Adaptive Imaging Preliminary: Speckle Correlation Analysis.

Slides:



Advertisements
Similar presentations
Noise in Radiographic Imaging
Advertisements

A Crash Course in Radio Astronomy and Interferometry: 2
Foundations of Medical Ultrasonic Imaging
ACHIZITIA IN TIMP REAL A SEMNALELOR. Three frames of a sampled time domain signal. The Fast Fourier Transform (FFT) is the heart of the real-time spectrum.
1 Small-scale Mobile radio propagation Small-scale Mobile radio propagation l Small scale propagation implies signal quality in a short distance or time.
7. Radar Meteorology References Battan (1973) Atlas (1989)
THE AUSTRALIAN NATIONAL UNIVERSITY Infrasound Technology Workshop, November 2007, Tokyo, Japan OPTIMUM ARRAY DESIGN FOR THE DETECTION OF DISTANT.
Hospital Physics Group
ECE 501 Introduction to BME
Ultrasonic Nonlinear Imaging- Tissue Harmonic Imaging.
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods Tomography Part 3.
Probabilistic video stabilization using Kalman filtering and mosaicking.
Image Enhancement in the Frequency Domain Part I Image Enhancement in the Frequency Domain Part I Dr. Samir H. Abdul-Jauwad Electrical Engineering Department.
Ultrasound Imaging Atam Dhawan.
Ultrasonic Nonlinear Imaging- Tissue Harmonic Imaging
EEE440 Modern Communication Systems Wireless and Mobile Communications.
Chapter 4 Mobile Radio Propagation: Small-Scale Fading and Multipath
Despeckle Filtering in Medical Ultrasound Imaging
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasonography.
Inputs to Signal Generation.vi: -Initial Distance (m) -Velocity (m/s) -Chirp Duration (s) -Sampling Info (Sampling Frequency, Window Size) -Original Signal.
For 3-G Systems Tara Larzelere EE 497A Semester Project.
Review Doppler Radar (Fig. 3.1) A simplified block diagram 10/29-11/11/2013METR
Lecture 1 Signals in the Time and Frequency Domains
Antennas The primary elements of a synthesis array M. Kesteven ATNF 25/September/2001.
Resident Categorical Course
EE 6332, Spring, 2014 Wireless Communication Zhu Han Department of Electrical and Computer Engineering Class 3 Jan. 22 nd, 2014.
1 Ultrasonic Elasticity Imaging. 2 Elasticity Imaging Image contrast is based on tissue elasticity (typically Young’s modulus or shear modulus).
Dr A VENGADARAJAN, Sc ‘F’, LRDE
Ping Zhang, Zhen Li,Jianmin Zhou, Quan Chen, Bangsen Tian
Wireless Communication Technologies 1 Outline Introduction OFDM Basics Performance sensitivity for imperfect circuit Timing and.
Building Three-Dimensional Images Using a Time-Reversal Chaotic Cavity
Advanced Imaging 1024 Jan. 7, 2009 Ultrasound Lectures History and tour Wave equation Diffraction theory Rayleigh-Sommerfeld Impulse response Beams Lecture.
Speckle Correlation Analysis1 Adaptive Imaging Preliminary: Speckle Correlation Analysis.
Detection of Obscured Targets: Signal Processing James McClellan and Waymond R. Scott, Jr. School of Electrical and Computer Engineering Georgia Institute.
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Sarah Gillies Ultrasound Sarah Gillies
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
1 What is small scale fading? Small scale fading is used to describe the rapid fluctuation of the amplitude, phases, or multipath delays of a radio signal.
Adaphed from Rappaport’s Chapter 5
Digital Beamforming. Beamforming Manipulation of transmit and receive apertures. Trade-off performance/cost to achieve: –Steer and focus the transmit.
Chapter 6: Real-Time Image Formation. display B, M, Doppler image processing Doppler processing digital receive beamformer system control keyboard beamformer.
Statistical Description of Multipath Fading
Z bigniew Leonowicz, Wroclaw University of Technology Z bigniew Leonowicz, Wroclaw University of Technology, Poland XXIX  IC-SPETO.
3.7 Adaptive filtering Joonas Vanninen Antonio Palomino Alarcos.
A 5-Pulse Sequence for Harmonic and Sub-Harmonic Imaging
XBSM Analysis - Dan Peterson Review of the optics elements: Pinhole (“GAP”), FZP, Coded Aperture Extracting information from the GAP what is the GAP width?
RADAR ANTENNA. Functions of Radar Antenna Transducer. Concentrates the radiated energy in one direction (Gain). Collects echo energy scattered back to.
1 EMLAB EM wave propagation. 2 EMLAB Impulse response Time Radio Propagation : physical model 안테나에서 나온 신호는 지형지물과 반사, 투과, 산란을 거치면서 다양한 진폭과, 시간 지연을 갖는 신호들로.
TISSUE HARMONIC IMAGING (THI)
EEE 441 Wireless And Mobile Communications
NUR INANI BINTI MOHD RUZLAN A Tissue harmonic imaging is a technique in ultrasonography that provides images of better quality as compared to conventional.
1 Opto-Acoustic Imaging 台大電機系李百祺. 2 Conventional Ultrasonic Imaging Spatial resolution is mainly determined by frequency. Fabrication of high frequency.
Mobile Radio Propagation - Small-Scale Fading and Multipath
Biointelligence Laboratory, Seoul National University
Shadowing.
Chapter 35-Diffraction Chapter 35 opener. Parallel coherent light from a laser, which acts as nearly a point source, illuminates these shears. Instead.
Lecture on Ultrasonography
Degradation/Restoration Model
Outline Introduction Signal, random variable, random process and spectra Analog modulation Analog to digital conversion Digital transmission through baseband.
Advanced Wireless Networks
Optical Coherence Tomography
Review of Ultrasonic Imaging
Digital Beamforming.
Fourier Optics P47 – Optics: Unit 8.
Ultrasound Despeckling for Contrast Enhancement
Unit 2 Particles and Waves Interference
Improved Spread Spectrum: A New Modulation Technique for Robust Watermarking IEEE Trans. On Signal Processing, April 2003 Multimedia Security.
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

