Traceability and Legal Metrology

Slides:



Advertisements
Similar presentations
49 th CGSIC Meeting in Savannah, Georgia, September GPS Time and Frequency Transfer Activities at NIST Victor Zhang National Institute of Standards.
Advertisements

PRINCIPLES OF A CALIBRATION MANAGEMENT SYSTEM
Effectively applying ISO9001:2000 clauses 6 and 7.
ISO 9000 Quality Standards ISO 9000 describes quality assurance elements in generic terms that can be applied to any business. It treats an enterprise.
Ensuring Better Services and Fair Value “Introduction and roadmap to implementation of ISO in Zambia’s water utilities” Kasenga Hara March 2015.
Judah Levine, Pulsar-Timing-2005: 1 Coordinated Universal Time (UTC) Judah Levine JILA –- Time and Frequency Div. NIST and Univ. of Colorado Boulder, Colorado.
Copyright Alan Rowley Associates Steps to an Accurate Result Select a method and validate it as suitable for the purpose envisaged. Establish that.
General Concepts of Time and Frequency Metrology
CMCs and the BIPM Key Comparison Database Raul Fernando Solís Betancur 2012.
Twinning Project RO2006/IB/EN/09 1 Saxony-Anhalt State Environmental Protection Agency Wolfgang GarcheBukarest Wolfgang Garche Saxony-Anhalt.
Primary and Derived Measures Terminology Prioritization 1.
CMCs and the BIPM Key Comparison Database Raul Fernando Solís Betancur 2015.
Quality Management System
Statistical Process Control
Traceability to the SI: What does it mean in practice? Ms. Sally Bruce and Ms. Dana Leaman National Voluntary Laboratory Accreditation Program NIST.
Chapter 11 Quality Control.
Quality Assurance/Quality Control Policy
Senior Lecturer in Pathology
1 Practical Interpretation of Unbroken Chain in Metrological Traceability as to VIM 3 Center for Measurement Standards (CMS) Industrial Technology Research.
ISO 9000 Certification ISO 9001 and ISO
APG15-4 Inf. Session 기반표준본부 시간센터 UTC and Leap Second in Rep. of Korea UTC and Leap Second in Rep. of Korea Dai-Hyuk Yu Center for Time and Frequency Division.
Discussion Topics For SIM Time and Frequency Metrology Working Group SIM Time and Frequency Metrology Working Group Workshop and Planning Meeting Panama.
1 Time Scales UT0, UT1, EAL, TAI, UTC Ricardo José de Carvalho National Observatory Time Service Division February 06, 2008.
Ensuring of Traceability
ISO/IEC A review of the new standard for laboratory accreditation
BSJ National Time Realization. Background The Bureau of Standards, Jamaica embarked on a Time and Frequency Program Its aim was to establish a national.
Time and Frequency activities at CENAMEP Panama, 2015.
Your measurement technology partner for global competitiveness CSIR National Metrology Laboratory The National Metrology Laboratory of South Africa Feedback.
Description of Work Objectives WP 8 serves to define the technical content of the VI for contact thermometry. It will contain a selection of guides to.
Parallel and Distributed Simulation Synchronizing Wallclock Time.
Distributing UTC(NIST) to Industrial Time and Frequency Users Michael Lombardi NIST Time and Frequency Division Distributing UTC(NIST) to Industrial Time.
National Time Service Center. CAS Time Standard and Ensemble Pulsar Time Scale Ding Chen, George & Bill, Dick, PPTA team 2011 年 5 月 9 日, Beijing.
43rd CGSIC - Timing subcommittee What future for UTC ? Włodzimierz Lewandowski Bureau International des Poids et Mesures Sèvres.
SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006 ISO/TS 16949:2002 certification – Meeting the requirements of.
Principles behind ISO/IEC 17025
Quality control & Statistics. Definition: it is the science of gathering, analyzing, interpreting and representing data. Example: introduction a new test.
Forum FM308: Policy Changes Tuesday June 5 10:00 a.m. AIHA Laboratory Accreditation Policy Changes—ISO “How Do Changes Affect My Laboratory?” “What.
The Inter-American Metrology System (SIM) Common-View GPS Comparison Network Michael A. Lombardi and Andrew N. Novick National Institute of Standards and.
Traceability: What is traceability? Competent transfer of traceability from a Standard to the device under test. CALIBRATION CERTIFICATES : ISO/IEC 17025:2005.
18 September How does my phone figure out my location? Why GPS of course! Which is to say the array of ~30 US Satellites is used.
It was found in 1946 in Geneva, Switzerland. its main purpose is to promote the development of international standards to facilitate the exchange of goods.
Overview of Time and Frequency Metrology
Kenneth Johnston US Naval Observatory Presented by Jules McNeff International Committee on GNSS Provider’s Forum Bangalore, India 4 September 2007 GPS.
TIMING APPLICATIONS OF GPS High Energy Transmission with High Precision GPS Time Gaurav Sharma John Hannah Vivekanand Sivaraman.
Petra Spitzer, Ralf Eberhardt, Janine Giera PTB, Germany Barbara Werner, Anett Czysch ZMK, Germany Traceable calibration procedures in chemistry Example:
1 Analytical Forensic Metrology Analytical Forensic Metrology 4 th International Conference on Forensic Research & Technology September 29, 2015 Jerry.
? Training removes doubt, instills confidence, and lays the foundation for everyone’s skill and experience level.
© A2LA Understanding the Accreditation Hierarchy - Who’s Who and What’s What ? NCSL International Conference Orlando, FL August 7, 2008 by Roxanne.
Visit us at E mail: Tele:
DOCUMENTATION ISO/IEC 17025:2005 Documentation.
Table 1. The Five Medical Laboratory Case Studies Jan Martin et al. Transnational Evaluation of a New Continuing Professional Development Activity for.
You’ve Found The Cure For Cancer… But Can’t Reproduce Your Results! Could Your Test Equipment Be The Weakest Link?
The Traceability of Conductivity Meter and Cell Calibration in the Low Conductivity Ranges Jun Bautista, Director, Genzyme Metrology Genzyme Corporation,
The Use of GPS Disciplined Oscillators as Primary Frequency Standards for Calibration Laboratories NCSLI – August 2008 Michael Lombardi National Institute.
Timing applications of GPS Time for GPS Vivekanandanmurugan Sivaraman Gaurav Sharma John Hannah.
Traceability in Chemical Measurement Comparable results are needed in order to avoid duplicating measurements which cost time and money. Comparable results.
ISO/IEC
Proficiency Testing as a quality improvement tool
External Validation of Quality Programs
Managing Your Company Calibration Program
Results of the questionnaire for remote calibration in T&F field
Gaurav Sharma John Hannah Vivekanand Sivaraman
Gaurav Sharma John Hannah Vivekanand Sivaraman
Chapter 11 Quality Control.
External Validation of Quality Programs
Presented By: T.M Ramashidzha Metrology Scientist
Primary and Derived Measures Terminology Prioritization
Presentation transcript:

