Immediate Use Steam Sterilization Roger Hermen, CSPDT, CRCST Sterile Processing Educator.

Slides:



Advertisements
Similar presentations
Infection Control Program
Advertisements

Unit Aseptic Techniques
Validating Sterilization of Medical Devices
1 The Short Story on Long Cycles Why the length of your steam sterilization cycle may effect your sterility assurance.
CESO Education Day November 30, 2004 Shawn Kenny Manager, Central Processing Department University Health Network – Toronto General Hospital.
Cleaning, Packaging & Sterilization of Surgical Instruments
“Getting it right, every step of the way”
1 Clean? Disinfect? Sterilize? What does it all mean?? DA 116 Infection Control.
STEAM Sterilization Issues (Extended Cycles & Flashing)
Continuing Education Contact Hours
By Dr. Shahzadi Tayyaba Hashmi DNT 356. Infection control Infection control is a way to minimize the transmission of microbes in the dental office The.
Disinfection and Sterilization of Patient-Care Equipment
Copyright © 2013 Wolters Kluwer Health | Lippincott Williams & Wilkins Chapter 6 Infection Control: Clinical Procedures.
3.03 Understand support services
Clean? Disinfect? Sterilize? What does it all mean??
Infection Control AHS II Unit F. Standard Precautions Sometimes called “Universal” precautions Sometimes called “Universal” precautions Used to break.
Good hygienic practices
STERLIZATION AND CROSS INFECTION CONTROL IN THE DENTAL PRACTICE:
Instrument Processing DA 116 Infection Control
Instrument Processing DA 116 Infection Control. Instrument Contamination Levels: 1. _______________ 2. _____________________ 3. _____________________.
What is the autoclave thing again. Sanitization: fancy word for Cleaning Removing of foreign material Dirt, body fluids, lubricants Using detergents,
STERILIZATION HEALTHCARE Steam Sterilizers Operation and Preventative Maintenance.
Flash Sterilization By Colleen Landers Registered Nurse Consultant Surgical Suites & Sterile Processing Department.
Cleaning, Packaging & Sterilization of Dental Instruments
Sterilization Methods
Autoclave. What is an Autoclave? Autoclave is a pressurized device designed to heat aqueous solutions above their boiling point at normal atmospheric.
8.02 Aseptic Techniques Implement aseptic technique to maintain equipment Images courtesy of google images.
FLASHTITE INSTRUCTIONS FOR USE Copyright 2010 © Case Medical Inc.
3.03 Understand support services Introduction
Infection Control.
Physical and Chemical Control of Microorganisms. Control of Microorganisms by Physical and Chemical Agents.
By Vickie Kirkley, RN. Director of Materials Management/ Central Supply/ Central Sterile Job Requirements Career / Job Requirements Bachelor of Science.
Department of Quality and Regulatory Affairs Barbara Ann Karmanos Cancer Center The Karmanos Cancer Center Regulatory Readiness (for Non Clinical Staff)
INFECTION CONTROL IN DENTISTRY Dr. Shahzadi Tayyaba Hashmi
CENTRAL STERILE PROCESSING
O BJECTIVES (39 Q UESTIONS ) Develop and review infection prevention and control policies and procedures Collaborate with public health agencies in.
SUR 111 LAB #8. Skill Assessments 10-1 Assembly of the Instrument Set 10-2 Hand Signals 10-3 Loading, Passing, and Unloading the Knife Handle 10-4 Instrument.
$1 Million $500,000 $250,000 $125,000 $64,000 $32,000 $16,000 $8,000 $4,000 $2,000 $1,000 $500 $300 $200 $100 Welcome.
Environmental Cleaning Tool Kit
ASEPTIC TECHNIQUE IN HEALTHCARE. MICROBIAL GROWTH FOLLOWING FACTORS INFLUENCE MICROBIAL GROWTH: TEMPERATURE PH, OR THE VALUES USED IN CHEMISTRY TO EXPRESS.
Preparation of Supplies for Sterilization. Objectives Explain the steps involved in cleaning, disinfecting, wrapping, and packaging sterile packs and.
Sterile Processing What’s in it for Infection Control?
Infection Control Lesson 2:
Felicia Krizman BSN, CRCST  Infection control/ Patient Safety specialist/ Central supply supervisor.
Wet Packs: Strategies for Prevention and Resolution
NANCY CHOBIN, RN, AAS, ACSP, CSPM, CFER CHOBIN & ASSOCIATES CONSULTING HOT TOPICS IN STERILIZATION AND DISINFECTION.
SteriTite Instructions for use
Housekeeping.
Decontamination and Cleaning
2017 SPRING EDUCATION CONFERENCE
3.03 Understand support services PP2
Sterilization Types & Control
به نام خدایی که در این نزدیکی است
3.03 Understand support services
Infection Control in the GI Setting
Basics of Cleaning, Sterilization and High Level Disinfection (HLD)
NYC APIC Monthly Meeting Questions and Answers
3.03 Understand support services
Infection Control Lesson 3: Surgical Asepsis
Sterilization process
3.03 Understand support services
Sterilization September 2018.
Sterilization Standards & Processes
CS Cheat Sheet.
3.03 Understand support services PP2
Infection Prevention Update
3.03 Understand support services
Sterile Technique.
3.03 Understand support services
3.03 Understand support services
Presentation transcript:

Immediate Use Steam Sterilization Roger Hermen, CSPDT, CRCST Sterile Processing Educator

Objectives  To review the process for Immediate Use Steam Sterilization  Discuss Competency Requirements for Immediate Use Steam Sterilization

Times sure have changed…. In the beginning…… OR cleaned, assembled and wrapped own instruments OR cleaned, assembled and wrapped own instruments Limited choices for sterilization Limited choices for sterilization If nothing melted and indicators turned, everything was ok!If nothing melted and indicators turned, everything was ok!

Sterilization Concerns & Issues  Process is complex  Requires staff to consistently follow all the necessary steps each time to ensure the sterility of the instruments to the point of use  No margin for error  Improper technique can result in the use of contaminated instruments in surgery resulting in serious consequences including surgical site infections (SSI)

SSI Statistics (CDC, 2011)  SSIs affect 2 to 5 % of all patients undergoing surgery 300,000 patients each year 300,000 patients each year  3 % mortality rate  3 % mortality rate  Increase of hospital length of stay 7 to 10 days  Equates to increase costs of $3,000 to $29,000 per SSI with upwards of $10 billion annually

2011 Multi-Society Position Statement  Stressed the importance of education for the personnel directly involved in reprocessing of surgical instruments to the point of use  Surgical and sterile processing personnel should be knowledgeable regarding the different types of steam sterilization cycles, standards and practices, cleaning, decontamination and rinsing processes, and correct aseptic transfer from the sterilizer to the point of use (AORN, 2011a)  Surgical and sterile processing personnel should be knowledgeable regarding the different types of steam sterilization cycles, standards and practices, cleaning, decontamination and rinsing processes, and correct aseptic transfer from the sterilizer to the point of use (AORN, 2011a)

Current Standards & Recommended Practices Best Practices are evidenced based and developed through sound principles, scientific data (research) and expert opinions  AAMI (Association for the Advancement of Medical Instrumentation)  AORN  CDC

The Joint Commission  Broadening its scrutiny and examining healthcare facility's sterilization practices  Looking at the entire sterilization process (cleaning and decontamination, sterilization, storage or return to the sterile field)  Asking staff whether they followed manufacturers' instructions and recommendations for sterilization and wrapping/packaging

Manufacturers Obligation for Reusable Medical Devices  Manufacturers responsibility to provide complete and comprehensive written instructions on Handling Handling Cleaning Cleaning Disinfection Disinfection Testing Testing Packaging Packaging Sterilization Sterilization

HC Personnel Responsibility for Reusable Medical Devices  Obtain and review manufacturers current data and recommendations from each manufacturer  Ensure that they have the necessary resources to follow manufacturers instructions thoroughly OR Workroom Binder OR Workroom Binder OneSource On Line Subscription OneSource On Line Subscription

Sterilization A process designed to remove or destroy all viable forms of microbial life, including bacterial spores Steam (Gravity/Pre-Vac) Steam (Gravity/Pre-Vac) ETO Gas ETO Gas Gas Plasma Hydrogen Peroxide (Sterrad) Gas Plasma Hydrogen Peroxide (Sterrad)

Immediate Use Steam Sterilization “FLASH” The process of sterilizing patient care items for immediate use Pre-Vac cycles provide the shortest allowable times to obtain a sterilized item to transfer to the surgical field Pre-Vac cycles provide the shortest allowable times to obtain a sterilized item to transfer to the surgical field

Immediate-use steam sterilization should only be used…. When a one-of-a-kind instrument has been contaminated and needs to be replaced to the sterile field immediately When a one-of-a-kind instrument has been contaminated and needs to be replaced to the sterile field immediately When an item has dropped on the floor and is needed to continue a surgical procedure When an item has dropped on the floor and is needed to continue a surgical procedure When specific instruments are needed for an emergency procedure When specific instruments are needed for an emergency procedure When there is no other sterilization alternative When there is no other sterilization alternative

