NA62 SILICON PIXELS DETECTOR VIBRATIONS MEASUREMENT 1Vittore Carassiti - INFN FE NA62 GTK Meeting , CERN
Vittore Carassiti - INFN FE2 NA62 GTK Meeting , CERN MEASURING INSTRUMENT
Vittore Carassiti - INFN FE3 NA62 GTK Meeting , CERN THE DETECTOR CONSTRAINTS THREE POINTS BY 50 μm THICKNESS KAPTON FOIL
Vittore Carassiti - INFN FE4 NA62 GTK Meeting , CERN MEASURED VIBRATIONS Tested partsDeflection (μm) RMS Velocity (mm/s) RMS Acceleration (g) RMS Filter (KHz) Room floor20--1 Test bench20--1 Supports20--1 Vessel no flow Vessel From 1 to 8 m^3/h 20 (no vibrations inside) 6,40,
Vittore Carassiti - INFN FE5 NA62 GTK Meeting , CERN PROTECTIVE RUBBER TEST TWO TYPES OF RUBBER HAVE BEEN TESTED IN LIQUID NITROGEN (3 DAYS) : 3140 RTV COATING (SILICONE RUBBER) 186 SILICONE ELASTOMETER (BICOMPONENT) NO CRACKS ; NO LOSS OF ELASTICITY
Vittore Carassiti - INFN FE6 NA62 GTK Meeting , CERN NEXT STEPS MAKING A BOARD FOR THE WIRE BONDS TEST FRONTBACK
Vittore Carassiti - INFN FE7 NA62 GTK Meeting , CERN NEXT STEPS OPTIMIZATION BY SIMULATIONS OF THE NITROGEN GUIDE CHANNELS TEST AIMING TO VERIFY THE EFFECTIVE TEMPERATURE DISTRIBUTION ON THE DETECTOR DESIGN OF THE NITROGEN SUPPLY SYSTEM