OxCart: Oxygen Concentrator and Analyzer for the Developing World Matt Amdahl, Ryan Le, Dan Nelson, Jay Patel, Abe Segura Department of Bioengineering,

Slides:



Advertisements
Similar presentations
QUICK QUIZ 28.1 You would like to construct an RC circuit with a time constant that is equal to 1 second (in other words, the current from the battery.
Advertisements

1 Schedule of classes in PRC I need your availability on - Wednesday - Thursday - Friday afternoon.
A Comparison of Early Versus Late Initiation of Renal Replacement Therapy in Critically III Patients with Acute Kidney Injury: A Systematic Review and.
MicroCART Micro processor C ontrolled A erial R obotics T eam Abstract MicroCART is a group of EE/CprE students tasked with developing an autonomous helicopter.
Entrepreneurship and Product Development in Bioengineering at Rice University ‘Unique Feature’ Maria Oden.
The Effect of Gas Quality on Appliance Design Dr Martin Searle Baxi Group European Forum Gas 2007.
Design and Computer Modeling of Ultracapacitor Regenerative Braking System Adam Klefstad, Dr. Kim Pierson Department of Physics & Astronomy UW-Eau Claire.
Oxygen Concentrators.
EXHAUST GAS RECIRCULATION IN DIESEL ENGINE
AME Senior Design Group E3: Hewlett-Packers Concept Design Review April 24, 2008 Caitlin Feely Ben Nelessen JJ Rees Jamie Rowland Patrick Tennant.
Alan Young, Dr. Weidong Zhou and Dr. Hongjun Yang Department of Electrical Engineering, The University of Texas at Arlington, Arlington, Texas Figure.
Fine Tuning and Setting PSA Presented by: Rastislav Vajdák Senior Test Engineer Oxymat-Slovakia, s.r.o.
Objective Background Fabrication Testing Conclusion Stephanie Moran, Ryan Rosario, Zachary Stauber, Bethany Tomerlin, Juan Carlos Ybarra Dispersing carbon.
NEW EDUCATIONAL LAB: MEASUREMENT AND UNCERTAINTY EVALUATION OF NANOFLUID PARTICLE CONCENTRATION USING VOLUMETRIC FLASK METHOD M. Kostic,
FINAL PRESENTATION P The Team Austin Frazer Role: Lead Engineer - Analysis Major: Mechanical Engineering Eileen Kobal Role: Lead Engineer – Mixtures.
Sensor Systems for CPI Sensor Transmitter temperature sensors
1 Live Sound Reinforcement Audio measurements. 2 Live Sound Reinforcement One of the most common terms you will come across when handling any type of.
Acknowledgments Summary of MAVs Design Criteria Design Solution Conclusions and Future Work Energy Harvesting for Micro-Air Vehicles Testing Harvesting.
University of Pennsylvania Department of Electrical and Systems Engineering ABSTRACT: Quantifying and measuring certain aspects of a golf swing is a helpful.
Mist Generator Kelly Toy Jon Cappel Joe Bothwell Abdikarim Mahamud Advisor – John Webster Client – Jack Jiang, MD, PhD tents/06.00/ asp.
Classification of Instruments :
Orit Skorka and Dileepan Joseph University of Alberta, Canada Reducing Crosstalk in Vertically- Integrated CMOS Image Sensors.
ME 270 Final Project Presentation Operational Amplifiers.
Infant Respiratory Rate Monitor Group 11 Sean Ermer Tim Lau Leo Li Mentor: Dr. Daniel Moran.
Second Chance Automated External Defibrillator Pads Defib Task Force Lisa Jiang, Justin Lin, Joanna Nathan, Carl Nelson, Brad.
Solar Powered Battery Charger
The goal of this project is to provide inexpensive electricity to citizens of third world countries who have limited access to electricity. The scope of.
Problem This project was funded by Mr. and Mrs. Charles Fox and Beyond Traditional Borders, which is made possible by a grant to Rice University from the.
MODEL-BASED DETERMINATION OF HYDROGEN SYSTEM EMISSIONS OF MOTOR VEHICLES USING CLIMATE- CHAMBER TEST FACILITIES Dr. Martin Weilenmann.
Senior Project – Electrical Engineering Biofeedback Training System for Improved Athletic Performance Matt Statton Advisor – Professor Helen Hanson.
PACER Flight Readiness Review (FRR) for the Grambling Ozone Detector (G. O. D.) Cassandra Hendon, Derrien Green, Demetrius Norman.
