CDS Users Meeting 21-22 Sep. 2005 In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 1 The in-flight monitoring.

Slides:



Advertisements
Similar presentations
Richard Young Optronic Laboratories Kathleen Muray INPHORA
Advertisements

THROUGHPUT OF A NORMAL INCIDENCE SPECTROMETER DESIGN IN DIFFERENT WAVELENGTH RANGES Luca Teriaca Max Planck Institut für Sonnensystemforschung.
Analysis of the Visible Absorption Spectrum of I 2 in Inert Solvents Using a Physical Model Joel Tellinghuisen Department of Chemistry Vanderbilt University.
The synthetic emission spectra for the electron non-thermal distributions by using CHIANTI Elena Dzifčáková Department of Astronomy, Physics of the Earth.
Estimating the Chromospheric Absorption of Transition Region Moss Emission Bart De Pontieu, Viggo H. Hansteen, Scott W. McIntosh, Spiros Patsourakos.
ALFVEN WAVES IN A POLAR CORONAL HOLE FROM HINODE/EIS OFF LIMB OBSERVATIONS Bemporad A. & Abbo L. –
Hinode / EIS Solar 24, Dec 2008 Dr Peter Young, NRL/GMU Active Region Diagnostics with EIS Dr Peter Young Dr Peter Young Naval Research Laboratory George.
RHESSI/GOES Xray Analysis using Multitemeprature plus Power law Spectra. J.McTiernan (SSL/UCB)
RHESSI/GOES Xray Analysis using Multitemeprature plus Power law Spectra. J.McTiernan (SSL/UCB) ABSTRACT: We present spectral fits for RHESSI and GOES solar.
The EUV spectral irradiance of the Sun from minimum to maximum Giulio Del Zanna Department of Space and Climate Physics University College London Vincenzo.
Ingolf E. Dammasch ROB/SIDC Brussels, Belgium Solar UV Spectroscopy with SUMER on SOHO (extended version for 11 Oct 2007)
The Non-Flare Temperature and Emission Measure Observed by RHESSI and SXI J.McTiernan (SSL/UCB) J.Klimchuk (NRL) Fall 2003 AGU Meeting.
Sunspot Studies in Oslo Nils Brynildsen, Per Maltby, Terje Fredvik and Olav Kjeldseth-Moe Institute of Theoretical Astrophysics University of Oslo.
Analysis of Differential Expression T-test ANOVA Non-parametric methods Correlation Regression.
Simulations to test He-like forbidden-to-intercombination line ratio modeling for O, Ne, Mg, S, Si and Fe. June 28, 2005 M. Walter-Range Swarthmore College.
Properties of Prominence Motions Observed in the UV T. A. Kucera (NASA/GSFC) E. Landi (Artep Inc, NRL)
CDS Session NAM Dublin April Active Region Loop Studies in Oslo Olav Kjeldseth-Moe and Terje Fredvik Institute of Theoretical Astrophysics University.
Symposium in Honour of Philippe Lemaire's Retirement Four Solar Cycles of Space Instrumentation Institut d’Astronomie Spatiale, Orsay M.C.E. Huber Radiometry.
03/09/2007 Earthquake of the Week
The Analysis of Variance
Richard Young Richard Young Optronic Laboratories Kathleen Muray Kathleen Muray INPHORA Carolyn Jones Carolyn Jones CJ Enterprises.
1 Seventh Lecture Error Analysis Instrumentation and Product Testing.
CALIBRATION METHODS.
Figure 1 Figure 8 Figure 9Figure 10 Altitude resolved mid-IR transmission of H 2 O, CH 4 and CO 2 at Mauna Loa Anika Guha Atmospheric Chemistry Division,
Analysis of Coronal Heating in Active Region Loops from Spatially Resolved TR emission Andrzej Fludra STFC Rutherford Appleton Laboratory 1.
Modeling the Solar EUV irradiance
Variation of EUV solar irradiances along the cycle Vincenzo Andretta 1, Giulio Del Zanna 2, Seth Wieman 3 1 INAF – Osservatorio Astronomico di Capodimonte,
Results of the WMO Laboratory Intercomparison of rainfall intensity gauges Luca G. Lanza University of Genoa WMO (Project Leader) DIAM UNIGE September.
SOHO/CDS CDS Users’ Meeting, Sep 2005 Dr Peter Young, CCLRC/RAL Element Abundance Results from CDS Dr Peter Young Dr Peter Young SOHO/CDS Project Scientist.
Goal 1 Notes Physical Science By Nancy Booth. Physical Science I. Applied Science vs. Pure Science II. Technology III. What is Physical Science? IV. Problems.
Evidence Based Medicine
Pretest 1. True or False: Compounds have fixed compositions TRUE!
