ROTATING SCHEIMPFLUG TOPOGRAPHIC PARAMETERS IMPORTANT IN DISTINGUISHING NORMAL FROM KERATOCONIC CORNEAL MORPHOLOGICAL FEATURES Clayton Falknor, MD, Orkun.

Slides:



Advertisements
Similar presentations
Aberrometry and the Tear Film — Understanding new methods —
Advertisements

Placido Based Corneal Topography
Asymmetric Keratoconus Study: Novel Tomographic Parameters to enhance sensitivity to detect abnormalities in eyes with normal axial curvature maps from.
TOPOGRAPHY ORBSCAN S.A.A. Mortazavi MD.
Ömür Uçakhan-Gündüz, MD Betül N. Bayraktutar, MD
Corneal topography orbscan
Swept Source Optical Coherence Tomography for Evaluation of Posterior Corneal Changes after Refractive Surgery Dr. Tommy Chung Yan Chan Dr. Vishal Jhanji.
Comparison of corneal astigmatism measured with 3 devices Mariko Shirayama, M.D, Li Wang, M.D, PhD, Mitchell P. Weikert, M.D, Douglas D. Koch, M.D. Cullen.
Correlation Of Corneal Contour With Higher Order Ocular Aberrations (HOA) In Indian Eyes Manish Chaudhary MS Amit Gupta MS Advanced Eye Centre, Postgraduate.
Accelerated CXL Combined with Same Day PTK and Single Intrastromal Ring Segment Implantation for Keratoconus Yakov Goldich, Noa Avni-Zauberman , Randall.
Agreement Between Corneal Topography with the Dual-Scheimpflug Placido Corneal Topographer, Scheimpflug Corneal Topographer, and Scanning-Slit Topography.
Corneal Biomechanics, Pachymetric Progression Profile and Corneal Volume: Indices for Detecting Ectasia and Screening Refractive Candidates Renato Ambrósio.
Purpose We studied the interdevice agreement of three systems in the different measurements of the corneal and anterior segment biometric parameters. The.
NORMAL CORNEA The highest diopter of human eye,about 43D at corneal apex Average radius curvature of 7.8 mm Is not absolutely transparent,scatters about.
CORNEAL TOPOGRAPHY Wide acceptance as a clinical examination procedure with the advent of modern laser refractive surgery Measure a greater area of the.
Corneal Iron Ring After Hyperopic LASIK
Application of Quantitative Method for Prospectively Identifying Risk for Ectasia After LASIK in Normal, Suspect, and Ectatic Eyes Mujtaba A. Qazi, MD.
David Zadok MD, Isaac Avni MD
E-Poster ALC Canedo, MD 2010 Rio de Janeiro CornealTomograhyandBiomechanics Study Group Study Group The impact of reference body selection for calculating.
Dr. Amin Hossein Rahgozar
Repeatability of Anterior and Posterior Corneal Higher- Order Aberrations in 4 mm, 6 mm and 8 mm diameters measured by Pentacam System Jadwiga Wojtowicz.
Dept. of Ophthalmology, Friedrich-Alexander University Erlangen-Nuremberg, Erlangen, Germany. Outcome after big-bubble deep anterior lamellar keratoplasty.
Financial Disclosure: Authors do not have financial or commercial interest in mentioned equipments Dr Arce is Consultant & Territory Manager for Latin.
Medical Director - R & D Consultant, Ziemer Group AG, Port, Switzerland Speaker, Bausch & Lomb do Brasil Territory Manager.
Evaluation of Corneal Parameters and Spherical Aberration After DSAEK Measured with Pentacam System Orkun Muftuoglu, Pawan Prasher, R. Wayne Bowman, Steven.
TRANSPALPEBRAL TONOMETER APPLICATION DURING INTRAOCULAR PRESSURE EVALUATION IN THE PATIENTS WITH REFRACTION ANOMALY BEFORE AND AFTER KERATOPHOTOREFRACTIVE.
Keratoconus Dr. Abdullah S. Al Yousef. Definition A non-inflammatory eye condition in which the normally round dome-shaped cornea progressively thins.
Evaluation of Interface Reflectivity and Corneal Aberrations following DSAEK Hamid Khakshoor, MD Professor of Ophthalmology Mashhad University of Medical.
Characterization of Higher Order Ocular Aberrations (HOA) In ‘Normal’ Versus Myopic Eyes To Study The Effect Of Myopia On Higher Order Aberrations Amit.
Evaluation of Zonal Equivalent Keratometry Readings After LASIK Timmy Kovoor, MD Orkun Muftuoglu, MD V.Vinod Mootha, MD Steven Verity, MD R. Wayne Bowman,
ASCRS 08 Changes Of Higher Order Aberrations After Excimer Laser Treatment For Moderate Myopia by Means of Preoperative Wavefront Aberration Levels using.
The authors have no financial interest in the subject matter of this poster Yinfei Xu, Peter S. Hersh, MD, David S. Chu, MD Institutional Affiliations.
Prospective Study of Intrastromal Corneal Ring Segments for Keratoconus Scott X. Stevens, MD Bend Ophthalmology, LLC Author has no financial interest.
