Laboratory Certification Update Part 2 Common Findings KWWOA Louisville April 15, 2015 Presented by Frank Hall, Laboratory Certification Coordinator.

Slides:



Advertisements
Similar presentations
CAP/POCT 2003 Carol Riley-Hunte, RRT Senior Education Specialist Bayer HealthCare.
Advertisements

Analyze-Immediately Parameters pH Analyze Immediately Within 15 minutes of collection: pH Residual Chlorine Temperature Chlorine Dioxide Dissolved Oxygen.
Analyze Immediately Temperature. Hold Times IMMEDIATELY! Between minutes, as close to sample collection time as possible - note both time of collection.
Why Lab. Certification? Laboratory certification tries to insure that the laboratory is qualified and capable of analyzing the sample and obtaining quality.
Quality is a Lousy Idea-
LABORATORY CERTIFICATION & DATA QUALITY MICHAEL W. MILLER, Ph.D. NJ-DEP Office of Quality Assurance
Contracting for Laboratory Services Ann Mullin Cleveland US Geological Survey Presented to the Environmental Protection Agency Science Advisory Board April.
Annual Management Review June Flowers, Flowers Chemical Labs Florida Society of Environmental Analysts - FSEA St. Petersburg, FL, May 2009.
UNDERSTANDING ENVIRONMENTAL LABORATORY/QC REPORTS Maya Murshak – Merit Laboratories, Inc.
Whole Effluent Toxicity
Chain of Custody Records Proper Documentation Techniques Dr. Richard Medina Environmental Testing and Consulting, Inc.
Wastewater Laboratory Certification Overview 1. What is laboratory certification? “Laboratory certification is a process that provides formal recognition.
Legal Defensibility in the Laboratory To Protect and Enhance Kentucky’s Environment KWWOA April 17, 2013.
Determine Scope of Assessment Select Assessment Team Notify Laboratory & Request Records Determine Team Assignments Sample Receipt and Storage Assessment.
Louisiana Department of EnvironmentalQuality LDEQ CAM Plan Overview LDEQ’s 27 th Annual Conference on the Environment Cajundome Convention Center Lafayette,
World Health Organization
Preparing for Accreditation Jerry L. Parr Catalyst Information Resources, L.L.C. TNRCC Trade Fair May 7, 2002.
Lecture 8. Quality Assurance/Quality Control The Islamic University of Gaza- Environmental Engineering Department Environmental Measurements (EENV 4244)
Importance of Quality Assurance Documentation and Coordination with Your Certified Laboratory Amy Yersavich and Susan Netzly-Watkins.
DOW Laboratory Certification Program Update – 2015 Presentation to: KWWOA April 15, 2015 Department for Environmental Protection Energy & Environment Cabinet.
OPTIMISING CHEMICAL ANALYSIS FOR THE WFD BY QUALITY ASSURANCE PROCEDURES & ACCREDITATION An Overview OPTIMISING CHEMICAL ANALYSIS FOR THE WFD BY QUALITY.
Supplementary Training Modules on Good Manufacturing Practice
Supplementary Training Modules on Good Manufacturing Practice
Quality Assurance/Quality Control Policy
Standard Operating Procedures l At l Call for help at Orlando Office Tallahassee.
Quality Assurance Plans (QAP)
QC | Slide 1 of 16 June 2006 Good Practices for Quality Control Laboratories P art 4: Inspecting the laboratory Supplementary Training Modules on Good.
1 Standard Operating Procedures Their Development and Use with Quality Assurance Project Plans.
Field Analysis Quality Control
