1 / 9 ASTM D19 Method Validation Procedures William Lipps Analytical & Measuring Instrument Division July, 2015.

Slides:



Advertisements
Similar presentations
Radiochemical Methods and Data Evaluation
Advertisements

Why Lab. Certification? Laboratory certification tries to insure that the laboratory is qualified and capable of analyzing the sample and obtaining quality.
Quality is a Lousy Idea-
LABORATORY CERTIFICATION & DATA QUALITY MICHAEL W. MILLER, Ph.D. NJ-DEP Office of Quality Assurance
Validation of screening methods (2002/657/EC)
UNDERSTANDING ENVIRONMENTAL LABORATORY/QC REPORTS Maya Murshak – Merit Laboratories, Inc.
Assessing Laboratory Quality – Systematic Bias Robert O. Miller Colorado State University Fort Collins, CO.
Mentoring Session Technical Assistance Committee Method Modifications.
Measurement of Dissolved Oxygen with a Luminescence-based Oxygen Quenching Sensor By Cary B. Jackson, Ph.D. Hach Company.
Interpreting Your Lab Report & Quality Control Results
World Health Organization
Summary 1 l The Analytical Problem l Data Handling.
Instrumental Methods: Intro  Types of Instrumental Methods  Fundamental Components of an Instrument  Instruments Measure Voltages and Currents! (“Machines”
Analytical Chemistry Definition: the science of extraction, identification, and quantitation of an unknown sample. Example Applications: Human Genome Project.
QA/QC FOR ENVIRONMENTAL MEASUREMENT Unit 4: Module 13, Lecture 2.
Chemometrics Method comparison
Field Analysis Quality Control
Validation of Analytical Method
Method Selection and Evaluation Method Selection and Evaluation D. Kefaya EL- Sayed Mohamed Prof. Of Clinical Pathology (Clinical Chemistry), Mansoura.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
LC-MS/MS Analysis of Naphthenic Acids in Environmental Waters Coreen Hamilton, Million B. Woudneh & Guanghui Wang Presented at Workshop on Analytical Strategies.
Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.
HD 2007 Rule Diesel Fuel Sulfur Testing and Sampling Methods and Requirements US EPA Office of Transportation and Air Quality November 20, 2002.
How to Select a Test Method Marlene Moore Advanced Systems, Inc. June 15, 2010.
QA/QC and QUALIFIERS LOU ANN FISHER CITY OF STILLWATER, OK
Wastewater Laboratory Certification Presented at: KWWOA April17,
Quality Assurance How do you know your results are correct? How confident are you?
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
Data Analysis: Quantitative Statements about Instrument and Method Performance.
William Telliard U.S. EPA Office of Science and Technology
Wenclawiak, B.: Fit for Purpose – A Customers View© Springer-Verlag Berlin Heidelberg 2003 In: Wenclawiak, Koch, Hadjicostas (eds.) Quality Assurance in.
Industrial Technology Institute Test Method Validation & Verification H.P.P.S.Somasiri Principal Research Scientist / SDD-QAD /QM Industrial Technology.
Control Charts and Trend Analysis for ISO 17025
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
Copyright © 2015, TestAmerica Laboratories, Inc. All rights reserved. 1 EPAs New MDL Procedure What it Means, Why it Works, and How to Comply Richard Burrows.
Uncertainty and error in measurement
Achieving Method Compliance with Innovative Technologies Tony Francis, PhD Principal, SAW Environmental TCEQ – Environmental Trade.
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Chapter 5 Quality Assurance and
Determinations and Interpretations of FTIR Detection Limits
The Method Update Rule Does this affect me? HOW?
Determination of TKN by Subtraction using ASTM D and ASTM D
Quality is a Lousy Idea-
Clean Water Act Methods Overview of EPA’s CWA Method Activities
The 2015/2016 TNI Standard and the EPA MDL Update
Simple Mathematics and How It Impacts You
Overview of EPA Method 1631, Revision E By Roy W
William Lipps Analytical & Measuring Instrument Division August 2017
EPA Method Equivalency
September 2017 Phillip Godorov Caitlin Farrell
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
ASTM International Interlaboratory Study Program (ILS)
Added Monrovia and South Bend to the slide.
Have Instrument, make method; How new methods are made and validated
Metals Subgroup Agenda
Quality is a Lousy Idea-
Update on ASTM and Standard Methods method development activities
EPA Method Equivalency
What it Means, Why it Works, and How to Comply
Deyuan Kong and Sara Mcmillen, Chevron Energy Technology Company USA
METHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS
Why Use Them? By: Marcy Bolek – Alloway
Environmental Laboratory Certification Program (ELCP)
Method Selection and Evaluation
Satish Pradhan Dnyanasadhana College, Thane. Department of Chemistry S
Multi-laboratory Method Validation
TCEQ Environmental Trade Fair Water Quality Division
Andy Eaton, PhD, BCES Technical Director Joint Editorial Board-Chair
2019 AWOP National Meeting Office of Ground Water and Drinking Water
California Water Boards PFAS Efforts
Presentation transcript:

1 / 9 ASTM D19 Method Validation Procedures William Lipps Analytical & Measuring Instrument Division July, 2015

This presentation is the opinion of the author. The author is a volunteer member of ASTM

All methods begin with an idea

An idea can be introduced by anyone at a sub committee meeting Determine if new standard is needed Identify and gather key stakeholders Appoint a Task Group Chair Register a Work Item Subcommittee decides on title and scope

By registering as a Work Item, ASTM: Provides a tracking number, “WK5432” Alerts other members Initiates a time table and process

In this presentation we are referring to “Standardized ” Methods Technique Scientific Principle, such as GCMS Method Adaptation of a technique to a measurement problem ASTM Method A method of known precision issued by ASTM

Validation of methods that measure the same analyte as other methods needs to establish: Equivalency Same result as approved method (interference free) Same QC Same detection (wet chemistry) Same extraction (GC) Existing Method New

Validation of the method must include Calibration MDL DOC Single Lab Study (matrices) Precision Spike Recoveries Multi-Lab study Precision

Validation of a New Method Preliminary Literature Search Design Phase Development Phase Validation Phase Evaluation Phase

The design phase occurs before significant amounts of data are collected: Literature search Draft “method” Scope Summary Technique Matrices Concentration Range (estimate) Optional vote at Subcommittee level

The Development phase collects preliminary data Single or two lab studies Proof of concept “preliminary data” Vote at Subcommittee level

The Validation phase includes collection of the following information: Selectivity Correctly ID analyte in matrices Calibration Technique and model Repeatability At a range of concentrations In numerous matrices Bias Compare to known matrices Spike samples Compare to other techniques Ruggedness What can change results

Before an inter-laboratory study is carried out the draft must pass subcommittee balloting Evaluation Phase Prefer 9 labs Prefer 9 matrices for CWA Minimum 3 matrices for SDWA 3 Youden Pairs (optional)

Validation of ASTM D7781 Test Method For Nitrite-Nitrate in Water by Nitrate Reductase Literature Search - justification Analytical methods using nitrate reductase have been previously reported,,, however, these investigators limited testing to surface and ground waters without evaluating complex matrices such as wastewater

A specific “technique” was chosen This new ASTM nitrate method is a discrete analyzer method. Discrete analyzers are defined in the standard

Selectivity was verified by showing nitrite and nitrate are recovered equally

Comparison with another method and evaluation of preservatives was made Sample #Commercial Laboratory Results, Analysis Method EPA (mg NO 3 +NO 2 -N/L) Analysis by Reductase (mg NO 3 +NO 2 - N/L) Non- Preserved < < Lost

Since the method is for an existing EPA pollutant (nitrate plus nitrite) with established sampling and preservation guidelines, D19.05 did not perform a holding time study.

Potential interferences were evaluated (partial list) Species Added (mg/L) Unspiked Sample Result (mg/L) Spiked Sample Results (mg/L) Spike Added (mg/L) % Recovery Cl F-F Br < PO SO <

Inter-Laboratory study carried out after successful subcommittee ballot 10 laboratories Four different discrete analyzer manufacturers One manufacturer had several different models

Each lab calculated reduction efficiency Lab Mg/L NO2-N (Found) Mg/L NO3-N (found) % Efficiency *

Each lab evaluated an LCS and control limits were established average Standard Deviation Lower Limit (99% CI)92.90 Upper Limit (99% CI)1093

Single Lab precision was plotted

Multiple Lab precision was plotted

Sample Matrix Mean (mg/L) Standard Deviation No. of Laboratories No. of Results Multi- laboratory %RSD High TDS (500 ppm) % High TOC (2 ppm) % ERA #698 WS % USGS N %

Sample MatrixMean (mg/L) Standard Deviation No. of Results Single Operator %RSD WW treatment % treatment plant effluent # % treatment plant effluent # % Paper Mill waste stream effluent % metal finisher wastewater effluent % commercial laundry wastewater effluent % ERA #507 Hardness % Confined Animal Feeding Operation (CAFO) effluent % Low Nutrient Seawater %

ASTM method validation procedures produce methods of defined precision and accuracy

Thank You, for more information contact me