Dynamics to determine calibration intervals SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006.

Slides:



Advertisements
Similar presentations
Microspectrophotometry Validation. Reasons for Changing Instruments Reduced reliability. Limited efficiency. Limited availability and cost of replacement.
Advertisements

Supplier SQM Participation. 2 | MDT Confidential What is SQM? Stands for Supplier Quality Managment –Formally referred to as SPACE and SPICE Is a system.
All activities involved in keeping a system’s equipments working are the part of maintenances management. Maintenance of machine involves the efforts.
The Role of the SAAMF (South African Automotive Measurement & Testing Forum) SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive.
Establish Verification Procedures (Task 11 / Principle 6)
Corrective & Preventive Action Programme l Corrective and preventive action managed by one programme l Closely linked to the internal audit programme l.
Laboratory Personnel Dr/Ehsan Moahmen Rizk.
Lecture 8. Quality Assurance/Quality Control The Islamic University of Gaza- Environmental Engineering Department Environmental Measurements (EENV 4244)
What is Data ? Data is a numerical expression of an activity. Conclusions based on facts and data are necessary for any improvement. -K. Ishikawa If you.
Soft. Eng. II, Spr. 2002Dr Driss Kettani, from I. Sommerville1 CSC-3325: Chapter 9 Title : Reliability Reading: I. Sommerville, Chap. 16, 17 and 18.
ISO 9001 Interpretation : Exclusions
©Ian Sommerville 2004Software Engineering, 7th edition. Chapter 22 Slide 1 Verification and Validation.
Quality Assurance. Function Quality program Quality Assurance –Covers everything from raw materials and GMP verification through finished-product release.
Quality Assurance/Quality Control Policy
Automotive Metrology Laboratory
Chapter 10 Verification Procedures. Objective In this module, you will learn: u How to define verification u What functions are part of HACCP plan verification.
JENN SHAFNER BRIAN KROUSE CLINT KEHRES. Pre ISO 9000  The BS 5750 standard required factories to document manufacturing procedures.  BS 5750 was known.
Project co- financed by Asean European Committee for Standardization Implementing Agency Project co-financed by European Union Module 7 GMP Workshop Kuala.
Planning an Internal Audit JM García Merced. Brainstorm.
BRC Food Safety Quality Management System Training Guide
Systems Thinking and Systems Engineering Verification & Validation 14 March 2013 Francois Christophe Galina Medyna Eric Coatanéa.
World Health Organization
Determining Sample Size
Introduction to Equipment
Introduction to ISO New and modified requirements.
Introduction to Software Quality Assurance (SQA)
Verification and Validation Yonsei University 2 nd Semester, 2014 Sanghyun Park.
Torrington, Hall & Taylor, Human Resource Management 6e, © Pearson Education Limited 2005 Slide 22.1 Protection from Hazards Conflict between needs for.
WHY CONFORMITY ASSESSMENT?. What is conformity assessment?  Conformity assessment is the name given to processes that are used to demonstrate that a.
Equipment Management Audience – Local lab responsibilities.
Quality Control Project Management Unit Credit Value : 4 Essential
Quality WHAT IS QUALITY
Automotive Test & Measurement Case Studies SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006.
ZHRC/HTI Financial Management Training Session 9: Stores and Supplies Management.
B. Neidhart, W. Wegscheider (Eds.): Quality in Chemical Measurements © Springer-Verlag Berlin Heidelberg 2000 J. Green Why do we need Good results? 1 Chemical.
10/07/2009 Matthias Schuetzeberg Slide 1 Calibration Management In Food & Beverage.
Important informations
Your measurement technology partner for global competitiveness CSIR National Metrology Laboratory The National Metrology Laboratory of South Africa Feedback.
Presented by: David Beamish DeFelsko Corporation
Project Specification part 2 BTEC National. Project boundaries or scope The boundaries or scope of a project are what the project aims to achieve. The.
Chapter 3 Definitions, Goals, and Objectives. Maintenance Defined Airline: “Those actions required for restoring or maintaining an item in a serviceable.
©Ian Sommerville 2000Software Engineering, 6th edition. Chapter 19Slide 1 Chapter 19 Verification and Validation.
Risk Management & Corporate Governance 1. What is Risk?  Risk arises from uncertainty; but all uncertainties do not carry risk.  Possibility of an unfavorable.
Uncertainty of measurement & Technical Specifications SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006.
SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006 ISO/TS 16949:2002 certification – Meeting the requirements of.
Information Security 14 October 2005 IT Security Unit Ministry of IT & Telecommunications.
Cost Accounting - an overview Cost accounting is the art or process of recording, analyzing and classifying of expenditure for the purpose of product costing.
1 A Seminar On Pharmaceutical Outsourcing A Seminar On Pharmaceutical Outsourcing.
UNCERTAINTY OF MEASUREMENT Andrew Pascall Technical Director Integral Laboratories (Pty) Ltd
Metrology & Accreditation – Their Role in the Global Market Presented at the 2009 NCSLI Conference San Antonio, Texas 30 July 2009 by Roxanne Robinson.
You’ve Found The Cure For Cancer… But Can’t Reproduce Your Results! Could Your Test Equipment Be The Weakest Link?
1. Our Presentation Topic: Importance Of Validation & Qualification In Pharmaceutical Industries Presented By: Md. Tanjir Islam (Group C) 2.
Overview of Instrument Calibration Presents by NCQC, India.
QUALITY ASSURANCE MODULE 3. Definitions Quality Assurance - A process/effort that ensures that processes are followed, that the processes of doing right.
Chapter 10 Verification Procedures.
Quality Brian Russell.
World Health Organization
MAINTENANCE ENGINEERING
Hardware & Software Reliability
CSC 480 Software Engineering
Calibration Web Innovations 2017.
Instrument Maintenance
Quality Control Procedures & Raw Materials
Understanding Service Centers
Construction Management
Chapter 10 Verification Procedures.
An Introduction to Quality Assurance in Analytical Science
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Measurements & Error Analysis
Quality Control Procedures & Raw Materials
Presentation transcript:

