Introduction Since 1996 [1], it has been clear that optical computed tomography (CT) represents a viable alternative to MRI scanning of dosimeter gels.

Slides:



Advertisements
Similar presentations
Image Reconstruction.
Advertisements

The Wave Nature of Light
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
Fire Protection Laboratory Methods Day
Topic 11.3 Diffraction.
Diffraction of Light Waves
Computed Tomography II
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Introduction Radiosensitive polyacrylamide gels (PAG) provide a method of mapping dose distributions in 3D with sub-millimetre resolution [1]. Largely.
Lecture 5.1 Scanning Electron Microscopy (SEM)
Optimisation of X-ray micro-tomography to perform low-dose imaging of highly-dosed gels P.M.Jenneson, E.C.Atkinson, P.Wai and S.J.Doran In 1993, Maryanski.
Introduction A Fricke gel was the first dosimetric system to offer the possibility of making 3-D radiation dose measurements in a totally non-destructive.
Introduction Three-dimensional gel dosimetry is a novel method for radiation measurement motivated by the need to verify experimentally the doses produced.
Uni S School of Electronics and Physical Sciences Department of Physics University of Surrey Guildford Surrey GU2 7XH, UK Optimisation of X-ray micro-tomography.
Amplitude Control: Closing the Loop in a Zero Path Length Difference Michelson Interferometer Michael G. Littman, Michael Carr , Laurent Pueyo, Jeremy.
Introduction Spectrophotometry has been the conventional technique for measuring optical density of irradiated gels [1]. Later, magnetic resonance imaging.
tomos = slice, graphein = to write
Introduction Following Gore’s observation [1] that the well known method of Fricke dosimetry [2] was ideally suited to a “readout” by Nuclear Magnetic.
Introduction The effect of air-cavities within the human body, for example in the head and neck regions, present possible sources of error when calculating.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Seeram Chapter 11: Image Quality
Dott. Dario Tresoldi CNR IPCF ME
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer.
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
S Dr. S. J. Doran Department of Physics, University of Surrey, Guildford, GU2 7XH, UK High-resolution measurements of radiation dose in 3-D using gel dosimetry.
M. D. A NDERSON Cancer Center Magnetic Resonance Imaging: 4.7 T, 40 cm Bruker Biospec James A Bankson, Ph.D. and John D Hazle, Ph.D. Small Animal Cancer.
Hg lamp Cylindrical lens, pinhole and filter  pseudo point-source Lens  parallel beam Scanning tank with matching medium Exposed gel Unexposed gel Diffuser.
Introduction A modification of the Fricke solution system to make it suitable for low-level dosimetric studies was proposed about three decades ago [1].
Evaluation of the Performance of the Fast Scanning Platform of an OCT System Malcolm Heard 1, Miguel Herrera 1, Geoffrey Ibbott 1 1 Department of Radiation.
10/17/97Optical Diffraction Tomography1 A.J. Devaney Department of Electrical Engineering Northeastern University Boston, MA USA
Other Applications using DMD By Walid. Introduction DMD, a dense array of hundreds of thousands of tiny switchable mirrors, whose pixel speed, contrast.
PHYS 1442 – Section 004 Lecture #22-23 MW April 14-16, 2014 Dr. Andrew Brandt 1 Cameras, Film, and Digital The Human Eye; Corrective Lenses Magnifying.
Integral University EC-024 Digital Image Processing.
Healthcare research in the Physics Department Dr S J Doran Lecturer in Magnetic Resonance Imaging Department of Physics School of Electronics and Physical.
Digital Image Fundamentals. What Makes a good image? Cameras (resolution, focus, aperture), Distance from object (field of view), Illumination (intensity.
Computed Tomography Q & A
S Optical CT scanning of PRESAGE TM polyurethane samples with a CCD-based readout system S J Doran 1*, N Krstajic 1, J Adamovics 2 and P M Jenneson 1 1.