W  eν The W->eν analysis is a phi uniformity calibration, and only yields relative calibration constants. This means that all of the α’s in a given eta.
A statistical study of C IV regions in 20 Oe-stars Dr Antonios Antoniou University of Athens, Faculty of Physics, Department of Astrophysics, Astronomy.
The Influence of the Return Current and the Electron Beam on the X-Ray Flare Spectra Elena Dzifčáková, Marian Karlický Astronomical Institute of the Academy.
The Examination of Residuals. Examination of Residuals The fitting of models to data is done using an iterative approach. The first step is to fit a simple.
The Influence of the Return Current and Electron Beam on the EUV and X-Ray Flare Emission E. Dzifčáková, M. Karlický Astronomical Institute of the Academy.
Comparison of Solar EUV Irradiance Measurements from CDS and TIMED/EGS W. T. Thompson L3 Communications EER, NASA GSFC P. Brekke ESA Space Science Department.
Optimising Cuts for HLT George Talbot Supervisor: Stewart Martin-Haugh.
Lecture 16 Section 8.1 Objectives: Testing Statistical Hypotheses − Stating hypotheses statements − Type I and II errors − Conducting a hypothesis test.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
6.1 Inference for a Single Proportion  Statistical confidence  Confidence intervals  How confidence intervals behave.
EVE Sun-as-a-Star Spectra Milligan et al “Fe Cascade” for SOL
Effective drift velocity and initiation times of interplanetary type-III radio bursts Dennis K. Haggerty and Edmond C. Roelof The Johns Hopkins University.
Modeling the UV/EUV and its relevance for PROBA2 observations Margit Haberreiter Physikalisch-Meteorologisches Observatorium Davos/World Radiation Center,
Flare Prediction and the Background Corona Coronal Diagnostic Spectrometer Wolter-Schwarzschild Type 2 telescope Two separate spectrometers- the Normal.
Dispersed fluorescence studies of jet-cooled HCF and DCF: Vibrational Structure of the X 1 A state.
Calculation of the Irradiance variations in the UV and extreme UV Margit Haberreiter PMOD/WRC, Davos, Switzerland IPC XI Sept 26 – Oct 15, 2010.
Variability Introduction to Statistics Chapter 4 Jan 22, 2009 Class #4.
Peterson xBSM Optics, Beam Size Calibration1 xBSM Beam Size Calibration Dan Peterson CesrTA general meeting introduction to the optics.
NAM and UKSTP 11 April 2003 Sensitivity Differences in Pre-Loss and Post-Recovery NIS/CDS Spectrum By Steven Chapman Centre for Astrophysics University.
Measurements and Their Analysis. Introduction Note that in this chapter, we are talking about multiple measurements of the same quantity Numerical analysis.
, Dan Peterson Apparent inconsistencies and other issues in the xBSM measurements of IBS Scans We have studied the pinhole and CodedAperture.
Lecture 8: Measurement Errors 1. Objectives List some sources of measurement errors. Classify measurement errors into systematic and random errors. Study.
IAS 20 June 2013 Celebrating the achievements of Alan Gabriel Laboratory spectroscopy Exploring the process of dielectronic recombination S. Volonte.
Spectrometer Throughput Dr Peter Young Consortium meeting, 3 March 2006.
On the spectroscopic detection of magnetic reconnection evidence with Solar B – I. Emission line selection and atomic physics issues P. F. Chen 1,2, H.
Figure 1 Figure 8 Figure 9Figure 10 Altitude resolved mid-IR transmission of H 2 O, CH 4 and CO 2 at Mauna Loa Anika Guha Atmospheric Chemistry Division,
Night OH in the Mesosphere of Venus and Earth Christopher Parkinson Dept. Atmospheric, Oceanic, and Space Sciences University of Michigan F. Mills, M.
Electron-impact excitation of Be-like Mg
The Value Premium and the CAPM
Significance Test for the Difference of Two Proportions
Section 11.1 Day 2.
ADAS workshop, Auburn University
Tests of significance: The basics
Introduction to Instrumentation Engineering
Atomic Absorption Spectroscopy
Studying Transition Region Phenomena with Solar-B/EIS
Product moment correlation
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
Presentation transcript:

