IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN99.699.7100.095.696.4100.082.985.099.6.

Slides:



Advertisements
Similar presentations
Introduction to Hypothesis Testing
Advertisements

Comparing Two Proportions (p1 vs. p2)
Commissioning an Anthropomorphic Spine and Lung Phantom for Remote Dose Verification of Institutions Participating in RTOG 0631 Douglas Caruthers, M.S.;
In the past few years the usage of conformal and IMRT treatments has been increasing rapidly. These treatments employ the use of tighter margins around.
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Slice Thickness Interpolation: The effect of interpolated slice thickness on the 2D vs 3D gamma results are shown in Table 3 for the QA data set only.
Algorithms used in heterogeneous dose calculations show systematic error as measured with the Radiological Physics Center’s anthropomorphic thorax phantom.
Evaluation of the characteristics of TLD LiF:Mg.Ti-100 Powder: A Measure of Consistency Between Multiple Batches of Powder Paola Alvarez,Jose Francisco.
Background:  IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer.  It has.
Chapter 7 Introduction to Sampling Distributions
The Dependence of the Apparent Diffusion Coefficient on Voxel Location and Calculation Method Lars Ewell 1, Naren Vijayakumar Meeting of the American.
Business Statistics: A Decision-Making Approach, 6e © 2005 Prentice-Hall, Inc. Chap 6-1 Introduction to Statistics Chapter 7 Sampling Distributions.
PY 427 Statistics 1Fall 2006 Kin Ching Kong, Ph.D Lecture 6 Chicago School of Professional Psychology.
Innovation/Impact: By designing a simulated human shaped (anthropomorphic) plastic phantom with targets, organs at risk (OAR) and heterogeneities, the.
CHAPTER 19: Two-Sample Problems
Introduction Modern radiation therapies such as intensity-modulated radiation therapy (IMRT) and volume modulated arc therapy (VMAT) demand from dose calculation.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
Respected Professor Kihyeon Cho
Introduction Ion recombination is approximately corrected for in the Task-Group-51 protocol by P ion, which is calculated by a two-voltage measurement.
Patient Plan Results: Table 3 shows the ratio of the Pinnacle TPS calculation to the DPM recalculation for the mean dose from selected regions of interest.
Kelly Younge, Ph.DKelly Younge, Ph.D Don Roberts, Benedick Fraass, Daniel McShan, and Martha Matuszak University of Michigan, Department of Radiation Oncology,University.
Pion test beam from KEK: momentum studies Data provided by Toho group: 2512 beam tracks D. Duchesneau April 27 th 2011 Track  x Track  y Base track positions.
David F. Lewis 1, Andre Micke 1, Xiang Yu 1 and Maria F. Chan 2 1. Advanced Materials Group, Ashland Inc., Wayne, NJ; 2. Department of Medical Physics,
Chapter 8 Introduction to Hypothesis Testing
MGT-491 QUANTITATIVE ANALYSIS AND RESEARCH FOR MANAGEMENT OSMAN BIN SAIF Session 19.
Blue: Histogram of normalised deviation from “true” value; Red: Gaussian fit to histogram Presented at ESA Hyperspectral Workshop 2010, March 16-19, Frascati,
Chap 6-1 A Course In Business Statistics, 4th © 2006 Prentice-Hall, Inc. A Course In Business Statistics 4 th Edition Chapter 6 Introduction to Sampling.
Prof. of Clinical Chemistry, Mansoura University.
© 2008 McGraw-Hill Higher Education The Statistical Imagination Chapter 10. Hypothesis Testing II: Single-Sample Hypothesis Tests: Establishing the Representativeness.
Chapter 4 analysis of variance (ANOVA). Section 1 the basic idea and condition of application.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Introduction Commercial implementations of the convolution/superposition (C/S) method make several approximations, which can lead to dose calculation inaccuracies.
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
Two-Way ANOVA Ch 13. In a 2-way design, 2 factors (independent variables) are studied in conjunction with the response (dependent) variable. There is.