By J DAVID MANOHAR (08A31A0232). NECESSITY OF SUPERCONDUSTOR  Damage from short circuit is constant threat in power systems  All the power systems components.
ENERGY EFFICIENCY TECHNOLOGIES AND ENERGY SAVINGS POTENTIALS FOR COLD ROOMS.
Improving Oxygen Concentrator Use in the Developing World
ECE-498 Matt Statton Advisor: Professor Hanson.  Goals of athletic training ◦ Muscle hypertrophy through stimulation  Muscle fatigue during a specified.
Solar Powered Blood Pressure Assist Device Laura Allen 1, James Berry 2, Casey Duckwall 2, David Harris 1 Advisor: Dr. Franz Baudenbacher 2 Department.
Handheld Flow Calibrator for Aerosol Observing System Jonathan Ma, Dung (Yung) Hoang Nguyen, Mario Zuniga Department of Mechanical and Aerospace Engineering,
Outline Introduction & BackgroundSystem Description & DiagramSpecificationsDesign – Four StagesFuture Development 211/22/2006.
Testing Considerations Because of potential life threatening consequences resulting from device malfunction, it is critical that the device be fully tested.
Performance Study of Localization Techniques in Zigbee Wireless Sensor Networks Ray Holguin Electrical Engineering Major Dr. Hong Huang Advisor.
New Product Developments World Leaders in Combustion Management Solutions C.E.M.S., MCERTs & E.P.A. Exhaust Gas Analysers C.E.M.S., MCERTs.
The D50 Is a state of the art environmental monitor using only the best quality components available. It offers multiple sensors in a single unit with.
Lab 3 Static Calibration of Electronic Pressure Transmitters using Manometers February 1, 2013 Group 0 Miles Greiner Lab Instructors: Michael Goodrick.
New Product Developments World Leaders in Combustion Management Solutions C.E.M.S., MCERTs & E.P.A. Exhaust Gas Analysers C.E.M.S., MCERTs.
Port Fuel Injection VS. Direct Fuel Injection The Basics of DFI The main focus of DFI is to effectively and precisely control the fuel-to-air ratio. To.
Resporite Elisa Fung Lindsay Kugler BeiXian Tang Gabriela Vargas Little Patients, Big Demands.
TRB Planning Applications Conference May 2009, Houston,TX A Caveat on O-D Matrix Estimation/Adjustment: Deviations from a seed matrix and Simultaneous.
 Oxygen Analyzer Jay Patel #19 April 27,  Senior Design Project [1]
COMMUNICATION SYSTEM EEEB453 Chapter 5 (Part III) DIGITAL TRANSMISSION Intan Shafinaz Mustafa Dept of Electrical Engineering Universiti Tenaga Nasional.
Authors: Noor Mohammed Dr. M.Z. Hossain Dr. Sumon Saha
 Oxygen Analyzer Team 7: Justin Akiyama, Andrew Do, Sapphire Lopez, Jay Patel, Marvin Patel, Nirav Shah.
A Testbed for FTC robot Components Sig Johnson April, 2016.
 Oxygen Analyzer Team 7: Marvin Patel, Jay Patel, Sapphire Lopez, Nirav Shah, Justin Akiyama, Andrew Van Do.
Kaeser Compressors, Inc.
Dual-Chamber Autonomous Bioreactor for the growth of 3-D Cell Cultures in Microgravity Austin James, Kevin King, Isaac Phillips, Cody Tapscott Problem.
Intrapulmonary Percussive Ventilation Carl W. Israel*, Tim Weaver*, James Grotting*, Joshua Gess^, Ryan Winter^, Dr. Paul King * *Department of Biomedical.
PATIENT OXYGEN PROCESSING FROM AIR. BY; CHARLES MUTONG’WA WANJALA WEBUYE DISTRICT HOSPITAL, BUNGOMA COUNTY.
Types of Sensors Used in Multi-Point Fuel Injection System
Thermistor Temp. (°C) Thermometer Temp. (°C) Thermistor Temp. vs. Thermometer Temp. Figure 1. Salter Labs ThermiSense ® [3] Figure 2. Thermistors measuring.
 Oxygen Analyzer Team 7: Justin Akiyama, Andrew Do, Sapphire Lopez, Jay Patel, Marvin Patel, Nirav Shah.
Diode Circuit Analysis 2
K. Warmuzinski, M. Tanczyk, M. Jaschik, A
Kaeser Compressors, Inc.