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 1 The in-flight monitoring and validation of the CDS NIS radiometric calibration J Lang, DH Brooks, AC Lanzafame, R Martin, CD Pike and WT Thompson CDS Users Meeting The Coseners House 21/22 September 2005

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 2 Status of CDS Radiometric Calibration Calibration as presently implemented in the CDS software is as defined by Lang et al (2002) in “The Radiometric Calibration of SOHO”, ISSI Scientific Report SR-002, ed. A Pauluhn, MCE Huber and R von Steiger, p105. Radiometric calibration built on: –laboratory calibration using a secondary standard source –joint observations with the NASA LASP rocket –joint observations with the NASA SERTS rocket –intercalibration observations with SUMER –papers by Landi et al and Del Zanna et al The burn-in corrections for NIS and GIS are also detailed in the ISSI paper.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 3 Present Work The present work is to look at the calibration as a function of time by measuring line ratios. Three classes of line ratio are investigated. –branching ratios (Griffin and McWhirter) –density and temperature insensitive line ratios (Neupert and Kastner) –density and/or temperature sensitive line ratios Data – use NISAT_S (ID7, variation 2) which gives the full NIS spectra and builds up a 20” by 240” raster using the smallest slit (2” by 240”). This has been run at least once a week on QS and at least once a month on AR and CH. Area of Sun observed checked using EIT Fe XI 195 Å. The corrected and calibrated data were averaged over the 10 mirror positions in the raster. To avoid any possible edge effects the lowest 15 and highest 9 pixels were ignored and the remaining 120 were divided into 4 equal segments along the slit. Only one segment was used in this work.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 4 Monitoring NIS-2O III 2s 2 2p 2 1 D – 2s2p 3 1 P (525.8 Å) and 2s 2 2p 2 1 S – 2s2p 3 1 P (597.8 Å) Branching ratio with well-known A values and good wavelength span Å is blended with 3/2-3/2 member of 3s 2 3p 2 2 P - 3s3p 3 2 D multiplet of Ca VIII. Blend corrected using adjacent 3/2- 5/2 component at Å. The correction is density sensitive. However, average correction is 3 % so even 50 % uncertainty in correction has a very small effect. In graphs, points just have fitting uncertainties, while theory is shown as solid line with dashed line having theory uncertainties, burn-in uncertainties and calibration uncertainties. Quiet Sun observations are shown. The corresponding results for AR are about 10 % higher which could indicate a blend affecting the Å line. The agreement between theory and observations is reasonable. If the blend also affects QS observations, this would account for the ratio being slightly high. The loss of attitude period is easily identified. The ratio is slightly lower after the loss of attitude. However, the agreement between theory and measurement has been maintained for six years.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 5 Monitoring NIS-2O II 2s 2 2p 3 4 S – 2s 2 2p 2 3s 4 P (539 Å) and 2s 2 2p 3 2 D – 2s 2 2p 2 3s 2 P (616 Å) The ratio of the intensity of the 3/2-5/2 transition (539.1 Å) or the blend of the 3/2-3/2 and 3/2-1/2 components (539.6 Å) of the 4 S - 4 P transition to the blend of the 5/2-3/2 and 3/2-3/2 components of the 2 D – 2 P multiplet at Å gives a good wavelength span but is temperature sensitive. O II has maximum population in ionisation balance at K and is zero by K and K.