The MPPC Study for the GLD Calorimeter Readout Introduction Measurement of basic characteristics –Gain, Noise Rate, Cross-talk Measurement of uniformity.

Slides:



Advertisements
Similar presentations
Scintillator Tile Hadronic Calorimeter Prototype (analog or semidigital) M.Danilov ITEP(Moscow) CALICE Collaboration LCWS04, Paris.
Advertisements

New test result of MPPC M.Taguchi(kyoto).
1 Construction of the hadronic calorimeter prototype for ILC (CALICE collaboration) or experience with Geiger mode operating multipixel photodiodes (SiPM)
The Multi-Pixel Photon Counter for the GLD Calorimeter Readout Jul Satoru Uozumi University of Tsukuba, Japan 1.Introduction 2.Recent.
Study of the MPPC Performance - contents - Introduction Fundamental properties microscopic laser scan –check variation within a sensor Summary and plans.
Performance of MPPC using laser system Photon sensor KEK Niigata university, ILC calorimeter group Sayaka IBA, Hiroaki ONO, Paul.
Study of Photon Sensors using the Laser System 05/7/12 Niigata University, Japan Sayaka Iba, Editha P. Jacosalem, Hiroaki Ono, Noriko.
Michele Faucci Giannelli TILC09, Tsukuba, 18 April 2009 SiW Electromagnetic Calorimeter Testbeam results.
Introduction on SiPM devices
MPPC R&D status Kobe Univ. CALICE collaboration meeting Yuji SUDO Univ. of Tsukuba ~ contents ~ Introduction Linearity curve Recovery time.
Performance test of STS demonstrators Anton Lymanets 15 th CBM collaboration meeting, April 12 th, 2010.
GLD Calorimeter Status Oct 学術創成会議 S. Uozumi Shinshu University FJPPL meeting held at the end of September. Preparation underway toward the ECAL.
H.-G. Moser Max-Planck-Institut for Physics, Munich CALOR 06 Chicago June 5-9, 2006 Silicon Photomultiplier, a new device for low light level photon detection.
CALICE Meeting DESY ITEP&MEPhI status report on tile production and R&D activities Michael Danilov ITEP.
MPPC Radiation Hardness (gamma-ray & neutron) Satoru Uozumi, Kobe University for Toshinori Ikuno, Hideki Yamazaki, and all the ScECAL group Knowing radiation.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
SiPM: Development and Applications
Scintillation hodoscope with SiPM readout for the CLAS detector S. Stepanyan (JLAB) IEEE conference, Dresden, October 21, 2008.
The Scintillator ECAL Beam Test at FNAL Adil Khan International Linear Collider Workshop 2010 LCWS10 & ILC10, Beijing, China CALICE Scintillator ECAL group.
The Scintillator ECAL Beam Test at FNAL K. Kotera, Shinshu-u, 1st October 2009 CALICE Scintillator ECAL group; Kobe University, Kyungpook University, the.
Light Calibration System (LCS) Temperature & Voltage Dependence Option 2: Optical system Option 2: LED driver Calibration of the Hadronic Calorimeter Prototype.
Future Beam Test Plans of the GLD Calorimeter Aug 学術創成会議 Satoru Uozumi (Shinshu) for the GLD calorimeter group We are planning to have two beam.
The CMS Electromagnetic Calorimeter Roger Rusack The University of Minnesota On behalf of the CMS ECAL collaboration.
MPPC update including plastic connector T2K experiment collaboration meeting 2007/4/18 (Wed) S.Gomi T.Nakaya M.Yokoyama ( Kyoto University ) T.Nakadaira.
R&D of MPPC for T2K experiment PD07 : Photosensor Workshop /6/28 (Thu) S.Gomi T.Nakaya M.Yokoyama H.Kawamuko ( Kyoto University ) T.Nakadaira.
Study of the MPPC performance - R&D status for the GLD calorimeter readout – Nov 6-10.
Study of the Multi-Pixel Photon Counter for ILC calorimeter Satoru Uozumi (Kobe University) Atami Introduction of ILC and MPPC The MPPC performance.
MEG II 実験 液体キセノンガンマ線検出器に用いる 光検出器 MPPC の 実装に向けた最終試験 家城 佳 他 MEG II collaboration + 九大の方々.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK)
Study of the MPPC for the GLD Calorimeter readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group May 29 – Jun 4 DESY Introduction.
Fine Pixel CCD for ILC Vertex Detector ‘08 7/31 Y. Takubo (Tohoku U.) for ILC-FPCCD vertex group ILC vertex detector Fine Pixel CCD (FPCCD) Test-sample.
