CARB Continuous PM 2.5 Monitoring Activity (BAM’s) Reggie Smith California Air Resources Board
Continuous PM2.5 Monitor Network ARB Ambient Air Monitoring Sites –13 Sites with BAM-1020 Monitors (17 units) –3 Special Purpose Sites (6 units) –61 total samplers 4 Collocated BAM-1020 Sites 9 Sites with FRM and BAM
Standard CARB BAM Configuration PM 10 Sampling Inlet (BX-802) Sharp Cut Cyclone (BX-807) Smart Heater Met Translator Option (BX-964) Firmware v2.58 Gast Rotary Pump (BX-121) Volumetric Flow Control
Data Analysis SLOPE: Indicator of the bias of the BAM compared to the FRM INTERCEPT: Offset between the BAM and FRM r 2: Correlation coefficient of the linear regression between the two instruments
BAM-1020 Accuracy
BAM-1020 Precision
Network Summary Overall BAM Network Performance –All BAM/FRM data regressed as one data set –Slope of –Intercept of 3.07 –r 2 of 0.940
BAM Maintenance Bi-Weekly Leak and Flow Checks are absolutely necessary to ensure proper operation Failed leak and flow checks are associated with poor FRM correlation in ARB’s experience Adhere to AQSB SOP 400 (BAM Standard Operating Procedures)
CARB Zero Test Sample Zero air using BX-302 HEPA Filter 76 consecutive hours (First 4 hours removed) Background set to Min/Max Zero test range of 14 ug/m 3 No value exceeding 20 ug/m 3 Any errors during test, re-run test Field zero test annually (24 hour test)
BAM Datalogger Review Review of the internal datalogger is important Reveals offsets between the BAM and the external data acquisition system Review of Q tot (total volume sampled) hourly values helps detect flow problems Detailed error codes are stored in the internal datalogger data
Conclusions The BAM-1020 Monitor is well-suited for PM-AQI, Prescribed Fire and Ag-Burn Forecasting, diurnal profiling, quantifying short term events, and characterizing atmospheric dynamics Is it good enough for regulatory determinations?
Next Steps Evaluate data using U.S. EPA’s Criteria Designation of Equivalence Methods for Continuous Surveillance of PM 2.5 Determine how good is good? –Slope: /- 5% –Intercept: 3 ug –Correlation: r 2 =.940 (r =.97)
Summary Contact Information –Voice: (916) – CARB BAM Website –