Corneal Topography using Orbscan : Basics and interpretation
World Cornea Congress VI Michael K. Smolek & Panagiotis Kompotiatis Department of Ophthalmology LSU Eye Center of Excellence, New Orleans The authors.
Evaluation of Long-Term Keratoconus Progression Using Scanning-Slit Topography First Author: David Galarreta, MD Contributing Authors: Nestor Cortes, MD,
Rio de Janeiro Corneal Tomography and Biomechanics Study Group Dr. Ambrósio is consultant for Oculus Optikgeräte GmbH (Wetzlar, Germany) Agreement of Artificial.
ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI ASCRS Boston April Vision Sana Research Group Vision Sana Research Group.
Copyright restrictions may apply JAMA Ophthalmology Journal Club Slides: Thickness and Asymmetry After DSAEK Dickman MM, Cheng YYY, Berendschot TTJM, van.
Comparison of Central Corneal Thickness and Peripheral Corneal Thickness using Sheimpflug system, Optical Coherence Tomography and Ultrasound Pachymetry.
Analysis of Ocular Wavefront Aberrations in Post Penetrating Keratoplasty Eyes with Two Different Hartmann-Shack Aberrometers Adriana S. Forseto 1, MD;
The Effect of Corneal Anterior Surface Eccentricity on Astigmatism after Cataract Surgery Choul Yong Park MD 1 Sung Jun Lee MD 1 Prabjot Channa MD 2 Roy.
Scheimpflug and OCT analysis of posterior keratoconus
1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro 2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli Scheimpflug topography, Scanning-Slit.
Evaluation of Efficacy and Safety of Intracorneal Ring Segment (Intacs SK) in Keratoconus Abdulrahman Al-Muammar, MD, FRCSC I have no financial interest.
Evaluation of Corneal Tomography in Primary Pterygium Sahil Goel, MD (Presenting Author), Murugesan Vanathi MD *The authors have no financial interests.
Ocular functional optical zone following hyperopic LASIK/PRK: Analysis based on polychromatic retinal image quality Mitchell P. Weikert, MD Li Wang, MD,
M. Escaf, J. Mendoza, L. Izquierdo jr., M. Henriquez Instituto de Ojos Oftalmo Salud Lima, Peru There is no financial interest in this poster.
Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD Noo Ne Eye Hospital, Seoul,
Ocular Response Analyzer Waveform Analysis in the Ectatic Corneas: Correlation of the New Corneal Biomechanics Parameters and Severity of Keratoconus Kristin.
Financial Disclosure: Medical Director–Galilei R&D Consultant, Ziemer Group AG, Port, Switzerland Consultant & Territory Manager for.
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
Authors: Leonidas Traipe Magdalena Rios Edison Villagra Allister Gibbons Marcelo Hidalgo Claudia Goya Authors have no financial interest.
Wavefront-guided Ablation Retreatment in Myopic Eyes Engy M Mohamed MD, Orkun Muftuoglu MD, R.Wayne Bowman MD, V. Vinod Mootha MD, H. Dwight Cavanagh MD,PhD,
Risk Factor Analysis of Topographic Progression in Keratoconus Seong Joon Ahn, MD 1,2, Mee Kum Kim MD, PhD 1,2, Won Ryang Wee, MD, PhD 1,2 1 Department.
Corneal shape and corneal aberrations after MicroIncision Cataract Surgery (MICS) NOCHEZ Y, BUREL B, MAJZOUB S, PISELLA PJ C.H.U.
CORNEAL EPITHELIAL IMAGING
17100 Contact lens fitting after corneal CXL
Corneal High-Order Aberrations in Keratoconus
LSU Eye Center, New Orleans, LA
Comparison of Single-Scheimpflug and Dual-Scheimpflug Pachymetry Measurements and Effect of Decentration World Cornea Congress VI April 7-9, 2010.
LASIK vs ASA Case Studies
MP Weikert, M Shirayama, L Wang, DD Koch
Mehdi Ghajarnia, MD Mark Mifflin, MD John A. Moran Eye Center
Comparison of corneal powers obtained from four different devices
Carlos G. Arce, MD Associate Researcher and Ophthalmologist
Introduction and Overview
Sun Woong Kim, M.D.1, Hae Jung Sun, M.D.1,
Dr Ciara Bergin, PhD Dr Ivo Guber, Dr Kate Hashemi, Dr Francois Majo
Age-related difference in Lasik Treatment
Presentation transcript:

ROTATING SCHEIMPFLUG TOPOGRAPHIC PARAMETERS IMPORTANT IN DISTINGUISHING NORMAL FROM KERATOCONIC CORNEAL MORPHOLOGICAL FEATURES Clayton Falknor, MD, Orkun Muftuoglu, MD, Steven Verity, MD, James P. McCulley, MD Some of the authors have received consultant reimbursement from Alcon Labs, Inc. None of the authors have financial interest in the subject matter of this poster.

