
Types of Analyses Single Collector –Uses only one collector or detector –This is the case with a Daly or EM Advantages: –Need only one detector –Peaks.
4-1 Chap. 7 (Optical Instruments), Chap. 8 (Optical Atomic Spectroscopy) General design of optical instruments Sources of radiation Selection of wavelength.
Seminar Author: Bojan Hiti Mentor: doc. dr. Matjaž Kavčič Determination of trace impurities on Si wafers with x-ray fluorescence.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
X-ray Polarimetry with gas proportional counters through rise-time K. Hayashida, T. Horikawa, Y. Nakashima, H. Tsunemi (Osaka University, Japan)
Materials Analysis and Research Laboratory (MARL) Rooms Town Engineering part of the Materials Science and Engineering Department of the College.
What Is Going On Here? Standards for Glass Analysis Satellite Lines? Beta Lines?
EDS Energy Dispersive Spectroscopy
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
FRANK LABORATORY OF NEUTRON PHYSICS ION BEAM ANALYSIS STANCIU-OPREAN LIGIA SUPERVISOR DR. KOBZEV ALEXANDER.
Quantitative X-ray Spectrometry in TEM/STEM
1 K. Overhage, Q. Tao, G. M. Jursich, C. G. Takoudis Advanced Materials Research Laboratory University of Illinois at Chicago.
Core Community Specifications for Purchase and Acceptance of Electron Beam Instruments John J. Donovan Department of Chemistry CAMCOR (Center for Advanced.
Pushing the Envelope for Beam Sensitive Samples and Trace Element Sensitivity and Accuracy John Donovan (541)
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Other modes associated with SEM: EBIC
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
Detection Limits N no longer >> N B at low concentration What value of N-N B can be measured with statistical significance? Liebhafsky limit: Element is.
Mass spectrometry (Test) Mass spectrometry (MS) is an analytical technique that measures masses of particles and for determining the elemental composition.
Reminders for this week Homework #4 Due Wednesday (5/20) Lithography Lab Due Thursday (5/21) Quiz #3 on Thursday (5/21) – In Classroom –Covers Lithography,
Interferences: Pathological and Otherwise John Donovan Micro Analytical Facility CAMCOR (Center for Advanced Materials Characterization in ORegon) University.
The Power of Iteration John Donovan Micro Analytical Facility CAMCOR (Center for Advanced Materials Characterization in ORegon) University of Oregon Eugene,
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Center for Materials for Information Technology an NSF Materials Science and Engineering Center Substrate Preparation Techniques Lecture 7 G.J. Mankey.
X-Ray Microanalysis – Precision and Sensitivity Recall… wt.fraction I = I SiKα (unknown) / I SiKα (pure std.) K-ratio I = [I SiKα (unknown ) / I SiKα (std.)
X-ray microanalysis in the electron microscope
Secondary Ion Mass Spectrometry A look at SIMS and Surface Analysis.
Basics of Ion Beam Analysis
Characterization of Nanomaterials 1- Scanning Electron Microscopy (SEM) It is one of the most widely used techniques in the characterization of the morphology,
MAN versus Off-peak Background Measurements
Spatial Resolution and minimum detection
(Chapter 7) CHEM–E5225 Electron microscopy
Low-Loss and No-Loss Spectra
Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University.
Auger Electron Spectroscopy (AES)
Electron probe microanalysis EPMA
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
ANALYTICAL CHEMISTRY deals with methods for determining the chemical composition of samples. Dr Seemal Jelani ENVR-303 6/16/2018.
What is XPS? XPS (x-ray photoelectron spectroscopy) is also known as ESCA (electron spectroscopy for chemical analysis). XPS provides chemical information.
NANO 230 Micro/NanoFabrication
Techniques for Improving Quantitative Analysis of Mineral Glasses
Instrument Parameters in WDXRF
Electron probe microanalysis EPMA
Preface: What’s EPMA all about? How does Geology 777 work?
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron Probe Microanalysis EPMA
Electron probe microanalysis EPMA
Electron probe microanalysis EPMA
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Geology 2217 Lab. 1. Recalculation of chemical analyses.
Gain measurements of Chromium GEM foils
Presentation transcript:

