Argon Enhanced Electrosurgery
Characteristics of argon gas (helium, neon, freon etc. also possible, but.....) inertnon-explosive heavier than air not expensive
Why Argon gas ? Argon gas combined with an activated high-voltage electrode ionises and becomes a conductor of energy Argon gas creates an oxygen free environment reducing tissue necrosis and generating less smoke
Argon Features Creates uniform conductive plasma bridge between electrode and tissue Conventional electrosurgeryArgon-enhanced electrosurgery
Argon cutting improved spark rain dry cutting edge less smoke
Argon cutting conventional cutting argon enhanced cutting
Clinical Benefits Reduced smoke Improved visualization Conventional electrosurgery Argon-enhanced electrosurgery
Argon fulguration direct coagulation to exposed tissue better identification of bleeders cooling of tissue
Argon fulguration via controlled spark rain conventional fulguration Argon fulguration
Clinical Benefits Flexible, thinner eschar –Less charring of tissue –Less rebleeding Conventional electrosurgery Argon-enhanced electrosurgery
Clinical Benefits Directable flow of argon gas which clears surgical site of fluids and smokeDirectable flow of argon gas which clears surgical site of fluids and smoke –Establish direct coagulation effect onto exposed tissue Not on top of blood –Reduced blood loss Rapid identification of bleeders
Conventionalelectrosurgery Argon enhanced electrosurgery
Argon gas faster visualisation of bleeders better tissue selective dehydration Reduction in blood loss: less necrosis improved vessel sealing Reduction in post-operative infections:
Less necrosis: spray coagulation: more superficial better control on size of coagulation area less tissue adhesion to electrode less smoke
Clinical Benefits Efficient noncontact coagulation –Less tissue adhesion to electrode –Ability to coagulate directly to bone surface –Ability to better coagulate adipose tissue
Clinical Benefits Time saving –Reduced blood loss –Coagulate large areas much faster –More controlled delivery of energy
Hardware & Accessories
ES Generator Force Argon ll Cable / Interface connections Gas Regulator Interface Cable Argon to Generator (5 & 4 pin) Serial Port connection from Argon regulator (2 if 2 bottles used)
Peritoneal Pressure Tubing –Introducing argon into the peritoneal cavity will increase the intra- peritoneal pressure. –The intra-peritoneal pressure can be monitored by the V ALLEYLAB F ORCE A RGON ™ II System when operated in its low flow (laparoscopic) setting. –E2520PT – sterile, single use peritoneal pressure tubing, 10/case
Included with F ORCE A RGON ™ II Unit –One V ALLEYLAB F ORCE A RGON ™ II Unit –One Gas Regulator –One E Adapter –One ESU Cable (to connect F ORCE A RGON ™ II to a V ALLEYLAB Generator) –One 110V Power Cord –One 230V Power Cord –One User’s Guide (English) –One Service Manual (English)