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 6 Monitoring NIS-2O II 2s 2 2p 3 4 S – 2s 2 2p 2 3s 4 P (539 Å) and 2s 2 2p 3 2 D – 2s 2 2p 2 3s 2 P (616 Å) The ratios I(539.1 Å)/I(539.6 Å) before and after the loss of spacecraft attitude are constant but different. Comparison with the other two ratios indicates the change can be ascribed to an increase in the intensity of the Å line. The ratio I(539.6 Å)/I(616.3 Å) shows little scatter. The difference between theory and measurements is attributed to the atomic data, most likely the 2-3 effective collision strengths. There is a slight fall in the ratio in the first year of operations. However, the ratio remains the same after the attitude-loss of the spacecraft albeit with a larger scatter. This indicates that the calibration has been maintained over the six years of observations.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 7 Monitoring NIS-2O IV 2s 2 2p 2 P 3/2 – 2s2p 2 2 P 3/2 (554.5 Å) and 2s 2 2p 2 P 1/2 – 2s2p 2 2 S 1/2 (608.4 Å) or 2s2p 2 2 D – 2p 3 2 P (617.0 Å) I(554.5 Å)/I(608.4 Å) is temperature sensitive: RatioTe K (0) K (Peak IB) K (0) Ratio agrees with theory. Loss-of attitude gap is obvious, but the ratio changes although still within the uncertainties. The Å 2s 2 2p 2 P – 2s2p 2 2 S line is in the wings of a very strong line and the problem is the burn-in correction compounded by the non-Gaussian shape of the post- recovery lines. The spectra moved on the detector following the attitude-loss and this put the Å line where the strong Å Mg IX line previously caused detector burn- in. The Å line is the most difficult NIS-2 burn-in case. The I(554.5 Å)/I(617.0 Å) ratio is more temperature dependent than the first ratio. The Å feature is a blend of O IV and O II. The O II line is 2s 2 2p 3 2 D 3/2 – 2s 2 2p 2 3s 2 P 1/2 and was corrected using the intensities of the other members of the multiplet at Å. This ratio shows that agreement between predicted and measured ratio is satisfactory and also that there was no change over the the loss-of attitude period. The radiometric calibration at Å will stand as the problem is burn-in.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 8 Monitoring NIS-2 Second OrderSi XI 2s2p 1 P - 2p 2 1 D (604.1 Å) and 2s 2 1 S - 2s2p 1 P (2*303.4 = Å) In NIS-2 the Si XI 2s2p 1 P - 2p 2 1 D (604.1 Å) line is seen in 1st order and the 2s 2 1 S - 2s2p 1 P resonance line in 2nd order at 2  = Å, in the wings of the second order He II 304 Å line. For this ratio at high electron density ( cm -3 ) theory and a theta-pinch measurement agree, in the regime where stepwise excitation of 2p 2 1 D via 2s2p 1 P predominates and gives density sensitivity. The solar ratio is independent of density. The measured ratio agrees with theory within the uncertainties. Most of the data points lie below the theoretical ratio. The loss-of-attitude made no discernable change to the ratio, i.e. to the relative calibration of NIS2 in first and second order. There is some variation with time for the later data well after the loss-of-attitude. This may be a burn-in effect as the Å line moved to where the red wing of the strong second order He II Å line was prior to the loss-of-attitude.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 9 Monitoring NIS-2 and NIS-1 Intercalibration Si XI 2s2p 3 P 2 – 2p 2 3 P 1 (371.5 Å) and 2s 2 1 S – 2s2p 3 P 1 ( Å) The 3 P – 3 P multiplet is observed in active region spectra and the Å line is the only unblended member. The ratio I(371.5 Å)/I(580.9 Å) is insensitive to plasma conditions for active regions. The measured and theory ratios are shown. Lang et al pointed out that there could be difficulties with the atomic model for triplet lines at theta-pinch densities (> cm -3 ). The figure shows the measured active region ratios. The measurements lie below the theory, with some agreeing within the uncertainties. There is no trend over the six year period and no apparent change before and after the period of loss-of-attitude of SOHO. The Si XI Å line could be affected by a blend from O II and this is still being investigated.