High Sensitivity EPMA: Past, Present and Future John Donovan CAMCOR University of Oregon (541) camcor.uoregon.edu

The Past: EPMA/SEM (from Goldstein, et. al. 1988) : Comparison of EDS to WDS, Equal Beam Current, pure Si and Fe, A (0.01 nA), 25 keV Comparison of EDS to WDS, Optimized Conditions, 15 keV, 180 seconds counting time: EDS : 2 x A (2 nA) to give 2K cps spectrum to avoid sum peaks WDS : 3 x A (30 nA) to give 13K cps on Si spectrometer (< 1 % dt)

More recent data for EPMA/SEM:

“ The detection limit cannot be reduced indefinitely by accumulating more counts, however, because systematic errors in the background correction eventually become significant. - Stephen Reed 4 Accuracy (not precision) in characterizing the continuum becomes the limiting factor...

Other Artifacts: “Holes” in the Continuum PPM wt.

Reality Check: Accuracy at the 400 PPM Level? Note: Blank level (C level ) can be non-zero

PET LPET 5 Spectrometers Still… We Need to Improve Sensitivity as well..

(normal PET crystal) 1 ppm 170 ppm NIST SDD EDS is Dead! 5 Spectrometers

WDS Analysis of Hg (polymer door frames from suspected Mexican facility) TakeOff = 40.0 KiloVolt = 20.0 Beam Current = 50 Un 6 std-flex Results in Elemental Weight Percents ELEM: Hg Pb Cr TIME: AVER: SDEV: Detection limit at 99 % Confidence ELEM: Hg Pb Cr AVER: PPM of Hg easily detected, with 48 PPM sensitivity Checked with EDS -count 200 sec - several nA -no Hg found Checked with WDS - count 20 sec - 50 nA - Hg found Why?

Polymer door frames (2000s) from suspected Mexican facility, Check with EDS for 100 sec, 20 keV, 50 nA... nothing sec counting time

Still nothing sec counting time

Hg peaks barely visible… 1000 sec counting time, with Be window inserted (to remove C and O)

15 keV, 200nA, 600 sec on-peak, 600 second off peak, Ti Ka, LPET + PET (aggregate intensities) So what exactly can WDS do on a “typical” quantitative analysis?

15 keV, 200nA, 300 sec on-peak, 300 second off peak, Zr La, LPET + PET (aggregate intensities)

Every sample is beam sensitive -at a sufficiently high beam current... Usually thermally insulating samples (e.g., non conductors…) Classical beam sensitive samples (e.g., alkali, hydrous glasses) Orientation dependent intensity changes over time (e.g., apatites) Trace element measurements (high beam currents, long counting) Use alternating on and off-peak measurements (constant delta) Extrapolate to zero time intensities Use a “blank” correction to apply a systematic error offset SiO2 GlassSiO2 Quartz

F K  in VG2 Glass (1800 secs total count time)

Correcting for Intensity Loss (and Gain) Results in Oxide Weight Percents ELEM: Na2O SiO2 Al2O3 MgO TiO2 MnO P2O5 Cl FeO K2O CaO O H2O SUM AVER: SDEV: SERR: %RSD: VOL%: DEV%: VOLF: LINEAR LINEAR LINEAR LINEAR LINEAR

But Not Always What You Expect!

Hyper-exponential Loss Results in Oxide Weight Percents ELEM: Na2O SiO2 Al2O3 MgO TiO2 MnO P2O5 Cl FeO K2O CaO O H2O SUM AVER: SDEV: SERR: %RSD: VOL%: DEV%: VOLF: QUADRA LINEAR LINEAR LINEAR LINEAR Two exponential processes with different decay constants overlapping in time (?)