Presentation Overview
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
In Burst experiments, the read gradient is much larger then the other imaging gradients, hence, only translation along this gradient direction will cause.
Introduction Magnetic resonance (MR) imaging is recognised as offering potential benefits in the delineation of target volumes for radiotherapy (RT). For.
Fig. 3 shows how a dose distribution that is initially relatively sharp becomes blurred with time, until after 16 hours, virtually no useful information.
Ion Beam Analysis of Gold Flecks in a Foam Lattice F E Gauntlett, A S Clough Physics Department, University of Surrey, Guildford, GU2 7XH, UK.
Mach-Zehnder Interferometer for 2-D GRIN Profile Measurement Instrument Overview Sample Preparation Measurement Uncertainty University of Rochester Gradient-Index.
WP3. OPTIMISATION SEDENTEXCT ANNUAL REPORT WORKPACKAGE-SPECIFIC INFORMATION WORKPACKAGE 3 (NKUA - Kostas Tsiklakis)
Technical development of a high resolution CCD-based scanner for 3-D gel dosimetry: (I) Scanner construction S J Doran, K K Koerkamp*, M A Bero, P Jenneson,
: Chapter 11: Three Dimensional Image Processing 1 Montri Karnjanadecha ac.th/~montri Image.
Identification of a 3D dosimeter best-suited for use by the RPC M. Heard, G. Ibbott, D. Followill, R. White, E. Jackson, M. Salehpour.
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali,
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Conclusions Despite being able to acquire successfully high resolution images of radiation dose, image quality is limited by hardware problems. Upgrading.
Lithography in the Top Down Method New Concepts Lithography In the Top-Down Process New Concepts Learning Objectives –To identify issues in current photolithography.
Part No...., Module No....Lesson No
The Microscope and Forensic Identification. Magnification of Images A microscope is an optical instrument that uses a lens or a combination of lenses.
S Department of Physics, University of Surrey, Guildford, GU2 7XH, UK
Modelling and Simulation of Passive Optical Devices João Geraldo P. T. dos Reis and Henrique J. A. da Silva Introduction Integrated Optics is a field of.
Purpose N-isopropylacrylamide (NIPAM) polymer gel dosimeters were employed to verify the dose distribution of clinical intensity modulated radiation therapy.
SARDAR PATEL INSTITUTE OF TECHNOLOGY E.NO : Guide By:- V.N.Thakkar.
DESCRIPTION OF PIXIRAD  The PIXIRAD Imaging Counter is an INFN-Pisa Spin-off  It works in photon counting  Pulse discrimination with two thresholds.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Light and Optics  The Electromagnetic Spectrum  Interference, Diffraction, and Polarization Wave Properties of Light.
Date of download: 9/20/2016 Copyright © 2016 SPIE. All rights reserved. (a) Photographs of the anthropomorphic phantom used in this study and (b) the inner.