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 10 Monitoring NIS-2 and NIS-1 Intercalibration O III 2s 2 2p 2 3 P – 2s 2 2p3s 3 P (374 Å) This multiplet is observed as three lines and the ratios formed are shown. The weighted means of the ratios are all higher than theory. The ratios with the 1-2 or the sum of the 1-0 and 2-1 components in the numerator are higher than the ratio between the lines, it could be argued that these lines have unknown blends.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 11 Monitoring NIS-2 and NIS-1 Intercalibration O III 2s 2 2p 2 3 P – 2s 2 2p3s 3 P (374.1 Å, blend of 0-1,2-2,1-1 lines) and 2s 2 2p 2 1 D 2 – 2s2p 3 1 D 2 (599.6 Å) The ratio I(374.1 Å)/I(599.6) is temperature sensitive and slightly density sensitive, the densities shown in the figure being cm -3, cm -3 and cm -3. The temperature of maximum ion population in ionisation balance is K. Below K and above K the populations are zero. Thus a range of values from about to 0.3 is expected. The spread of values is as expected apart from one point. The majority of measured ratios lie within a much narrower range. There is no trend in the ratio over the six year period and no apparent change before and after the period of loss-of-attitude of the spacecraft.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 12 Monitoring NIS-2/NIS-1 Intercalibration Ne V 2s 2 2p 2 3 P – 2s2p 3 3 P (2-3, 2-2, Å) and 2s 2 2p 2 3 P – 2s2p 3 3 S (2-1, Å) The theory ratio is sensitive to temperature. It is insensitive to density in the range from cm -3 to cm -3 and then increases with density. The temperature of maximum ion population in ionisation equilibrium is K and zero below K and above K. The experimental point is from Lang (1988) from a theta-pinch measurement at a density of cm -3. If uncertainties of around 30 % in the theory are taken into account, then experiment and theory agree within the uncertainties.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 13 Monitoring NIS-2/NIS-1 Intercalibration Ne V 2s 2 2p 2 3 P – 2s2p 3 3 P (2-3, 2-2, Å) and 2s 2 2p 2 3 P – 2s2p 3 3 S (2-1, Å) From the theory a ratio from about 0.12 to 0.50 is expected. For the period before the loss of attitude of SOHO the ratios are above that expected from theory for the temperature of maximum ion population in ionisation equilibrium and many are above the temperature at which the ion population should be zero. However, there is still agreement within the uncertainties. The results taken before the loss-of-attitude of SOHO indicate the usefulness of this temperature sensitive line ratio for calibration purposes. After the loss-of-attitude the results are more sparse and varying. This is a direct result of the loss of resolution of NIS-1 after the loss-of-attitude. The adjacent Fe XIII lines are difficult to resolve. However, the results indicate that there is no obvious change in the ratio after the loss-of- attitude.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 14 Monitoring NIS-1 Fe XI 3s 2 3p 4 3 P – 3s3p 5 3 P 2-2 at Å and 1-2 at Å These lines form a branching ratio. In a critical evaluation Shirai et al gave A(2-1)/A(1-2) = ( )/( ) = In papers not included in that work Bhatia and Doschek gave A(2- 1)/A(1-2) = ( )/( ) = 3.16 and Fritzsche et al gave A(2-1)/A(1-2) = ( )/( ) = 3.20 Although the ratios are the same there are factors of two differences in the A values. There is no discernable change in the ratio after the loss-of- attitude or with time over the six years of observation. The weighted mean of the ratios before and after the loss-of-attitude are 2.23±0.09 and 2.41±0.14. The spread in this branching ratio is greater than the spread for ratios dependent on plasma conditions. This ratio has also been observed by SERTS. Again, there is a similar scatter in the results. SERTS-89 SERTS-91SERTS-93 AR 3.28± ± ±0.72 QS 2.46± ±0.87 Off limb4.00±1.88

