Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20 MAMMOGRAPHY QA Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20 PURPOSE OF QC ASSURES OPTIMUM PATIENT CARE by avoiding unnecessary repeat exposures due to artifacts, processing problems and equipment failure Rev. Spring 2009/2010
EQUIPMENT DESIGN & OPERATION STANDARDS (DEDICATED UNITS) MQSA National Statistics FEDERAL LEGISLATION OCT. 1, 1994 EQUIPMENT DESIGN & OPERATION STANDARDS (DEDICATED UNITS) INTENSIVE QUALITY CONTROL PROGRAM Certification statistics, as of March 1, 2010 12,623 Certified facilities with FFDM2 units / Accredited FFDM units 5,402 / 7,963 Total annual mammography procedures reported, as of March 1, 20101 38,057,377 Percent of inspections with no violation 80.2% WHAT DOES MQSA REPRESENT? __________________ 1 This number is an aggregate of the total number of procedures performed annually as reported by facilities to
MAMMO & Quality Control Quality control is an equipment and processing evaluation program in diagnostic imaging centers. QC involves testing, record keeping and evaluation of imaging equipment and processing QC Team includes Radiographer/Mammographer, Physicists and Radiologist QA involves people – scheduling, reports, waiting time and follow up exams “double reads” See Bushong – CH. ____ PG ____
Elements of QC pg 333 FREQUENCY TASKS Daily Weekly Monthly/Quarterly Darkroom Clean Processor QC – sensitometry/densitometry and DEVELOPER temp Screen Cleaning VISUAL CHECK Phantom Images Repeat Analysis (should be less than 2% repeat rate - diagnostic rad = 5%)
MAMMO QC MONTHLY Checks SEMIANNUALLY DARKROOM FOG SCREEN/FILM CONTACT COMPRESSION NOT MORE THAN 40 PSI (25 PSI is average)
QUALITY CONTROL TOOLS PHANTOM DIGITAL THERMOMETER SCREEN CLEANER LINT FREE CLOTHS FIXER RETENTION TESTING KIT CHARTS
Hypo Retention Checks for what? Fixer left on film – “Archival Properties”
WIRE MESH Checks for what How long should you wait after loading a cassette to use it? 15 min & placed in an upright position
Check for leaks/cracks? Types of Safelights? Wratten 6b GBX Check for leaks/cracks? How often? Bulb Wattage?
For those still using film… clean crossovers daily Use “dedicated” processor for only mammo
Not calcifications in the breast VISUAL CHECKLIST MONTHLY FIXER RETENTION REPEAT ANALYSIS DARKROOM FOG SEMI-ANNUAL FILM/SCREEN CONTACT COMPRESSION (25 PSI) These specks were produced by flakes trapped between the film and the screen – Not calcifications in the breast
PHANTOM IMAGES 4/5 masses 3/4specks 4/6 fibers should be seen WEEKLY - to assure film density, contrast, uniformity and image quality are optimum – Using the SAME cassette and technical factors each time 4/5 masses 3/4specks 4/6 fibers should be seen (see pg 500 review book)
REPEAT ANALYSIS Mammography should be less than 2% Number of repeated film Repeat Rate = --------------------------- X 100 Total # of films Diagnostic Radiology = less than 5% Students = less than 10 %