Graduation Project-II submitted to:
IE-432 Design Of Industrial Experiments
Created by Art Kay Presented by Peggy Liska
The capture of CO2 from flue gas using adsorption combined with membrane separation Prof. Krzysztof Warmuzinski Polish Academy of Sciences Institute.
Electrical traditional Chinese Instrument - Xun
Wireless Electrocardiogram System Senior Project
Paper-based Anemia Diagnosis for Use in Low-Resource Settings
Presentation transcript:

OxCart: Oxygen Concentrator and Analyzer for the Developing World Matt Amdahl, Ryan Le, Dan Nelson, Jay Patel, Abe Segura Department of Bioengineering, Rice University, Houston TX Team RedOx Objective Concentrator Design Objectives To design and build: - A low-purity, high-flow oxygen concentrator. - A reliable, inexpensive oxygen analyzer. CriteriaGoals Oxygen Flow> 15 L/min, 40% O 2 Power Consumption< 400W Longevity> 5 years Size< 2.5’ x 2.5’ x 8’ Cost< $1000 Motivation for OxCart Sensor Design Objectives CriteriaGoals Range21 – 70% O 2 Accuracy< 5% absolute error Power Demand< 10 W Cost< $50 Concentrator Design, Testing and Results Analyzer Design, Testing and Results Conclusions Air O2O2 Closed Air Pressurization Time (seconds) Purge Time (seconds) Maximum %O 2 Reached (%) Average Pressure (psi) Closed Acknowledgements & References We would like to thank Dr. Maria Oden, Dr. David Hilmers, Dr. John Graf, Dr. Gary Woods, Carlos Amaro, Joe Gesenhus the Oshman Engineering Design Kitchen (OEDK), Rice University’s Dept. of Bioengineering, Beyond Traditional Borders (BTB), NASA, Wyle Labs, Lockheed Martin, and the Bioastronautics Contract. 1.Duke, T., et al. Improved oxygen systems for childhood pneumonia: a multihospital effectiveness study in Papua New Guinea. Lancet, 2008; 372: Howie, S.R.C., et al. Meeting oxygen needs in Africa: an options analysis from the Gambia. Bull World Health Organ, 2009; 87:  Achievements Prototype testing indicates both systems are capable of achieving the design objectives.  Recommendations A top-top equalization step might increase the efficiency of the concentrator, helping decrease cost. The analyzer may be modified to rely on smaller, less expensive button zinc-air batteries.  Benefits The OxCart will offer healthcare providers a cheaper and more reliable respiratory treatment option than current standards.  2 million children die each year from acute respiratory illness. 98% of all deaths occur in developing regions. 1  Oxygen therapy reduces mortality from pneumonia by 35%. 2  Current oxygen delivery methods in developing areas are Prohibitively expensive (upwards of $1000) Prone to malfunction Fail to measure oxygen concentration  Current portable oxygen analyzers cost upwards of $200.  Stand-alone component  Relies on existing zinc-air battery technology  Battery voltage dependent on ambient oxygen concentration StepLeft BedRight Bed 1Venting/PurgingPressurizing 2 Venting/PurgingProducing 3 PressurizingVenting/Purging 4 ProducingVenting/Purging (Top) Fig 3. CAD drawing of OxCart. (Right) Fig 4. Current prototype of concentrator The OxCart uses a Pressure Swing Adsorption (PSA) system:  System contains several beds full of 5A Zeolite  5A Zeolite preferentially adsorbs nitrogen when pressurized Thus purifying the air, allowing for higher concentrations of oxygen  Depressurizing and purging zeolite regenerates it.  Varying cycle time leads to varying %O 2 concentrations.  As cycle time increases past a certain point, the output oxygen purity begins to decrease, but oxygen recovery will increase.  We are testing to find the maximum oxygen purity attained at each cycle time, time it takes to reach this maximum purity, and output flow O2O2 Fig 2. The Four steps of a PSA cycle. Table 1. Bed states during each step of PSA cycle. (Top) Graph 2. Comparison of our analyzer to commercial sensor (Right) Fig 1. Circuit diagram of analyzer Table 2. The data represented here is only sample data. Multiple tests under similar pressurization and purge times have been run. However, there are varying results because of varying qualities of zeolite as well as pressure drops.  3 ways to relate battery voltage to O 2 concentration Difference between final and baseline signal Difference between final and initial signal Amplified differential signal between 2 batteries  Analyzer calibrated and tested between 21% - 93% O 2  Logarithmic model relates voltage and O 2 concentration  Initial tests show analyzer capable of accuracy within 1% Graph 1. The graph shows how closely our data for the differential voltage signal between two batteries correlates with a best fit log line.