Debra Waller NJDEP-Office of Quality Assurance
WWLC Standard Operating Procedures Presented by Frank Hall, Laboratory Certification Coordinator.
DOW Laboratory Certification Program Update – 2013 Presentation to: KWWOA April 17, 2013 Department for Environmental Protection Energy & Environment Cabinet.
By: Farzad Dadgari Soil and Environmental Specialist SWHISA.
Laboratory Technical Issues Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Environmental & Public Protection Cabinet To Protect.
Fundamentals of Writing an Effective QAP Presentation to: KWWOA April 15, 2013 Department for Environmental Protection Energy & Environment Cabinet To.
Quality WHAT IS QUALITY
DOW Laboratory Certification Program Update – 2014 Presentation to: KWWOA April 9, 2014 Department for Environmental Protection Energy & Environment Cabinet.
How to Select a Test Method Marlene Moore Advanced Systems, Inc. June 15, 2010.
Microbiology Overview KWWOA April 09, 2014 Department for Environmental Protection Energy & Environment Cabinet.
Quality Assurance Plan (QAP) Department for Environmental Protection Environmental & Public Protection Cabinet To Protect and Enhance Kentucky’s Environment.
QA/QC and QUALIFIERS LOU ANN FISHER CITY OF STILLWATER, OK
Understanding Your QC Presentation to: KWWOA April 15, 2015 Department for Environmental Protection Energy & Environment Cabinet To Protect and Enhance.
Laboratory QA/QC An Overview.
Wastewater Laboratory Certification Presented at: KWWOA April17,
Understanding the Audit Process Presentation to: KWWOA Department for Environmental Protection Environmental & Public Protection Cabinet To Protect and.
Why do we need to do it? What are the basic tools?
Field-Only Certification Presented by Frank Hall, Laboratory Certification Coordinator.
Regulatory Issues in Laboratory Management
WHAT IS THE CHEROKEE NATION? Cherokee Nation Air Quality Data Management Concepts for Quality Data Collection Ryan Callison.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
Proposed Laboratory Certification Rules Revision 15A NCAC 2H.0800 Laboratory Certification April 16, 2015 Dana Satterwhite Gary Francies.
George E. Detsis Manager, Analytical Services Program United States Department of Energy Office of Health, Safety and Security Office of Sustainability.
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Method Validation of VINDTA 3C (TA/DIC) Marine Scotland Science (MSS) Experience and Concerns Pamela Walsham.
Field Equipment Calibration PH, Temperature, Conductivity, Turbidity, Dissolved Oxygen, Free & Total Chlorine Gregory Langland CCSF.
The Method Update Rule Does this affect me? HOW?
BSB Biomanufacturing CHAPTER 4 GMP – Documentation Part I (SOP)
Quality is a Lousy Idea-
EMSA 22 Lab Module 1 Field Equipment Calibration
Overview of Quality Assurance
Instrument Maintenance
EPA Method Equivalency
Quality is a Lousy Idea-
An Introduction to Quality Assurance in Analytical Science
Observations from California’s On-Site Assessment Unit
EPA Method Equivalency
EPA Region 10 Alternate Test Procedures and Method Update Rule
Topics in Microbiology Quality Assurance Project Plan Essentials
Why Use Them? By: Marcy Bolek – Alloway
Panhandling & Other Lab Procedures Illinois EPA Checks
Presentation transcript:

Laboratory Certification Update Part 2 Common Findings KWWOA Louisville April 15, 2015 Presented by Frank Hall, Laboratory Certification Coordinator

Overview

Applications Field Procedures Chain of Custody (C-O-C) Sample Storage / Integrity DOC/MDL Microbiology WET Organics Metals Inorganics QA/QC Proficiency Test PT

Applications

QAP –Not addressing all areas Scope Organization Charts –If using the templates Make certain you are deleting out the portions that don’t apply to you Make certain you are completing all the required areas SOP –Must follow the method as written –Analyst must be performing the analysis as written in the SOP Completeness –Make sure all attachments are available Method-Analyte list request QAP SOP DOC/MDL PT

Field Parameters

Calibration –All calibrations for field parameters is “daily, in use” –pH 95% - 105% slope 0.1 S.U. accuracy Must bracket readings –Conductivity In range of use

Field Parameters QC –Calibration Verification –Blank –Duplicate –QCS PT –pH –Chlorine –Conductivity –Turbidity Records

Field Parameters QC –Calibration Verification Daily for pH, conductivity, chlorine, & turbidity –Blank Daily for conductivity, chlorine, & turbidity –Duplicate 1/ every 20 sample for pH, conductivity, chlorine, turbidity, DO, & temperature (may be over multiple days) –QCS Quarterly for conductivity, chlorine, & turbidity PT –Annually for pH, conductivity, chlorine, & turbidity

Chain of Custody (C-O-C)

C-O-C A C-O-C is a legal document! Gaps in C-O-C –Relinquished/Received by Missing information –Sampler –Date –Time –Preservative –Number and type of bottles –Sample ID –Parameters requested Verification upon receipt by the laboratory

Sample Storage / Integrity

Reagent logs –Every reagent needs to be tracked –Every time a reagent is used it needs to be tracked –Need a certificated of analysis for all purchased reagents Standard logs –Every standard needs to be tracked –Every time a standard is used it needs to be tracked –Need a certificated of analysis for all purchased standards

Sample Storage/Integrity Temperature logs –Ovens Thermometer in sand Recorded each day in use Adjusted for correction factor Documentation of corrective action –Refrigerator Thermometer in liquid Recorded each day in use Adjusted for correction factor Documentation of corrective action

Sample Storage/Integrity Internal sample tracking –Procedures must be in place to assure that sample remains secure –Procedures must be in place to assure that sample preservation conditions are met and maintained If you didn’t write it down, you didn’t do it!

DOC/MDL

Must be performed annually DOC must be performed by every analyst –4 replicates Must be at the same concentration MDL must be performed or verified for every instrument –Perform on 2-3 non consecutive days –New criteria proposed in the new MUR

Microbiology

Signatures on the BARF –Need to be signed after the Expired Colilert/Colilert-18 comparators Must utilize a SW PT for drinking water E.coli enumeration (not WW PT) Insufficient pre-warming for Colilert-18 presence/absence testing –7 minutes for sample load of 10 or more is typically not sufficient to bring those samples up to o C. Most labs have had to increase to 8 or 9 minutes.

WET

Routine chemical and physical analyses associated with WET testing –Must be treated the same as a normal chemical analysis Must calibrate meters Must analyze associated QC Must be documented

Organics

Manual integrations –Allowable –Must be documented –Print original chromatograph and the new chromatograph –Give brief reason for manual integration

Metals

Digestion –Must be performed if turbidity is not < 1 NTU Turbidity analysis must be preformed utilizing an approved method and must be documented. –The digestion must be documented Linear Dynamic Range (LDR) –If not performed then analysis results can only be performed up to 90% of calibration range Spectral Interference Check –Must be performed to assure that various metals are not creating interferences with the analytes of concern.

Inorganics

Gravimetric Methods –Balance must be verified, in range of use each day –Reporting limit is a function of minimum weight requirements and volume Titration Methods –Titrant must be verified –Reporting limit is a function of titrant concentration and volume Colormetric Methods –Spectrometer must be verified annually –Calibration must be verified

Inorganics Potentialmetric Methods (probes) –Probe maintenance –Probe storage –Calibration must be verified –Documentation Physical Properties –Method defined parameters Must be analyzed exactly as written

QA/QC

QC must be documented QC frequency must be defined QC acceptance criteria must be established and reviewed Both passing and failing QC must be charted –Even if analysis results for the failed QC are not used Analyst should know what the purpose of each type of QC. –Kevin Stewart is giving a presentation on QC later today. RLS analysis is required for all WW parameters at the time of calibration or at least quarterly QC is a requirement not a recommendation –WW must address all of the 12 mandatory elements for every method

PT

Send PT results to: Recommend performing these in 1 st or 2 nd quarter to allow to adequate time for repeats if necessary DOW only allows clerical revisions for the PTs submitted with the initial application PT samples must be analyzed the same as all samples.

Questions?? Comments?? Franklin C. Hall, Jr. Laboratory Certification Coordinator DOW (502) extension 4959