Dynamics to determine calibration intervals SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006

Slide 2 © CSIR Calibration Myths A calibration certificate implies that an instrument meets its accuracy specification, Calibration means that the instrument was adjusted to ensure it meets its accuracy specification, It is sometimes more expensive to have an instrument calibrated than to replace it every year with a new one, Only instruments which can be adjusted require calibration; therefore e.g. LIG thermometers don’t need periodic calibration, Re-calibration of an instrument only applies to instruments where the manufacturer has specified a calibration interval, Calibration interval is dictated by the calibration service provider and the user is bound by this, Safety regulations stipulate 1 year cal intervals

Slide 3 © CSIR What is the aim of calibration? To ensure that a piece of test or measuring equipment can perform the test or measurement within the required level of accuracy To maintain a balance between cost and the risk of a measurement being performed outside of the required accuracy Excessive cost – zero risk Zero cost – excessive risk Reasonable cost – acceptable risk Risk of a “safety critical” component erroneously being found to comply with the technical specification requirements may be very different to the risk associated with say the thickness of a carpet erroneously being found to comply with the specifications

Slide 4 © CSIR What is a calibration interval? It is the interval between calibrations, chosen to ensure the test or measuring instrument continues to perform at the required level of accuracy At the end of the interval an accuracy must be predictable with a certain level of confidence It is NOT an interval dictated by the calibration service provider It is NOT an interval determined by the freq of quality audits It is NOT an interval determined by budgetary considerations It is NOT an interval determined by the availability of the instrument (taken out of service for cal) It should form part of an integrated calibration plan (not be arbitrarily chosen) If managed correctly it can reduce the costs of calibration

Slide 5 © CSIR What factors influence a calibration interval? The type (quality/stability) of instrument Selection of appropriate instrument when purchasing very important

Slide 6 © CSIR What factors influence a calibration interval? (2) The accuracy required Higher accuracy usually implies higher risk High accuracy instruments being used for low accuracy applications

Slide 7 © CSIR What factors influence a calibration interval? (3) The calibration service provider Do they always use the same calibration procedure? Do they adjust the instrument? Do they ensure compliance as an output of calibration? Is the same calibration service provider used for regular calibrations (important factor to consider when building up a history) Historical performance of the instrument Can include factors such as environment, operator use, etc The use/abuse of the instrument

Slide 8 © CSIR Who is responsible for determining the calibration interval? The “user”, in conjunction with the calibration service provider This responsibility CANNOT be delegated to the calibration service provider without careful consideration and contract review This CANNOT be made the responsibility of the quality department This, by necessity, implies that instrument “users” must be more intimately involved with their instruments Financial constraints should not influence the determination of calibration intervals Calibration intervals should however be technically optimised to reduce costs to the minimum required to maintain the desired level of accuracy

Slide 9 © CSIR Suggested ways of determining a calibration interval Industry – instruments on the factory floor Each calibration, check for conformance to the manufacturer’s accuracy specifications Ensure calibration service providers adjust instruments, if necessary, to ensure they conform Calibration service providers must record “before” and “after” adjustment values in order for the user to take remedial action if required This method assumes the accuracy of the instrument, as purchased, is fit for the intended purpose Track when adjustment is required – if no adjustment required lengthen interval, if adjustment required shorten the calibration interval This method should however not exceed a predetermined appropriate maximum interval even if no adjustment is required as this increases risk

Slide 10 © CSIR Ways of determining a calibration interval (2) Test Labs – instruments used within a laboratory Calibration intervals can be determined based on actual drift

Slide 11 © CSIR Use of confidence checks to reduce risk Even a calibration interval selected on one of the two methods discussed is meaningless when:- The instrument fails but the fault is not instantly noticeable, The instrument is abused unknowingly, The instrument is adjusted without prior knowledge “In-between” checks can reduce the risk associated with these scenarios Frequency and thoroughness of check depends on the risks associated and to some degree on available resources Confidence checks are NOT a substitute for calibration or “cheap do-it-yourself” calibration Confidence checks do NOT lengthen calibration intervals Confidence checks REDUCE risk!

Questions??