Characterisation of Bessel and Gaussian beam illumination modes on LSM imaging quality of early chick embryo development Ricardo F Bango Da Cunha Correia1,
Analysis of optical IPM data
Theoretical consideration for x-ray phase contrast mammography
A. Nisbet 1,2, A. Dimitriadis 1,2,3, A.L. Palmer 1,4, C.H. Clark 2,3
Computed Tomography (C.T)
Spot-Projector System for the Measurement of Intra-Pixel Response
Presentation transcript:

Introduction Since 1996 [1], it has been clear that optical computed tomography (CT) represents a viable alternative to MRI scanning of dosimeter gels. Both laser- and CCD-based methods allow high-resolution images to be obtained. We are pursuing the CCD route, because it allows the generation of several hundred slices in a time similar to that required for a single slice on a laser-based system. This will allow full 3-D datasets to be obtained in a realistic scan-time. PRESAGE TM is a novel dosimeter with a number of very attractive properties and the overall aim of this project was to investigate what modifications are required to the previous prototype scanner to enable successful scanning of PRESAGE TM. This poster demonstrates the following areas of our work: the first imaging of a PRESAGE TM dosimeter using a CCD-based system; the modifications needed to the scanner to improve the image quality further; the resolution capabilities of this prototype scanner. Optical CT scanning of PRESAGE TM polyurethane samples with a CCD-based readout system Acknowledgements NK is supported by a PhD studentship from the EPSRC. The authors thank Mr C Bunton for technical help in construction of the scanner. Experiments performed using PRESAGE TM PRESAGE TM samples were supplied by Heuris Pharma in the form of cylinders of diameter 7 cm. These were irradiated using 6 MV X- rays (30 Gy) from a linac in two patterns: a square field 2.7 × 2.7 cm 2 ; a grid pattern using a purpose built lead collimator The results of these scans are displayed in Fig. 1 and show considerable grounds for optimism. In particular, the in-plane resolution in the image of the grid phantom is excellent (~7 pixels across a 2 mm spot). However, there are a number of issues that must be resolved before obtaining useful images. The limiting factor for both spatial resolution and dose sensitivity is the poor contrast / artefact ratio. This occurs for three main reasons: Ring artefacts arise from the scanner’s diffuser screen [2]; Refraction artefacts arise from the matching liquid. It is necessary to match very precisely (within 0.1%) the refractive index of the dosimeter. This is done using different ratios of two components in a mixture. If the final index is slightly wrong then bands appear at the sample edges [2]. More seriously, if the mixture is not homogeneous (incomplete mixing of the two components), then irregular light and dark fringes are seen in the projections. It is also possible that some minor non-uniformities may exist in the PRESAGE TM itself. Absolute attenuation values cannot yet be measured, because the CCD camera is not yet quantitatively calibrated. The first of these effects has already been corrected by redesigning the scanner to dispense with the need for the diffuser screen and we are optimistic that the ideas we are currently pursuing will alleviate the other problems, too. References [1] JC Gore et al., Phys. Med. Biol. 41, 2695 (1996) [2] SJ Doran et al., Phys. Med. Biol. 46, 3191–3213 (2001) Figure 1: (a) Optical CT image of a 2.7 × 2.7 cm 2 square field, 30 Gy at the top surface of the dosimeter; (b) schematic arrangement of irradiation protocol for the grid phantom, with 30 Gy at the top of the lead collimator; (c) optical-CT image of the grid phantom S Figure 2: (a) New ultra-bright LED light source; (b) New optical arrangement S J Doran, N Krstajic, J Adamovics, P Jenneson Department of Physics, University of Surrey, Guildford, GU2 7XH, UK Heuris Pharma, Skillman, NJ, USA * Cross-section X-ray beam ~30 Gy at lead surface PRESAGE TM a c b Modifications to the scanner Three key design features have been changed since DOSGEL ’01: The mercury lamp has been replaced with an ultra-bright LED (nominally 1 W) — see Fig. 2a. The focusing arrangement is changed so acquisition no longer uses a diffuser screen — see Fig. 2b The tank has been redesigned so that optical glass is used rather than perspex. * LED light source Scanning tank CCD chip camera lens ab Characterisation of the scanner From a series of experiments imaging phantoms made from thin wires (similar to the “needle phantoms” of Oldham et al.), we were able to obtain a profile through a wire of known thickness (0.25 mm) and hence, by constrained deconvolution, to make a first estimate of the modulation transfer function of our system. Note that by altering the position of the camera lens, we are able to look at a very wide range of fields-of-view and so the MTF will change. The calculation below is for a field of view of ~4 cm. Position on profile / mm Image intensity ~ optical density Line-pairs / mm MTF abc Figure 3: (a) Images of cross-section of wire phantom, with enlargement inset. The streak artefacts relate to the fact that the wires are completely opaque to light, rather than partially absorbing like a gel (for which this artefact would not be present). The thinnest wire has diameter 0.25 mm. (b) Profile through thinnest wire, with (inset) an enlargement of the central region, showing excellent resolution. (c) First measurement of MTF.