Scintillator ECAL for ILC for Calorimeter Review DESY May2007 Tohru Takeshita (CALICE-Shinshu) Idea implementation current status future mile stones.
1 Development of Multi-Pixel Photon Counters (1) S.Gomi, T.Nakaya, M.Yokoyama, M.Taguchi, (Kyoto University) T.Nakadaira, K.Yoshimura, (KEK) Oct
1 MPPC update S.Gomi, T.Nakaya, M.Yokoyama, M.Taguchi, (Kyoto University) T.Nakadaira (KEK) Nov KEK.
Scintillator-based ILC detector R & D Status and Milestone DongHee Kim Kyungpook National University Joint Korea-Japan Collaboration KPS meeting, Spring.
Development and Study of the Multi Pixel Photon Counter
GLD-CAL and MPPC Based on talks by T. Takeshita and H. K. Kawagoe / Kobe-U 2005-Sep-16 MPPC
Study of the MPPC for the GLD Calorimeter readout Satoru Uozumi (Shinshu University) Feb Beijing Introduction Basic performances Future.
R&D status of the Scintillator- strip based ECAL for the ILD Oct LCWS14 Belgrade Satoru Uozumi (KNU) For the CALICE collaboration Scintillator strips.
MPPC status M.Taguchi(kyoto) T2K ND /7/7.
Development of Multi-Pixel Photon Counters(MPPC) Makoto Taguchi Kyoto University.
Performance of SiPM Ryuhei Nakamura (Kobe Univ.) GLC calorimeter group (KEK, Kobe, Konan, Niigata, Shinshu, Tsukuba) Contents ・ Introduction ・ Test results.
Status of photon sensor study at Niigata University -- SiPM and MPPC -- Photon sensor mini workshop 05/9/16 University Niigata University.
Development of Multi-Pixel Photon Counters and readout electronics Makoto Taguchi High Energy Group.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group (KNU, Kobe, Niigata, Shinshu, ICEPP.
Multipixel Geiger mode photo-sensors (MRS APD’s) Yury Kudenko ISS meeting, KEK, 25 January 2006 INR, Moscow.
R&D of Calorimeter using Strip/Block Scintillators with SiPM
SiPM for CBM Michael Danilov ITEP(Moscow) Muon Detector and/or Preshower CBM Meeting ITEP
5-9 June 2006Erika Garutti - CALOR CALICE scintillator HCAL commissioning experience and test beam program Erika Garutti On behalf of the CALICE.
Systematics on ScECAL 1 st prototype DESY Apr , CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278.
F Don Lincoln, Fermilab f Fermilab/Boeing Test Results for HiSTE-VI Don Lincoln Fermi National Accelerator Laboratory.
Performance of new MPPC Nov. 21 Korea-Japan joint meeting Takashi Maeda Hideki Yamazaki Yuji Sudo (University of Tsukuba) --- Contents ---
Performance of Scintillator-Strip Electromagnetic Calorimeter for the ILC experiment Satoru Uozumi (Kobe University) for the CALICE collaboration Mar 12.
Jet Energy Measurement at ILC Separation of jet particles in the calorimeter is required for the PFA  Fine granular calorimeter is necessary. Particle.
The Multi-Pixel Photon Counter for the GLD Calorimeter Readout Jul Satoru Uozumi University of Tsukuba, Japan for the GLD Calorimeter.
Study and Development of the Multi-Pixel Photon Counter for the GLD Calorimeter Satoru Uozumi (Shinshu, Japan) on behalf of the GLD Calorimeter Group Oct-9.
Upgrade of the MEG liquid xenon calorimeter with VUV-light sensitive large area SiPMs Kei Ieki for the MEG-II collaboration 1 II.
Silicon Photomultiplier Development at GRAPES-3 K.C.Ravindran T.I.F.R, OOTY WAPP 2010 Worshop On behalf of GRAPES-3 Collaboration.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group Kobe Introduction Performance.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
M.Taguchi and T.Nobuhara(Kyoto) HPK MPPC(Multi Pixel Photon Counter) status T2K280m meeting.
Future Beam Test Plans of the Calorimeter Group Aug 学術創成会議 Satoru Uozumi (Shinshu) for the GLD calorimeter group We are planning to have two beam.
Performance of 1600-pixel MPPC for the GLD Calorimeter Readout Jan. 30(Tue.) Korea-Japan Joint Shinshu Univ. Takashi Maeda ( Univ. of Tsukuba)
Development of Multi-Pixel Photon Counters (1)
Progress report on SiPM development and its applications
ITEP&MEPhI status report on tile production and R&D activities
On behalf of CEPC calorimeter working group
R&D of MPPC for T2K experiment
R&D of MPPC in kyoto M.taguchi.
The MPPC Study for the GLD Calorimeter Readout
Presentation transcript:

