B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun

Slides:



Advertisements
Similar presentations
DIGITAL RADIOGRAPHY.
Advertisements

Technique Guidance Systems
Image Reconstruction.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
UNDERSTANDING RAW Joe Sukenick DigiQuest
Primary Exposure Factors IV
Chapter 18 Radiographic Exposure
Quality Assurance and Digital Radiography
Digital Radiography.
Resident Physics Lectures
RAD 350 Chapter 17Digital Rad Tech. Spatial Resolution – ability to distinguish small items in close proximity with near the same atomic mass density Spatial.
Resident Physics Lectures
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
Digital Image Manipulation I By Professor Stelmark.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Computed Tomography III
FILM ARTIFACTS.
Optical Density and Brightness
Unsharpness Calculations and Resolution By Professor Stelmark.
X-Rays and CT Scans.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Intensifying Screens Kyle Thornton DMI 50B.
How the Camera Works ( both film and digital )
Seeram Chapter 11: Image Quality
Quality Control Rad T 110.
Digital Image Characteristic
Unit III Creating the Image. Unit III Creating the Image.
Computers in Imaging Q & A Robert Metzger, Ph.D..
Quality Control.
Dr Mohamed El Safwany, MD.
Image Quality Radiographic Resolution.
RAD 254 Digital Imaging Basic Elements of Digital Imaging CR/DR.
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
Digital Radiography and PACS By Professor Stelmark.
 The image that the digital camera sensor captures is based on the light reflected or emitted from a subject and how much the sensor is exposed to that.
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Things they never tell you when the equipment is installed REV 11/2008
Review CR & DIGITAL IMAGING (1) 2012 – RT 244 wk 15
RAD 254 Chapter 16 Image Quality Film factors Geometric factors Subject factors.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Sensitometry By Prof. Stelmark. The study of the relationship between the intensity of exposure of the film and the blackness after processing is called.
Radiographic Quality Visibility and Sharpness
Radiographic Equipment
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
Film Optical Density Spectral Matching & sensitometry.
BME5002 Image Processing Lecture 1 John G. Harris 1/26/01.
Image Production and Evaluation NOTE: QA AND QC ARE USED INTERCHANGABLY IN APPLETON AND LANGE.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Part No...., Module No....Lesson No
X-ray SNR in 3 steps. I ∆I. X-ray transmission SNR Review Let N = average number of transmitted x-rays N = N 0 exp [ - ∫  dz ] Emission and transmission.
Objectives Identify the image quality characteristics that apply to all medical imaging modalities Understand the concept of image optimization Review.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
DIGITAL IMAGING.
Course Title : Image Recording
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Digital Radiographic Image Processing and Manipulation Chapter 7.
By Prof. Stelmark. Digital Imaging In digital imaging, the latent image is stored as digital data and must be processed by the computer for viewing on.
Direct Digital Radiography or Direct Capture Radiography
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
History of digital radiology 1980 Kinos L.T. ET all developed a portable radiographic X-ray camera made of tantalum and aluminum. Digital radiography.
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS &
Exposure Factors or Prime Factors
Chapters 21 & 22.
מה צריך רנטגנאי לדעת על דימות דיגיטלית?
RECORDED DETAIL.
Image Production and Evaluation
RECORDED DETAIL.
Digital radiography.
Presentation transcript:

B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun

I MAGING WITH F ILM The greatest challenge of film radiography is to get the range of exposure to fit into the dynamic range of the film. If the exposure falls outside the latitude, there will be little or no image contrast formed.

I MAGING WITH F ILM There are two conditions that contribute to receptor exposure outside the film latitude: Error in setting the correct exposure factors. Some regions of the body have wide dynamic of exposure that exceeds the latitude of the film. Using a wider latitude film can correct this but results in lower image contrast.

T HE A DVANTAGE OF A W IDE D YNAMIC R ANGE Here we can see the advantages of a digital receptor that has a wide dynamic range. Even when there is a wide range of exposure coming from the body (wide histogram) and exposure at different levels (exposure errors) they still fit within the dynamic range.

D IGITAL I MAGE C ONTRAST In a digital image, contrast is represented by different pixel values. A typical digital receptor will have a linear relationship between exposure and resulting pixel value.

