Page  1 A new method to measure 10 B uptake in lung adenocarcinoma in hospital BNCT D. Bolognini, E.M. Donegani, A. Mattera, M. Prest, S. Hasan E. Capelli,

Slides:



Advertisements
Similar presentations
Lesson 16 Nuclear Medicine. What is Nuclear Medicine? Diagnosis and Treatment of Disease using small amounts of radio-nuclides (radiopharmaceuticals)
Advertisements

RapidArc plan verification using ArcCHECK™
16 th ICNCT June 14-19, 2014 Helsinki, Finland Design and simulation of an optimized photoconverter for e-linac based neutron source for BNCT research.
Boron Neutron Capture Therapy
Click to edit Master title style IEEE short course Calorimetry Erika Garutti (DESY Hamburg, Germany) Michele Livan (University of.
1 Activation Analysis A comparison between FLUKA and FISPACT results Pavia, Gabriele Firpo Reactor and Safety Dept. Phone:
1 Current Compton Imaging Effort for Homeland Security at LLNL Address one of the outstanding problems for Homeland Security: “Faint” sources in the midst.
PHYSICS IN NUCLEAR MEDICINE: QUANTITAITVE SPECT AND CLINICAL APPLICATIONS Kathy Willowson Department of Nuclear Medicine, Royal North Shore Hospital University.
The LHCb Inner Tracker LHCb: is a single-arm forward spectrometer dedicated to B-physics acceptance: (250)mrad: The Outer Tracker: covers the large.
experimental platform
A novel therapy for liver metastases: a concrete hope after the first human treatment T.Pinelli*, A.Zonta +, S. Altieri*, S.Barni ++, A.Braghieri*, P.Pedroni*,
Laser Induced breakdown spectroscopy in Water for elemental analysis.
H. Koivunoro1, E. Hippelänen1, I. Auterinen2, L. Kankaanranta3, M
To evaluate the effect of bevacizumab (BV) on local tumor response and lung metastatic potential in boron neutron capture therapy (BNCT), referring to.
Determination of gamma dose component in thermal column of Pavia Triga reactor by using alanine ESR detectors Saverio Altieri Department of Physics University.
G. Bartesaghi, 11° ICATPP, Como, 5-9 October 2009 MONTE CARLO SIMULATIONS ON NEUTRON TRANSPORT AND ABSORBED DOSE IN TISSUE-EQUIVALENT PHANTOMS EXPOSED.
16 th International Congress on Neutron Capture Therapy June 14-19, Helsinki, Finland Improvement of a PGNAA Facility for BNCT in THOR C. K. Huang 1, H.
Nuclear Forensics: Neutron Activation & Radiography
VCI - V. Conti February, A TIME OF FLIGHT DETECTOR FOR THERMAL NEUTRONS FROM RADIOTHERAPY LINACS V. Conti, G.Bartesaghi, D.Bolognini, M.Prest,
Neutron autoradiography in nuclear track detectors: simultaneous observation of cells and nuclear tracks from BNC reaction by UV C sensitization of polycarbonate.
Inter-comparison project for boron concentration determination at INFN-University of Pavia (Italy) and CNEA (Argentina) A. Portu 1,2, I. Postuma 3,4, M.A.
Determination of uranium concentration in Jordanian oil shale using fission fragments techniques with CR-39 Khalid M. Abumurad Department of Physics, Yarmouk.
A General Purpose Brachytherapy Software Simulation + Analysis (isodose calculation) 2/10/2002 Geant4 Workshop CERN Susanna Guatelli Univ. and INFN Genova.
Parameter optimization for the determination of BSH in whole blood by 10 B-NMR Keijiro Saito K. Yoshino, M. Muto, A. Ishikawa and H. Ohki Department of.
Radiation Damage in Bipolar Transistors Caused by Thermal Neutrons I. Mandić, V. Cindro, G. Kramberger, E. S. Krištof, M. Mikuž, D. Vrtačnik Jožef Stefan.
Design and construction of BNCT irradiation
Possible calibration methods for the final LXe calorimeter A. Papa 01/20/2004.
Spectrophotometry.
AIR CORE SCINTILLATION DOSIMETER SUMMARY We have shown that Cerenkov light can be reduced to a negligible level in scintillation dosimetry by using an.
1 Data Acquisition What choices need to be made?.
N. Gubanova 1, V. Kanygin 2, A. Kichigin 3, S. Taskaev 4 1 Institute of Cytology and Genetics, Novosibirsk, Russia 2 Novosibirsk State Medical University,