The Radiological Physics Center’s Anthropomorphic Quality Assurance Phantom Program Carrie F. Amador, Nadia Hernandez, Andrea Molineu, Paola Alvarez, and.
1 A Comprehensive Study on the Heterogeneity Dose Calculation Accuracy in IMRT using an Anthropomorphic Thorax Phantom S Davidson 1, R Popple 2, G Ibbott.
Outcome Measures of Triple Board Graduates: Marla J. Warren, MD,MPH; David W. Dunn, MD; Jerry L. Rushton, MD,MPH. Section of Child Psychiatry.
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
Development and Implementation of a Remote Audit Tool for High Dose Rate (HDR) 192 Ir Brachytherapy Using Optically Stimulated Luminescence Dosimetry Kevin.
S Scarboro 1,2, D Cody 1,2, D Followill 1,2, P Alvarez 1, M McNitt-Gray 3, D Zhang 3, L Court 1,2, S Kry 1,2 * 1 UT MD Anderson Cancer Center, Houston,
Xoft Routine Per Article Manufacturing Testing of the Axxent ® Source S. Axelrod, T.W. Rusch, M. Powell; Xoft, Inc., Fremont, CA  Purpose: To select sources.
Local Predictability of the Performance of an Ensemble Forecast System Liz Satterfield and Istvan Szunyogh Texas A&M University, College Station, TX Third.
K. Pulliam, MS 1,2., D Followill, PhD 2., L Court, PhD 2., L Dong, PhD 3., M Gillin, PhD 2., K Prado, PhD 3., S Kry, PhD 2 1 The University of Texas Graduate.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Conclusions Despite being able to acquire successfully high resolution images of radiation dose, image quality is limited by hardware problems. Upgrading.
 An exposure-response (E-R) analysis in oncology aims at describing the relationship between drug exposure and survival and in addition aims at comparing.
1 Chapter 8 Interval Estimation. 2 Chapter Outline  Population Mean: Known  Population Mean: Unknown  Population Proportion.
Karolina Kokurewicz Supervisors: Dino Jaroszynski, Giuseppe Schettino
Radiation Therapy Trials - Quality Assurance:  patient safety  adherence to protocol constraints  uniformity of patient treatments  efficient review.
Purpose N-isopropylacrylamide (NIPAM) polymer gel dosimeters were employed to verify the dose distribution of clinical intensity modulated radiation therapy.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Chapter 9 Introduction to the t Statistic
Characterization of proton-activated implantable markers for proton range verification using PET J. Cho1, G. Ibbott1, M. Kerr1, R. A. Amos2, F. Stingo1,
Introduction Out-of-hospital cardiac arrest (OHCA) is the sudden cessation of the heart in an out of hospital setting. In the United States, the incidence.
The magnitude of H&N IMRT dose delivery errors from three possible failure modes: beam quality, symmetry, and MLC position Jackie Tonigan, M.S. Advisor:
Optimisation of Patient Protection for Radiography
CONCLUSION INTRODUCTION RESULTS METHODS AND MATERIALS
Introduction to Sampling Distributions
Association between two categorical variables
Sample Size Estimation
Left Posterior Superior Right Anterior Inferior
Template Matching Can Accurately Track Tumor Evaluation of Dose Calculation of RayStation Planning System in Heterogeneous Media Huijun Xu, Byongyong Yi,
A. Nisbet 1,2, A. Dimitriadis 1,2,3, A.L. Palmer 1,4, C.H. Clark 2,3
Fig. 4. Percentage of passing rate between clinical and 544 plans.