Adaptive Imaging Preliminary: Speckle Correlation Analysis

Speckle Formation Speckle results from coherent interference of un- resolvable objects. It depends on both the frequency and the distance. sample volume transducer

Speckle Second-Order Statistics The auto-covariance function of the received phase-sensitive signals (i.e., before envelope detection) is simply the convolution of the system’s point spread function if the insonified region is –macroscopically slow-varying. –microscopically un-correlated.

Speckle Second-Order Statistics The shape of a speckle spot (assuming fully developed) is simply determined by the shape of the point spread function. The higher the spatial resolution, the finer the speckle pattern, and vice versa.

Speckle Statistics The above statements do not hold if the object has structures compared to or larger than the ultrasonic wavelength. Rician distribution is often used for more general scatterer distribution. Rayleigh distribution is a special case of Rician distribution.

van Cittert-Zernike Theorem A theorem originally developed in statistical optics. It describes the second-order statistics of the field produced by an in-coherent source. The insonification of diffuse scatterers is assumed in-coherent. It is different from the aforementioned lateral displacement.

van Cittert-Zernike Theorem The theorem describes the spatial covariance of signals received at two different points in space. For a point target, the correlation of the two signals should simply be 1. For speckle, correlation decreases since the received signal changes.

van Cittert-Zernike Theorem The theorem assumes that the target is microscopically un-correlated. The spatial covariance function is the Fourier transform of the radiation pattern at the point of interest.

van Cittert-Zernike Theorem radiation pattern correlation

van Cittert-Zernike Theorem The theorem states that the correlation coefficient decreases from 1 to 0 as the distance increases from 0 to full aperture size. The correlation is independent of the frequency, aperture size, …etc.

van Cittert-Zernike Theorem In the presence of tissue inhomogeneities, the covariance function is narrower since the radiation pattern is wider. The decrease in correlation results in lower accuracy in estimation if signals from different channels are used.

van Cittert-Zernike Theorem distance correlation

van Cittert-Zernike Theorem Channel Time (Range) RF Signals

van Cittert-Zernike Theorem (Focal length 60mm vs. 90mm)

van Cittert-Zernike Theorem (16 Elements vs. 31 Elements)

van Cittert-Zernike Theorem (2.5MHz vs. 3.5MHz)

van Cittert-Zernike Theorem (with Aberrations)

Lateral Speckle Correlation correlation coefficient displacement L/2

Lateral Speckle Correlation Assuming the target is at focus, the correlation roughly decreases linearly as the lateral displacement increases. The correlation becomes zero when the displacement is about half the aperture size. Correlation may decrease in the presence of non-ideal beam formation.

