RADIOGRAPHIC EQUIPMENT Adler & Carlton Ch. 8 & Bushong Ch. 1
Modern radiology Radiography Fluoroscopy Two basic types of x-ray examinations Radiography Fluoroscopy
Radiography exam
Radiographic Room
The transformation from a physicist toy to a valuable, large scale medical specialty 1907 Snook transformer – high voltage 1913 Coolidge x-ray tube – air evacuated, separate selection of voltage and current Standardizing x-ray tube beam output
THE X-RAY TUBE The X-Ray tube is the single most important component of the radiographic system. It is the part that produces the x-rays
THE X-RAY TUBE ANODE + CATHODE -- GLASS ENCASED IN STURDY STEEL HOUSING PRIMARY COMPONENTS ANODE + CATHODE --
XRAY TUBE HOUSING MADE OF LEAD AND STEEL TO ABSORB ANY STRAY RADIATION TO PREVENT LEAKAGE RADIATION FROM THE TUBE
TUBE HOUSING MADE OF LEAD & STEEL
How “X-rays” are created SEE: MAN MADE RADIATION (PG How “X-rays” are created SEE: MAN MADE RADIATION (PG.111) Adler & Carlton TO PRODUCE X-RAYS YOU NEED: A SOURCE OF ELECTONS A FORCE TO MOVE THEM QUICKLY SOMETHING TO STOP THEM SUDDENLY
How “X-rays” are created Power is sent to x-ray tube via cables mA (milliamperage) is sent to filament on cathode side. Filament heats up – electrons “boil off” Negative charge
How “X-rays” are created Positive voltage is applied to ANODE Negative electons = attracted across the tube to the positive ANODE. Electrons “slam into” anode – suddenly stopped. X-RAY PHOTONS ARE CREATED
How “X-rays” are created Electron beam is focused from the cathode to the anode target by the focusing cup Electrons interact with the electrons on the tungsten atoms of target material PHOTONS sent through the window PORT – towards the patient
More on the tube later……..
X-RAY TABLE
Radiographic tables Are designed to support the patient during a radiographic exam Comfort is not the primary concern Foam pads should be used if the patient will be required to be on the table for longer than 10 minutes
Tabletop Must be uniformly radiolucent to easily permit x-ray to pass through. Carbon fiber is used because it is strong and very little x-ray photons are absorbed. Usually tabletops are flat however some are curved
Tabletop Most tabletops are floating, some are motor-driven The brakes can be released usually by the technologist hand or foot The brakes are electromagnetic Floating table tops save significant amounts of time and strain on the technologist
Tables Tables are fixed or tilting Fixed rooms are designed for diagnostic radiographic work only The table can usually be raised or lowered to accommodate the patient and the technologist.
Table top technique Performing imaging using just the cassette, plate or digital image receptor Also called “non- grid” technique
Fluoroscopy exam
FLUOROSCOPY IMAGES IN MOTION
Fluoroscope 1898 by Thomas Edison
Radiation Injury 1904 the first recorded x-ray fatality in the US – Clarence Dally, Edison’s assistant Early injuries took the form of skin damage, loss of hair and anemia Snook transformer & Coolidge tube reduced the injuries Why?
Tables Tilting rooms are designed for both diagnostic and fluoroscopic work Tilting models usually tilt to 90 degrees in one direction and 15 – 30 degrees in the other direction Tilting models include ancillary equipment; footboard, shoulder support, handgrips, compression bands
Table Movement Longitudinal Transverse Vertical Tilt or Angling
Fluoroscopy tables: the tube is under the table, image capture is above the patient
Tube Supports Designed to help technologists with various tube locations for creative imaging. Tube suspension systems are available in 5 versions: ceiling mounted, floor-to-ceiling, floor, mobile and c-arm.
Tube Movement Longitudinal Transverse Vertical Angling or Rolling Rotating Telescoping
Improvement in radiography Reducing exposure time to reduce blur Dentist William Rollins began using a diaphragm to improve image quality First application of collimation and filtration 1921 Potter-Bucky grid – improving image contrast
Dr. William Rollins Dr. William Rollins was a Boston physician and dentist who was the first to use collimation and filtration in the late 1800s. After receiving radiation burns to his hand in x-ray experiments in 1898, he used leather and aluminum filters when he made x-ray exposures of his patients' teeth.
The ‘BUCKY’ The Potter-Bucky is the device in the table or chest board that holds the film cassette. The ‘bucky’ is like a drawer that opens and closes to insert and remove the film cassette.
TABLE OR UPRIGHT BUCKY TRAY
Radiographic grid ?
Bucky cross hatched Grids are used to “clean up” scatter radiation from the patient To improve contrast on the radiograph Potter-Bucky are usually a focused moving grid
CASSETTES W/ GRID CAPS
Radiation shields
Viewing images View boxes Computer monitors
COLLIMATOR ATTACHES DIRECTLY BELOW THE X-RAY TUBE SERVES AS A X-RAY BEAM LIMITING DEVISE CONTROLS THE SIZE AND SHAPE OF X-RAY FIELD
Cone collimator
ALWAYS KEEP THE COLLIMATED AREA SMALLER THAN THE SIZE OF THE CASSETTE
X-ray tube controls Displays Tube angle Distance to table top bucky Collimator controls Tube lock controls High voltage cables
CASSETTE or FILM HOLDER The CASSETTE is used to hold the film during examinations. It consist of front and back intensifying screens, and has a lead (Pb) backing. The cassette is light tight
CR image capture PSP Plate PHOTOSTIMULABLE PHOSPHOR PLATE
FILM direct exposure & screen-film or film-screen SIZES 14 X 17 14 X 14 11 X 14 10 X 12 8 X 10
Film Sizes Standard “inches”: 8” x 10” 10” x 12” 11” x 14” 14” x 17” Metric: 18cm x 24cm 24cm x 30cm 30cm x 35cm 35cm x 43cm
DR(DDR) Collection element TFT (Thin film transistor) Photodiode CCD (charge-coupled device)
CONTROL CONSOLE GIVES THE TECHNOLOGIST CONTROL OF THE X-RAY MACHINE TECHNIQUE SELECTION Located OUTSIDE of the Radiographic Room
The Control Console The control console is device that allows the technologist to set technical factors (mAs & kVp) and to make an exposure. Only a legally licensed individual is authorized to energize the console.
“Technique” kVp , mAs (mA x s) What is set at the control panel How the “image” is created on the “film” or Image receptor (digital) kVp controls the “ENERGY” of the beam The Higher kVp – more penetrating Ranges is 50 -110 in Diagnostic x-ray
+ 30% + 50 % mas
kVp Changes
Darkroom
What is in the Darkroom?
Safe Light 15 Watts Red filter Must be 3-6 feet from counter top or feed tray of processor Used to be amber or orange filter
UNLOAD EXPOSED FILM
FILM ID PRINTER
FILM BIN - STORAGE
Transport System (Rollers) Entrance Crossover Feed Tray Squeegee Deep Racks Dryer Turnaround
Analog processor
CR processor What does a DDR processor look like?
Other x-ray stuff…. Positioning phantoms Pixie
Dosimeter An instrument that detects and measures exposures to ionizing radiation Personal vs Field survey instruments
Densitometer Measures optical density on a radiograph
Step wedge or penetrometer Test tool made from aluminum Accurately-calibrated filter strip that provides a stepped range of exposures
Other x-ray stuff…. Positioning sponges Lead markers Gurney