The MPPC Study for the GLD Calorimeter Readout Introduction Measurement of basic characteristics –Gain, Noise Rate, Cross-talk Measurement of uniformity with microscopic laser Summary and plans 2006/10/31 Takashi Maeda Institute of Physics, University of Tsukuba for KEK-DTP photon sensor group for the GLD Calorimeter group

Sampling calorimeter with Pb/W - scintillator sandwich structure with WLSF readout Particle Flow Algorithm (PFA) needs particle separation in the calorimeter Fine granularity with strip/tile scintillators Huge number of readout channels –~10M (ECAL) + 4M (HCAL) ! Used inside 3 Tesla solenoid Need a new photon sensor which is compact and low-cost, can operate in a strong magnetic field GLD (Global Large Detector) Calorimeter … a candidate detector for ILC (International Linear Collider) particles readout absorber plate 1 cm x 5cm x 2 mm EM-scintillator-layer model

~ 1 mm 20~100  m Depletion region ~ 2  m ~ 8  m Substrate 1600 pixels 400 pixels substrate p + p-p- Guard ring n - Al conductor p+p+ n+n+ Si Resistor Bias voltage (~70V) The Multi-Pixel Photon Counter (MPPC) …novel photon sensor being developed by Hamamatsu Photonics (HPK)

Requirements for the GLD Calorimeter Gain: ~ at least 10 5, preferably 10 6 Dynamic range: up to ~1000 p.e. (need > 2500 pixels) – to measure EM shower maximum Single Photon Detection Efficiency: ~ 30 % – to identify MIP signals Noise rate : < 1 MHz (threshold = 0.5 p.e.) Good uniformity, small cross-talk Timing Resolution ~ 1 nsec Sensor area: 1.5 x 1.5 mm 2 – to place a larger number of pixels Should be stable against bias voltage / temperature / time

Characteristics of the 1600-pixel MPPC Evaluate performance as a function of bias voltage –Gain, Noise Rate, Cross-talk probability –Photon Detection Efficiency, Linearity (measurements still ongoing) Temperature dependence is also measured –MPPC performance is known to be sensitive to temperature Thermostatic Chamber Green LED MPPC