VIEW on RAD HARDNESS TESTS of STS FEE IN MEPHI. Simakov A.B. – Head of “Special Microelectronics” Lab.
The improvement of the energy resolution in epi-thermal region of Bonner sphere using boric acid solution moderator H. Ueda1, H. Tanaka2, Y. Sakurai2.
Beam test possibilities at JINR and Fermilab V. Pronskikh Fermilab 02/15/2012.
Institute of Experimental and Applied Physics Czech Technical University in Prague IEAP – CTU Prague 1 Current status and future development of neutron.
Institute of Nuclear Engineering and Science National Tsing Hua University Institute of Nuclear Engineering and Science National Tsing Hua University BNCT.
M Asnal *, T Liamsuwan 2 and T Onjun 1 1 Sirindhorn International Institute of Technology, Thammasat University 2 Nuclear Research and Development Division,
ATOMIC ABSORPTION SPECTROSCOPY (AAS) Atomization: It is the conversion of molecules to their component atoms in gaseous state using a source of heat (flame).
PHYSICS 225, 2 ND YEAR LAB NUCLEAR RADIATION DETECTORS G.F. West Thurs, Jan. 19.
An electron/positron energy monitor based on synchrotron radiation. I.Meshkov, T. Mamedov, E. Syresin, An electron/positron energy monitor based on synchrotron.
Progress with GaAs Pixel Detectors K.M.Smith University of Glasgow Acknowledgements: RD8 & RD19 (CERN Detector R.&D. collaboration) XIMAGE (Aixtron, I.M.C.,
FrontierScience G.P.1 MATRIX an innovative Pixel Ionization Chamber for Online Monitoring of Hadrontherapy Treatments Giuseppe Pittà.
1 MPGD2009 Advancements of labelled radio-pharmaceutics imaging with the PIM-MPGD J. Donnard a, N.Arlicot b,
Photoneutron Distributions around 18 MV X-ray Radiotherapy Accelerators using Nuclear Track Detectors Fazal-ur-Rehman, H. Al-Ghamdi, M. I. Al-Jarallah.
Overview of the BNCT neutron beam line facility in NRI Rez (Prague) Most pictures taken from presentation of J.Burian/NRI – Milano09 1 NRI = Nuclear research.
Institute of Experimental and Applied Physics Czech Technical University in Prague IEAP – CTU Prague 1 BrightnESS T4.1 Meeting Activity at IEAP CTU in.
BNCT, a binary radiotherapy at cellular level
Progress with GaAs Pixel Detectors
Boron Neutron Capture Therapy: tecnica e sorgenti
RHD-DoPET CNAO test beam result (may 2014)
Low Energy Target Irradiation at LERF
Sensitivity of Hybrid Resistive Plate Chambers to Low-Energy Neutrons
Application of Čerenkov Radiation for Oxygenation and Surface Dose Assessment during External Beam Radiotherapy Rongxiao Zhang1, Adam Glaser2, Tatiana.
Triple GEM detectors : measurements of stray neutron.
Very High Energy Electron for Radiotherapy Studies
solid state vs gas detectors Scintillator detector
ADvanced MOnolithic Sensors for
Diagnostics of FRIBs beam transport line
Cross-section Measurements of (n,xn) Threshold Reactions
GLAST LAT tracker signal simulation and trigger timing study
Neutron detectors for the NMX instrument
Gamma-ray Spectroscopy technique is commonly used in Planetary Exploration Missions.
1. Introduction Secondary Heavy charged particle (fragment) production
Function and Structure in
Design of A New Wide-dynamic-range Neutron Spectrometer for BNCT with Liquid Moderator and Absorber S. Tamaki1, I. Murata1 1. Division of Electrical,
HE instrument and in-orbit performance
O. Svoboda, A. Krása, A. Kugler, M. Majerle, J. Vrzalová, V. Wagner
Sheng Yang a, Yen-Wan Hsueh Liu b
Nanoparticles for enhancing the effectiveness of proton therapy
Aging test of glass and MRPC
Presentation transcript:

Page  1 A new method to measure 10 B uptake in lung adenocarcinoma in hospital BNCT D. Bolognini, E.M. Donegani, A. Mattera, M. Prest, S. Hasan E. Capelli, P. Chiari A. Zanini G. Giannini, E. Vallazza F. Basilico, P. Mauri F. Borasio P. Cappelletti, M. Frigerio, S. Gelosa, A. F. Monti