Reducing Treatment Time and MUs by using Dynamic Conformal Arc Therapy for SBRT Breath-Hold Patients Timothy Miller, Sebastian Nunez Albermann, Besil Raju,
Choice of Methods and Instruments
Chi Square (2) Dr. Richard Jackson
CHAPTER 18: Inference about a Population Mean
Presentation transcript:

IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN HN HN GU GYN GYN Average: Introduction The gamma (γ) metric introduced by Low et al (1) is a quantitative method of comparing two dose distributions using a combined distance-to-agreement (DTA) and dose difference acceptance criteria. The use of this metric in IMRT quality assurance (QA) has become routine. However, γ calculations are potentially sensitive to a number of factors not commonly taken into account. As noted in the literature (2), γ can be affected by noise in the evaluated and reference distributions. Specifically, γ is underestimated (i.e. better agreement) in the presence of noise in the evaluated distribution while noise in the reference distribution adds noise to γ in proportion to the normalized dose noise. Noise is inherent to film measurement and processing as well as computational methods such as Monte Carlo. Additionally, the resolution (i.e. number of sample points) of both distributions could affect γ values (3). The more points contained in the evaluated distribution, the greater the probability of finding a point that meets the acceptance criteria. This is important clinically because film digitization can be performed at a variety of resolutions and using the highest resolution could cause an underestimation of gamma values (overestimation of agreement). The recommendation of Low et al (2) is that the measured and calculated distributions represent the reference and evaluated distributions, respectively. However, in practice, there is not a standard clinical assignment for the distributions. Since gamma is not symmetric with respect to which distribution is designated as the reference, the sensitivity to noise and resolution may have a dependence on this assignment. Due to the lack of literature evaluating the clinical impact of noise, resolution, and assignment of reference distribution, we evaluated the effect of these parameters on the percentage of pixels with passing gamma values for six clinical IMRT QA plans. Results Resolution study: The data from our resolution study is summarized in Table 1(a) and 2(a). We found that increasing the film image resolution increased the % of pixels passing when the film was designated as the evaluated distribution. The opposite trend was observed (i.e. % passing pixels decreased for higher film resolution) when the film was designated as the reference distribution. Noise study: The data from our noise study is summarized in Table 1(b) and Table 2(b). In general, we found that increasing the noise in the film image increased the % of passing pixels when the film was designated as the evaluated distribution. The opposite trend was observed, but with a greater magnitude of change, when the film was designated as the reference distribution. We believe that this difference in magnitude is due to the fact that % of passing pixels cannot be inflated past 100% and thus there is greater room for this value to decrease than increase. We found that noise had a dramatic effect on the gamma distribution, as evidenced by Figure 1, and generally had a greater impact on the results of gamma comparisons than image resolution. Conclusions Based on our results, we have found that the percentage of pixels passing in gamma analysis for IMRT QA is sensitive to both the resolution and presence of noise in the film image, with image noise being the more dominant factor. The sensitivity to noise and resolution is also dependent on the choice of reference distribution. Designating the measured (film) distribution as the reference appears to make the gamma comparison less sensitive (more robust) to the effects of both resolution and noise, and is thus recommended for clinical practice. It is necessary to have a good understanding of these factors (i.e. image noise, image resolution, and choice of reference) as well as how your software handles these factors in order to thoughtfully design IMRT QA protocols and guidelines such that delivery errors are not masked by factors that can artificially inflate gamma passing rates. References 1) D. A. Low et al, Med Phys 25 (5), (1998). 2) D. A. Low et al Med Phys 30 (9), (2003). 