Immediate-use steam sterilization should only be used….  When the device manufacturer’s written instructions on cycle times, exposure times, temperature settings and dry times are followed  When items are dissembled and thoroughly cleaned with detergent and water to remove soil, blood, body fats and other substances  When lumens are brushed and flushed under the water with cleaning solution and rinsed thoroughly  When items are placed in a closed sterilization container in a manner that allows steam contact and aseptic transfer to the operating room

Clinical Competency In order to maintain competency, each member of the staff that performs flash sterilization should receive competency based training and should be evaluated on an annual basis

Cleaning All items must be sent through a mechanical washer in SPD prior to any IUSS that prepares the item for safe handling by removing, usually with detergents, of all adherent visible soil from the surfaces, crevices, serrations, joints and lumens of instruments/devices/equipment

Intra-operative Cleaning  Instruments are cleaned and or irrigated with Sterile Water as they are used during surgery  This prevents blood and debris from drying on instruments  Normal Saline will cause pitting

Cleaning Points  Follow manufacturers recommendations  Apply Personal Protective Equipment (PPE)

Cleaning Points  Instruments with multiple parts should be disassembled for cleaning  Soft-bristle brushes should be used to clean the instruments, with particular attention paid to ratchets, serrations, and box locks (joints)  Manual cleaning should be done while holding the instrument and the brush under the water level to prevent aerosolization of bacteria

Cleaning Agent  Necessary to make the cleaning process more efficient and to protect the surgical instruments' finish  All detergents should be measured carefully and used in accordance with the manufacturers' instructions

Cleaning Agent Asepti-Zyme Instrument Cleaner  Enzyme formulated to remove and dissolve proteinacious matter  ½ to 1 ounce in 1 Gallon of Water

Wrapping  Transport the cleaned item to the prep and pack area  Choose appropriate packaging material Single Wrap Single Wrap Rigid Container/Tray Rigid Container/Tray  Place Chemical Indicator inside of each package or tray

Wrap

Transport items back to the O.R. for further processing in steam sterilizer

Steam Sterilizer  Place item into Sterilizer  Choose the appropriate cycle to start  Complete ALL Documentation for the appropriate Sterilizer

Steam Sterilization Log Loading   Patient Label with Name and medical record (MR) number   Initials of person Loading the sterilizer   Room #   Implant & BI Lot #   The reason for flash sterilization   Load description, including each individual item in the load   Load Cycle (i.e. 1, 2, 3, 4)   It is imperative that all information is entered onto this form with each cycle   At end of cycle all BIs that have been run in the load must be given to SPD so they can incubate the BI to assure that the BI has passed   Results of the Quality Control Tests (i.e. Bowie-Dick test, biological indicator) will be recorded on biological sheet in SPD after the BI has passed incubation

Steam Sterilization Log Unloading  Verify exposure parameters on the Sterilizer Gauge & Recording Temperature of degrees F Minimum exposure time for nonporous, non-lumened items only is 3 minutes Minimum exposure time for metal items with lumens and porous items (eg. rubber, plastic) sterilized together: 1. dynamic air removal (prevacuum) cycle: 4 min  Name of person unloading contents

Unloading Contents   After completion of the cycle, the wrapper and the item within will be hot   Personnel must wear sterile gloves and may use sterile towels as "potholders" when removing items from the sterilizer   Once in the room, the wrapper can be opened by the circulator   The sterile items may then be removed by the scrub person and taken to the sterile field

Quality Control Biological Indicator Bacterial spores in a carrier that serves to demonstrate whether sterilization conditions were met Performed Daily and with loads containing implants

Implants   Loads containing implants will be quarantined until the result of the biological indicator test is available   If this is not possible then the exception form for premature release of sterilized sets and implantable devices/trays must be completed

Quality Control Chemical Indicator A device employed to monitor one or more process parameters of the sterilization cycle. Consists of chemical/ink dye that provides a distinct color change when exposed to the sterilization process Consists of chemical/ink dye that provides a distinct color change when exposed to the sterilization process

Quality Control Mechanical Indicator A device that indicates sterilization cycle conditions Including time, temperature and pressure

Bowie Dick Test Diagnostic test of a Pre-Vac steam sterilizer’s ability to remove air from the chamber Performed Daily on Pre-Vac Sterilizers

Guidelines for Completion of Competency  OR staff that utilize the Immediate Use Steam Sterilizer will need to be signed off by the appropriate person  Competency for Immediate Use Steam Sterilization will be kept in the employee file

Questions