Fluoroscopy Q&A Robert L. Metzger, Ph.D..

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

Image Reconstruction.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Radiation Safety Training for Fluoroscopy in Research Radiation Safety Office Indiana University Purdue University Indianapolis and Associated Facilities.
X-Ray Interaction with Matter & Human Biology
Digital Radiography.
Film Screen Q & A Robert Metzger, Ph.D.. RAPHEX General Question 2001 D10: Flood replenishment of a processor involves all of the following except: A.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Resident Physics Lectures
Analyzing the Image Density. Density Overall blackening of the image.
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
CVT 102.  Time  Distance  Shielding.
Radiation Exposure and Risk John Hamaty. Evaluation of Radiation Exposure Levels in Cine Cardiac Catheterization Laboratories American Association of.
Fluoroscopy Equipment
Intensifying Screens Kyle Thornton DMI 50B.
Quality Control Rad T 110.
RAD 350 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion Invented by Thomas Edison.
Fluoroscopy Notes Rad Tech 290.
Special Imaging Systems
Fluoroscopy Robert Metzger, Ph.D..
F luoroscopy. 2 Fluoroscopy system 3 Different fluoroscopy systems  Remote control systems Not requiring the presence of medical specialists inside.
Control of Scatter Radiation
Computers in Imaging Q & A Robert Metzger, Ph.D..
FLUOROSCOPY EQUIPMENT
Quality Control.
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
Dr Mohamed El Safwany, MD.
Image Quality Radiographic Resolution.
Fluoroscopy Equipment. Introduction Fluoro: is dynamic radiographic examination Fluoroscopy is primarily domain of the radiologist The role of radiographer.
Mammography Questions and Answers
Geometric Factors Focal Spot Object Film Object b a h c Film B A H C
Factors affecting CT image RAD
Exposure Factors or Prime Factors
Photon Tissue Interactions
Appendix H: Chapter 40: Fluoroscopy
What is being tested? (kVp, mA, mAs, mR?) What is the purpose of the HVL test? If an x-ray tube does not pass the HVL test what is done to correct the.
Radiographic Quality Visibility and Sharpness
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Radiographic Equipment
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
Image Production and Evaluation NOTE: QA AND QC ARE USED INTERCHANGABLY IN APPLETON AND LANGE.
RAD 150 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic (motion)
RAD 354 Chapt. 25 Fluoroscopy Fluoro’s primary function is to enable one to view the patient in “real time” with dynamic motion.
Examples of Good & Bad Practice: Effect of the wedge filter and field size in the skin dose distribution L 9.1.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Doses in Fluoroscopy factors influencing patient doses Colin Martin and David Sutton.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Image Quality and Patient Dose. X-ray tube design - basic principles Electrons generated by thermionic emission from a heated filament (cathode) Accelerating.
RADIOGRAPHIC QUALITY S. Guilbaud, Education Director School of Radiologic Sciences.
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
Exposure Factors or Prime Factors
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Resident Physics Lectures
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Fluoroscopic Unit Thomas Edison 1896.
Resident Physics Lectures
Pediatric Fluoroscopy
Fluoroscopic Image Intensifier Image production
Pediatric Fluoroscopy
Pediatric Fluoroscopy
Fluoroscopy Static Image Recording Systems
Image Production and Evaluation
Fluoroscopic Image Quality Considerations
Fluoroscopic Image Intensifier Image Production
Fluoroscopy: Static Image Recording Systems
Fluoroscopic Image Quality Considerations
Generation of X-rays Q&A
Presentation transcript:

Fluoroscopy Q&A Robert L. Metzger, Ph.D.

RAPHEX Diagnostic Question 2001 D23: A fluoroscope is switched to the magnification mode (9” to 6” II). This will INCREASE all of the following EXCEPT: A. Conversion gain B. Quantum mottle C. Patient dose D. Spatial resolution

RAPHEX Answer to 2001 Question D23 B. Quantum mottle Conversion gain is related to brightness gain which is the product of mnification gain and flux gain. In the 6” mode, the minification gain decreases. Quantum mottle remains about the same. The automatic brightness control boosts the radiation level in magnification mode, to maintain brightness, and in doing so boosts the patient dose. Spatial resolution improves because of apparent magnification.

RAPHEX Diagnostic Question 2001 D30: The reason for overframing in cine filming is to: A. Improve spatial resolution B. Increase the field of view C. Reduce patient dose D. Change the shape of the cine image E. Increase the framing rate

RAPHEX Answer for 2001 Question D30 A. Improve spatial resolution The FoV is decreased and the patient dose is increased. Overframing has no clinical effect on shape or framing rate.

RAPHEX Diagnostic Question 2002 D36: The fluoroscopic operating factors displayed on a monitor are 120 kVp and 10 mA. Which of the following is true? A.The skin entrance dose is unusually low B. The five-minute timer is broken C. The skin entrance dose is extremely high D. The display must be wrong E. The anti-scatter grid is not in the beam

RAPHEX Answer to 2002 Question D36 C. The skin entrance dose is extremely high. As the fluoroscopic beam is positioned over thick or more dense areas of a patient, the penetration of x-rays decreases. The factors of 120 kVp and 10 mA indicate that the image intensifier is not receiving enough transmitted radiation through the patient. At such high technique factors the patient’s skin entrance exposure is extremely high. A broken timer will not affect the technique, and a missing grid would tend to decrease the factors.

RAPHEX Diagnostic Question 2002 D37: The maximum vertical resolution in 1p/mm for 23-cm diameter image intensifier coupled to a 1024 line TV system is _____. A. 0.8 B. 1.5 C. 2.2 D. 3.1 E. 4.0

RAPHEX Answer to 2002 Question D37 B. 1.5 Of the 1024 lines in standard TV only about 980 lines are used to actually trace out the image. It takes two lines to make a line pair, so there are 980/2 =4 490 line pairs. However, small objects are generally not perfectly aligned between the TV lines, so the effective resolution is obtained by multiplying by the Kell factor, which takes into account the random positioning of small objects in the TV field. The Kell factor is generally about 0.7, so the effective resolution is about 490 X 0.7 = 343 line pairs. For a 23-cm (230 mm) input, the resolution would be 343/230 =1.5 lp/mm.

RAPHEX Diagnostic Question 2003 D40: _____ is the persistence of an image on the fluoroscopic TV monitor. With motion, the image becomes blurred because of this persistence. A. Lag B. Vignetting C. Flare D. Veiling glare E. Blooming

RAPHEX Answer for 2003 Question D40 A. Lag Vignetting is the brightness change from the center to the edge of the II. The center is brighter than the edge. Flare and veiling glare are related parameters that indicate a degradation in contrast due to light scattering in the image intensifier. Blooming is the excessive brightness that occurs at the outside edge of a highly attenuation object.

RAPHEX Diagnostic Question 2003 D39: In cardiac catheter procedures, all of the following changes would reduce the patient’s dose EXCEPT: A. Reducing the frame rate B. Increasing the f number of the lens C. Increasing the aperture of the lens D. Increasing conversion gain of the II E. Increasing the source-to-image receptor distance (SID)

RAPHEX Answer for 2003 Question D39 B. Increasing the f number of the lens. Dose is directly proportional to frame rate. Large f-number lenses are less efficient at gathering light. Larger apertures permit more light to reach the cine film. A higher conversion gain produces more light for less radiation. Larger SIDs utilize geometry to reduce the patient’s radiation dose.