3) N. L. Childress et al Med Phys 32 (2), (2005). Methods and Materials Clinical Gamma Analysis: For each clinical IMRT plan, a hybrid QA plan was created in Pinnacle 3 using a solid water phantom and delivered with a Varian Clinac 21EX. A transverse dose plane was measured using Kodak EDR2 radiographic film, and gamma analyses were performed using OmniPro-I’mRT software (IBA Dosimetry, Germany), Six arbitrary clinical IMRT quality assurance (QA) plans representing a variety of treatment sites (3 Head and Neck, 1 Genitourinary, and 2 Gynecology) were chosen for this study. Resolution and Noise Studies: For the image resolution study, each of the six films was digitized at 71, 142, and 285 dpi with a VIDAR VXR-16 Dosimetry Pro (VIDAR Systems Corporation). For the noise study,1 and 2% standard deviation local Gaussian noise was added to the digitized 71 dpi films. Gamma comparisons for both studies was performed at 5%/3mm, 3%/3mm, and 2%/2mm acceptance criteria. The overall results were analyzed for changes in the percentage of pixels passing per 1% increase in image noise and doubling of resolution, respectively. For both studies, the film was used as the evaluated distribution per our clinical practice. Assignment of Reference Distribution: The noise and resolution studies were repeated using the Low et al. recommendation of film as the reference distribution. The effects of image resolution and noise on the gamma dose distribution comparison method Jessie Huang, Kiley Pulliam, David Followill, and Stephen Kry (1) The University of Texas MD Cancer Center, Houston, TX (2) The University of Texas Health Science Center Houston, Graduate School of Biomedical Sciences (a) Reference distribution Evaluated distribution Average change in % pixels passing per 1% increase in noise 5%/3mm3%/3mm2%/2mm plan film with noise 0.1 (99.7)1.1 (95.8)4.5 (82.9) film with noise plan-7.7 (99.2)-10.1 (92.7)-10.2 (75.0) (b) Reference distribution Evaluated distribution Average change in % pixels passing per x2 increase in resolution 5%/3mm3%/3mm2%/2mm planfilm0.1 (99.7)1.0 (95.8)3.3 (82.9) filmplan-0.1 (99.2)-0.1 (92.7)-0.3 (75.0) (a) Table 1. All data from the (a) noise and (b) resolution study in which the treatment plan is the reference and the film is the evaluated distribution using the OmniPro-I’mRT software for all QA plans at each acceptance criteria. Table 2. Change in % of passing pixels, averaged over all QA plans, when the (a) resolution is doubled, (b) the noise is increased by 1%, and (c) when noise is increased by 1% and there is a 3mm mis-registration error. Additionally, the second row of data in each table shows the effect of the assignment of the reference distribution. Shown in parenthesis is the baseline average % passing pixels (71dpi, no added image noise) averaged over all QA plans. IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 71 dpi142 dpi285 dpi71 dpi142 dpi285 dpi71 dpi142 dpi285 dpi HN HN HN GU GYN GYN Average: (b) Discussion Since image noise is a dominant factor influencing the results of gamma analysis, we further wanted to investigate if the presence of image noise could cause a failing comparison to pass, i.e. if image noise could inflate the % passing pixels enough to change the outcome of gamma analysis. To do this, we introduced an image mis-registration in our comparisons (3mm shift). The results are listed in Table 2(c). Notably, we found that increasing the image noise by 1% was enough to increase the % passing pixels by an average of 3% (3%/3mm, film = evaluated distribution). For one of the patients in this study, the % passing pixels increased from 85% to 100% with the addition of 2% image noise (3%/3mm, film = evaluated distribution). Thus, image noise can potentially affect the overall results of gamma analysis. Reference distribution Evaluated distribution Average change in % pixels passing per 1% increase in noise 5%/3mm3%/3mm2%/2mm plan film with noise + 3mm shift 1.0 (97.8)3.4 (92.6)8.8 (77.9) film with noise + 3 mm shift plan-7.1 (95.7)-8.7 (87.2)-8.4 (68.3) (c) Contact: Figure 1: Representative gamma maps showing the effect of increased noise in the film image using the OmniPro-I’mRT software. The film is the evaluated distribution in (a) and (b) and reversed as the reference distribution in (c) and (d). All film images digitized at a resolution of 71dpi and were evaluated at a 2%/2mm criteria. Images (a) and (c) have no noise added while images (b) and (d) show the change in the distribution of passing pixels (blue) when 1% noise is added. (a) (b) (c) (d) 0% added noise1% added noise (a)(b) (c)(d) Support Work supported by PHS grant CA10953 awarded by NCI, DHHS