Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

TV/TVM Tissue Doppler … Introducing Tissue Doppler What is it ?
Doppler Echocardiography
Doppler Echocardiography
Foundations of Medical Ultrasonic Imaging
Doppler echocardiography & tissue doppler Dolly mathew.
Doppler Ultrasound: Principles
Specular reflectorquasi-specular reflector quasi-Lambert reflector Lambert reflector Limiting Forms of Reflection and Scatter from a Surface.
7. Radar Meteorology References Battan (1973) Atlas (1989)
إعداد : أ. بلسم فهد صوفي 1 Ultrasound in Medicine Ch.3 Ultrasound pictures of the body.
01: Introduction to Ultrasound George David, M.S. Associate Professor of Radiology.
إعداد : أ. بلسم فهد صوفي،،،المصدر:محاضرات د.حنان 1 Ultrasound in Medicine Ch.4 Ultrasound pictures of the body.
Resident Physics Lectures
Hospital Physics Group
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
Basic Physics of Ultrasound
ECE 501 Introduction to BME
Specular reflectorquasi-specular reflector quasi-Lambert reflector Lambert reflector Limiting Forms of Reflection and Scatter from a Surface.
Doppler Effect(1) Stationary source Stationary observer Moving source
- Doppler Ultrasound & Obstetrical Ultrasound
Ultrasound Imaging Atam Dhawan.
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Development of an Ultrasound Lab Laura Wade April 4 th Z.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Doppler Radar From Josh Wurman NCAR S-POL DOPPLER RADAR.
The Doppler Shift Brought to you by McCourty-Rideout enterprises- tuned into your frequency.
Doppler Radar From Josh Wurman Radar Meteorology M. D. Eastin.
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
Doppler Effect The Doppler effect is the apparent change in the frequency of a wave motion when there is relative motion between the source of the waves.
Display of Motion & Doppler Ultrasound
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
Resident Categorical Course
Pegasus Lectures, Inc. COPYRIGHT 2006 Volume II Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th.
US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS.
Reflection & Transmission
Sound Overview The Facts of Sound The Ear and Sound Sound Vocabulary Musical Instruments and Sound.
THE DOPPLER EFFECT. When the source of a sound is moving towards you, the pitch sounds higher than that of the source. When the source moves away from.
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Chapter 14 Sound. Sound waves Sound – longitudinal waves in a substance (air, water, metal, etc.) with frequencies detectable by human ears (between ~
Sarah Gillies Ultrasound Sarah Gillies
Ultrasound Imaging Basic Principle: sound wave pulse emitted at >20 kHZ, some reflected, and some transmitted. Reflection from single signal – A-scan Reproduced.
Introduction to the Physics Echocardiography Introduction to the Physics Echocardiography Jose L. Rivera, M.D. January 9, 2010.
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
The Doppler Effect.
Doppler Effect - Applications
Ultrasound Physics Reflections & Attenuation ‘97.
Doppler Physics Waves from a static source Wave peaks evenly spaced around the source at 1 wavelength intervals.
Artifacts Ultrasound Physics George David, M.S.
Blood Flow Detection > by the Doppler principle > by Electromagnetic principles © D. J. McMahon 2014 rev
Performance Issues in Doppler Ultrasound 1. 2 Fundamental Tradeoffs In pulsed modes (PW and color), maximum velocity without aliasing is In pulsed modes,
Ultrasound Learning Objectives: Describe the properties of ultrasound;
Doppler Effect A Doppler effect is experienced whenever there is relative motion between a source of waves and an observer. A Doppler effect is experienced.
Ultrasound Physics Image Formation ‘97.
Doppler Effect and Redshift
Doppler Spectral Analysis
Lecture on Ultrasonography
Sound in medicine Lect.10.
Reflections & Attenuation
Ultrasound.
Ultrasound Physics Image Formation ‘97.
Ultrasound.
Doppler Radar Basics Pulsed radar
Doppler Effect.
Diagnostic Medical Sonography Program
Medical Ultrasound: Image Optimisation
Louis Allard, MSc, Guy Cloutier, PhD  Journal of Vascular Surgery 
DOPPLER ARTIFACTS.
Ultrasonic Imaging Using Contrast Agents
Presentation transcript:

Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital & Research Centre Biomedical Physics Department Riyadh, Kingdom of Saudi Arabia Doppler Ultrasound

Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer or the sound source is moving with respect to each other. when sound source is moving towards the detector/observer (R), the sound appears to have a higher frequency (shorter wavelength) when sound source is moving away from the detector/observer (L), the sound appears to have a lower frequency (longer wavelength) if the sound source is moving perpendicular (90  ) to the detector/observer, there is no change in frequency or in wavelength and thus no Doppler Effect is observed Both source and detector Source moving with are stationary respect to detector

