Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA.

Slides:



Advertisements
Similar presentations
Tutorial 3 Refractor assignment, Analysis, Modeling and Statics
Advertisements

Flatness, Parallelism, & Profile
Ion Beam Analysis techniques:
Saeedeh Ghaffari Nanofabrication Fall 2011 April 15 1.
X-ray Diffraction. X-ray Generation X-ray tube (sealed) Pure metal target (Cu) Electrons remover inner-shell electrons from target. Other electrons “fall”
Internal – External Order We described symmetry of crystal habit (32 point groups) We also looked at internal ordering of atoms in 3-D structure (230 space.
Are your X-ray powder diffraction patterns any good? Are your X-ray powder diffraction patterns any good?
Rutherford Backscattering Spectrometry
First of all, do you know any methods to check chemical composition? Or how you know what is what? First of all, do you know any methods to check chemical.
Chem Single Crystals For single crystals, we see the individual reciprocal lattice points projected onto the detector and we can determine the values.
GG450 April 22, 2008 Seismic Processing.
Understanding Flatness, Parallelism, & Profile in Calypso Last Updated: 9/12/2014Understanding Flatness, Parallelism, & Profile in Calypso1.
Catalysis and Catalysts - XPS X-Ray Electron Spectroscopy (XPS)  Applications: –catalyst composition –chemical nature of active phase –dispersion of active.
1 EE 542 Antennas and Propagation for Wireless Communications Array Antennas.
Chemical Analysis with the SEM Qualitative analysis Quantitative analysis Mapping of locations of elements.
Radiation Exposure, Dose and Relative Biological Effectiveness in Medicine Background Image:
Lesson 5 Representing Fields Geometrically. Class 13 Today, we will: review characteristics of field lines and contours learn more about electric field.
Rutherford Backscattering Spectrometry
Ge 116 Module 1: Scanning Electron Microscopy
1 Stratified sampling This method involves reducing variance by forcing more order onto the random number stream used as input As the simplest example,
Qualitative, quantitative analysis and “standardless” analysis NON DESTRUCTIVE CHEMICAL ANALYSIS Notes by: Dr Ivan Gržetić, professor University of Belgrade.
etc… Analysing samples with complex geometries Particles Inclusions
Thin Film Quantitation of Chemistry and Thickness Using EPMA John Donovan Micro Analytical Facility CAMCOR (Characterization of Advanced Materials in Oregon)
Ion Beam Analysis of Gold Flecks in a Foam Lattice F E Gauntlett, A S Clough Physics Department, University of Surrey, Guildford, GU2 7XH, UK.
Unit 2, Part 3: Characterizing Nanostructure Size Dr. Brian Grady-Lecturer
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron-impact inner shell ionization cross section measurements for heavy element impurities in fusion reactors Jingjun Zhu Institute of Nuclear Science.
Reflection and Refraction of Light
Chapter 12 Atomic X-Ray Spectroscopy
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
Adventures of an Igneous Petrologist in a Probe Lab John Fournelle University of Wisconsin Madison, Wisconsin.
The Nature of Light and the Laws of Geometric Optics
Sphere Standards and Standard Spheres Dr. Richard Young Optronic Laboratories, Inc.
Energy-Dispersive X-ray Microanalysis in the TEM Anthony J. Garratt-Reed Neil Rowlands.
Other modes associated with SEM: EBIC
Ion Beam Analysis of Gold Flecks in a Foam Lattice F E Gauntlett, A S Clough Physics Department, University of Surrey, Guildford, GU2 7XH, UK.
NANO 225 Micro/NanoFabrication Electron Microscopes 1.
PROJECTIONS OF SOLIDS & SECTIONS OF SOLIDS
Reminders for this week Homework #4 Due Wednesday (5/20) Lithography Lab Due Thursday (5/21) Quiz #3 on Thursday (5/21) – In Classroom –Covers Lithography,
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA Monte Carlo Simulations.
Crystallography and Diffraction. Theory and Modern Methods of Analysis Lectures Electron Diffraction Dr. I. Abrahams Queen Mary University of London.
The physics of electron backscatter diffraction Maarten Vos AMPL, RSPHYSSE, Australian National University, Acton 0200, Canberra Aimo Winkelmann Max Planck.
GG 450 Feb 27, 2008 Resistivity 2. Resistivity: Quantitative Interpretation - Flat interface Recall the angles that the current will take as it hits an.
Gravity and Motion. Gravity is what gives the universe its _________ A universal force that acts on _________ the objects in the universe Every particle.
Electron probe microanalysis Low Voltage SEM Operation Modified 9/23/10.
Förster Resonance Energy Transfer (FRET)
The probe Some material used from:. EPMA - electron probe microanalysis Probe signals.
2015 Cameca Users Meeting KαKα K L M Fluorescence and absorption… A sample of olivine has Fe, Mg and Si. Fe K α = 6.4 keV Binding energies… Mg K = 1.30.
Extrapolation Techniques  Four different techniques have been used to extrapolate near detector data to the far detector to predict the neutrino energy.
Spatial Resolution and minimum detection
X-ray spectroscopy Repetition, exercises and more
Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University.
Electron probe microanalysis EPMA
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
A. WAVE OPTICS B. GEOMETRIC OPTICS Light Rays
NANO 230 Micro/NanoFabrication
Monte Carlo studies of the configuration of the charge identifier
Electron probe microanalysis EPMA
Preface: What’s EPMA all about? How does Geology 777 work?
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Data Analysis in Particle Physics
Electron Probe Microanalysis EPMA
Electron probe microanalysis EPMA
Electron Probe Microanalysis EPMA
Proximity correction in electron beam lithography
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron probe microanalysis EPMA
Electron probe microanalysis
Lesson 4: Application to transport distributions
Presentation transcript:

Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA

What’s the point? EPMA is traditionally done for bulk material. What are the issues for particles? How precise/accurate are such analyses?

Bulk vs particle Normal EPMA assumes that the electron beam is exciting a homogeneous volume, i.e. there is no difference either laterally or vertically. Thus, the matrix correction is being applied in a uniform matter, and there is one applicable  (  z) profile for each element. However, particles create many difficulties, including: Their size (small) Their venue (location, surroundings) Their surface (not flat nor smooth)

The Good, the Bad, the Ugly Most operators of electron microprobes, using WDS, understand the difficulties of trying to do EPMA on non-flat surfaces (particles). However, the real problem is that there are probably 100 EDS systems on SEMs for every 1 WDS on an electron microprobe. And with SEMs and EDS systems so easy to operate, a large number of users make major errors without knowing it, assuming whatever the software spits out as a composition is really what it is. After these slides, you should know the difference between the GOOD, the BAD, and the UGLY… particle analyses.

Size: “Mass Effect” Goldstein et al, 1992, p. 479, 481 Consider a 2 um diameter sphere (say of NIST glass K412). Mass effect/error: electrons escape from sides of small particles if the E0 is great enough so that electrons scatter out of the body before using up all their energy.

Size: Absorption Effect Absorption effect of non-flat upper surface: there is a different path length from the normal flat geometry. In this case, the emergent x-rays will travel ~50% shorter distance in the material, and thus have ~50% of the absorption, and all else being equal, will yield a K-ratio twice as large as it would be if it were larger and had a flat polished surface. Goldstein et al, 1992, p. 479, 481

Size - Detector-sample Geometry EDS (shown to right): Variable effect of geometry of trajectory between beam impact area on non- uniform surface and the location of the detector. K-ratios from point 1 could be significantly different than those from point 3, despite the sphere being of uniform composition. WDS: Could be as bad, with up to 5 spectrometers positioned at 5 different orientations Goldstein et al, 1992, p. 479, 481

EDS vs WDS For all the problems cited, EDS has an advantage over WDS for “particle analysis” in that the spectrometer is in a constant position relative to the sample and beam “landing spot”, whereas WDS spectrometers would be at different positions, creating various different absorption path lengths to each spectromter. WDS 1 WDS 2

How Do People Try to Do EPMA of Particles? 1. Take whatever values come off the EDS with no correction other than built in ‘normalization to 100 wt%’; 2. Overscan the sample; 3. Put the beam on a top “flat” surface and hope for the best; 4. Do #3 but apply Armstrong and Buseck’s geometric correction factors; 5.Peak-to-background method; 6.Do Monte Carlo simulations with DTSA-2 or PENEPMA and compare with actual K-ratios (time consuming); 7.If large enough, mount, polish flat and do traditional WDS

Two approaches Traditional approach: (a) “overscan” and (b) normalize numbers – but this is not very good (above left table) Armstrong and Buseck (1975) developed a procedure based upon a regular geometric shape factor, where the different path length and other effects could be used. Method is based on bracketing particle and beam overscanning during collection of spectrum by EDS and modeling of electron path and x-ray propagation out through several shapes – sphere, hemisphere, squared pyramid, and rectangular, tetragonal, cylindrical and right triangular prisms. Correction factors are given in terms of predicted k-factors for pairs of particular elements, vs particle thickness along e beam. This is not easy, takes much trial and error, but apparently can yield fairly good results (see table above). Goldstein et al, 1992, p. 488, 489

Armstrong-Buseck correction Goldstein et al, 1992, p. 490 Here is another useful way to view a lot of data: error histograms. The top one shows the “conventional” particle analysis procedure to be very inaccurate, whereas using their geometric- correction approach, the errors shrink greatly. Though there still are problems, which the analyst must recognize and report with any writeup.

