Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.

Slides:



Advertisements
Similar presentations
Topic 8. Gamma Camera (II)
Advertisements

Instruments for Radiation Detection and Measurement Lab # 4.
Chapter 8 Planar Scintigaraphy
Fysisk institutt - Rikshospitalet 1. 2 Overview Gamma camera Positron emission technology (PET) Computer tomography (CT) Proton therapy Electrical impedance.
1. 2  Ray Imaging u Inject a pharmaceutical labelled with a radioactive nuclide  Detect where  -rays are coming from u Local concentration in patient.
Nuclear Medicine Spring 2009 FINAL. 2 NM Team Nuclear medicine MD Nuclear medicine MD Physicist Physicist Pharmacist Pharmacist Technologist Technologist.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Gamma Camera Technology
Computed Tomography RAD309
GAMMA RAY SPECTROSCOPY
Main detector types Multi Pixel Photon Counter (MPPC) and Charge Coupled Devices (CCDs) How does it work? 1. Photon hits a pixel producing electron hole.
Radiation Safety level 5 Frits Pleiter 02/07/2015radiation safety - level 51.
Lecture 2-Building a Detector George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
Instruments for Radiation Detection and Measurement
Planar scintigraphy produces two-dimensional images of three dimensional objects. It is handicapped by the superposition of active and nonactive layers.
Photomultiplier Tube. What is it? Extremely sensitive detector of light in the ultraviolet, visible and near infrared Multiplies the signal produced by.
Gamma Spectroscopy HPT TVAN Technical Training
Lecture 11  Production of Positron Emitters, Continued  The Positron Tomograph.
Instruments for Radiation Detection and Measurement Lab # 3 (1)
Lens ALens B Avg. Angular Resolution Best Angular Resolution (deg) Worst Angular Resolution (deg) Image Surface Area (mm 2 )
1 Light Collection  Once light is produced in a scintillator it must collected, transported, and coupled to some device that can convert it into an electrical.
Photon detection Visible or near-visible wavelengths
Innovation is in our genes. 1 Siemens Medical Solutions Molecular Imaging What are SPECT basics?
SCINTILLATION COUNTER. PRINCIPLE When light radiations strike fluorescent material it produces flashes of light called scintillations. These are detected.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
Techniques for Nuclear and Particle Physics Experiments By W.R. Leo Chapter Eight:
Introduction to nuclear medicine technology NMT 231 Aya Ahmed Saeed.
Nuclear Medicine Principles & Technology_I
Lecture 3-Building a Detector (cont’d) George K. Parks Space Sciences Laboratory UC Berkeley, Berkeley, CA.
1 Nuclear Medicine SPECT and PET. 2 a good book! SR Cherry, JA Sorenson, ME Phelps Physics in Nuclear Medicine Saunders, 2012.
CAT scanners and gamma cameras Unit 16 Waves. Learning objectives Describe how CAT scanners can produce a much more detailed image than conventional X-rays.
Chapter V Radiation Detectors.
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Lecture 12  Last Week Summary  Sources of Image Degradation  Quality Control  The Sinogram  Introduction to Image Processing.
Acquisition time6 min1 min 12 s Collimator height25 mm (Anger)12 mm (HiSens) Detector1 layer, 1 pixel / hole3 layers, 1 pixel / hole3 layers, 4 pixels.
Nuclear Medicine Physics and Equipment 243 RAD 1 Dr. Abdo Mansour Assistant Professor of radiology
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
DIGITAL RADIOGRAPHY.
Chapter-2 The Planar Imaging Gamma Camera
Introduction to Gamma Camera
Chapter-2 The Planar Imaging Important points in chapter 2 (chapter 13 from the book) The gamma camera (the basic principles of the gamma camera) The types.
Nuclear Medicine Introduction
PET Imaging Positron Emission Tomography
Muhammad Musaddiq.
Chapter-5 Positron emission tomography (PET)
Gamma Camera Technology
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
Nuclear Medicine Physics
Image quality and Performance Characteristics
Image quality and Performance Characteristics
Gamma Camera.
PAN-2013: Radiation detectors
Lecture 1: Development of the Analog Gamma Camera
Instrument Parameters in WDXRF
Application of Nuclear Physics
Scintillation Counter
When this happens... We see this... 12/6/2018 Steve Wotton.
Lecture 1—Basic Principles of Scintillation
Oct 10, 2018 Muhammad Qasim Abdul Wali Khan University, Department of Physics, Mardan, Pakistan by.
Gamma Camera & Basic Principles
Function and Structure in
Gamma Camera Ilker Ozsahin Oct
HE instrument and in-orbit performance
3/9/2019 L13-L14 1 Instruments for Radiation Detection & Measurement PRINCE SATTAM BIN ABDUL AZIZ UNIVERSITY COLLEGE OF PHARMACY Nuclear Pharmacy (PHT.
Radioactivity B. Sc. -III Feb Dr. Wagh G. S. M. Sc. M. Phil. Ph
Photomultiplier (PMT) Tubes
Assist. Prof. Dr. Ilker Ozsahin Oct
X-ray detectors single photon detectors
PHYS 3446 – Lecture #17 Wednesday ,April 4, 2012 Dr. Brandt
Presentation transcript:

Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist

Methods of Analysis  Once tracer has traced – need some method of analysing distribution Imaging Gamma Camera, PET Camera Compartmental Analysis Sample Counter

Radiation Detectors  Converts incident photon into electronic signal  Most commonly used detectors are scintillation  Photon interacts with crystal to convert incident photon into light photons  PMT changes light into electrical signal  Electrical signal recorded and analysed

Imaging Equipment  The Gamma Camera  Basic principle hasn’t changed since 1956!