Keratoconus  Essential to identify keratoconus prior to keratorefractive surgery  Keratoconus is characterized by:  Non-inflammatory, progressive corneal disease  Stromal thinning and anterior bulging of cornea  Irregular astigmatism and myopia  Potentially severe corneal scarring  Keratoconus is identified by:  Fleisher ring (corneal iron line at base of cone)  Vogt striae (stromal stress lines within cone)  Scissoring of retinoscopic reflex  Apical scarring and/or subepithelial fibrosis  Central or paracentral steepening on topography  Subclinical (forme fruste) keratoconus is difficult to identify

Identification of keratoconus  Traditional method to identify subclinical keratoconus is corneal topography  Placido disk-based  Measures slopes of anterior corneal surface only  Axial curvature method subject to misalignment of corneal apex and corneal sighting point May lead to misleading maps, eg normal eyes may show asymmetric bow-tie or inferior steepening  Contribution of posterior corneal surface important  Orbscan (Bausch and Lomb, Salt Lake City, Utah, USA) Slit-scanning beam combined with Placido ring technology Posterior surface recreated with triangulation algorithms Not all images include central cornea  Pentacam (Oculus Optikgeraete GmbH, Germany)

Pentacam Comprehensive Eye Scanner  Rotating Scheimpflug camera  Monochromatic slit light source rotates with camera  slit images per acquisition  Eye movement monitoring by 2 nd camera  Less than 0.6mm decentration  Rotates 180º in 2 seconds  All images include central cornea  Corneal elevation data independent of visual axis and corneal apex acam_aao_2006.pdf

Purpose  Evaluate parameters obtained by Pentacam important in distinguishing keratoconus from normal  Pentacam parameters to detect keratoconus  Pachymetry  Progression index of corneal thinning  Corneal volume within fixed diameter  Simulated keratometry  AC volume, depth and angle  Posterior elevation over best-fit sphere (option of toric ellipsoid)  Zernike HOA of anterior and posterior surfaces  Corneal variance indices ISV (index of surface variance) IVA (index of vertical asymmetry) IHA (index of height asymmetry) IHD (index of height decentration) Rmin (radius minimum) KI (keratoconus index) CKI (center keratoconus index)

Patients  Keratoconus (Diagnosed clinically with topography support)  108 eyes of 54 patients (34 men, 20 women)  Inclusion: distorted keratometry mires, abnormal retinoscopic reflex, Vogt’s striae, Fleischer’s ring, corneal scarring, available topography maps  Exclusion: prior corneal surgery, extensive corneal scarring  Controls (normals presenting for keratorefractive surgery)  72 eyes of 36 patients (16 men, 20 women)  Inclusion: underwent pre-operative screening for keratorefractive Sx, normal corneal exam, available topography maps  Exclusion: prior ocular surgery or trauma, suspicion for keratoconus or pellucid marginal degeneration by topography  Age-matched  Keratoconus group 36.4 ± 11  Control group 43 ± 14

Posterior corneal elevation  Mean posterior elevation  Keratoconus 98.7 ± 46.3 µm  Control 11.8 ± 6.12 µm  Difference significant (p<0.001) With Orbscan IIz, posterior elevation optimal cutoff point to discriminate keratoconus and keratoconus suspect versus normal corneas was 40 µm (Rao et al & Fam et al)

Pentacam keratoconus parameters

Corneal variance indices

Summary of results  Mean posterior corneal elevation significantly higher in keratoconus compared to controls  For cut-off of 35 µm, sensitivity 93% & specificity 95%, comparable to Orbscan  Progression index minimum, average, and maximum all significantly different in keratoconus vs. controls  Other significant parameters:  All corneal variance parameters (all based on anterior surface) Most sensitive: ISV, Abr, IVA, KI, IHD Most specific: CKI, KI, IVA  Pachymetry at pupil center and thinnest, flat and steep keratometry, AC depth, corneal volume of central 7mm diameter  Not significantly different:  Keratometry axis, AC volume, AC angle

Zernike analysis  Both anterior and posterior elevation data decomposed into Zernike higher-order aberration polynomials  Real differences between keratoconus and controls within the third through sixth orders  Trefoil, coma, fourth-order astigmatism, spherical aberration all differ both anteriorly and posteriorly  For both anterior and posterior surfaces, vertical coma most important HOA

Conclusions  Pentacam is useful for identifying keratoconus  Both anterior and posterior corneal surface parameters are important  Advantages over other instruments  Measures central corneal zone  Elevation data independent of reference axis  Zernike analysis of anterior and posterior corneal HOA