Gain measurement ・ 30 ℃ ・ 25 ℃ ・ 20 ℃ ・ 15 ℃ ・ 10 ℃ ・ 0 ℃ ・ - 20 ℃ d S : ADC sensitivity = 0.25 pC/ADCcount A : Amp gain = 63 e : electron charge = 1.6 x C C : Pixel capacitance V 0 : Geiger-mode starting voltage Pedestal 1 pix. fired 2 pix. fired 70V, 20 ℃

C, V 0 vs. Temperature V 0 = aT +b C looks not sensitive to temperature, at least under < 20 o C V 0 is linear to temperature a = (5.67 ± 0.03) x10 -2 V/ o C b = 66.2 ± 0.1 V V 0 =aT+b

Noise Rate … rate of avalanche signals induced by thermal electrons V bias – V 0 (T) [V] ・ 30 ℃ ・ 25 ℃ ・ 20 ℃ ・ 15 ℃ ・ 10 ℃ ・ 0 ℃ ・ - 20 ℃ Lower temperature  Lower noise rate 1MHz

Cross-talk Cross-talk probability looks stable with temperature in V bias – V 0 < 2.5V. The cross-talk to adjacent pixels is caused by photons created in an avalanche. Cross-talk probability is measured from dark noise rates : ・ 30 ℃ ・ 25 ℃ ・ 20 ℃ ・ 15 ℃ ・ 10 ℃ ・ 0 ℃ ・ - 20 ℃ V bias – V 0 (T) [V]

Using a microscopic laser system we perform scan within a pixel pixel-by-pixel scan to see the variation of Gain Hit probability Cross-talk 1 pixel Measurement of uniformity in the sensor

Measurement with Microscopic Laser System 1600 pixel MPPC Introduced by KEK-DTP YAG Laser, = 532 nm (green) Pulse width ~ 2 nsec, rate ~ 8 kHz Spot size ~ 1  m Light yield ~ 0.5 p.e. (not calibrated) Can perform precise pinpoint scan with the well-focused laser ~25  m Laser spot

Hit fraction vs. Bias Voltage Inject laser to center of a pixel. The hit fraction depends on bias voltage, but is stabilized in V bias > 70 V. Pedestal 1 pix. fired 2 pix. fired (cross-talk) Hit fraction

Uniformity within a Pixel Fraction of sensitive region ~ 20% Variation within a sensitive region ~9.2% (RMS) 1 pixel The shape of sensitive region is not changed with bias voltage Bias voltage ・ -71.0V ・ -70.0V ・ -69.5V ・ -69.0V Hit probability

Gain Uniformity within a Pixel Higher gain in central part Gain variation in a sensitive region ~ 2.7% (RMS) Gain (x10 5 ) y-point (1  m pitch) x-point (1  m pitch) Edge of the sensor V bias = 70.0 V

Cross-talk Variation within a Pixel Shape of the cross-talk probability depends on bias voltage Edge part shows larger cross-talk Bias voltage ・ -71.0V ・ -70.0V ・ -69.5V ・ -69.0V Sensitive region in a pixel Pedestal 1 pix. fired 2 pix. fired (cross-talk)

Pixel-by-pixel Scan - Hit fraction edge of the sensor Variation ~3.2% 20 x 20 pixels Sensor

Pixel-by-pixel Scan - Gain 3.2(x10 5 ) 3.8 (x10 5 ) edge of the sensor Edge pixels have higher gain Strange structure is seen, reason unknown Variation ~2.4%

Summary We are evaluating the MPPC performance from viewpoint of the GLD calorimeter readout use –Gain, Noise rate, Cross-talk are acceptable The MPPC properties are sensitive to V bias -V 0 (T) and temperature –Lower Noise rate and Cross-talk with lower temperature The MPPC properties are observed to be uniform within a sensor. Measure photon detection efficiency and Linearity Perform same measurements for new MPPC samples and evaluate device-by-device variation (We just have been provided new samples by HPK) Plans