Doppler ultrasound is used primarily to identify and evaluate blood flow in vessels velocity and waveform information can be used to evaluate stenoses, resistance and vessel patency Doppler Frequency Shift (f D ) where f 0 is the original frequency and f is the frequency of the returning echo (from RBC) f0f0 f (RBC) maximum shift is obtained when  = 0° minimum shift is obtained when  = 90° (recall, imaging of strongest echoes) the shift is comparatively small and typically is between kHz (audible sound) the shift is positive when the RBC is moving towards the transducer and is negative when the RBC is moving away from the transducer doppler frequency shift (f D ) = change in frequency = (f 0 - f)

Flow Speed (v) f0f0 f (RBC) It can be shown, that f D = (f 0 - f) = 2 v f 0 cos  c and re-arranging the above equation we have v = c f D 2 f 0 cos  where f 0 is the original frequency, f is the frequency of the returning echo (from RBC), v is the speed of the interface (RBC), c is the speed of sound in soft tissue (1540 m/s), f D is the Doppler shift and  is the angle between transducer and RBC flow direction The velocity equation can be simplified to v (cm/s) = 77 f D (kHz) 2 f 0 (MHz) cos  ASIDE The factor 77 is only valid when v, f 0 and f D are given in the units shown above.

the values of v calculated from the observed f D are only as accurate as the estimated Doppler angle  one reason why sonography is combined with Doppler techniques is to estimate the Doppler angle with a cross-sectional image of the vessel walls, the sonographer can adjust a flow direction indicator which, when combined with an indication of the beam direction, can yield the Doppler angle the instrument uses this value of angle to convert Doppler shift to flow speed reflector speed  doppler shift  operating frequency (f 0 )  doppler shift  doppler angle  doppler shift  NOTE The moving reflector can be a tissue boundary (blood vessel wall or heart wall) or blood cells in circulation

high Q transducers are desirable for Doppler because they produce a narrow range of ultrasound frequencies (f 0 ) Doppler systems tend to run at lower frequencies than B-mode systems because resolution is not as important and there is a need to minimize attenuation (blood is a weak scatterer) the range and frequency distribution of flow velocities can be displayed as a spectrum Spectral broadening is the result of a mixture of velocities in the sample and produces a shaded area below the peak velocity value Flow Velocity Waveform (Spectrum) Operational Techniques

Continuous Doppler in continuous Doppler, one transducer (high Q) continually transmits and another transducer (low Q) continuously receives the frequency of the two signals are subtracted to give the Doppler shift, which is in the audio range continuous wave Doppler is inexpensive and does not suffer from aliasing artifacts but lacks depth resolution and provides little spatial information continuous wave Doppler is good for measuring fast flow and deep lying vessels depth gain compensation is not used in continuous wave Doppler

Pulsed Doppler pulsed Doppler allows both velocity and depth information (ranging) to be obtained the Doppler information is only provided for a specific area pulsed Doppler uses a longer pulse length than B-mode, typically up to 15 mm long Doppler information is displayed audibly and graphically as a waveform (spectrum) aliasing artifacts result in errors in estimating velocity the use of lower frequencies allow higher velocities to be measured without aliasing Dublex scanning involves displaying Doppler data on real time (B-mode) images and allows velocity and position information to be obtained simultaneously most of the time is spent in Doppler mode, the B-mode image being updated once a second

Continuous Doppler Pulsed Doppler

Color Flow Doppler Color Doppler is a hybrid that combines anatomic information obtained using B-mode system with flow information obtained using pulsed Doppler analysis pulse length in color Doppler is typically 2 mm colors (blue and red) are assigned dependent on motion (toward or away) from the transducer turbulence (i.e.: variations in flow direction) can vary between green and yellow the depth of each color varies with the velocity of flow, stationary tissues appear gray information is provided over a large area and superimposed on a gray scale image color Doppler can detect flow in vessels too small to see by imaging alone spectral analysis may also be obtained using commercial color Doppler systems modern instruments incorporate both color Doppler and spectral Doppler

Power Doppler Power Doppler is a signal processing method that relies on the total strength of the Doppler signals (amplitude) and ignores directional (phase) information the power (also known as energy) mode of signal acquisition is dependent on the amplitude of all Doppler signals, regardless of the frequency shift this dramatically improves the sensitivity to motion (i.e.: slow blood flow) at the expense of directional and quantitative flow information compared to conventional color flow imaging, power Doppler produces images that have more sensitivity to motion and are not affected by the Doppler angle and aliasing is not a problem as only strength of the frequency shifted signals are analyzed greater sensitivity allows detection and interpretation of very subtle and slow blood flow frames rates though tend to be slower and a significant amount of “flash artifacts” occur, which are related to color signals arising from moving tissues, patient motion or transducer motion Power Doppler does not mean more power to the patient, but in fact the power level are typically the same as those in a standard color flow procedure

Circle of Willis Submucosus fibroid, note small vessels inside the tumor