Lateral Speckle Correlation 14.4 mm Array

Lateral Speckle Correlation

Lateral Speckle Correlation: Implications on Spatial Compounding

Speckle Tracking Estimation of displacement is essential in many imaging areas such as Doppler imaging and elasticity imaging. Speckle targets, which generally are not as ideal as points targets, must be used in many clinical situations.

Speckle Tracking From previous analysis on speckle analysis, we found the local speckle patterns simply translate assuming the displacement is small. Therefore, speckle patterns obtained at two instances are highly correlated and can be used to estimate 2D displacements.

Speckle Tracking Displacements can also be found using phase changes (similar to the conventional Doppler technique). Alternatively, displacements in space can be estimated by using the linear phase shifts in the spatial frequency domain.

Speckle Tracking Tracking of the speckle pattern can be used for 2D blood flow imaging. Conventional Doppler imaging can only track axial motion. Techniques using phase information are still inherently limited by the nature of Doppler shifts.

Adaptive Imaging Methods: Correlation-Based Approach

Sound Velocity Inhomogeneities transducer array v1 v2 v3 point of interest body wallviscera

Sound Velocity Inhomogeneities Velocity (m/sec) water 1484 blood 1550 myocardium 1550 fat 1450 liver 1570 kidney 1560

Sound Velocity Inhomogeneities Sound velocity variations result in arrival time errors. Most imaging systems assume a constant sound velocity. Therefore, sound velocity variations produce beam formation errors. The beam formation errors are body type dependent.

Sound Velocity Inhomogeneities Due to beam formation errors, mainlobe may be wider and sidelobes may be higher. Both spatial and contrast resolution are affected. no errorswith errors

Near Field Assumption Assuming the effects of sound velocity inhomogeneities can be modeled as a phase screen at the face of the transducer, beam formation errors can be reduced by correcting the delays between channels. beam formation correction geometric delay velocity variations aligned

Correlation-Based Aberration Correction No Focusing

Correlation-Based Aberration Correction Transmit Focusing Only

Correlation-Based Aberration Correction Transmit and Receive Focusing

Correlation-Based Aberration Correction Wire: Before CorrectionWire: After Correction

Correlation-Based Aberration Correction Diffuse Scatterers: BeforeDiffuse Scatterers: After

Correlation Based Method Time delay (phase) errors are found by finding the peak of the cross correlation function. It is applicable to both point and diffuse targets.

Correlation Based Method The relative time delays between adjacent channels need to be un-wrapped. Estimation errors may propagate.

Correlation Based Method Two assumptions for diffuse scatterers: –spatial white noise. –high correlation (van Cittert-Zernike theorem). filtercorrelator xx

Correlation Based Method Correlation using signals from diffuse scatterers under-estimates the phase errors. The larger the phase errors, the more severe the underestimation. Iteration is necessary (a stable process).

Alternative Methods Correlation based method is equivalent to minimizing the l2 norm. Some alternative methods minimize the l1 norm. Correlation based method is equivalent to a maximum brightness technique.

Baseband Method The formulation is very similar to the correlation technique used in Color Doppler.

Baseband Method CORDIC I Q I Q acc. Q sign bitsign control

One-Dimensional Correction: Problems Sound velocity inhomogeneities are not restricted to the array direction. Therefore, two- dimensional correction is necessary in most cases. The near field model may not be correct in some cases.

One-Dimensional Correction: Problems

Two-Dimensional Correction Using 1D arrays, time delay errors can only be corrected along the array direction. The signal received by each channel of a 1D array is an average signal. Hence, estimation accuracy may be reduced if the elevational height is large. 2D correction is necessary.

Two-Dimensional Correction Each array element has four adjacent elements. The correlation path between two array elements can be arbitrary. The phase error between any two elements should be independent of the correlation path.

Full 2D Correction (1,1)(1,3)(1,2)corr (3,1)(3,3)(3,2)corr (2,1)(2,3)(2,2)corr

Row-Sum 2D Correction (1,1)(1,3)(1,2)corr (3,1)(3,3)(3,2)corr (2,1)(2,3)(2,2)corr

Correlation Based Method: Misc. Signals from each channel can be correlated to the beam sum. Limited human studies have shown its efficacy, but the performance is not consistent clinically. 2D arrays are required to improve the 3D resolution.