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 15 Monitoring NIS-2 Fe XII 3s 2 3p 3 4 S – 3s3p 4 4 P, 3/2-1/2 at Å and 3/2-5/2 at Å The intensity ratio I(346.9 Å)/I(364.5 Å) is shown for quiet Sun observations. The ratio is insensitive to electron density and temperature for the quiet Sun and active regions. The pre-loss data are lower and less variable than the later data. In active region spectra the Si XI 2s2p 3 P 1 – 2p2 3 P 1 line will blend with the Å line. Young et al estimate a 7% contribution and Landi et al <5%. Del Zanna et al say the line is blended in active regions but not in the quiet Sun. The blend was not observed in the quiet Sun spectra of Brooks et al. The lines are weak and hard to fit, especially for the post-loss data. SERTS data are also variable. Many of the post-loss results agree with theory and it is difficult to conclude that a calibration change is needed post-loss. SERTS results SERTS-89 SERTS-91 SERTS-93 AR 0.27± ± ±0.07 QS 0.53± ±0.06 Off-limb 0.55±0.11

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 16 Monitoring NIS-2 Mg VIII 2s 2 2p 2 P J – 2s2p 2 2 P J (311.8 Å, Å, Å and Å) Ratios involving the lines of the multiplet are shown. For pre- loss spectra, Lanzafame et al found that the 315 Å line was blended and the fitted pre-loss data here have been reduced by a factor 1.5. Brooks and Warren found a smaller factor. I(313.7 Å)/I(317.0 Å) branching ratios agrees with theory and observations before and after loss-of-attitude agree. I(315.0 Å)/I(311.8 Å) branching ratio, data are sparse and the Å line is difficult to fit post-loss. For the density sensitive line ratio I(315.0 Å)/I(317.0 Å), the results are lower than theory but there is no change with time, although the scatter is greater post-loss. For I(311.8 Å )/I(317.0 Å ) the post-loss Å line is difficult to fit. Active region results show that this ratio differs before and after the loss-of-attitude and the ratio is outside the uncertainties post-loss. The conclusions are that the pre-loss sensitivity at 315 Å should be increased and that the post-loss sensitivity at Å should also be increased.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 17 Monitoring NIS-2 Mg VIII 2s 2 2p 2 P 1/2 – 2s2p 2 2 P 1/2 (313.7 Å) and 2s 2 2p 2 P 3/2 – 2s2p 2 2 S 1/2 ( Å) This line ratio is independent of temperature and density in active region spectra. Active region line ratio measurements are shown. The pre-loss data yield a steady ratio while the post loss data show larger variation and are higher with most points agree with theory within the uncertainties. It is difficult to draw conclusions. The Å line is difficult to fit, particularly post-loss. Ascribing the change to the Å line is inconsistent with the ratios just discussed. If the change is in the Å line, it has to reduce the intensity which rules out blending. It could well be that the post-recovery responsivity needs to be changed at Å.

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 18 Monitoring NIS-2 Fe XVI 2p 6 3s 2 S – 2p 6 3p 2 P (1/2-3/2 at Å and 1/2 -1/2 at Å) The active region ratio I(335.4 Å)/(360.7 Å) is shown. The theory ratio is insensitive to electron temperature and density for active region conditions. The Å line is blended with Mg VIII and Fe XII. The corrections for each average at 2 % and thus the uncertainty in the corrected ratio is insensitive to the uncertainty in the correction. The agreement with theory is within the uncertainties. There is a 17 % increase in the weighted means of the ratios taken before and after the the loss-of-attitude period. However, this cannot be attributed directly to changes in responsivity. As noted by Lang et al. the wide-slit burn-in is not well known for these lines and could well account for the differences in the ratios before and after the the loss-of-attitude. SERTS Results: SERTS ±0.5 SERTS ±0.3 SERTS ±0.3

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 19 Discussion and conclusions The responsivities of NIS-2 and NIS-1 are shown along with the line pairs used in the present work. For NIS-2, the O II, O III and O IV lines give good wavelength coverage. For data taken before the loss-of- attitude, all ratios agree with theory within the uncertainties, validating the calibration. For post-loss data, there are small changes in O II and O III ratios while the O IV ratio involving the Å line increases slightly with time. For NIS-2 in second order, the measurement of the Si XI Å line could be affected by burn-in post-recovery. However, data taken before the loss-of-attitude validates the calibration. For NIS-2/NIS-1 inter-comparison, line ratios for Si XI, O III and Ne V validate the calibration. The ratios also indicate that the loss-of-attitude did not cause any change in the inter-calibration

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 20 Discussion and conclusions For NIS-1, the results from Fe XI and Fe XII are found to be variable as in other measurements. Both could be affected by unknown blends. The Mg VIII ratios reveal some problems. The calibration post-loss at Å could well be in error. The calibration pre-loss at 315 Å is also causing problems and needs revising. The same is true at Å. Difficulties with the Fe XVI lines at Å and Å are most probably a result of incomplete burn-in correction..

CDS Users Meeting Sep In-flight monitoring and validation of the CDS NIS radiometric calibration Lang et al. Page 21 Discussion and conclusions The line ratios measured in the present work indicate that overall, apart from three specific wavelengths, the relative radiometric calibration is validated, yielding results which agree with theory within the uncertainties and being maintained over the years of observation. Problems with the burn-in correction account for difficulties at Å, Å (seen in second order), Å and Å. It can be considered that the line ratios used here relate the calibrations of NIS-1, NIS-2 second order and NIS-2 low wavelengths to the long wavelength end of NIS-2 The radiometric inter-calibration of CDS and SUMER was done for lines in the longer wavelength range of NIS-2 for the first 4.6 years of operations and indicate that the long wavelength part of the NIS-2 calibration was maintained. The present results thus extend this comparison to the whole NIS wavelength range. The line ratios method has been extended successfully to use ratios which depend on the plasma conditions.