The Present “state-of-the-art”: MultiPoint Backgrounds: Combined Qualitative and Quantitative acquisition ThSiO4 (Pb free, i.e., “blank”) Th Mz1 and Mz2

Fluorescence from Si K  Characteristic Fluorescence from Al Continuum ~100 PPM Si

Back To The Future: A proposal for a TEPNA instrument The Transmission Electron Probe Nano Analyzer integrates several new detection technologies to optimize compositional characterization with a target spatial resolution of ~10 nm for “as deposited” films and particles in the range of tens to hundreds of nanometers in thickness, while still attached to electron opaque substrates.

50 nm Bi2Te3, 30 keV, 20 nm beam

Thin films are a trace element problem... ~50 nm Fe, Nb, Se film on Si wafer (20 keV, 30 nA) Is that my signal? Nope… Si sum peak!

18 hour integration at 30 nA can provide significant sensitivity... Newbury ID “blunders” are still here… Si sum peak identified as Sn… Nb peak not identified...

Even though the Nb Ka peak is visible...

Like thin films, nano-particles also present a sensitivity problem... WO 3 nano-particles on Si Si sum peak is not 3 sigma, but neither are the W peaks!

The increase in signal as a 10  m diameter beam is scanned over a region containing a monolayer of Hf atoms deposited on a Si substrate using a Cameca SX50 EPMA WDS Monolayer Detection Demonstrated nm thick, 10 um dia. 100 nm thick, 0.01 um dia. Requires a 1000 fold improvement in sensitivity! (1,000,000 x fewer atoms but 1000 x thicker film)

How do we improve sensitivity 1000 fold? Utilizing high energy emission lines with higher fluorescent yields e.g., Nb L  = 3.5%, Nb K  = 74% (20-30 fold improvement) Energy filtering of Be exit windows for high energy emission lines (?) Why not do it now? Goldstein et. al. 1992

Highest effective fluorescent yields are found for element emission lines whose absorption edges are higher than 8 keV Zn K  is keV Nb K  is keV In 2 cm of Ar 37% of Zn K  trans. 86% of Nb K  trans. In 2 cm of Xe.05% of Zn K  trans. 59% of Nb K  trans. While still retaining soft x-ray sensitivity!

Other sensitivity improvements are possible... Small FC and/or large area crystals (3 to 4 fold improvement) Multiple WDS spectrometers in “aggregate” mode - 2 to 5 fold improvement using only software Increased counting time/beam current in electron “transmission mode” - 30 keV beam through 100 nm of FeS2 loses ~30 eV of energy - assume 2 to 5 fold improvement by increasing time/current Reduced continuum signal using “faraday cup” TEM grid holders - preliminary measurements show a 30% reduction in continuum

An electron beam instrument that integrates several innovations to optimize compositional characterization with a target spatial resolution of ~20 nm for samples in the range of tens to hundreds of nanometers in thickness on various electron opaque substrates. Transmission Electron Probe Nano Analyzer (TEPNA) The TEPNA complements existing analytical techniques by providing an unmet need for quantitative compositional analysis conveniently intermediate between that currently achieved by wavelength dispersive x-ray (WDX) electron probe micro analysis (EPMA) and energy dispersive x-ray (EDX) analytical electron microscopy (AEM).

EPMA vs. LA-ICP-MS Larger error bars for EPMA reflect actual small scale compositional variation.

Current/Future Capabilities of High Sensitivity EPMA/TEPNA WDS: Bulk Analysis: presently single digit PPM sensitivity (and accuracy) -AFTER correction of various continuum artifacts, e.g., “blank” Thin Film/Particle Analysis: feasible now for major/minor elements -with typical ~1 um beam diameters on samples >50 nm thick -requires more sensitivity for <50 nm beam TEPNA: Transmission Electron Probe Nano Analyzer - utilize high energy nm electron beam (transmission mode) - high fluorescent yield lines (> 8 keV) - tandem gas flow/SDD photon counters (full energy sensitivity) - large area/small FC crystals/spectrometers - aggregate intensities in software