Displaced Phase Screen Model Sound velocity inhomogeneities may be modeled as a phase screen at some distance from the transducer to account for the distributed velocity variations. The displaced phase screen not only produces time delay errors, it also distorts ultrasonic wavefronts.

Displaced Phase Screen Model Received signals need to be “back-propagated” to an “optimal” distance by using the angular spectrum method. The “optimal” distance is determined by using a similarity factor. phase screen

Displaced Phase Screen Model

TSC + BP Time-shift compensation with back-propagation

Abdominal Wall Measurements

Displaced Phase Screen Model After the signals are back-propagated, correlation technique is then used to find errors in arrival time. It is extremely computationally extensive, almost impossible to implement in real-time using current technologies.

Wavefront Distortion Measurements on abdominal walls, breasts and chest walls have shown two- dimensional distortion. The distortion includes time delay errors and amplitude errors (resulting from wavefront distortion).

Phase Conjugation phase screen at face of transducer displaced phase screen ff phase

Phase Conjugation

No aberration At 0 mm At 60 mm At 40 mm At 20 mm

Phase Conjugation Simple time delays result in linear phase shift in the frequency domain. Displaced phase screens result in wavefront distortion, which can be characterized by non-linear phase shift in the frequency domain.

Phase Conjugation Non-linear phase shift can be corrected by dividing the spectrum into sub-bands and correct for “time delays” individually. In the limit when each sub-band is infinitesimally small, it is essentially a phase conjugation technique.

End 4/13/2005

Some of the Recent Developments

Real-Time In Vivo Imaging[15]

Real-Time In Vivo Imaging

Distribution of time delay corrections

Clinical Imaging Using 1-D Array [16]

Clinical Imaging Using 1-D Array Before CorrectionAfter Correction

Clinical Imaging Using 1-D Array Before CorrectionAfter Correction

Clinical Imaging Using 1-D Array Channel DataComplex Scattering Structures

Real Time Adaptive Imaging with 1.75D, High Frequency Arrays [17] 1D and 2D Least Squares Estimation

Real Time Adaptive Imaging with 1.75D, High Frequency Arrays Before Correction After Correction

Real Time Adaptive Imaging with 1.75D, High Frequency Arrays Before Correction After Correction

Real Time Adaptive Imaging with 1.75D, High Frequency Arrays Original1 iteration4 iterations

Real Time Adaptive Imaging with 1.75D, High Frequency Arrays OriginalReceive Only

2D Correction Using 1.75d Array On Breast Microcalcifications [18]

2D Correction Using 1.75d Array On Breast Microcalcifications

(also with a 60% brightness improvement)

2D Correction Using 1.75d Array On Breast Microcalcifications

(a)1D (b)1D with correction (c)1.75D (d)1.75D with correction

Adaptive Imaging Methods: Aperture Domain Processing Parallel Adaptive Receive Compensation Algorithm

Single Transmit Imaging Fixed direction transmit, all direction receive

Measuring Source Profile

Removing Focusing Errors

Focusing Errors No Aberrations With Aberrations

Single Transmit Imaging No Aberrations With Aberrations

PARCA No Correction With Correction

Simplifications: 1. DFT vs. Single Transmit Imaging 2. Weighting vs. Complex Computations

DFT vs. Single Transmit Imaging Single Transmit ImagingDFT

Adaptive Weighting

Frequency Domain Interpretation of the Aperture Data Aberrated Speckle Coherent Incoherent *P.-C. Li and M.-L. Li, “Adaptive Imaging Using the Generalized Coherence Factor”, IEEE UFFC, Feb., 2003.

Coherence Factor (CF) A quantitative measure of coherence of the received array signals. Coherent sum (DC) Total energy (times N) N: the number of array channels used in beam sum C(i,t) : the received signal of channel i The larger, the better?

Determination of the Optimal Receive Aperture Size Classify “object types” Unwanted Sidelobes Object of Interest Enhance Optimize the receive aperture size Suppress

Experimental Results: Tissue Mimicking Phantoms Range Azimuth Dynamic range: 60 dB –40  40  0 XX XX Original Adaptive Receive Aperture 28.6 mm 96.2 mm