Page  2  BNCT  The INFN PhoNeS project  An innovative approach to Boron Imaging  Neutron Autoradiography  Measurements and results  Conclusion Outline

Page  3 Boron Neutron Capture Therapy BNCT could become useful for highly-fatal tumours: Extended (e.g. liver, lung) Radioresistant (melanoma) Localized near vital organs (glioblastoma) 10 B (n,α,γ) 7 Li binary technique high LET particles

Page  4 High intensity thermal-epithermal neutron beam (5·10 8 n/cm 2 s 1 ) < 10 keV energy range Gamma contamination < 2 · Gy cm 2 / n Low toxicity High specificity Non-persistence in the bloodstream BSH BPA Difficulties for BNCT applications

Page  5 Neutron Autoradiography  Nuclear reactor  Solid state track detector (CR 39)  Sometimes cryogenic temperature during the irradiation  Optic microscope for identification and count  Nuclear reactor  Solid state track detector (CR 39)  Sometimes cryogenic temperature during the irradiation  Optic microscope for identification and count  Hospital-based neutron source  Real time detector  Hospital-based neutron source  Real time detector

Page  6 PhoNeS ( PhotoNeutronSource ) Neutron flux in the treatment cavity: Φ = (1.3 ± 0.1) x 10 7 n cm -2 s -1 γ-contamination: reduced of 97%

Page  7 Real-time Neutron Autoradiography 0.5 cm CERROBEND photonconversion 6 cm PMMA thermalization IRON BOX Samples Detector Neutron flux: Φ = (1.5 ± 0.1) x 10 5 n cm -2 s -1

Page  8 The imaging detector Silicon Dimensions(9.5 x 9.5) cm 2 Hamamatsu Microstrip Silicon Detector768 strips Readout pitch242 μm Read-out by3 self-triggering TAA1 ASICs Counts vs position

Page  9 The Data Acquisition System INTER-BUNCH acquisition UNDEPLETED silicon detector to detect α-particles from boronated samples down to 25 ng of 10 B 18 MeV electrons Bunches of 5 μs Frequency Hz

Page  10 SPECIMEN SUPPORT -plastic support (liquid solution) evaporation -glass slide (blood samples) coagulation process to prevent crystallization and self-absorption of the α-particles by the sample itself KNOWN VOLUME Drops of 5 μl The samples SOLUTION HOMOGENEITY

Page  11 Calibrating with BSH 500 MU/min –18 MV photons Varian CLINAC 2100 C/D, Sant’Anna Hospital (Como) Integrated Spot Content (AU) vs Concentration (ppm)

Page  12 The first ever kinetic curves Special thanks to Dr.Sauerwein URINEBLOOD rescaled at the max value obtained by Si-autoradiography Blood and urine of a patient administered BSH, taken at different times with respect to the administration.

Page  13 Ex-vivo lung perfusion Human lung affected by adenocarcinoma: therapetic resection perfusion of blood and BPA Blood samples taken at different times: to estimate 10 B uptake in collaboration with “S.Luigi” Hospital in Orbassano (TO), Italy

Page  14 Kinetic curves Blood samples doped with BPA Comparison between real-time autoradiography to MS analysis The absorption takes place only in the first 90mins

Page  15 CR 39 Healthy tissue Unhealthy tissue Special thanks to Dr. Giannini 10 B concentration in tumour and healthy tissues of resected lobe High-resolution images to reproduce the features of the tissues C T / C H ~

Page  16 Conclusions  Neutron Autoradiography in a hospital environment successfully performed - Linac  Development of an hospital-base neutron source to perform Boron Imaging with a 1D spatial resolution  Real-time detector to perform pharmacokinetic studies results during the irradiation  Blood samples from a perfused lung affected by adenocarcinoma kinetic curves - 10 BPA down to concentrations of 5ppm  Evaluation of 10 B concentration in tumour and healthy tissues Si-autoradiography – CR 39  Looking forward to make BNCT possible in every radiotherapic unit

Page  17 A new method to measure 10 B uptake in lung adenocarcinoma in hospital BNCT D. Bolognini, E.M. Donegani, A. Mattera, M. Prest, S. Hasan E. Capelli, P. Chiari A. Zanini G. Giannini, E. Vallazza F. Basilico, P. Mauri F. Borasio P. Cappelletti, M. Frigerio, S. Gelosa, A. F. Monti Thank you!