Computed Tomography Q & A

Slides:



Advertisements
Similar presentations
ARTIFACTS IN COMPUTED TOMOGRAPHY
Advertisements

IMAGE GENERATION IN CT.
Computed Tomography Principles
CT Scanning: Dosimetry and Artefacts
Computers and Computed Tomography
Image Reconstruction.
Computed Tomography
Image Quality Bushong Chapter 7.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Computed Tomography Stewart C. Bushong
Image Artifacts Chapter 8 Bushong.
X-ray diagnostics and Computed tomography. 15th century head examination.
Computed Tomography II
Spiral CT Bushong Chapter 5.
Seeram Chapter 13: Single Slice Spiral - Helical CT
Module H Computed Tomography Physics, Instrumentation, and Imaging.
MULTISLICE CT.
Advanced Biomedical Imaging
CT Multi-Slice CT.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Computed Tomography RAD309
Special Imaging Techniques Chapter 6 Bushong. Dynamic Computed Tomography (DCT) Dynamic scanning implies 15 or more scans in rapid sequence within one.
tomos = slice, graphein = to write
Chapter 2 Stewart C. Bushong
Computed Tomography
Conventional and Computed Tomography
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Seeram Chapter 11: Image Quality
COMPUTED TOMOGRAPHY (I) RAD 365 CT - Scan
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Quality Control.
SPECIAL IMAGING/ADVANCED IMAGING INTRODUCTION TO DIGITAL RADIOGRAPHY
Image Quality Radiographic Resolution.
Seeram Chapter 7: Image Reconstruction
Mammography Questions and Answers
PRINCIPLE AND BASIC PHYSICS OF COMPUTED TOMOGRAPHY
Computed Tomography Physics, Instrumentation, and Imaging
Factors affecting CT image RAD
Fig. 3 shows how a dose distribution that is initially relatively sharp becomes blurred with time, until after 16 hours, virtually no useful information.
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
▪ History ▪ Equipment ▪ Image Production/Manipulation.
Module C Computed Tomography Physics, Instrumentation, and Imaging.
LEC (1) Design/layout by E Bashir, CAMS, King Saud University, Ref: Seeram, Karthikeyan, internet.
Image Quality The capacity to define, measure, and assess image quality is a primarily responsibility of a CT Technologist.
COMPUTED TOMOGRAPHY - I RAD 365 CT - Scan
CT IMAGE RECONSTRUCTION  Hounsfield envisioned dividing a slice into a matrix of 3-dimensional rectangular boxes (voxels) of material (tissue). Conventionally,
Module D Computed Tomography Physics, Instrumentation, and Imaging.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Part No...., Module No....Lesson No
Single-Slice Rebinning Method for Helical Cone-Beam CT
Part No...., Module No....Lesson No
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Basic Principles of CT Scanning. CT CT - Computed Tomography CAT Scan - Computerized Axial Tomography.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Page 1 Physics Case of the Day - Thursday During routine QA testing of a computed tomography scanner, the physicist recorded the data below for the tube.
CT Multi-Slice CT.
Computed Tomography Basics
CT PRINCIPLES AYMAN OSAMA.
Computed Tomography Image Reconstruction
Physics Testing for ACRIN 6678 Protocol version 3.0 QIBA Group 1C
Wednesday Case of the Day
Basic principles Geometry and historical development
Presentation transcript:

Computed Tomography Q & A Phillip W Patton, Ph.D.

B. Back-projection algorithm C. Tomographic slice D. Primary beam 1) The measured x-ray transmissions from a single CT fan beam through a patient is called a A. Filter B. Back-projection algorithm C. Tomographic slice D. Primary beam E. Projection

1) Answer E. A projection is a profile of transmitted x-ray intensities through the patient at any given location of the tube.

A. kV B. mA C. Scan time D. Section thickness E. Number of sections 2) Anode heat loading on a CT x-ray tube increases with all the following EXCEPT A. kV B. mA C. Scan time D. Section thickness E. Number of sections

D. Section thickness does not directly affect x-ray heat loading. 2) Answer D. Section thickness does not directly affect x-ray heat loading.

A. HU of blood vessels B. Required kVp C. Required mA D. Patient dose 3) Use of intravascular contrast when performing a single CT section will significantly increase the A. HU of blood vessels B. Required kVp C. Required mA D. Patient dose E. Image noise

3) Answer A. Intravenous contrast increases the density and atomic number of blood and tissues. This increases x-ray attenuation and thereby the resultant HU value.

A. Variable for different section thicknesses 4) CT collimators are A. Variable for different section thicknesses B. Not necessary for helical scans C. Usually made out of plexiglass D. Bow-tie shaped E. Cooled using fans

4) Answer A. The collimators are located at the x-ray tube and have a variable width (1 to 10 mm), which defines the CT section thickness

5) The CT image display contrast A. Must be selected prior to the x-ray exposures B. May be altered after the CT scan C. Does not modify the appearance of the CT image D. Can be used to change the HU values of image data E. None of the above

5) Answer B. Changing the display contrast alters the appearance of the CT image, but not the reconstructed image data.

6) Partial volume artifacts in CT are generally reduced when A. Section thickness increases B. Scanning time is increased C. Image matrix size increases D. Fifth-generation scanners are used E. Small focal spot sizes are used

6) Answer C. a larger image matrix size improves spatial resolution and hence is likely to reduce “volume averaging” known as the partial volume effect.