Scintillation Imaging  Administration of Isotope

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake

Scintillation Imaging  Localisation and Uptake Enhanced contrast between Organ of Interest and rest of body

Scintillation Imaging  Imaging distribution Gamma-rays emitted by radiopharmaceutical Collimator ‘selects’ only those rays travelling at right angles to face of camera Scintillation events in crystal recorded

Early Scintillation Study

Components of a Modern Gamma Camera  The components of a modern gamma camera Lead Shield Collimator Lightguide PMTs Electronics Crystal

The Collimator  The collimator consists of:  a lead plate  array of holes  It selects the direction of the photons incident on the crystal  It defines the geometrical field of view of the camera

The Collimator  In the absence of collimation:  no positional relationship between source – destination  In the presence of collimation:  all γ-rays are excluded except for those travelling parallel to the holes axis – true image formation Patient Detector

Types of Collimators  Several types of collimator:  Parallel-Hole  Converging  Diverging  Pin-Hole

Energy Ranges of Collimators Type of Collimator Energy Range Typical Nuclide Low Energy (LE) keV Tc-99mTl-201 Medium Energy (ME) keV In-111 High Energy (HE) 300 – 400 keV I-131

The Scintillation Crystal  First step of image formation  Photon detected by its interaction in the crystal  γ-rays converted into scintillations

Scintillation  Can be thought of as “partial ionisation”  Electrons excited and gain energy  As electrons fall back to ground state, photons emitted  Use of doping (eg NaI:Tl) creates smaller gaps

Scintillation Crystal Properties   High stopping efficiency   Stopping should be without scatter   High conversion of γ-ray energy into visible light   Wavelength of light should match response of PMTs   Crystal should be transparent to emitted light   Crystal should be mechanically robust   Thickness of scintillator should be short

Properties of NaI(Tl) Scintillator   The crystal – NaI(Tl)   emits light at 415 nm   high attenuation coefficient   intrinsic efficiency: 90% at 140 keV   conversion efficiency: 10-15%   energy resolution: keV at 150 keV

Disadvantages of NaI(Tl) crystal  NaI(Tl) crystal suffers from the following drawbacks:  Expensive (~£50,000 +)  Fragile  sensitive against mechanical stresses  sensitive against temperature changes  Hygroscopic  encapsulated in aluminium case

Lightguide and Optical Coupling  Lightguide acts as optical coupler  Quartz doped plexiglass (transparent plastic)  The lightguide should:  be as thin as possible  match the refractive index of the scintillation crystal  Silicone grease to couple lightguide, crystal and PMT  No air bubbles trapped in the grease

The Photomultiplier Tube  A PMT is an evacuated glass envelope  It consists of:  a photocathode  an anode  ~ 10 dynodes

The Photomultiplier Tube  Photocathode of PMT emits 1 photoelectron per ~ 5 – 10 photons  Photoelectron accelerated towards first dynode  Dynode emits 3 – 4 secondary e - per photoelectron  Secondary e - accelerated towards next dynode  Multiplication factor ~ 10 6  Output of each PMT proportional to the number of light photons

PMT Properties  The photocathode should  be matched to blue light  have high quantum efficiency  High stability voltage supply: ~1kV

Positional and Energy Co-ordinates  PMT signals processed  spatial information – X and Y signals  energy information – Z signal  Z signal – the sum of the outputs of all PMTs  proportional to the total light output of the crystal  Light output proportional to the energy of incident gamma  Pulse height analyser accepts or rejects the pulse

Pulse Height Analysis  Z-signal goes to PHA  PHA checks the energy of the γ-ray  If Z-signal acceptable  γ-ray is detected  position determined by X and Y signals  20% window still includes 30% of scattered photons

Determining the Position of Events

Image Acquisition Techniques  Static-(Bones, Lungs)  Dynamic-(Renography)  Gated-(Cardiac)  Tomography  SPECT  PET  List Mode-(Cardiac)

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display 1

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display 1 1

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display 1 1 1

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display

Static Imaging  Camera FOV divided into regular matrix of pixels  Each pixel stores number of gamma rays detected at corresponding location on detector  Typical Matrix Sizes: 256 2, 128 2, 64 2 CameraComputer MemoryImage Display

Dynamic Imaging  Series of sequential static frames  E.g. 90 frames each of 20s duration  Image rapidly changing distribution of activity within the patient  Used in Renography

Dynamic Imaging Analysis ROIs Curves showing changing renal activity over time Split Renal Function

Gated Imaging  Several frames acquired covering the cardiac cycle  Acquired over many cycles