Traceability and Legal Metrology

4 Parts of a Measurement Device Under Test (quartz, rubidium, and cesium) Reference (cesium oscillator or transfer standard) Method (measurement system and procedure) Uncertainty Analysis (statistics and data reduction)

ISO definition of traceability The property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties.

The Traceability Chain The traceability chain is a series of comparisons between the device under test to a reference. The final comparison in the chain is made using the International System (SI) units as a reference. Each comparison is a link in the chain. The uncertainty of each comparison (link) must be known and documented.

The Traceability Chain (cont.) National metrology institutes (NMIs) like NIST provide the ultimate measurement references for their countries. The intent of all NMIs is to realize the SI units as closely as possible. Although the goal is to establish traceability to the SI, this is often done by comparing to an NMI that in turn compares its references to the SI.

Time and Frequency SI Units Second (s) standard unit for time interval intrinsic, defined using cesium 133 atom one of 7 base SI units Hertz (Hz) standard unit for frequency (s-1) events per second one of 21 SI units derived from base units

The Traceability Pyramid

The Importance of Traceability Establishing traceability to national and international standards provides evidence that measurements are being made correctly Traceable measurements are essential elements of: quality control systems trade agreements allowing compatible products to be manufactured all over the world

The Traceability Problem Technical usually not a difficult technical problem, especially if requirements are modest Legal requires repeatable, reliable procedures requires documentation of traceability chain for quality control systems may be contractual requirement or a prerequisite for doing business

ISO Guide 17025 - 5.6.2.1 For calibration laboratories, the program for calibration of equipment shall be designed and operated so as to ensure that calibrations and measurements made by the laboratory are traceable to the SI (Systeme International) units of measurement.

Laboratory Accreditation Calibration laboratories that seek accreditation through a program such as NVLAP must comply with ISO 17025 and document the traceability chain for each measurement. The traceability chain must link the measurement back to the SI unit, and the documentation must include the uncertainty of each link of the chain.

Types of Time and Frequency Information Date and Time-of-Day records when an event happened Time Interval duration between two events Frequency rate of a repetitive event

Everything Relates to the Second By counting seconds, we can determine date and time-of-day By counting events per second, we can measure frequency

SI Definition of the Second The duration of 9,192,631,770 periods of the radiation corresponding to the transition between two hyperfine levels of the ground state of the cesium-133 atom.

Uncertainties of physical realizations of the base SI units

Coordinated Universal Time (UTC) About 50 NMIs do a continuous comparison of their standards (over 200 atomic oscillators are involved). The average value of these standards is used to create the UTC time scale. The BIPM in France handles the data and computes the uncertainties. BIPM Circular T shows difference between UTC and UTC(k), or the UTC maintained by each NMI.

BIPM Circular T (www.bipm.fr)

Low accuracy measurements that require traceability

High accuracy measurements that require traceability

Other Areas of Metrology that require Traceable Frequency

Ways to establish traceability to the SI through NIST NIST controlled broadcasts WWV/WWVH, WWVB, ACTS, ITS NIST monitored broadcasts LORAN-C, GPS NIST controls/monitors measurements Frequency Measurement Service

Traceability chain for signals controlled by a NMI

The Distribution of UTC(NIST)

Traceability chain for signals monitored by a NMI

GPS Monitoring More complex than monitoring ground based signals because: Each orbiting satellite carries its own frequency standards Satellites can only be received when they fly over monitoring station

NIST GPS Data Archive

NIST GPS Data Archive

GPS Traceability Chain A

GPS Traceability Chain B

NIST traceable signals and links

Measuring 1 x 10-8

Measuring 1 x 10-9

Measuring 1 x 10-10

Measuring 1 x 10-11

Time and Frequency is well suited for remote calibrations Since frequency standards are sensitive to shipment and environmental changes, calibrations are often made at the customer’s site Radio reference (often GPS) is used as transfer standard Calibrations can be monitored by an NMI using a modem or Internet link

NIST Frequency Measurement and Analysis Service Measurements are made remotely at customer’s site NIST supplies everything needed Measurement process is automated NIST downloads data by modem, performs uncertainty analysis, sends calibration reports Provides traceability to NIST with uncertainty of 2 x 10-13 per day

NIST Frequency Measurement and Analysis Service (cont.) NIST Service ID Number 76100S 5 oscillators can be calibrated at once Calibrations are automated Recognized by NVLAP Replacement parts sent when necessary using overnight delivery service

FMAS Customer Map (labs with cesium oscillators in red)

NIST FMAS Display

Calibration Reports Monthly calibration reports are sent to each NIST FMAS customer. Report is compliant with ISO Guide 25/17025 and includes a statement of uncertainty. The report defines the calibration period as one day (24 hours). The report annotates all situations where no data were recorded or where the device under test was out of tolerance.

Reporting of Measurement Errors and Outages The monthly calibration report annotates the following situations: No data were recorded GPS reception problems GPS broadcast errors Measurement System Errors including hardware or software failures Device under Test errors Device under Test substitutions/changes

FMAS Specifications

NIST Policy on Traceability For more information about traceability, visit the NIST Traceability Web Site: http://nist.gov/traceability/