Other approach: Peak/Background It is known that whereas geometric effects can create large differences in measured x-ray intensities, the ratio of a characteristic x-ray to the nearby background (=continuum of same energy range) intensity is less sensitive to geometric effects. Thus, an ‘automatic internal normalization factor’ can be created for both a standard and an unknown. This has been shown to work well. The issue then is to find software that has this implemented. This approach is discussed in Goldstein et al, 2nd edition, pp (It is also used in biological EPMA, where the beam can destroy the cellular material, so mass “disappears”.)

Other approaches: Peak/Background Results from using P/B approach for 2 minerals. Excellent results for the pyrite, better but not within the 1% error we would like to see. Goldstein et al, 1992, p. 491

Other approaches: Monte Carlo Simulations Another approach is to create a similar geometric shape, give it a composition which is what you believe the unknown to be (based upon initial EPMA results). In the model, you place the beam on a particular spot corresponding to where the beam hits it, and then “run the simulation” to create x-ray counts (and similarly for a standard), and then compare the resulting simulated K- ratio with the experimental K-ratio. If the K-ratios match, then the composition is correct; if not, modify the composition and run another simulation. Iterate until the simulated and the experimental K-ratios match.

Other approaches: Monte Carlo Simulations Nicholas Ritchie at NIST has upgraded the DTSA-II EDS program to include Monte Carlo simulation of EDS in various simple shapes

DTSA-II You then compare the EDS spectrum of the simulated material with the actual experimental acquired EDS spectrum

PENEPMA Another high powered EPMA simulation program is PENEPMA, developed by Salvat and Llovet at U of Barcelona. It includes secondary fluorescence, though is a bit more complicated than DTSA-II. It can be used to create X-ray spectra in complicated geometries.

Other Approaches to Particle Analysis Another use of SEM to identify particles is electron backscatter diffraction (EBSD) which can use the diffraction pattern (Kikuchi lines) to match a database of crystal lattice data to ID a particular phase, though not the exact chemical composition. For example, it can distinguish hematite from magnetite, but not the amount of say Ti or Mn present.

EBSD of Fe-oxide particles Electron backscatter diffraction (EBSD) allows diffraction (”Kikuchi”) patterns to be collected in an SEM, upon a sample surface tilting (70°) sharply away from the electron beam, collected with a special phosphor or cc detector, and then the pattern is compared with a set of possible matching crystal structures (all computerized, of course). A best fit (MAD < 0.5-1) is calculated if one of the structures can match for any possible orientation direction.

Here 6 bands were used to match a particle from the smoke stack of a steel mill, to hematite, with a fit (MAD) of.55

Other concerns regarding particles If very fine (e.g. powder), difficult to disperse on a substrate, so different particle sizes and potentially different compositions are clearly separate Difficult to apply conductive coat to side walls and underhangs, so that insulating materials will charge in high vacuum. Need to use rotating/swiveling coating attachment.

Dispersion Technique-1 Ingredients: A small container with tight-fitting lid (here, 8 oz) Can of compressed air Mount with sticky tape (here, 1 cm Al SEM mount) Optional: something for mount to sit in/on so surface facing up (here: a large nut) Step 1: Place a small amount of the powder/ small grains to be dispersed, along wall. Above, amount shown is really too much, use only a couple of grams of fine powder.

Dispersion Technique-2 Step 2: Have already poked a small hole in the top, just large enough for the compressed air nozzle to stick through Step 3: Poke nozzle thru hole and give very brief (1/10 second) blast. Wait seconds before opening the top to investigate (10 seconds for coarse particles, 30+ seconds for fine powder)

Dispersion Technique-3 Results: nice dispersal of < um fly ash particles on carbon tape. If not dispersed, they would all be in a thick mass and difficult to clearly see the individual particles.

Inclusions = particles Plan View Cross section In a polished mount, there is a large flat surface, but we would consider small inclusions (at or less than the incident electron interaction volume) to be “particles”. Here, instead of being surrounded on the sides by air (or vacuum), and on the bottom by some substrate, the complication of having x-rays potentially from 2 masses is obvious.

Conclusion EPMA of particles is increasingly being applied in many situations; is not exactly straight forward; does yield results with potentially large inaccuracies/ errors; can only be considered qualitative, unless several careful steps are included