D IGITAL I MAGE C ONTRAST We have seen that this relationship extends over a relatively wide range of exposures. This is contrasted with the non-linear relationship seen with film.

O PTIMUM E XPOSURE IN D IGITAL R ADIOGRAPHY The wide dynamic range and linear response of digital receptors is like a double edged sword. The advantage is a wide range of exposures and exposure errors will still produce an image with good contrast.

O PTIMUM E XPOSURE IN D IGITAL R ADIOGRAPHY So, what is the problem? There are two that we can see here. Even though good contrast with low exposure is obtained. Due to the low exposure, we have high noise.

O PTIMUM E XPOSURE IN D IGITAL R ADIOGRAPHY With film the image would be light. The other problem is that exceptionally good images are obtained with high exposure ( very low noise). With film the image would be dark or over exposed

O PTIMUM E XPOSURE IN D IGITAL R ADIOGRAPHY In general, the challenge is to make sure the exposure factors are set for optimum image quality.

M ONITORING E XPOSURE L EVELS One of the challenges is to know when the digital image is properly exposed since it is unlike film where under and over exposure is obvious. Each manufacturer of digital receptors will provide an approach to calculate the image exposure information.

M ONITORING E XPOSURE L EVELS Some will display the “S” number. This displayed value generally indicates the speed of the receptor that would match the actual exposure.

M ONITORING E XPOSURE L EVELS A low exposure would result in a high calculated S number (like S=1000) and a high exposure would produce a low number (like s=50). This is opposite of screen-film speed.

M ONITORING E XPOSURE L EVELS The operator should determine the appropriate range of values for optimum exposure and monitor the values. The method varies by manufacturer and clinical procedure.

ACR E XPOSURE F ACTOR C HART Fuji (S Number) Agfa (LGM value Kodak (Exposure Index) Detector Exposure Estimate mR Indication & Action >1000<1.45<1250<0.20Repeat Underexposed QC exception Underexposed QC review Acceptable < > > >4.0 Overexposed Repeat

D IGITAL E XPOSURE Proper exposureOver exposure

D IGITAL R ADIOGRAPHY I MAGE Q UALITY Like all x-ray images, there are 5 specific quality characteristics. Spatial detail Detail Contrast Noise Artifacts We will now see how 3 of these, contrast, detail and noise are effected by the operation of digital radiography.

D IGITAL R ADIOGRAPH C ONTRAST Contrast sensitivity of digital procedures and image contrast depend upon several factors. Two are: X-ray beam spectrum Scattered radiation Similar to film.

D IGITAL R ADIOGRAPH C ONTRAST What is different is the ability to adjust and optimize contrast after the image is recorded. This is done through the digital processing of the image and then the adjustment of the window when the image is being viewed.

D IGITAL R ADIOGRAPHIC D ETAIL Visibility of detail is reduced and limited by the blurring that occurs at different stages of the imaging process. Some are common to both film and digital radiography.

C OMMON S OURCES OF B LURRING Common sources of blurring; The focal spot (depends upon size and object location.) Motion if present The receptor (generally due to light spreading within the fluorescent or phosphor screen)

D IGITAL S OURCES OF B LURRING Additional blurring is the result of dividing the image into pixels. The size of a pixel (amount of blurring) is the ratio of the image size (image size relative to the anatomy) and the matrix size.

D IGITAL S OURCES OF B LURRING Pixel size is another factor that must be considered because it limits the detail of the image x 1280 on a 18cm x 24 cm image would be sharper than on a 35 cm x 43 cm image. Generally, digital images lack the spatial detail of film images due to the digital processing.

N OISE IN D IGITAL R ADIOGRAPHS The most predominate source of noise in digital imaging is quantum noise associated with the random distribution of the x- ray photons received by the image receptor.

N OISE IN D IGITAL R ADIOGRAPHS The level of noise depends upon the amount of receptor exposure used to produce the image. With digital imaging, it can be adjusted over a wide range due to the wide dynamic range.

N OISE IN D IGITAL R ADIOGRAPHS Noise is controlled by using the appropriate exposure factors.

D IGITAL I MAGE N OISE UnderexposedProper exposure